DIAZYME 25-OH VITAMIN D ASSAY; DIAZYME 25-OH VITAMIN D CONTROL SET

K133410 · Diazyme Laboratories · MRG · Mar 14, 2014 · Clinical Chemistry

Device Facts

Record IDK133410
Device NameDIAZYME 25-OH VITAMIN D ASSAY; DIAZYME 25-OH VITAMIN D CONTROL SET
ApplicantDiazyme Laboratories
Product CodeMRG · Clinical Chemistry
Decision DateMar 14, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1825
Device ClassClass 2

Intended Use

The Diazyme 25-OH Vitamin D Assay is intended for use in clinical laboratories for the quantitative determination of total 25-OH Vitamin D in human serum and plasma on automated chemistry analyzer. Measurement of 25-hydroxyvitamin D (25-OH-D) is for the assessment of vitamin D sufficiency. For in vitro diagnostic use only. The Diazyme 25-OH Vitamin D Control Set is intended for use as quality controls for the Diazyme 25-OH Vitamin D Assay Kit only. For in vitro diagnostic use only.

Device Story

Direct competitive colorimetric immunoassay for quantitative determination of total 25-OH Vitamin D in human serum and plasma. Principle: alpha-complementation of beta-galactosidase; competition between enzyme donor-25-OH Vitamin D conjugate, anti-Vitamin D antibody, and sample 25-OH Vitamin D. Higher sample concentration yields higher beta-galactosidase activity. Uses nitro-phenyl-beta-galactoside (NPG) substrate; reaction product measured at 415 nm. Used in clinical laboratories on automated chemistry analyzers (e.g., Roche Modular P). Healthcare providers use results to assess vitamin D sufficiency. Benefits include automated, high-throughput quantification of vitamin D status.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A (n=80 per specimen, 20 days); total CVs 2.1-8.8%. Linearity established 7.6-147.8 ng/mL per CLSI EP6-A. LoB/LoD/LoQ determined per CLSI EP17-A (LoQ 7.6 ng/mL). Interference study (CLSI EP7-A2) showed <10% deviation for common substances. Method comparison (n=98) vs. predicate showed correlation coefficient 0.984, slope 1.005, intercept -0.21. Reference range study (n=157) established 15.0-45.9 ng/mL (2.5th-97.5th percentile).

Technological Characteristics

Direct competitive colorimetric immunoassay. Reagents: sheep monoclonal anti-25-OH Vitamin D antibody, enzyme donor-Vitamin D conjugate, enzyme acceptor, NPG substrate. Energy source: automated chemistry analyzer (photometric detection at 415 nm). Sample types: serum, K3-EDTA plasma, Lithium-Heparin plasma. Standards: CLSI EP05-A2, EP6-A, EP07-A2, EP17-A. Storage: 2-8°C.

Indications for Use

Indicated for quantitative determination of total 25-OH Vitamin D in human serum and plasma to assess vitamin D sufficiency in clinical laboratory settings. For prescription use only.

Regulatory Classification

Identification

A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.

Special Controls

*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and (2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ k133410 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 | Submitter's name: | Diazyme Laboratories | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's address: | 12889 Gregg Court<br>Poway, CA 92064 | | Name of Contact Person: | Dr. Abhijit Datta<br>Director, Technical Operations<br>Diazyme Laboratories<br>12889 Gregg Court<br>Poway, CA 92064<br>Email: abhijit.datta@diazyme.com<br>Phone: 858-455-4762<br>Fax: 858-455-2120 | | Date the Summary was Prepared: | November 04, 2013 | | Name of the Device | Diazyme 25-OH Vitamin D Assay Kit<br>Diazyme 25-OH Vitamin D Control Set | | Trade Name: | Diazyme 25-OH Vitamin D Assay<br>Diazyme 25-OH Vitamin D Control Set | | Common/Usual Name | Vitamin D Assay | | Device Classification Name | Vitamin D Test System | | Product code: | MRG - Vitamin D Test System<br>JJX – Single (specified) Analyte Controls (Assayed and<br>Unassayed) | | Panel: | Chemistry (75) | | Submission Type | 510k | {1}------------------------------------------------ | Regulation Number | 21 CFR 862.1825 - Vitamin D Test System<br>21 CFR 862.1660 - Quality Control material (Assayed<br>and Un-assayed) | |----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Class | II (Assay)<br>I (Control) | | Predicate Device: | The Diazyme 25-OH Vitamin D Assay Kit and Control<br>Set is substantially equivalent to the currently marketed<br>LIAISON® 25-OH Vitamin D TOTAL Assay<br>(k112725, k071480). | | Manufacturing Address | Diazyme Laboratories<br>12889 Gregg Court<br>Poway, CA 92064<br>USA | | Establishment Registration | 2032900 | # DESCRIPTION OF THE DEVICE The Diazyme 25-OH Vitamin D Assay is a direct competitive colorimetric immunoassay for the quantitative determination of total 25-OH vitamin D in serum and plasma. The assay is based on the principle of α-complementation of the enzyme β-galactosidase and the competition between an enzyme donor-25-OH Vitamin D conjugate, an anti-Vitamin D antibody and the 25-OH Vitamin D content of a serum sample. Samples with higher 25-OH Vitamin D concentrations produce higher ß-galactosidase activities and vice versa. A nitro-phenyl-0-galactoside derivative (NPG) is used as the enzyme substrate. The reaction's product has maximum absorbance at 415 nm. The 25-OH Vitamin D concentration of a specimen is proportional to the measured Bgalactosidase activity. Five calibration levels are needed for each run. Calibrators are treated exactly the same as patient samples. The Diazyme 25-OH Vitamin D Control Set (2 levels) is intended for use with the Diazyme 25-OH Vitamin D Assay kit only. Controls are treated exactly the same as patient samples. The quality controls assist laboratory users in verification steps ensuring that the assay reagents are functioning correctly. Users are instructed to verify the calibration curve with the controls. {2}------------------------------------------------ ## INDICATIONS FOR USE The Diazyme 25-OH Vitamin D Assay is intended for use in clinical laboratories for the quantitative determination of total 25-OH Vitamin D in human serum and plasma on automated chemistry analyzer. Measurement of 25-hydroxyvitamin D (25-OH-D) is for the assessment of vitamin D sufficiency. For in vitro diagnostic use only. The Diazyme 25-OH Vitamin D Control Set is intended for use as quality controls for the Diazyme 25-OH Vitamin D Assay Kit only. For in vitro diagnostic use only. | LIAISON® 25-OH Vitamin D TOTAL Assay | Diazyme 25-OH Vitamin D Assay | |--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | (predicate k112725)<br>Kit can be only used for the 25-OH Vitamin D<br>quantification on the Diasorin Liaison® analyzer. | Kit can be used for the 25-OH Vitamin D quantifi-<br>cation on the Roche Modular P Chemistry Analyz-<br>er and similar chemistry analyzer systems. | | <b>Magnetic Particles:</b><br>Magnetic particles coated with antibody against<br>25-OH Vitamin D, protein, phosphate buffer, <<br>0.1% sodium azide. | <b>Dilution buffer:</b><br>Sheep Antibody against 25-OH Vitamin D, pro-<br>teins, phosphate buffer, < 0.1% sodium azide. | | <b>Assay Buffer:</b><br>Buffer with 10% ethanol, surfactants and preserva-<br>tives. | <b>Reagent R1:</b><br>Proprietary dissociation solution with surfactants<br>and preservatives. | | <b>Conjugate:</b><br>25-OH Vitamin D conjugated to an isoluminol<br>derivative, in phosphate buffer with 10% ethanol,<br>EDTA, surfactant and preservatives. | <b>Reagent R2:</b><br>Enzyme Donor-Vitamin D conjugate, proteins,<br>surfactants and preservatives. | | <b>LIAISON Starter 1:</b><br>Catalyst in 4% NaOH. | <b>Reagent R3:</b><br>Enzyme Acceptor, proteins, surfactants and pre-<br>servatives. | | <b>LIAISON Starter 2:</b><br>0.12% peroxide solution. | N/A | | <b>LIAISON Wash/System Liquid:</b><br>Phosphate buffer solution (10x concentrate). Pre-<br>servative: sodium azide. | N/A | | <b>LIAISON 25-OH Vitamin D diluent:</b><br>Human serum with buffer salts and <0.1% sodium<br>azide. | N/A | | Calibrator Set (k071480) | Calibrator set | | 1 x 1.0 mL Calibrator 1<br>1 x 1.0 mL Calibrator 2 | 1 x 1.0 mL Calibrator 1<br>1 x 1.0 mL Calibrator 2<br>1 x 1.0 mL Calibrator 3<br>1 x 1.0 mL Calibrator 4<br>1 x 1.0 mL Calibrator 5 | Table 1 Summary of Assay Kit Components {3}------------------------------------------------ | Control set (k071480) | Diazyme Control set | |----------------------------------------|----------------------------------------| | 1 x 1.0mL Control 1 | 1 x 1.0mL Control 1 | | 1 x 1.0mL Control 2 | 1 x 1.0mL Control 2 | | Serum-based controls | Serum-based controls | | Liquid form | Liquid form | | Stable for at least 12 months at 2-8C. | Stable for at least 12 months at 2-8C. | | Assigned range is mean +/- 25%. | Assigned range is mean +/- 25%. | # PERFORMANCE TESTING SUMMARIES ON THE ROCHE MODULAR-P CHEMIS-TRY ANALYZER # Precision Study The precision of the Diazyme 25-OH Vitamin D Assay was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) EP5-A guideline. A total of 11 precision levels were used in the study: - . Two serum controls (containing 23.1 ng/mL and 45.7 ng/mL). - Nine serum samples distributed across the dynamic range of the assay. ● Controls and samples were measured daily over the span of 20 days, using three lots of reagents and one chemistry analyzer. A total of 40 independents run were performed on each specimen. Each run produced two measurements. A total of 80 data points were obtained per specimen. The mean value (Mean), standard deviation, within-run imprecision and total imprecision are calculated and summarized in the following tables: | 25-OH Vitamin D (ng/mL) | | | Within-run | | Between-run | | Total | | |-------------------------|----|---------|------------|------|-------------|------|---------|------| | Specimen | n | Mean | SD | CV | SD | CV | SD | CV | | | | (ng/mL) | (ng/mL) | (%) | (ng/mL) | (%) | (ng/mL) | (%) | | Control #1 | 80 | 23.1 | 1.47 | 6.4 | 1.04 | 4.5 | 1.68 | 7.3 | | Control #2 | 80 | 45.7 | 2.06 | 4.5 | 1.67 | 3.7 | 2.12 | 4.6 | | Sample #1 | 80 | 22.6 | 1.19 | 5.3 | 1.11 | 4.9 | 1.45 | 6.4 | | Sample #2 | 80 | 31.7 | 1.42 | 4.5 | 1.59 | 5.0 | 1.81 | 5.7 | | Sample #3 | 80 | 40.6 | 1.42 | 3.5 | 1.59 | 3.9 | 1.66 | 4.1 | | Sample #4 | 80 | 48.6 | 2.32 | 4.8 | 1.71 | 3.5 | 2.41 | 4.9 | | Sample #5 | 80 | 55.8 | 2.14 | 3.8 | 1.73 | 3.1 | 2.34 | 4.2 | | Sample #6 | 80 | 65.4 | 2.03 | 3.1 | 1.79 | 2.7 | 2.42 | 3.7 | | Sample #7 | 80 | 69.7 | 2.02 | 2.9 | 1.99 | 2.9 | 2.55 | 3.7 | | Sample #8 | 80 | 92.8 | 2.52 | 2.7 | 2.02 | 2.2 | 3.40 | 3.7 | | Sample #9 | 80 | 134.6 | 2.97 | 2.2 | 2.69 | 2.0 | 3.87 | 2.9 | | Very low<br>Sample #1 | 80 | 9.4 | 1.22 | 13.0 | 0.98 | 10.4 | 1.31 | 14.0 | | Very low<br>Sample #2 | 80 | 11.2 | 1.58 | 14.2 | 0.88 | 7.9 | 1.55 | 13.9 | # Linearity/Reportable Range {4}------------------------------------------------ To establish the linearity of the 25-OH Vitamin D assay, a study design was used based on the CLSI protocol EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approach: Approved Guideline. Eleven levels of linearity were prepared by diluting a high serum sample containing 147.8 ng/mL of 25-OH Vitamin D with Vitamin D-depleted serum (0 ng/mL, Seracare Life Sciences). These samples were tested with the Diazyme 25-OH Vitamin D assay, in triplicates. The results were processed using the EP Evaluator Software (Version 8.0) parameterized to an allowable systematic error of 8.9%. The assay was found to be linear between 7.6 ng/mL and 147.8 ng/mL. # LoB/LoD/LoO The Limit of Blank (LoB), the Limit of Detection (LoD) and the Limit of Quantitation (LoQ) of the Diazyme 25-OH Vitamin D assay on microplate were determined according to CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation. The following are the limits determined with the Diazyme 25-OH Vitamin D Assay: LoB = 2.0 ng/mL LoD = 3.5 ng/mL LoQ = 7.6 ng/mL # Analytical specificity # Interference Study The Diazyme 25-OH Vitamin D Assay was subjected to an interference study according the CLSI EP7-A2 protocol. The following substances normally present in the blood produced less than 10% deviation when tested at levels equal to the concentrations listed below: | Interference | Concentration | |-------------------------|---------------| | Conjugated Bilirubin | 40 mg/dL | | Free Bilirubin | 40 mg/dL | | Hemoglobin | 100 mg/dL | | Ascorbic Acid | 176 mg/dL | | Triglycerides | 750 mg/dL | | Uric Acid | 20 mg/dL | | Biotin | 2 mg/dL | | Human Serum Albumin | 9 g/dL | | N-Acetyl Cysteine Amide | 1663 ng/mL | | Ampicillin | 1000 ng/mL | | Cyclosporine C | 105 ng/mL | | Cefoxitin | 660 ng/mL | | Acetylsalicylic Acid | 1000 ng/mL | {5}------------------------------------------------ | Rifampicin | 64 ng/mL | |---------------|-----------| | Acetaminophen | 200 ng/mL | | Ibuprofen | 500 ng/mL | | Theophylline | 100 ng/mL | ## Cross Reactivity Cross-reactivity of the Diazyme 25-OH Vitamin D Assay was determined by adding Vitamin D metabolites to serum pool samples. Based on the results in the table below, the assay did not cross react with Vitamin D2 and Vitamin D3 and the assay recovers both 25-OH Vitamin D2 and 25-OH Vitamin D3 similarly. Cross-reactivity with various Vitamin D metabolites is summarized in the table below: | Compound | Concentration tested (ng/mL) | Cross-reactivity | |-------------------------|------------------------------|------------------| | 25-OH Vitamin D3 | 44.0 | 100% | | 25-OH Vitamin D2 | 44.0 | 92.3% | | Vitamin D3 | 44.0 | 1.0% | | Vitamin D2 | 44.0 | 2.9% | | 1,25-(OH)2 Vitamin D3 | 2.9 | 2.5% | | 1,25-(OH)2 Vitamin D2 | 2.9 | -1.5% | | 24R,25-(OH)2 Vitamin D3 | 41.0 | 5.1% | | 3-epi-25-OH Vitamin D3 | 42.0 | 61.7% | | 3-epi-25-OH Vitamin D2 | 42.0 | 55.1% | *% Cross-reactivity = (Corrected Assay Value /Concentration Spiked)* 100 No significant cross-reactivity (4.1%) was found for Paricalcitol (Zemplar®) up to 25ng/mL. #### Comparison Studies #### Method Comparison Human serum samples were tested with the Diazyme 25-OH Vitamin D Assay and the obtained results were compared to the predicate method. A total of 98 unaltered serum samples were used in this experiment. Using this study, we found that the Diazyme 25-OH Vitamin D Assay correlated with the predicate method with the following results: | Deming Regression Analysis | 95% Confidence Interval | |----------------------------|-------------------------| | Slope | 1.005 (0.969 to 1.041) | | Intercept | -0.21 (-2.15 to 1.73) | | Correlation Coefficient | 0.984 (0.976 to 0.989) | | Range | 9.5-140.9 | {6}------------------------------------------------ # Matrix Comparison To evaluate the effect of anticoagulants, the Diazyme 25-OH Vitamin D Assay was used to measure the 25-0H Vitamin D concentrations of matched sets of serum, K3-EDTA plasma and Li-Heparin plasma. The reported values for each sample and for each matrix were obtained from single measurements. The total number of matched sets tested was 66. In order to cover the claimed measuring range for each matrix, seven spiked patient samples were included in the study. Linear regression of the "Li-Heparin plasma versus Serum" data yielded the following results: y = 0.9657 x - 0.6596 and R2 = 0.9736. Linear regression of the "K3-EDTA plasma versus Serum" data vielded the following results: y = 0.9948x - 0.7057 and R2 = 0.9866. # Reference Range Study To determine a reference range for the Diazyme 25-OH Vitamin D Assay, the 25-OH Vitamin D serum concentrations of a US population of 157 apparently healthy individuals were measured with the Diazyme method. The individual patient serum samples used in this study were obtained from certified commercial sources. Forty seven (47) samples from Pennsylvania (Northern U.S.) were collected from an FDA Licensed Donor Center with informed consent. Fifty six (56) samples from Tennessee (Central U.S.) and Fifty four (54) samples from Texas (Southern U.S.) were collected according to an IRB approved protocol. All participating individuals met the following inclusion conditions: - The age of all individuals was within the 21-80 years old range. . - Individuals were from three different geographical locations: 47 from Pennsylvania . (Northern US), 56 from Tennessee (Central US) and 54 from Texas (Southern US). - All samples were collected during the months of October and November (fall season). ● - The studied population consisted of 72 light skin individuals (46%) and 85 dark skin in-. dividuals (54%). - 155 individuals (98.7%) did not take any artificial Vitamin D supplements. Two individ-. uals (1.3%) did take some Vitamin D supplements but did not exceed the daily dose of 2000 IU. - All 157 individuals did not have any family history of parathyroid or calcium regulatory . disease. - . All 157 individuals did not have any history of kidney disease. GI disease, calcium-levels related disease, thyroid disease, parathyroid disease, calcium related disease, seizures, chronic disease or bariatric surgery. - All 157 individuals were not currently taking any medications that are known to affect . absorption or catabolism of Vitamin D (including cholesterol absorption inhibitors such as Vytorin®, Inegy™ or Zetia; anticonvulsants such as Neurontin, Depakine® and Trileptal: glucocorticoids such as Cortisol, Prednisone and Dexamethasone; HAART (AIDS treatment) or antirejection medications. {7}------------------------------------------------ Analysis of the reference range study data vielded the following results: - Lowest 25-OH Vitamin D concentration: 12.6 ng/mL. . - Highest 25-OH Vitamin D concentration: 51.4 ng/mL. . - . Median 25-OH Vitamin D concentration: 25.6 ng/mL - . Observed range (2.5th to 97.5th percentile): 15.0 to 45.9 ng/mL. ## Conclusion The Diazyme 25-OH Vitamin D assay has a linear range of 7.6 - 147.8 ng/mL for serum samples. For samples with Vitamin D levels ranging from 23.1 ng/mL to 134.6 ng/mL; within-run CVs were 1.8 to 8.8% and total CVs were 2.1 to 8.8%. For 98 tested serum samples, Deming regression yielded a correlation coefficient between the Diazyme 25-OH Vitamin D assay and the Diasorin Liaison® 25-OH Vitamin D Total assay was 0.984, the slope was 1.005, and the v intercept was -0.21. Interference studies demonstrated that this assay was not significantly affected by triglycerides (up to 750 mg/dL), ascorbic acid (up to 176mg/dL), free bilirubin (up to 40 mg/dL), conjugated bilirubin (up to 40 mg/dL) and hemoglobin (up to 100 mg/dL). The Diazyme 25-OH Vitamin D Assay and the Diazyme 25-OH Vitamin D Control data presented and provided is complete and supports the basis for substantial equivalence to the predicate device. ## References - 1. Wacker M, Holick MF. Sunlight and Vitamin D: A global perspective for health. Dermatoendocrinol. 2013, 5, 51-108. - 2. National Osteoporosis Foundation. Prevention Vitamin D. http://www.nof.org/aboutosteoporosis/prevention/vitamind {8}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 14, 2014 DIAZYME LABORATORIES ABHIJIT DATTA DIRECTOR, TECHNICAL OPERATIONS 12889 GREGG COURT POWAY CA 92064 Re: K133410 Trade/Device Name: Diazyme 25-OH Vitamin D Assay; Diazyme 25-OH Vitamin D Control Set Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: Il Product Code: MRG, JJX Dated: February 10, 2014 Received: February 12, 2014 Dear Dr. Datta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {9}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. # Courtney H. Lias -S Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {10}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. 510(k) Number (if known) k133410 Device Name Diazyme 25-0H Vitamin D Assay and Diazyme 25-0H Vitamin D Control set Indications for Use (Describe) The Diazyne 25-OH Vitamin D Assay is intended for use in clinical laboratories for the quantitative determination of 25hydroxywiamin D (25-OH-D) in human serum and plasma on automated chemistry analyzer. Measurement of 25-hydroxy wiamin D (25-OH-D) is for assessment of vitamin D sufficiency. For in vitro diagnostic use only. The 25-0H Vitamin D Control Ser is intended for use as quality controls for Diazyne 25-0H Viamin D Assay Kit only. For in vitro diagnostic use only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. M Z .. H M B B B .. W ( ) FOR F D A USE ONEY ( M 4 .. ) M .. 1 M .. M .. M .. M .. M .. M .. M .. M .. M .. M .. M .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure) Image /page/10/Picture/14 description: The image shows the text "Yung FV.Chan -S" in a simple, sans-serif font. The letters "FV.C" are stylized with a pattern of concentric lines, giving them a textured appearance. The rest of the text is plain and easily readable. The overall impression is a logo or branding element with a slightly retro or vintage feel due to the patterned letters.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...