K133390 · Institute For Quality Resource Management · LRK · Oct 16, 2014 · Dental
Device Facts
Record ID
K133390
Device Name
NDX DORSAL APPLIANCE
Applicant
Institute For Quality Resource Management
Product Code
LRK · Dental
Decision Date
Oct 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ClearDream® is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Device Story
ClearDream is a custom-fabricated, prescription-only, intraoral mandibular repositioning device; used by adults during sleep to reduce snoring and mild to moderate OSA. Device consists of two PMMA acrylic trays (maxillary and mandibular) created from dental impressions; bilateral stainless steel expansion screws allow dentist-controlled mandibular advancement (up to 6.5mm). Mechanism: protrusive repositioning of mandible to decrease air turbulence, increase pharyngeal space, and maintain airway patency. Dentist uses adjustment key to titrate advancement; patient wears device nightly. Benefits: improved airway patency, reduced apnea occurrences, and patient comfort via mouth-opening capability.
Clinical Evidence
No clinical data provided. Per FDA Class II Special Controls Guidance Document for Intraoral Devices for Snoring and/or Obstructive Sleep Apnea, clinical studies are not requested for devices with intended use, design, and technology identical to a predicate device.
Technological Characteristics
Materials: Polymethyl methacrylate acrylic (PMMA) and surgical grade stainless steel. Design: Bilateral, two-piece intraoral appliance with expansion screws for titration. Standards: ISO 7405, ISO 10993 (parts 5 and 10), and ASTM standards for materials. Connectivity: None (mechanical). Sterilization: Not specified (custom dental appliance).
Indications for Use
Indicated for adults with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2014
National Dentex Corporation C/O Mr. Bradley J. Chott Director of Operations 160 Larkin Williams Industrial Court Fenton, MO 63026
Re: K133390
Trade/Device Name: ClearDream® Regulation Number: 21 CFR 872.5570 Regulation Name: Device, anti-snoring Regulatory Class: II Product Code: LRK Dated: October 1, 2014 Received: October 6, 2014
#### Dear Mr. Chott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Chott:
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
### Indications for Use
510(k) Number: K133390
Device Name: ClearDream®
Indications for Use:
The ClearDream® is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Prescription Use (Part 21 CFR 801 Subpart D)
______________________________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "Keller" in a stylized, sans-serif font. The word is in blue and is set against a white background. The font is bold and the letters are closely spaced together. The logo is simple and modern.
## 510(k) Number: K133390
| Manufacturer: | National Dentex Corporation (NDX)<br>3910 RCA Blvd. Ste. 1015<br>Palm Beach Gardens, FL 33410<br>Telephone: 636-600-4244<br>Fax: 636-600-4396<br>E-mail:bchott@kellerlab.com |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Brad Chott<br>Keller Laboratories fully owned subsidiary of NDX<br>160 Larkin Williams Industrial Court<br>Fenton, MO 63026<br>Telephone: 636-600-4244<br>Fax: 636-600-4396<br>E-mail:bchott@kellerlab.com |
| Date Prepared: | October 9, 2014 |
| Trade Name: | ClearDream® |
| Common Name: | Intraoral device for snoring and mild to moderate obstructive<br>sleep apnea (OSA) |
| Classification Name: | Device, anti-snoring |
| Regulation Number: | 21 CFR 872.5570 |
| Product Code: | LRK |
| Class: | II |
| Panel: | Dental |
| Predicate Device: | K050592, Somnomed MAS RXA |
#### Description of the device:
The ClearDream® is an intraoral device and is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. The product is a prescription-only, customized device for use by a single patient; dentist titratable mandibular repositioning device worn during sleep to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. Dental impressions must be used to construct the device. A bite registration reflecting the relationship between the patient's mandible (lower jaw) and maxilla (upper jaw) is taken at the dental office. The product consists of 2 pieces. Both pieces are constructed using polymethyl methacrylate acrylic (PMMA). One piece fits over the maxillary (upper jaw) dental arch and is retained by the maxillary teeth. The maxillary piece has blocks of acrylic located bilaterally on the buccal surfaces. One expansion stainless steel screw is encased in each one of the acrylic blocks. The mandibular (lower jaw) piece is fitted over the mandibular arch and is retained by the lower dental arch. The mandibular piece features bilateral buccally located flanges that engage the upper mechanisms. This engagement repositions the mandible protrusively to decrease air turbulence, increase pharyngeal space, and maintain airway patency. It also permits the patient to open their mouth allowing patient comfort. The ClearDream® appliance keeps the PMMA covering the lingual surfaces of the lower teeth to a minimum. The duration of the interval of wearing the device is 8 hours or less. The PMMA material used is identical to the predicate device, and therefore does not present any possible biocompatibility differences that have been tested on the PMMA. The ClearDream® is accompanied by an adjustment
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "keller" in blue font. The font is a sans-serif font, and the letters are all lowercase. The word is set against a white background. The logo is simple and modern.
key used by the dentist. The dentist uses the adjustment key to move the bilateral, buccally-located expansion stainless steel screws allowing forward mandibular advancement a maximum from baseline of 6.5 millimeters, thereby improving patency of the airway, decreasing air turbulence, and reducing obstructive apnea occurrences.
### Indications for use:
The ClearDream® is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
# Technological Characteristics:
The ClearDream® oral appliance has identical technical characteristics as the predicate device. Table 1 contains a basic description of the intended use, design attributes and technological characteristics demonstrating that the proposed and predicate device are identical in terms of how they achieve their intended use.
### Table 1 Comparison summary to legally marketed predicate device Somnomed MAS RXA:
| Intended Use | ClearDream® | Somnomed MAS RXA |
|----------------------------------------------------------|-------------|------------------|
| Intended as an intraoral device | Yes | Yes |
| Intended to reduce snoring | Yes | Yes |
| Treatment of mild to moderate<br>obstructive sleep apnea | Yes | Yes |
| Intended for nighttime use | Yes | Yes |
| Indicated for single patient multiuse | Yes | Yes |
| Indicated for use at home or sleep<br>laboratories | Yes | Yes |
| Indicated for adult population | Yes | Yes |
| Design Attributes | ClearDream® | Somnomed MAS RXA |
| Customized fit for each specific patient | Yes | Yes |
| Separate upper and lower tray pieces | Yes | Yes |
| Bilateral design to engages the upper<br>mechanisms | Yes | Yes |
| Works by mandibular advancement | Yes | Yes |
| Can be adjusted or refit | Yes | Yes |
| Lower jaw adjustment using supplied<br>components | Yes | Yes |
| Permits patient to breathe through<br>mouth | Yes | Yes |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "keller" in blue font. The font is a sans-serif font, and the letters are all lowercase. The word is set against a white background. The logo is simple and modern.
| Upper and lower trays disengage for<br>easy removal | Yes | Yes |
|------------------------------------------------------------------------|-------------|------------------|
| Cleaned and inspected daily by patient | Yes | Yes |
| Stainless steel expansion screws to<br>facilitate titration by dentist | Yes | Yes |
| Titration is done by the dentist | Yes | Yes |
| Material | ClearDream® | Somnomed MAS RXA |
| Trays constructed with polymethyl<br>methacrylate acrylic (PMMA) | Yes | Yes |
| Bilaterally adjustable stainless steel<br>hardware | Yes | Yes |
| Advancement mechanism constructed of<br>surgical grade stainless steel | Yes | Yes |
| Mandibular piece features bilateral<br>buccally located flanges | Yes | Yes |
#### Non Clinical Performance Data:
The ClearDream® conforms to the requirements of the FDA Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA published on November 12, 2002. The material composition of the ClearDream polymethyl methacrylate acrylic (PMMA) medical grade has been tested and found to comply with the Standard ISO 7405 and ISO 10993 part 5 and part 10 referenced per 21 CFR 872.5570 Intraoral devices for snoring and/or obstructive sleep apnea, product code LRK. The material characteristic of the ClearDream ® appliance, meeting ASTM standards, have been found to be substantially equivalent to the predicate device material characteristics. Included in the 510K under Performance Bench Testing is evidence of the ClearDream ® component material identity, chemical composition, material safety support the requirements noted in the Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea: Guidance for Industry and FDA issued November 12. 2002. The material composition, performance characteristics, technical characteristics, risks analysis, and biocompatibility of the ClearDream ® appliance and components have been addressed and found to comply with the guidance set forth in FDA Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA published on November 12, 2002 enabling the ClearDream® to meet the indications for use. A risk analysis was conducted and confirms the risks published in the FDA guidance document FDA Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA published on November 12, 2002 have been addressed and no additional risks were identified with the submission of the ClearDream®. The risk analysis did not identify any other specific risks to the ClearDream®.
#### Biocompatibility:
The materials and fabrication of the proposed and predicate device are identical. The material composition of the ClearDream polymethyl methacrylate acrylic (PMMA) medical grade has been
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Keller. The logo is in blue and features a stylized font. The logo is set against a white background and is surrounded by a thin blue border.
tested and found to comply with the Standard ISO 10993 part 5 and part 10 referenced per 21 CFR 872.5570 Intraoral devices for snoring and/or obstructive sleep apnea, product code LRK.
# Clinical Data:
None provided in accordance with the least burdensome provisions of the FDA Modernization Act of 1997. Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA issued November 12, 2002 item 8, Clinical testing states, the agency does not request clinical studies for new devices whose intended use is identical to that of the claimed predicate device, having similar design and technology.
# Substantial equivalence discussion:
National Dentex submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the ClearDream® is substantially equivalent in indications and design principles to the following predicate device Somnomed MAS RXA K050592
For the purposes of FDA's regulation of medical devices, the ClearDream® is substantially equivalent to the Somnomed MAS RXA device. Both the ClearDream® and the Somnomed MAS RXA are prescription-only, customized for one patient only for multiple use, dentist titratable mandibular repositioning devices intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. The material composition, technical and functional designs as well as indications for use of each product are identical. They both require dental impressions to allow patient specific customized and fabricated acrylic trays which fit onto the upper and lower teeth and are positioned in relation to each other by adjustment mechanisms. The subject device is manufactured using polymethyl methacrylate acrylic (PMMA) and stainless steel expansion screws that conform to the FDA Class I General Controls Guidance Document. The amount of mandibular advancement for ClearDream® is one millimeter more than the predicate, however the difference is well within the range of mandibular advancement in other previously cleared FDA intraoral devices for treatment of mild to moderate sleep apnea, and titration of mandibular advancement is the responsibility of the prescribing dentist. Therefore, the slight difference in mandibular advancement of the ClearDream® device as compared to the predicate does not raise any additional concerns. There are no new safety concerns raised by the design of the ClearDream® when compared to the predicate device.
The material composition, design specifications, technical features, nonclinical testing, and performance characterizes of the ClearDream® have been demonstrated to be safe and as effective and to perform as well as or better than the predicate device.
#### Conclusions:
The ClearDream® has the same intended use as the predicate device, same material composition, design features and technological characteristics that do not raise different types of questions of safety and effectiveness when compared to the predicate device. We believe that the analysis provided herein supports the claim that the ClearDream® OSA appliance intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.