← Product Code [OMP](/productcode/OMP) · K133276

# V.A.C. GRANUFOAM DRESSING, GRANUFOAM THIN DRESSING, GRANUFOAM ROUND DRESSING, GRANUFOAM SILVER DRESSING, WHITEFOAM DRESS (K133276)

_Kci USA, Inc. · OMP · Nov 19, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133276

## Device Facts

- **Applicant:** Kci USA, Inc.
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Nov 19, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and V.A.C. Simplicity Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute, extended and home care settings. When used on open wounds, they are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, they are also intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

## Device Story

V.A.C. Therapy Wound Dressings function as part of the V.A.C. Negative Pressure Wound Therapy (NPWT) System. The system applies controlled negative pressure to a wound site via a dressing, which is connected to a powered suction pump. This process removes exudate and infectious material, reduces edema, and promotes granulation tissue formation and perfusion, creating an environment conducive to healing. The system is used in acute, extended, and home care settings by clinicians or patients. The dressing maintains a closed environment over the wound or surgical incision. By managing the wound environment, the device assists in preparing the wound bed for closure and managing post-surgical drainage, potentially accelerating healing and improving patient outcomes.

## Clinical Evidence

No clinical study was conducted. Design validation was performed on healthy human volunteers to ensure the modified device meets user requirements, as no reliable bench-top method exists for these specific metrics.

## Technological Characteristics

The system consists of wound dressing components used with a powered suction pump. Materials are evaluated for biocompatibility per ISO 10993-1. Key performance characteristics include peel adhesion and moisture vapor transmission rate (MVTR). The system is designed for use in acute, extended, and home care environments.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- V.A.C. Therapy Wound Dressings ([K120033](/device/K120033.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left, followed by the letters "KCI" in a bold, sans-serif font. The letters are outlined in black, and the interior of the letters is white. There is a small circle after the I.

::

...
... ......................................................................................................................................................................

510(k) SUMMARY

- - -- -- -- --

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

..

NOV 1 9 2013

-K133276.page 1.of 2

' .

## V.A.C. ® Therapy Wound Dressings

| Submitter Information [21 CFR 807.929(a)(1)]                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                                   | KCI USA, Inc. (Kinetic Concepts, Inc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Address                                                                                | 6203 Farinon Drive<br>San Antonio, TX 78249                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Phone number                                                                           | 210-515-4059                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Fax number                                                                             | 210-255-6727                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Establishment<br>Registration Number                                                   | 1625774                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Name of contact person                                                                 | Melanie Avila                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Date prepared                                                                          | 10/23/2013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Name of the device [21 CFR 807.92(a)(2)]                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Trade or proprietary name                                                              | V.A.C. ® Negative Pressure Wound Therapy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common or usual name                                                                   | Negative Pressure Wound Therapy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification name                                                                    | Negative Pressure Wound Therapy Powered Suction Pump (and components)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation                                                                             | 878.4780                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code(s)                                                                        | OMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Legally marketed device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | V.A.C. Therapy Wound Dressings cleared for use under multiple 510(k)s for<br>the KCI V.A.C. Therapy Negative Pressure Wound Therapy Systems. The<br>most recent 510(k) was K120033.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device description<br>[21 CFR 807.92(a)(4)]                                            | V.A.C. Therapy Wound Dressings for use with KCI's V.A.C. NPWT System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for use<br>[21 CFR 807.92(a)(5)]                                           | The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and<br>V.A.C. Simplicity Negative Pressure Wound Therapy Systems are integrated<br>wound management systems for use in acute, extended and home care<br>settings.<br><br>When used on open wounds, they are intended to create an environment that<br>promotes wound healing by secondary or tertiary (delayed primary) intention<br>by preparing the wound bed for closure, reducing edema, promoting<br>granulation tissue formation and perfusion, and by removing exudate and<br>infectious material. Open wound types include: chronic, acute, traumatic,<br>subacute and dehisced wounds, partial-thickness burns, ulcers (such as<br>diabetic, pressure or venous insufficiency), flaps and grafts.<br><br>When used on closed surgical incisions, they are also intended to manage the<br>environment of surgical incisions that continue to drain following sutured or<br>stapled closure by maintaining a closed environment and removing exudates<br>via the application of negative pressure wound therapy. |

{1}------------------------------------------------

. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo with the letters "EKCI" in a stylized font. To the left of the letters, there are several curved lines that resemble a stylized "E". The letters "K", "C", and "I" are in a bold, outlined font. A small circle is located to the right of the "I".

- ਸ਼ਾਪੂ

... ...
... ...

#### 510(k) SUMMARY

-----

### V.A.C. ® Therapy Wound Dressings

Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]

| Characteristic       | V.A.C. ® Therapy Wound Dressing System (Modified Device) | V.A.C. ® Therapy Wound Dressing System (Predicate)                                                                                                              |
|----------------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indicated Wound Type | Same as predicate                                        | Chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts |
| Dressing             | Same as predicate                                        | Multiple dressing components                                                                                                                                    |

Performance Data [21 CFR 807.92(b)]
Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]
The V.A.C. ® Dressing System was evaluated to ensure conformance to the design specifications.
The following tests were conducted:
- Peel adhesion testing
- Moisture vapor transmission rate (MVTR) testing
- Biocompatibility testing according to ISO10993-1
- Study conducted on healthy human volunteers, under design validation (210 CFR 820.30(g)), to ensure that the proposed modified device meets user requirements

Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]
There was no clinical study conducted, however, a study was conducted on healthy human volunteers, under design validation (210 CFR 820.30(g)), to ensure that the proposed modified device meets user requirements as there is no reliable method for measuring these requirements on the bench.
Conclusions drawn [21 CFR 807.92(b)(3)]
Testing indicates that the modified V.A.C. ® Therapy Wound Dressing System is substantially equivalent in terms of both indications for use and fundamental scientific technology to the predicate product.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three curved lines representing arms or wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

KCI USA, Incorporated Ms. Melanie Avila Regulatory Affairs Project Manager 6203 Fairnon Drive San Antonio, Texas 78249

November 19, 2013

Re: K133276

Trade/Device Name: V.A.C. Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 23, 2013 Received: October 24, 2013

Dear Ms. Avila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Melanie Avila

・・・・・・

ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤ

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely vours.

# Joshua C. Nipper -S

For

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### INDICATIONS FOR USE

510(k) Number (if known): __ K.133276 Device Name: _ V.A.C. ® Drape Indications for Use:

The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and V.A.C. Simplicity Negative Pressure Wound Therapy Systems are Integrated wound management systems for use in acute, extended and home care settings.

When used on open wounds, they are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.

When used on closed surgical incisions, they are also intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

Prescription Use Over-The-Counter Use ਮ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page | of | |

(Posted November 13, 2003)

Image /page/4/Picture/10 description: The image shows the name "Jiyoung Dang -S" in a bold, sans-serif font. The text is black against a white background. There is some noise or artifacts around the letters, possibly due to the image quality or scanning process.

---

**Source:** [https://fda.innolitics.com/device/K133276](https://fda.innolitics.com/device/K133276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
