← Product Code [PEW](/productcode/PEW) · K133281

# NEOTRACT UROLIFT SYSTEM (K133281)

_Neotract, Inc. · PEW · Dec 20, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133281

## Device Facts

- **Applicant:** Neotract, Inc.
- **Product Code:** [PEW](/productcode/PEW.md)
- **Decision Date:** Dec 20, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5530
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The UroLift System is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men 50 years of age or older.

## Device Story

Implantable transprostatic tissue retractor system; treats BPH-related urinary outflow obstruction. System comprises delivery device and UroLift implant. Physician inserts delivery device into male urethra under direct visualization via standard cystoscopy; device displaces urethra toward prostatic capsule; implant deployed transversely through prostatic tissue. Implants secure retracted urethral position; maintains expanded urethral lumen; reduces fluid obstruction; improves LUTS. Used in clinical setting by physicians. Minor modifications to delivery device improve usability/reliability; implant component identical to predicate.

## Clinical Evidence

Bench testing only. Transit and functionality testing demonstrate performance equivalent to predicate. Usability testing confirms modified device utilized in same manner as predicate.

## Technological Characteristics

Implantable transprostatic tissue retractor system. Components: delivery device and UroLift implant. Mechanical deployment mechanism. Used with standard cystoscopy sheath and telescope. No software or electronic components described.

## Regulatory Identification

An implantable transprostatic tissue retractor system is a prescription use device that consists of a delivery device and implant. The delivery device is inserted transurethrally and deploys the implant through the prostate. It is designed to increase prostatic urethral patency by providing prostate lobe tissue retraction while preserving the potential for future prostate procedures and is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia in men.

## Special Controls

In combination with the general controls of the FD&C Act, the UroLift® System is subject to the following special controls:

- (1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
- (2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
- (3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity and device functionality over the requested shelf life.
- (4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
	- (A) Deployment testing must be conducted
	- (B) Mechanical strength must be conducted
	- (C) Resistance-to-degradation testing must be conducted
- (5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment
- (6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of
	- use. The following performance characteristics must be tested:
	- (A) Deployment testing must be conducted
	- (B) Implant migration must be conducted
- (7) Labeling must bear all information required for safe and effective use of the device, and must include:
	- (A) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device
	- (B) Information on the patient population for which the device has been demonstrated to be effective
	- (C) A detailed summary of the device technical parameters
	- (D) Information on how the device operates and the typical course of treatment
	- (E) An expiration date/shelf life
	- (F) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.

*Classification.* Class II (special controls). The special controls for this device are:(1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Mechanical strength must be conducted.
(iii) Resistance-to-degradation testing must be conducted.
(5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance environment.
(6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Implant migration must be conducted.
(7) Labeling must bear all information required for safe and effective use of the device, and must include:
(i) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device.
(ii) Information on the patient population for which the device has been demonstrated to be effective.
(iii) A detailed summary of the device technical parameters.
(iv) Information on how the device operates and the typical course of treatment.
(v) An expiration date/shelf life.
(vi) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.

## Predicate Devices

- NeoTract UroLift System ([K130651](/device/K130651.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| NeoTract, Inc.<br>Special 510(k) |                                                                                                                                                     | October 24, 2013<br>NeoTract UroLift System |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
|                                  | <b>510 (k) Summary</b>                                                                                                                              | K133281<br>pg 1 of 2                        |
| <b>Company Information</b>       |                                                                                                                                                     |                                             |
| <i>Manufacturer:</i>             | NeoTract, Inc.<br>4473 Willow Road, Suite 100<br>Pleasanton, CA 94588<br>Tel: 650 269 2552<br>Fax: 925 401 0683<br>FDA Registration No.: 3005791775 | DEC 2 0 2013                                |
| <i>Contact:</i>                  | Nancy E. Isaac, JD, MPH<br>Vice President, Clinical & Regulatory Affairs, Quality Assurance                                                         |                                             |

#### Device Information

| Trade Name:   | NeoTract® UroLift® System                          |
|---------------|----------------------------------------------------|
| Common Name:  | Implantable transprostatic tissue retractor system |
| Class:        | 2                                                  |
| Regulation:   | 21 CFR 876.5530                                    |
| Product Code: | PEW                                                |

### Intended Use

The UroLift System is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men 50 years of age or older.

# Predicate Device

NeoTract UroLift System, K130651

#### Device Description

The NeoTract UroLift System comprises two main components, the UroLift Delivery Device and UroLift Implant. Each delivery device comes pre-loaded with one UroLift Implant. The insertion of the UroLift Delivery Device into the male urethra is performed under direct visualization using standard surgical technique, using a standard cystoscopy sheath and telescope. 

{1}------------------------------------------------

The UroLift Delivery Device is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift Delivery Device is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. A UroLift Implant is then deployed transversely through the prostatic tissue. The implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving LUTS. This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant.

### Comparison with the Predicate Device

Minor modifications have been made to the UroLift Delivery Device to improve usability and reliability. The materials used remain the same, as well as the manner in which the implant is deployed from the delivery device. The UroLift Implant component is identical to that of the predicate device.

The modified UroLift System shares the same intended use and employs the same fundamental scientific technology as the legally marketed predicate device.

#### Performance Testing

Transit and functionality testing conducted on the modified NeoTract UroLift System demonstrate that the device meets the same performance requirements as the unmodified predicate device. Usability test results confirm that the modified device does not affect the usability of the device and that it can be utilized in the same manner as the unmodified predicate device.

# Conclusion

The modified NeoTract UroLift System is as safe and effective, has the same intended use, technological characteristics and principles of operation as the unmodified predicate device. The minor technological differences between the modified and unmodified predicate device do not raise any new questions of safety or effectiveness. Therefore, the modified NeoTract UroLift System is substantially equivalent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font. The text is black and the background is white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2013

Neotract, Inc. Nancy Isaac, J.D., M.P.H. VP, Clinical Affairs, Regulatory and Quality 4473 Willow Road, Suite 100 Pleasanton, CA 94588

Re: K133281

> Trade/Device Name: NeoTract UroLift® System Regulation Number: 21 CFR 876.5530 Regulation Name: Implantable Transprostatic Tissue Retractor System Regulatory Class: Class II Product Code: PEW Dated: December 4, 2013 Received: December 5, 2013

Dear Nancy Isaac,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 [OIR/IVD OPTION] and Part 809); medical device

Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

{3}------------------------------------------------

Page 2 - Nancy Isaac, J.D., M.P.H.

reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 |OIR/IVD OPTION] and Part 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/8 description: The image shows the name "Elaine Blyskun" in bold, black font. Below the name, the word "for" is written in a similar font. The rest of the image is filled with a pattern of small dots and lines, which may be noise or part of the original image. The text is clear and legible, while the rest of the image is more abstract.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use Statement

510(k) Number K133281

NeoTract UroLift® System Device Name

Indications UroLift System is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia for Use (BPH) in men 50 years of age or older.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Cleared by CDRH, Office of Device Evaluation (ODE)

Prescription Use X (Per 21 CFR 801. 109)

: '' '' ''

ﮩ

OR

Over-The-Counter Use

# Elaine Blyskun for Benjamin Fisher

---

**Source:** [https://fda.innolitics.com/device/K133281](https://fda.innolitics.com/device/K133281)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
