GENESIS II

K133434 · Dornier Medtech America, Inc. · JAA · Nov 26, 2014 · Radiology

Device Facts

Record IDK133434
Device NameGENESIS II
ApplicantDornier Medtech America, Inc.
Product CodeJAA · Radiology
Decision DateNov 26, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2

Indications for Use

The Dornier GENESIS II is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position.

Device Story

Genesis II is an image-intensified fluoroscopic X-ray system; utilizes X-ray generator, tube housing, image intensifier, and procedure table. System captures radiographic and fluoroscopic images for urology, gynecology, endoscopic, and contrast examinations. Operated by clinicians in clinical settings; patient positioned horizontally or vertically on table. X-ray tube and detector are fixed relative to each other; table and imaging unit move to position patient. Output images displayed on monitors for diagnostic review by healthcare providers. Device aids clinical decision-making by providing diagnostic-quality imaging for urological/gynecological procedures.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, and 60601-2-54 standards. Diagnostic image quality verified by a board-certified radiologist through evaluation of sample images, confirming performance is substantially equivalent to the predicate Genesis (K122768).

Technological Characteristics

Image intensified fluoroscopic X-ray system. Components: X-ray generator, tube housing, image intensifier, monitors, procedure table. Connectivity: DICOM (NEMA PS 3.1-3.20). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-3 (radiation protection), IEC 60601-1-6 (usability), IEC 60601-2-28 (X-ray source assemblies), IEC 60601-2-54 (X-ray equipment).

Indications for Use

Indicated for patients requiring general urological and gynecological examinations, endoscopic procedures, and contrast studies, including radiography and fluoroscopy imaging in horizontal or vertical positions.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple, clean, and easily recognizable. November 26, 2014 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Dornier MedTech America, Inc. % Mr. John Hoffer Vice President, Quality/Regulatory/Clinical 1155 Roberts Blvd. KENNESAW GA 30144 Re: K133434 Trade/Device Name: Genesis II Regulation Number: 21 CFR 892.1650 Regulation Name: Image intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: October 23, 2014 Received: October 23, 2014 Dear Mr. Hoffer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A Ochs for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K133434 Device Name Genesis II Indications for Use (Describe) The Dornier GENESIS II is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Dornier MedTech America, Inc.'s Genesis II | Sponsor Company: | Dornier MedTech America, Inc. | |-----------------------|-------------------------------------------------------------------------------------| | Sponsor Address: | 1155 Roberts Blvd. Kennesaw, GA 30144 | | Contact Person: | John Hoffer<br>Vice President, Quality/Regulatory/Clinical | | Email Address: | jhoffer@dornier.com | | Phone Number: | 770-514-6163 | | Fax Number: | 770-514-6291 | | Date Prepared: | October 23, 2014 | | Device Trade Name(s): | Genesis II | | Device Common Name: | Image Intensified Fluoroscopic X-ray System | | Classification Name | (21 CFR 892.1650, Product Code JAA) | | Predicate Device(s): | Dornier Genesis (K122768)<br>Dornier Gemini (K121656)<br>Dornier Urotract (K955019) | ## General Device Description: The Dornier Genesis II is an Image Intensified Fluoroscopic X-ray System. It is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position. The Genesis II consists of the following components: an x-ray generator and tube housing, image intensifier detector, monitors and procedure table. An X-ray cabinet contains system elements such as the X-ray generator, power electronics and electronics for the imaging chain. The Genesis II is a radiographic and fluoroscopy examination table with the X-ray tube housing mounted over the table on a fixed arm. An image intensifier is mounted underneath the patient table. While the X-ray tube and detector are fixed in their positions relative to each other when the system is in use, the table top and X-ray/detector unit can be moved in a variety of planes to position the patient in the desired imaging position. {4}------------------------------------------------ ## Indications For Use: The Dornier GENESIS II is intended for use in a wide field of applications, including all general examinations in urology and gynecology, as well as endoscopic and contrast examinations, imaging with radiography and/or fluoroscopy on patients in either the horizontal or vertical position. Genesis II has the same intended use and indications for use as the predicate Genesis system (K122768). ## Comparison of Technological Characteristics: The Genesis II is an Image Intensified Fluoroscopic X-ray System with an image intensifier image receptor system. Both the Genesis II and the predicate Genesis (K122768) consist of the following components: an X-ray generator and tube housing, image intensifier, monitors and procedure table. The X-ray cabinet contains system elements such as the X-ray generator, power electronics and electronics for the imaging chain. The only technological difference between the Genesis II and the predicate Genesis (K122768) is the substitution of an Image Intensifier for the predicate device's Flat Panel Detector. The image intensifier system in the Genesis II is identical to the solid state image intensifier system used in the Dornier Gemini Lithotripter system (K121656). Both imaging systems have demonstrated the ability to provide acceptable image quality appropriate for the intended uses of the device. Furthermore, the use of image intensifiers with uroradiology systems with substantially similar indications for use as those of the Genesis II have been previously cleared by FDA (e.g., Urotract I (K955019)). Therefore, the use of cleared image intensifier technology in the Genesis II does not raise any new questions of safety or effectiveness and meets the requirements for substantial equivalence. ## Non-Clinical Performance Data: The device is in compliance with the following consensus standards: - · IEC 60601-1: Medical electrical equipment. Part 1: General requirements for safety (2005) - · IEC 60601-1-2: Medical electrical equipment. Part 1-2: General requirements for safety; Electromagnetic compatibility-requirements and tests (2007) - IEC 60601-1-3: Medical electrical equipment. Part 1: General requirements for safety; general . requirements for radiation protection in diagnostic X-ray equipment (2008) - · IEC 60601-1-6: Medical electrical equipment. Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability (2006) - · IEC 60601-2-28: Medical electrical equipment. Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis (2010) - IEC 60601-2-54: Medical electrical equipment. Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy (2009) - NEMA PS 3.1-3.20 Digital Imaging and Communications in Medicine (DICOM) Set {5}------------------------------------------------ To verify and validate diagnostic quality of the device output image, a board certified radiologist evaluated a number of sample images obtained using the Genesis II and determined that the device output images were of acceptable diagnostic quality and that the performance of the proposed device is substantially equivalent to that of the cleared Genesis (K122768). ## Conclusion and Summary of Substantial Equivalence: Genesis II has the same intended use and indications for use as the previously cleared Genesis system (K122768). The only technological difference between the Genesis II and the predicate Genesis system is the substitution of an Image Intensifier for the predicate device's Flat Panel Detector. The image intensifier system in the Genesis II is identical to the solid state image intensifier system used in the Dornier Gemini Lithotripter system. The Genesis II has demonstrated the ability to provide acceptable image quality appropriate for the intended uses of the device. Based on intended use, technological characteristics, and performance data as discussed above, Dornier believes that the Genesis II and the predicate devices selected are substantially equivalent and that the differences between the devices do not raise new issues of safety or effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...