← Product Code [IYN](/productcode/IYN) · K133329

# UGEO WS80A DIAGNOSTIC ULTRASOUND SYSTEM (K133329)

_Samsung Medison Co., Ltd. · IYN · Nov 27, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133329

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Nov 27, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The UGEO WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal, Abdominal, Small Organs, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial) and Peripheral vessel.

## Device Story

UGEO WS80A is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via transducers and displays it in B-mode, M-mode, Color/Power Doppler (S-Flow), PW Spectral Doppler, Harmonic imaging, and 3D/4D imaging modes. Operated by clinicians in clinical settings, the system allows measurement of anatomical structures and provides analysis packages to assist in diagnosis. It features real-time acoustic output display (mechanical and thermal indices). The system processes received echoes to generate on-screen anatomical and fluid flow visualizations. Output is used by healthcare providers to inform clinical decision-making and diagnostic assessments. Benefits include non-invasive visualization of internal structures and fluid dynamics.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility, thermal, electrical, electromagnetic, and mechanical safety testing in accordance with IEC 60601-1, IEC 60601-2-37, and ISO 10993-1 standards.

## Technological Characteristics

Mobile ultrasound system; supports linear, curved, endocavity, and phased array transducers (1.0-20.0 MHz). Modes: B, M, PW Doppler, Color/Power Doppler, Tissue Harmonic Imaging, 3D/4D. Connectivity: Standard ultrasound I/O. Biocompatibility: ISO 10993-1. Safety: IEC 60601-1, IEC 60601-2-37, NEMA UD-2/UD-3. Software-controlled processing.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- ACCUVIX A30 Diagnostic Ultrasound System ([K112339](/device/K112339.md))
- UGEO HM70A Diagnostic Ultrasound System ([K130803](/device/K130803.md))
- ACCUVIX XG Diagnostic Ultrasound System ([K103397](/device/K103397.md))
- ACUSON S2000 Diagnostic Ultrasound System ([K130739](/device/K130739.md))
- iU22 Diagnostic Ultrasound System ([K093563](/device/K093563.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K133329

510(k) Premarket Notification

UGEO WS80ADiagnostic Ultrasound System

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E. Section 807.92.

## 1. Submitter's Information: 21 CFR 807.92(a)(1)

SAMSUNGMEDISON CO., LTD. 42, Teheran-ro 108-gil, Gangnam-gu, Seoul, Korea

Contact Person: Kyeong-Mi, Park Regulatory Affairs Manager

Telephone: 82.2.2194.1373 Facsimile: 82.2.556.3974

Data Prepared: September 2, 2013

#### 2. Name of the device:

Common/Usual Name: Diagnostic Ultrasound System and Accessories Proprietary Name: UGEO WS80A Diagnostic Ultrasound System Classification Names: Product Code FR Number Ultrasonic Pulsed Doppler Imaging System 892.1550 IYN Ultrasound Pulsed Echo Imaging System 892.1560 IYO Diagnostic Ultrasound Transducer 892.1570 ITX

#### 3. Identification of the predicate or legally marketed device:

- ACCUVIX A30 Diagnostic Ultrasound System(K112339) .
- UGEO HM70A Diagnostic Ultrasound System(K130803) -
- -ACCUVIX XG Diagnostic Ultrasound System (K103397)
- ACUSON S2000 Diagnostic Ultrasound System (K130739) -
- iU22 Diagnostic Ultrasound System (K093563) -
- * The proprietary name of predicate device (K130803) was changed to UGEO HM70A Diagnostic Ultrasound System from UGEO H70c Diagnostic Ultrasound System on FDA Databases .

510(k) Summary / Statement Certification

ATTACHEMENT 1

{1}------------------------------------------------

#### ··············································································································································································

## 4. Device Description:

The UGEO WS80A is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode. M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode: S-Flow), PW Spectral Dopler mode, Harmonic imaging, Tissue Doppler imaging, 3D imaging mode (real time 4D imaging mode) or as a combination of these modes. The UGEO WS80A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The UGEO WS80A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

#### 5. Intended Uses:

The UGEO WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal, Abdominal, Small Organs, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial) and Peripheral vessel.

## 6. Technological Characteristics:

The UGEO WS80Ais substantially equivalent with respect to safety, effectiveness, and functionality to the ACCUVIX A30 Diagnostic Ultrasound System (K112339) and UGEO HM70A Diagnostic Ultrasound System (K130803).

It is substantially equivalent with respect to safety, effectiveness, and functionality to the [3D Cine. HDVI and Volume NT | of SAMSUNG MEDISON's UGEO HM70A (K 130803) in regards to the device with [5D Cine'), FRV2), 5D NT31 and 2D NT+1].

It is substantially equivalent with respect to safety, effectiveness, and functionality to the [Syngo Auto OB Measurements] of Siemens's ACUSON S2000 (K130739) in regards to the device with [5D LB (Long Bone) 3).

It is substantially equivalent with respect to safety, effectiveness, and functionality to the High O Automatic Doppler Analysis] of Philips's iU22 (K093563) in regards to the device with [MP16].

All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

These are described in detail in the technological characteristics comparison table as below.

510(k) Summary / Statement Certification

ATTACHEMENT 1

{2}------------------------------------------------

# . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. ・・・

|                                                     | The subject<br>device | The predicate devices                                                                                                                                                          |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
|-----------------------------------------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Feature / Characteristics<br>、ジ<br>11               | UGEO WS80A            | ACCUVIX<br>A30<br>(K112339)                                                                                                                                                    | UGEO<br>HM70A<br>(K130803)                                                                                                                                                     | ACCUVIX<br>XG<br>(K103397)  . | ACUSON<br>S2000 -<br>(K130739) | iU22<br>(K093563)                                                                                                                                                              |  |  |  |
| Indication for Use                                  |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - Fetal                                             | ﺮ                     | 4                                                                                                                                                                              | -                                                                                                                                                                              | ﮯ                             | 2                              | ﮨﯿﮯ                                                                                                                                                                            |  |  |  |
| - Abdominal                                         | イ                     | ど                                                                                                                                                                              | V                                                                                                                                                                              | 1                             | 4                              | ﺮ ﺳ                                                                                                                                                                            |  |  |  |
| - Pediatric                                         |                       | V                                                                                                                                                                              | ﺳ                                                                                                                                                                              | 4                             | イ                              | 2                                                                                                                                                                              |  |  |  |
| - Small Organ                                       | ﯿﮯ                    | ﮯ                                                                                                                                                                              | ಳ                                                                                                                                                                              | ਪ੍ਰ                           | V                              | ﮨﮯ۔                                                                                                                                                                            |  |  |  |
| - Neonatal Cephalic                                 |                       | ど                                                                                                                                                                              | ﮨﮯ                                                                                                                                                                             | ど                             | ﮨﮯ                             | ﯿﮯ                                                                                                                                                                             |  |  |  |
| - Adult Cephalic                                    |                       | 4                                                                                                                                                                              | 1                                                                                                                                                                              | く                             | ←                              | 2                                                                                                                                                                              |  |  |  |
| - Trans-rectal                                      | ﻳﻢ                    | ん                                                                                                                                                                              | く                                                                                                                                                                              | ﮨﮯ                            | V                              | 1                                                                                                                                                                              |  |  |  |
| - Trans-vaginal                                     | ﻴﻢ                    | 2                                                                                                                                                                              | ﺮ ﻣﺤﻤﺪ ﺍﻟﻤﺴﺘﻌﻤﺎﺭ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟ  | ಲ್ಲ                           | V                              | ﻳﺴ                                                                                                                                                                             |  |  |  |
| - Musculo-skeletal                                  |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| (Conventional)                                      | ly                    | イ                                                                                                                                                                              | ﺮ ﺍﻟﻤﺴﺘﻤﺮ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﻌﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﻤﻞ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘﺸﺮﻳﻦ ﺍﻟﻤﺴﺘ  | V                             | V                              | ゾ                                                                                                                                                                              |  |  |  |
|                                                     |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - Musculo-skeletal<br>(Superficial)                 | r                     | く                                                                                                                                                                              | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | V                             | v                              | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |  |  |  |
| - Cardiac Adult                                     |                       | ﮯ                                                                                                                                                                              | -                                                                                                                                                                              | ど                             | V                              | ﯿﺴ                                                                                                                                                                             |  |  |  |
|                                                     |                       | イ                                                                                                                                                                              | ﺳ                                                                                                                                                                              | 2                             | イ                              | V                                                                                                                                                                              |  |  |  |
| - Cardiac Pediatric                                 |                       | 4                                                                                                                                                                              | V                                                                                                                                                                              | く                             | V                              | V                                                                                                                                                                              |  |  |  |
| - Peripheral vessel                                 | ريم                   |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| Scanhead Types                                      | ﯾﮯ                    | 2                                                                                                                                                                              | ﮨﮯ                                                                                                                                                                             | ﮨﮯ                            | 4                              | ﮨﮯ                                                                                                                                                                             |  |  |  |
| - Linear Array                                      | イ                     |                                                                                                                                                                                | -                                                                                                                                                                              | ﮯ                             | 4                              | ﻢ                                                                                                                                                                              |  |  |  |
| - Curved Linear Array                               |                       | ﮨﮯ                                                                                                                                                                             | 1                                                                                                                                                                              | ﮯ                             | く                              |                                                                                                                                                                                |  |  |  |
| - Endocavity                                        | 4                     | 4                                                                                                                                                                              |                                                                                                                                                                                |                               |                                | 1                                                                                                                                                                              |  |  |  |
| - Phased Array                                      |                       | ਪ                                                                                                                                                                              | ﻳﻢ                                                                                                                                                                             | ﮯ                             | く                              | V                                                                                                                                                                              |  |  |  |
| - Static Probes                                     |                       | ﮯ                                                                                                                                                                              | ﻣﯿﺮ                                                                                                                                                                            | ど                             | 1                              | ﯾﮯ                                                                                                                                                                             |  |  |  |
| Scanhead Frequency                                  |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| 1.0~20.0 MHz                                        | V                     | V                                                                                                                                                                              | V                                                                                                                                                                              | く                             | 4                              | ﮯ                                                                                                                                                                              |  |  |  |
| Modes of Operation                                  |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - B-mode                                            | ど                     | ﮯ                                                                                                                                                                              | ﮨﮯ                                                                                                                                                                             | イ                             | V                              | ﺮﻳﺴﻤﺎ                                                                                                                                                                          |  |  |  |
| - M-mode                                            | V                     | 2                                                                                                                                                                              | ど                                                                                                                                                                              | 2                             | く                              | バ                                                                                                                                                                              |  |  |  |
| - Pulsed wave (PW)<br>Doppler                       | y                     | ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー  | ﺮ ﺍ                                                                                                                                                                            | -                             | V                              | く                                                                                                                                                                              |  |  |  |
| - Continuous wave (CW)<br>Doppler                   |                       | -                                                                                                                                                                              | ﺃ                                                                                                                                                                              | ત્                            | V                              | 1                                                                                                                                                                              |  |  |  |
| - Color Doppler                                     | 2                     | 4                                                                                                                                                                              | V                                                                                                                                                                              | イ                             | V                              | V                                                                                                                                                                              |  |  |  |
| - Power Amplitude<br>Doppler                        | 1                     | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ﺮﻳﺲ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ   | V                             | ﮯ                              | V                                                                                                                                                                              |  |  |  |
| - Tissue Harmonic                                   | V                     | V                                                                                                                                                                              | く                                                                                                                                                                              | V                             | 4                              | ﺮﻳﻢ ﺍﻟﻤﺴﻠﺴﻞ ﺍﻟﻤﺴﺘﻘﺒﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ   |  |  |  |
| Imaging                                             | ಳ                     |                                                                                                                                                                                | ﭘﯿ                                                                                                                                                                             |                               | V                              |                                                                                                                                                                                |  |  |  |
| - 3D/4D imaging mode                                | l                     | 2                                                                                                                                                                              | V                                                                                                                                                                              | ﮯ                             |                                | ﯾﺴ<br>2                                                                                                                                                                        |  |  |  |
| - Combined modes                                    |                       | V                                                                                                                                                                              |                                                                                                                                                                                | V                             | ど                              |                                                                                                                                                                                |  |  |  |
| Safety & EMC Compliance                             |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - IEC 60601-1                                       | ﮨﮯ۔                   | v                                                                                                                                                                              | v                                                                                                                                                                              | v                             | -                              | -                                                                                                                                                                              |  |  |  |
| - UL 60601-1                                        |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - CSA C22.2 No.601.1                                |                       |                                                                                                                                                                                |                                                                                                                                                                                |                               |                                |                                                                                                                                                                                |  |  |  |
| - IEC 60601-2-37                                    | 1                     | 2                                                                                                                                                                              | V                                                                                                                                                                              | V                             | V                              | 4                                                                                                                                                                              |  |  |  |
| - IEC 60601-1-2<br>Acoustic Output Display Standard | V                     | ﮯ                                                                                                                                                                              | ﻴﺴ                                                                                                                                                                             | ど                             | ﮨﮯ۔                            | ﻤﯿﮟ                                                                                                                                                                            |  |  |  |

## <Technological Characteristics Comparison Table>

510(k) Summary / Statement Certification

## ATTACHEMENT 1

{3}------------------------------------------------

# ..............................................................................................................................................................................

·

・

## UGEO WS80ADiagnostic Ultrasound System

. , .

| Feature / Characteristics                  | The subject device<br>UGEO WS80A | ACCUVIX A30<br>(K112339) | UGEO HM70A<br>(K130803) | ACCUVIX XG<br>(K103397) | ACUSON S2000<br>(K130739) | iU22<br>(K093563) |
|--------------------------------------------|----------------------------------|--------------------------|-------------------------|-------------------------|---------------------------|-------------------|
| Track 3                                    | √                                | √                        | √                       | √                       | √                         | √                 |
| Patient Contact Materials                  |                                  |                          |                         |                         |                           |                   |
| Tested to ISO 10993-1                      | √                                | √                        | √                       | √                       | √                         |                   |
| Functionality                              |                                  |                          |                         |                         |                           |                   |
| - Quick Scan (Q Scan)                      | √                                | √                        | √                       | √                       |                           |                   |
| - Spatial Compound Imaging                 |                                  | √                        |                         | √                       |                           |                   |
| - SMDR (SMDR evo)                          | √                                | √                        | √                       | √                       |                           |                   |
| - Auto IMT+<br>(Auto IMT)                  | √                                | √                        | √                       | √                       |                           |                   |
| - Elastoscan                               | √                                | √                        | √                       | √                       |                           |                   |
| - Panoramic                                | √                                | √                        | √                       | √                       |                           |                   |
| - 3D Imaging<br>(Volume Data Acquisition)  | √                                | √                        | √                       | √                       |                           |                   |
| - 3D Imaging presentation                  |                                  |                          |                         |                         |                           |                   |
| 3D Cine/4D Cine                            |                                  | √                        | √                       | √                       |                           |                   |
| 5D Cine                                    | √(1)                             |                          |                         |                         |                           |                   |
| - 3D Rendering<br>MPR(Multi Planer Render) | √                                | √                        | √                       | √                       |                           |                   |
| - 3D XI<br>MSV(Multi Slice View)           | √                                | √                        |                         | √                       |                           |                   |
| Oblique View                               |                                  |                          |                         |                         |                           |                   |
| - 3D MXI<br>Volume Slice, Mirror View      | √                                | √                        | √                       | √                       |                           |                   |
| - 3D MagiCut                               | √                                | √                        | √                       | √                       |                           |                   |
| - Volume Calculation<br>(VOCAL, XI VOCAL)  | √                                | √                        | √                       | √                       |                           |                   |
| - XI STIC                                  | √                                | √                        | √                       | √                       |                           |                   |
| - HDVI                                     | √                                | √                        | √                       | √                       |                           |                   |
| - FRV                                      | √(2)                             |                          |                         |                         |                           |                   |
| - Volume NT/IT                             | √                                | √                        | √                       | √                       |                           |                   |
| - 5D NT                                    | √(3)                             |                          |                         |                         |                           |                   |
| - 2D NT                                    | √(4)                             |                          |                         |                         |                           |                   |
| - ADVR                                     | √                                | √                        | √                       | √                       |                           |                   |
| - 5D LB                                    | √(5)                             |                          |                         |                         | √(5)                      |                   |
| MPI                                        | √(6)                             |                          |                         |                         |                           | √(6)              |

ATTACHEMENT 1 ·

{4}------------------------------------------------

510(k) Premarket Notification

### 7. A brief discussion of the bench and non-clinical tests conducted on the subject device

The device has been evaluated for acoustic output, biocompatibility effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform to applicable medical device safety standards.

The UGEO WS80A and its application comply with voluntary standards as below:

- UL 60601-1, Safety requirements for Medical Equipment
- CSA C22.2 No. 601.1. Safety requirements for Medical Equipment
- IEC60601-2-37. Diagnostic Ultrasound Safety Standards
- EN/IEC60601-1,Safety requirements for Medical Equipment
- EN/IEC60601-1-2.EMC requirements for Medical Equipment
- NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- ISO10993-1. Biocompatibility
- ISO14971, Application of risk management to medical devices

#### Summary of Clinical Tests:

Not applicable. The subject of this submission, UGEO WS80A, did not require clinical studies to support substantial equivalence.

## 8. Conclusion

Intended uses and other key features are consistent with traditional practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the UGEO WS80Ato be as safe, as effective, and performance is substantially equivalent to the predicate devices.

### END of 510(K) Summary

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with outstretched arms, possibly representing care or protection.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002

November 27, 2013

Samsung Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K133329

Trade/Device Name: UGEO WS80A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 28, 2013 Received: October 29, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the UGEO WS80A Diagnostic Ultrasound System, as described in your premarket notification:

| Transducer Model Number |        |        |
|-------------------------|--------|--------|
| SC1-6                   | C2-6   | E3-12A |
| VR5-9                   | L3-12A | L5-13  |
| V4-8                    | V5-9   |        |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

ﯽ

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

for

Janine M. Morris Director. Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

SIO(k) Premarket Notification

UGEO WS80A Diagnostic Ultrasound System

## SECTION 1.3 INDICATIONS FOR USE

510(k) Number (if known): _K133329

Device Name: UGEO WS80A Diagnostic Ultrasound System

Indications for Use:

The UGEO WS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal, Abdominal, Small Organ, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial) and Peripheral vessel

Prescription Use (Part 21 CFR 80) Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sinh

(Division Sign Off) Division of Radiological Health Office of In I itro Diagnostic and Radiological Health

> K133329 510(k)_

Indication for use

page 1 of10

{8}------------------------------------------------

510(k) No .:

| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |                               |   |                                                   |     |     |                   |                      |                            |  |
|--------------------------------------------------------------------------------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|----------------------------|--|
|                                                                                                  | Clinical Application          |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                            |  |
| General<br>(Track I only)                                                                        | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)           |  |
| Ophthalmic                                                                                       | Ophthalmic                    |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Fetal/Obstetrics (See Note 3) | N | N                                                 | N   |     | N                 | Note 1               | Notes 2, 4, 7, 8, 11       |  |
|                                                                                                  | Abdominal (See Note 10)       | N | N                                                 | N   |     | N                 | Note 1               | Notes 2, 7, 8, 9, 11, 12   |  |
|                                                                                                  | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                            |  |
| Fetal Imaging<br>& Other                                                                         | Laparoscopic                  |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Pediatric                     |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Small Organ (See Note 5)      | N | N                                                 | N   |     | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11, 12 |  |
|                                                                                                  | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Adult Cephalic                |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Trans-rectal                  | N | N                                                 | N   |     | N                 | Note 1               | Notes 2, 7, 8, 11, 12      |  |
|                                                                                                  | Trans-vaginal                 | N | N                                                 | N   |     | N                 | Note 1               | Notes 2, 7, 8, 11, 12      |  |
|                                                                                                  | Trans-urethral                |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Musculo-skel. (Convent.)      | N | N                                                 | N   |     | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                                                                                                  | Musculo-skel. (Superfic.)     | N | N                                                 | N   |     | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                                                                                                  | Intra-luminal                 |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Other (spec.)                 |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                            |  |
| Cardiac                                                                                          | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                            |  |
|                                                                                                  | Other (spec.)                 |   |                                                   |     |     |                   |                      |                            |  |
| Peripheral<br>Vessel                                                                             | Peripheral vessel             | N | N                                                 | N   |     | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                                                                                                  | Other (spec.)                 |   |                                                   |     |     |                   |                      |                            |  |

N= new indication: P= previously cleared by FDA K122583: F= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note I: B+M, B+PW, B+C, B+PD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PPI+PW, B+TD+PW, B+TD+PW, B+C+M, Dral/Quad, B+E Note 2: Includes imaging for guidance of biopsy

Note 3. Includes infertility monitoring of follicle development

Device Name:UGEO WS80ADiagnostic Ultrasound System

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

**Note 7: Tissue Harmonic Imaging (THI)**

**Note 8: 3D imaging**

Note 8: JD immging
Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note II: Panoramic imaging

Note 12: ElastoScan

Concurrence of CDRH. Office of in I'uro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801 109)

Indication for use

page 2 of 10 .

{9}------------------------------------------------

510(k) No.:

中心

Device Name: SC1-6for use with UGEO WS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                     |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|---------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)    |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                     |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes2, 4, 7, 8, 11 |
|                           | Abdominal/(See Note 10)       | P                                                 | P | P   |     | P                 | Note I               | Notes2, 7, 8.9, 11  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                     |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                     |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                     |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                     |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                     |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                     |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                     |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                     |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                     |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                     |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                     |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                     |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                     |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                     |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                     |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                     |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                     |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                     |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                     |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                     |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                     |

N= new indication; P= previously cleared by FDA K 130803; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

induall Comments:
 Color Dopler includes Pover, (Ampinde) Doppler
 Note I: B+M, B+PW, B+DD, B+PD, B+PPI, B+PD+PW, B+PD+PW, B+PP++PW, B+PP++PW, B+TD+PW, B+C+M, Dwl/Qwd(B, 
 B+

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5. For example: thereid, parathyroid, breast, scrotum and penis in adult, pediating and neonatal patients

Note 6 Abdominal organs and peripheral vessel

Note 7 Tissue Harmonic Imaging (THI)

Note 8. 3D imaging

Note 9. Spaiial Compound Imaging

Note 10: Includes Renal. Gynecology/Pelvis

Note II: Panoramic imaging

Note 12: ElastoScan

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR B01.109)

Indication for use

page 3 0910

{10}------------------------------------------------

510(k) No.:

Device Name: C2-6 for use with UGEO WS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                    |                      |                      |
|----------------------|-------------------------------|---------------------------------------------------|---|-----|-----|--------------------|----------------------|----------------------|
| General              | Specific                      | B                                                 | M | PWD | CWD | Color<br>*Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)     |
| (Track I only)       | (Tracks I & 111)              |                                                   |   |     |     |                    |                      |                      |
| Ophthalmic           | Ophthalmic                    |                                                   |   |     |     |                    |                      |                      |
|                      | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                  | Note I               | Notes 2, 4, 7, 8, 11 |
|                      | Abdominal(See Note 10)        | P                                                 | P | P   |     | P                  | Note !               | Notes 2, 7, 8, 11    |
|                      | Intra-operative (See Note 6)  |                                                   |   |     |     |                    |                      |                      |
|                      | Intra-operative (Neuro.)      |                                                   |   |     |     |                    |                      |                      |
| Fetal Imaging        | Laparoscopic                  |                                                   |   |     |     |                    |                      |                      |
| &<br>Other           | Pediatric                     |                                                   |   |     |     |                    |                      |                      |
|                      | Small Organ (See Note 5)      |                                                   |   |     |     |                    |                      |                      |
|                      | Neonatal Cephalic             |                                                   |   |     |     |                    |                      |                      |
|                      | Adult Cephalic                |                                                   |   |     |     |                    |                      |                      |
|                      | Trans-rectal                  |                                                   |   |     |     |                    |                      |                      |
|                      | Trans-vaginal                 |                                                   |   |     |     |                    |                      |                      |
|                      | Trans-urethral                |                                                   |   |     |     |                    |                      |                      |
|                      | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                    |                      |                      |
|                      | Musculo-skel. (Convent.)      |                                                   |   |     |     |                    |                      |                      |
|                      | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                    |                      |                      |
|                      | Intra-luminal                 |                                                   |   |     |     |                    |                      |                      |
|                      | Other (spec.)                 |                                                   |   |     |     |                    |                      |                      |
|                      | Cardiac Adult                 |                                                   |   |     |     |                    |                      |                      |
| Cardiac              | Cardiac Pediatric             |                                                   |   |     |     |                    |                      |                      |
|                      | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                    |                      |                      |
|                      | Other (spec.)                 |                                                   |   |     |     |                    |                      |                      |
| Peripheral           | Peripheral vessel             |                                                   |   |     |     |                    |                      |                      |
| Vessel               | Other (spec.)                 |                                                   |   |     |     |                    |                      |                      |

N= new indication: P= previously cleared byFDA K130803; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note L: B+M, B+PW, B+C, B+PD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PP+PW, B+PP+PW, B+TD+PW, B+C+M, Dxd/Qual(B. B+C. B+PD, B+TD, B+PPI, B+PD), B+E

Note 2: Includes imaginy for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note S: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal palients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal. Gyzecology/Peivis

Note 11: Panoramic imaging

Note 12: ElastoScan

## Concurrence of CDRH, Office of In I'irro Diagnostics and Radiological Health (OIR) Prescription Use (Per 2) CFR 801.109)

Indication for use

page 4 0/10

{11}------------------------------------------------

510(k) No.:

## Device Name: E3-12Afor use with UGEO WS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation ("includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) | N                                                 | N | N   |     | N                 | Note 1               | Notes 7, 8       |
|                           | Abdominal(See Note 10)        | N                                                 | N | N   |     | N                 | Note 1               | Notes 7, 8, 12   |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                  | N                                                 | N | N   |     | N                 | Note 1               | Notes 7, 8, 12   |
|                           | Trans-vaginal                 | N                                                 | N | N   |     | N                 | Note 1               | Notes 7, 8, 12   |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication: P= previously cleared by FDA: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

ലംബ മാറ്റിം, B+C, B+PD, B+PD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PP+PW, B+PP+PW, B+C+M, Dml/Qmd(B, B+C, B+PD, B+TD, B+PPI, B+PD), B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrown and penis in adult, pediatric and neonated patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9, Spatial Compound Imaging

Note 10 Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

#### Concurrence of CDRH, Office of in Virro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)

Indication for use

page 5 0110

{12}------------------------------------------------

....

## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: VR5-9for use with UGEO WS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note I               | Note 2, 7, 8     |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note I               | Note 2, 7, 8, 12 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|…

---

**Source:** [https://fda.innolitics.com/device/K133329](https://fda.innolitics.com/device/K133329)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
