HEMOSIL D-DIMER HS 500 CONTROL

K133582 · Instrumentation Laboratory · GGN · Aug 15, 2014 · Hematology

Device Facts

Record IDK133582
Device NameHEMOSIL D-DIMER HS 500 CONTROL
ApplicantInstrumentation Laboratory
Product CodeGGN · Hematology
Decision DateAug 15, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Indications for Use

HemosIL D-Dimer HS 500 Controls are assayed, human-sourced controls intended for the quality control of the HemosIL D-Dimer HS 500 assay as performed on the ACL TOP Family System, in a clinical laboratory setting. The controls are intended for in vitro diagnostic use. Level 1 D-D HS 500 Control is intended for the assessment of the assay around the clinical cutoff for VTE (500 ng/mL FEU). Level 2 D-D HS 500 Control is intended for the assessment of precision and accuracy of the assay at abnormal D-Dimer levels (above the cut-off).

Device Story

Liquid control materials containing partially purified D-Dimer (derived from human fibrin digested with human plasmin); used for quality control of HemosIL D-Dimer HS 500 assay on ACL TOP family analyzers. Device provides known concentrations of D-Dimer to verify assay precision and accuracy in clinical laboratories. Healthcare providers use control results to validate instrument performance before patient sample testing; ensures reliability of D-Dimer measurements used in VTE assessment. Kit includes two levels: Level 1 (near clinical cut-off) and Level 2 (abnormal levels).

Clinical Evidence

No clinical data. Bench testing only. Precision and reproducibility studies conducted on ACL TOP 300/500/700 analyzers. Precision study (N=80 per level/instrument) showed within-run CV% 1.80-4.59% and total CV% 2.69-5.86%. Reproducibility study across three sites (N=80 per level/site) met pre-specified acceptance criteria (repeatability CV% ≤8%, total CV% ≤12% for Level 1; ≤10% for Level 2). Instrument-to-instrument variability was <10%.

Technological Characteristics

Latex-based immunoturbidimetric immunoassay. Contains partially purified D-Dimer from human fibrin digested with human plasmin, bovine serum albumin, buffer, stabilizers, and preservative. Liquid form. On-board stability 24 hours at 15-25°C; open-vial stability 1 month at 2-8°C or 3 months at -20°C. Analyzed on ACL TOP family systems.

Indications for Use

Indicated for quality control of the HemosIL D-Dimer HS 500 assay on the ACL TOP Family System in clinical laboratory settings. Level 1 assesses performance near the 500 ng/mL FEU VTE clinical cutoff; Level 2 assesses precision and accuracy at abnormal levels above the cutoff.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k133582 B. Purpose for Submission: To obtain clearance of a modified assayed control materials intended for the quality control of the HemosIL D-Dimer HS 500 assay as performed on the ACL TOP Family Systems. C. Measurand: D-dimer fragment of fibrin D. Type of Test: Latex based immunoturbidimetric assay E. Applicant: Instrumentation Laboratory, Inc. F. Proprietary and Established Names: HemosIL D-Dimer HS 500 Controls G. Regulatory Information: 1. Regulation section: CFR §864.5425 Multipurpose system for in vitro coagulation studies 2. Classification: Class II 3. Product code: GGN, Plasma, Coagulation Control 4. Panel: Hematology (81) H. Intended Use: 1. Intended use(s): D-Dimer HS 500 Controls are assayed human-sourced controls intended for the quality control of the HemosIL D-Dimer HS 500 assay as performed on the ACL TOP Family {2} Systems, in a clinical laboratory setting. The controls are intended for in vitro diagnostic use. Level 1 D-Dimer HS 500 Control is intended for the assessment of precision and accuracy of the assay around the clinical cut-off for VTE (500nglmL FEU). Level 2 D-Dimer HS 500 Control is intended for the assessment of precision and accuracy of the assay at abnormal D-Dimer levels (above the cut-off). 2. Indication(s) for use: Same as intended use. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: ACL TOP family of analyzers (k073377). I. Device Description: The Level 1 and Level 2 D-Dimer HS 500 Controls are prepared by means of a dedicated process and contain different concentrations of partially purified D-Dimer obtained by digestion of Factor XIIIa cross-linked human fibrin with human plasmin. The HemosIL D-Dimer HS 500 Controls kit consists of 2 liquid control levels: Level 1 (Cat. No.0020013110) and Level 2 (Cat. No.0020013120). Each control consists of 5 vials x1 mL of a liquid solution of D-Dimer partially purified from human fibrin digested with human plasmin containing bovine serum albumin, buffer, stabilizers and preservative. Use of both controls is recommended for a complete quality control program. The kit is only available in one (1) configuration (Cat No. 0020013100). J. Substantial Equivalence Information: 1. Predicate device name(s) and Predicate 510(k) number(s): HemosIL D-Dimer HS 500 Controls, Low Control (k090264) HemosIL D-Dimer HS 500 Controls, High Control (k090264) 2. Comparison with predicate: {3} | Similarities | | | | --- | --- | --- | | Item | Device HemosIL D-Dimer HS 500 Controls | Predicate HemosIL D-Dimer HS 500 Controls | | Intended Use/ Indications for Use | HemosIL D-Dimer HS 500 Controls are assayed, human-sourced controls intended for the quality control of the HemosIL D-Dimer HS 500 assay as performed on the ACL TOP Family Systems, in a clinical laboratory setting. The controls are intended for in vitro diagnostic use. Level 1 D-D HS 500 Control is intended for the assessment of precision and accuracy of the assay around the clinical cut-off for VTE (500 ng/mL FEU). Level 2 D-D HS 500 Control is intended for the assessment of precision and accuracy of the assay at abnormal D-Dimer levels (above the cut-off). | For the quality control of the D-Dimer HS 500 assay performed on the ACL TOP Family Systems | | Principle of Operation | Latex-based immunoturbidimetric immunoassay | Same | | Technology | The Controls are prepared by means of a dedicated process and contain different concentrations of partially purified D-Dimer obtained by digestion of Factor XIIIa cross-linked human fibrin with human plasmin. | Same | | Opened Stability (@ 2-8°C) | 1 month | Same | | Differences | | | | --- | --- | --- | | Item | Device HemosIL D-Dimer HS 500 Controls | Predicate HemosIL D-Dimer HS 500 Controls | | Kit Part # | 0020013100 | 0020500200 | | Delivery | Liquid | Lyophilized | | Composition | Level 1 D-D HS 500 Control 5 vials x 1 mL of a liquid solution of D-Dimer partially purified from human fibrin digested with human plasmin and stabilized with a crosslinker containing bovine serum albumin, buffer, stabilizers and preservative | Low D-D HS 500 Control 5 vials x 1 mL of a lyophilized solution of D-Dimer partially purified from human fibrin digested with human plasmin containing bovine serum albumin, buffer, stabilizers and preservative. | {4} | Differences | | | | --- | --- | --- | | Item | Device HemosIL D-Dimer HS 500 Controls | Predicate HemosIL D-Dimer HS 500 Controls | | | **Level 2 D-D HS 500 Control** 5 vials x 1 mL of a liquid solution of D-Dimer partially purified from human fibrin digested with human plasmin and stabilized with a crosslinker containing bovine serum albumin, buffer, stabilizers and preservative. | **High D-D HS 500 Control** 5 vials x 1 mL of a lyophilized solution of D-Dimer partially purified from human fibrin digested with human plasmin containing bovine serum albumin, buffer, stabilizers and preservative. | | | **Buffer** Hepes (50 mM), NaCl (100 mM), CaCl2 (2 mM), NaN3 (1g/L), BSA (10g/L) at a pH of 7.5 | **Buffer** Hepes (50 mM), NaCl (100 mM), Glycine 4%, NaN3 (1g/L), BSA (10g/L) at a pH of 7.5 | | On Board Stability (ACL TOP 15-25°C) | 24 hours | 8 hours | | Stability @ -20°C | 3 months | 2 months | ## K. Standard/Guidance Document Referenced: - CLSI EP05-A2 - CLSI EP15-A2 - CLSI EP25-A - CLSI H59-A ## L. Test Principle: The controls are intended for the quality control of the HemosIL D-Dimer HS 500 assay on Instrumentation Laboratory’s ACL TOP family analyzers. The controls contain different concentrations of partially purified D-Dimer obtained by digestion of human fibrin with human plasmin. In the D-Dimer assay the latex reagent coated with the monoclonal antibody specific for the D-Dimer domain agglutinates upon binding to the antigen present in the patient plasma. The degree of agglutination is directly proportional to the concentration of D-Dimer in the sample and is determined by measuring the decrease of transmitted light caused by the aggregates (turbidimetric immunoassay). ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: {5} # Precision study: The Sponsor tested 3 lots of controls using one (1) lot of HemosIL D-Dimer HS 500 assay on a representative member (1) of the ACL TOP Family (ACL TOP 500). Each control was analyzed for 20 days, with 2 runs a day, 2 replicates per run for each sample level $(N = 80/$ per level/ per instrument). Calibration was performed at the start of the study $(T = 0)$ with the kit's calibration plasma. Acceptance criteria were set at $\leq 8\%$ for within-run CV% and $\leq 12\%$ for total CV. This study was considered sufficient to support the precision claim as the $\% \mathrm{CV}$ for each tested variable meets the pre-specified acceptance criteria. Results are summarized in the table. | Control Level | Level 1 | Level 2 | | --- | --- | --- | | Grand Mean (ng/mL FEU) | 566 | 1730 | | Within Run (%CV) | 4.59% | 1.80% | | Run-to-Run (%CV) | 2.4% | 0.6% | | Day-to-Day (%CV) | 1.5% | 0.5% | | Lot-to-Lot (%CV) | 2.3% | 1.9% | | Total (%CV) | 5.86% | 2.69% | The variability between instrument platforms ACLTOP 300/500/700 was tested in a separate experiment, using three (3) lots of HemosIL D-Dimer HS Controls by three (3) operators using three (3) instruments and one (1) lot of HemosIL D-Dimer Reagent. In this study, each operator used a different lot of control material and a different instrument type, representing "worst case" scenario. The instrument to instrument variability was less than $10\%$ . The table below shows a summary of the results. | Statistics | Control Level 1 | Control Level 2 | | --- | --- | --- | | Overall Mean (ng/mL FEU) | 527.77 | 1842.02 | | Between Runs (%CV) | 3.22% | 0.40% | | Between Days (%CV) | 0.00% | 0.00% | | Between (Control Lot + Instruments) | 2.13% | 9.47% | | Repeatability (%CV) | 5.83% | 1.99% | | Total (%CV) | 6.99% | 9.69% | | Between Instruments Specification | ≤10% | ≤10% | | PASS/FAIL | PASS | PASS | # Reproducibility study A reproducibility study was performed in three (3) different locations using one (1) lot of HemosIL D-Dimer HS 500 controls (PN 0020013100) and one (1) lot of HemosIL D-Dimer HS 500 assay on an ACL TOP 700. At each site each control level was analyzed for 20 days, with 2 runs a day, 2 replicates per run for each sample level $(N = 80 / \text{per level} / \text{per site})$ . Calibration was performed at the start of the study $(T = 0)$ with the kit's calibrator. The results met pre-specified acceptance criteria of {6} repeatability $\mathrm{CV\%}\leq 8\%$ and total $\mathrm{CV\%}\leq 12\%$ for Level 1 Control and $\leq 10\%$ for Level 2 Control. For a summary of the results please see table below. | Control Level | Site 1 | | Site 2 | | Site 3 | | | --- | --- | --- | --- | --- | --- | --- | | | Level 1 | Level 2 | Level 1 | Level 2 | Level 1 | Level 2 | | Mean (ng/mL FEU) | 483.84 | 1648.69 | 520.91 | 1647.38 | 537.15 | 1750.15 | | Repeatability (CV %) | 2.15% | 1.49% | 3.81% | 1.50% | 3.10% | 1.25% | | Repeatability Spec. | ≤8% | ≤8% | ≤8% | ≤8% | ≤8% | ≤8% | | Repeatability | PASS | PASS | PASS | PASS | PASS | PASS | | Between run (CV%) | 0.79% | 0.29% | 1.42% | 0.17% | 0.51% | 0.90% | | Between day (CV%) | 0.61% | 0.89% | 0.79% | 0.72% | 1.58% | 0.83% | | Total (CV%) | 2.37% | 1.76% | 4.14% | 1.67% | 3.52% | 1.75% | # b. Linearity/assay reportable range: Not applicable # c. Traceability, Stability, Expected values (controls, calibrators, or methods): Value Assignment A value assignment study is performed 4 times on a minimum of two (2) ACL TOP units. For each time, the assay is calibrated using the Reagent House Standard and the Calibrator House Standard followed by analyzing 8 samples of the Control Test and 8 of the Control House Standard. In total 32 results are obtained for each material. The package insert acceptance range is $\pm 18\%$ of the target value for Level 1 and $\pm 10\%$ of the target value for Level 2 controls. Once produced, acceptance ranges for a specific lot are printed on the package insert sheet, i.e. every lot has individualized target value. The customer acceptance criteria are then $\pm 20\%$ of the target values for Level 1 and $\pm 20\%$ for Level 2 control. # Stability Real-time shelf life stability studies are ongoing. The study will be completed on an ACL TOP 700 analyzer using three (3) lots of controls and three (3) lots of the assay reagent. The three lots of assay reagent are to be used instead of one lot due to shorter stability of the assay reagent as compare to stability of controls. For on-board stability and in open-vial stability studies, two (2) lots of controls were tested using one (1) lot of the reagent. The stability claims were calculated using an acceptance range of $\pm 15\%$ compared to the unstressed baseline or compared against initial values $(T = 0)$ when placed on-board the instrument. The results support open vial stability for 1 month when open vial is re-stored at $2 - 8^{\circ}\mathrm{C}$ and 3 months when open vial is re-stored at $-20^{\circ}\mathrm{C}$ . Also, provided data support 24 hour on-board stability at $15 - 25^{\circ}\mathrm{C}$ . {7} d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Reference ranges are on each package insert based on each lot of control Value assignment N. Proposed Labeling: The revised labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10 {8} O. Conclusion: The information submitted in this premarket notification is complete and supports a substantial equivalence decision. 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...