SPIROL BLOCK (19 GA. OPEN END MODELS), (20 GA. OPEN END MODELS), (19 GA. CLOSED END MODELS), (20 GA. CLOSED END MODELS)
K133316 · Epimed International, Inc. · BSO · May 28, 2014 · Anesthesiology
Device Facts
Record ID
K133316
Device Name
SPIROL BLOCK (19 GA. OPEN END MODELS), (20 GA. OPEN END MODELS), (19 GA. CLOSED END MODELS), (20 GA. CLOSED END MODELS)
Applicant
Epimed International, Inc.
Product Code
BSO · Anesthesiology
Decision Date
May 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The catheter is designed to deliver anesthesia to manage perioperative pain and/or alleviate postoperative analgesia.
Device Story
Spirol® Block is a single-use, disposable peripheral nerve block catheter; 19 Ga. or 20 Ga. sizes; 14" to 24" length. Device consists of a polymer-coated stainless steel spring; available in open/closed-ended and styletted/non-styletted configurations. Internal radiopaque, echogenic coiled wire allows visualization via fluoroscopy or ultrasound. Physician inserts catheter via introducer needle to target peripheral nerve group; needle is withdrawn. Catheter connects to standard Luer-lock accessories (bacterial filter, extension set, pump/syringe) via included connector. Enables continuous or bolus delivery of anesthetic agents to manage perioperative/postoperative pain. Used in clinical settings by trained physicians. Benefits include kink resistance, structural integrity, and sustained analgesic delivery for up to 72 hours.
Clinical Evidence
Bench testing only. Performance testing included dimensional analysis, substance compatibility, kink testing, stiffness, tensile strength, connector attachment, flow rate, leakage resistance (aspiration/vacuum/pressure), corrosion resistance, and skive testing. Biocompatibility testing performed per ISO 10993-3, -5, -6, -10, and -11.
Technological Characteristics
Polymer-coated stainless steel spring catheter; 19 Ga. or 20 Ga. diameters; 14-24 inch lengths. Features internal radiopaque, echogenic coiled wire. Single-use, disposable. Connects via Luer-lock. Biocompatibility per ISO 10993. No electronic components or software.
Indications for Use
Indicated for placement of catheters next to nerve plexuses for continuous nerve block anesthesia or analgesia techniques in upper extremity, lower extremity, abdominal, and paravertebral locations for periods not exceeding 72 hours.
Regulatory Classification
Identification
An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for "Epimed - Designed by Experts for Experts". The logo is in black and white. The word "Epimed" is in a bold, sans-serif font, with a geometric shape to the left of the word. Below the word "Epimed" is the tagline "Designed by Experts for Experts" in a smaller font.
# 510(k) Summar
### Submitter Information:
| Name: | Epimed International, Inc. |
|-------------------|----------------------------------------------------------------------------|
| Address: | 141 Sal Landrio Drive<br>Crossroads Business Park<br>Johnstown, N.Y. 12095 |
| Telephone Number: | (518) 725-0209 |
| Fax Number: | (518) 725-0207 |
| Contact Person: | Preston H. Frasier<br>Manager - QA/RA |
| Telephone Number: | (518) 725-0209 Ext. 1300 |
| Email Address: | prestonf@epimedint.com |
| Date Prepared: | October 23rd, 2013 |
### Device Name & Classification:
| Device Trade Name: | Spirol® Block Continuous Peripheral Nerve Block Catheter |
|----------------------|----------------------------------------------------------|
| Common Name: | Peripheral Nerve Block Catheter |
| Classification Name: | Anesthesia Conduction Catheter, BSO (21 CFR 868.5120) |
### Predicate Device(s) Information:
| Predicate Device #1: | Epimed International's "Feth-R-Kath" (K981329)<br>• Catheter later renamed “Spirol®” by Epimed. |
|----------------------|-------------------------------------------------------------------------------------------------|
| Predicate Device #2: | Arrow International's “ARROW FlexBlock Continuous<br>Peripheral Nerve Block Catheter” (K122027) |
### Device Description:
Epimed's Spirol® Block Continuous Peripheral Nerve Block Catheter has the following characteristics:
- · 19 Ga. & 20 Ga., 14" to 24" in length.
- · Offered in open and closed end distal tip configurations.
- · Offered in styletted and non-styletted versions.
- · Internal radiopaque, echogenic coiled reinforced wire to allow visibility under fluoroscopy and/or ultrasound.
- · Catheters to be provided in sterile kit/set style configurations or as a standalone replacement catheter.
141 Sal Landrio Drive • Crossroads Business Park • Johnstown • NY • 12095 • USA 1-800-866-3342 = 1-518-725-0209 · Fax: 1-518-725-0207 · Website: www.epimedpain.com · cserve@epimedint.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "Epimed", with the tagline "Designed by Experts for Experts" underneath. The logo features a stylized letter "A" to the left of the word "Epimed". The text and logo are in black against a white background.
### Statement of Intended Use:
The catheter is designed to deliver anesthesia to manage perioperative pain and/or alleviate postoperative analgesia.
### Statement of Indications For Use:
The Spirol® Block Continuous Peripheral Nerve Block Catheter permits placement of catheters next to nerve plexuses for continuous nerve block anesthesia or analgesia techniques including upper extremity, lower extremity, abdominal and paravertebral locations for periods not exceeding 72 hours.
#### Technological Characteristics and Substantial Equivalence:
Epimed's Spirol Block Continuous Peripheral Nerve Block Catheter has identical Intended Use and Indications For Use as that of the Predicate Device #2 (K122027). The catheter itself is identical in design, manufacture and material composition to Predicate Device #1 (K981329).
The catheter is a single-use, disposable device consisting of a polymer coated stainless steel spring. The catheter is available in various lengths, as well as open-ended or closed-ended configurations; the closed-ended models having three eyes (injection ports) at the distal end. The Spirol® Block is also available in styletted or non-styletted versions, as well as 19 Ga. or 20 Ga. models. This catheter is identical in design, manufacture, and material composition to Predicate Device #1; the only difference being an expansion to the Intended Use / Indications For Use.
The catheter functions by being inserted by a trained physician into a patient by way of an introducer needle (17 Ga. needles recommended to be used to place 19 Ga. catheters. 18 Ga. needles recommended to be used to place 20 Ga. catheters). After placement to the desired peripheral nerve group, the introducer needle is withdrawn, leaving the catheter in place. An Epimed catheter connector (K051171), which is packaged with the device, can then be connected to the catheter which then allows for a standard Luer-lock connection point for a bacterial filter and/ or extension set and/or syringe/pump. Upon successful connection to the desired accessories, anesthetic agents may be delivered continuously or by bolus injection to the intended peripheral nerve group.
As the device is simple in design, there are few scientific concepts that apply; the catheter simply provides a conduit to deliver anesthesia. The polymer coated stainless steel spring has been proven to resist kinking, collapsing and shearing, and has the ability to withstand a minimum acceptable tensile force. The materials used to construct the device have been tested for compatibility with drugs used for performing peripheral nerve blocks. The catheter injection ports have been designed to remain patent and deliver injectants at a flow rate sufficient for the intended use. Lastly, the catheter and the connector with which it is packaged have been tested and proven to be compatible with each other.
141 Sal Landrio Drive . Crossroads Business Park . Johnstown . NY . 12095 . USA 1-800-866-3342 • 1-518-725-0209 • Fax: 1-518-725-0207 • Website: www.epimedpain.com • cserve@epimedint.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "Epimed", with the tagline "Designed by Experts for Experts" underneath. The logo features a stylized letter "E" with a triangle shape attached to the left side. The word "Epimed" is written in bold, sans-serif font. The tagline is written in a smaller, italicized font.
There are no technological differences among the subject and predicate devices as the subject device, as well as both predicates, are:
- Single-use, disposable devices consisting of a polymer coated stainless . steel spring.
- Available in various lengths, as well as open-ended or closed-ended . configurations; the closed-ended models having three eyes (injection ports) at the distal end.
- Available in styletted or non-styletted versions. .
- Available in 19 Ga. or 20 Ga. models. .
Each catheter functions by being inserted by a trained physician into a patient by way of an introducer needle. After placement to the desired location, the introducer needle is withdrawn, leaving the catheter in place. A catheter connector can then be connected to the catheter which then allows for a standard Luer-lock connection point for a bacterial filter and/ or extension set and/or syringe/pump. Upon successful connection to the desired accessories, anesthetic agents may be delivered continuously or by bolus injection to the intended peripheral nerve group.
As the technologic characteristics across all of the referenced devices are identical, there is no effect on substantial equivalence.
#### Nonclinical Testing:
The results of the performance testing performed, i.e. dimensional, substance compatibility, kink testing, stiffness, tensile strength, connector attachment, flow test, resistance to leakage during aspiration or vacuum, resistance to leakage under pressure, corrosion resistance and skive testing, demonstrate that the Spirol® Block Continuous Peripheral Nerve Block Catheter performs comparably to, and is substantially equivalent to Predicate Device #1 (K981329) and Predicate Device #2 (K122027).
Pre-Clinical Design Validation studies have been performed on the Spirol® Block Continuous Peripheral Nerve Block Catheter. Based on the evaluation results, the Spirol® Block has been proven to conform to physician's needs, is suitable for its intended use, meeting the requirements of the Design Inputs.
In accordance with the applicable sections of ISO 10993, the relevant patient contacting components have been shown to meet the necessary biocompatibility requirements. The biocompatibility of the materials used to manufacture the Spirol® Block have been proven substantially equivalent through laboratory testing performed and documented in the Spirol® Catheter Biocompatibility Assessment. The biocompatibility testing which has been performed includes the following:
- Cytotoxicity (per ISO 10993-5) .
- . Sensitization (per ISO 10993-10)
- Irritation or Intracutaneous Reactivity (per ISO 10993-10) .
- Acute Systemic Toxicity (per ISO 10993-11) catheter only .
- Sub-acute / Sub-chronic Toxicity (per ISO 10993-11) catheter only .
- . Genotoxicity (per ISO 10993-3) catheter only
- Implantation (per ISO 10993-6) catheter only .
141 Sal Landrio Drive . Crossroads Business Park . Johnstown . NY . 12095 . USA 1-800-866-3342 • 1-518-725-0209 • Fax: 1-518-725-0207 • Website: www.epimedpain.com • cserve@epimedint.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Epimed", with the tagline "Designed by Experts for Experts" underneath. The logo is in bold, black font and has a horizontal line on either side. The "E" in Epimed is stylized with a triangle shape on the left side.
Lastly, in accordance with ISO 10993-18 Biological Evaluation of Medical Devices - Part 18: Chemical Characterization of Materials) Annex C (Principles for Judging Toxicological Equivalency) part c) The materials comprising the Spirol® Block have already been established in a more invasive exposure (Epidural applications) than this less invasive proposed application (Peripheral applications).
#### Conclusions:
The results of bench and laboratory testing demonstrate that Epimed's Spirol® Block Continuous Peripheral Nerve Block Catheter is substantially equivalent to the cited predicate devices.
Pestur ferin
Preston H. Frasier Manager - QA/RA Epimed International, Inc. prestonf@epimedint.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
May 28, 2014
Epimed International, Inc. Preston Frasier Manager - QA/RA 141 Sal Landrio Dr. Crossroads Business Park Johnstown, NY 12095
Re: K133316
Trade/Device Name: Spirol Block Continuous Peripheral Nerve Block Catheter Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: Class II Product Code: BSO Dated: April 21, 2014 Received: April 28, 2014
Dear Mr. Frasier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Frasier
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely yours,
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D.
Clinical Deputy Director
ADAGRID/ODE/CDRII FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known) K133316
#### Device Name
The Spirol® Block Continuous Peripheral Nerve Block Catheter
#### Indications for Use (Describe)
The Spirol® Block Continuous Peripheral Nerve Block Catheter permits placement of catheters next to nerve plexuses for coninuous nerve block anesthesia techniques including upper extremity, lower extremity, abdominal and paravertebral locations for periods not exceeding 72 hours.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع المو . " Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/6/Picture/12 description: The image shows the text 'Todd D. Courtney-S' on the first line and '2014.05.28 12:45:48 -04'00' on the second line. The text appears to be a timestamp or a record of some event. The font is simple and clear, making the text easily readable. The overall impression is that this is a digital record or a label.
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration . Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.