K133218 · Spinal Elements, Inc. · OVE · Dec 10, 2014 · Orthopedic
Device Facts
Record ID
K133218
Device Name
Crystal ®; Mosaic ®; Vertu ®
Applicant
Spinal Elements, Inc.
Product Code
OVE · Orthopedic
Decision Date
Dec 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Crystal®: The Crystal® device is intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach and packed with autogenous bone. The Crystal® device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems). Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices. Mosaic®: The Mosaic® device is an interbody fusion device intended for spinal fusion procedures at one level (C3-C7) in sketally mature patients with degenerative disc disease (defined as neck pain with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone. Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices. The implant is designed to accommodate two, three, or four screws. The maximum number of screws should be used to ensure adequate fixation of the implant. Vertu®: The Vertu® device is a stand-alone interbody fusion device intended for spinal fusion procedures at one level ( C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach and packed with autogenous bone. Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices. The implant is designed to accommodate two screws. Two screws should be used to ensure adequate fixation of the implant.
Device Story
Intervertebral body fusion devices (Crystal, Mosaic, Vertu) for cervical spine surgery; box-shaped implants with holes for autograft; superior/inferior surfaces feature teeth for migration prevention; Mosaic and Vertu include fixation screws. Devices implanted via open, anterior approach by surgeons. Crystal requires supplemental anterior plate fixation; Mosaic and Vertu provide fixation via integrated screws. Plasma-sprayed commercial pure titanium coating added to surfaces to enhance performance. Devices stabilize spinal segment to facilitate fusion; benefit patients by addressing discogenic neck pain.
Clinical Evidence
No clinical studies were performed. Substantial equivalence supported by bench testing, including dynamic torsion (ASTM F2077-03), gravimetric wear assessment (ASTM F2025-06), particle characterization (ASTM F1877-89), static shear (ASTM F1044-05), static tension (ASTM F1147-05), and abrasion resistance (ASTM F1978-00).
Technological Characteristics
Materials: PEEK and plasma-sprayed commercial pure titanium coating (ASTM F1580). Form factor: Box-shaped interbody fusion implants with superior/inferior teeth. Mosaic/Vertu include fixation screws. No electrical/EMC components. Sterilization method not specified.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (neck pain of discogenic origin) at one level (C3-C7) of the cervical spine. Requires at least 6 weeks of prior non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized eagle or bird with three profiles of human faces incorporated into the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 10, 2014
Spinal Elements, Incorporated Julie Lamothe, Ph.D. Regulatory Affairs and Quality Assurance Director 3115 Melrose Drive, Suite 200 Carlsbad, California 92010
Re: K133218
Trade/Device Name: Crystal®; Mosaic®; Vertu® Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, OVE Dated: November 7, 2014 Received: November 10, 2014
Dear Dr. Lamothe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
Page 2 - Julie Lamothe, Ph.D
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K133218
Device Name Crystal®; Mosaic®; Vertu®
## Indications for Use (Describe)
Crystal®:
The Crystal® device is intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach and packed with autogenous bone.
The Crystal® device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
#### Mosaic®:
The Mosaic® device is an interbody fusion device intended for spinal fusion procedures at one level (C3-C7) in sketally mature patients with degenerative disc disease (defined as neck pain with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
The implant is designed to accommodate two, three, or four screws. The maximum number of screws should be used to ensure adequate fixation of the implant.
#### Vertu®:
The Vertu® device is a stand-alone interbody fusion device intended for spinal fusion procedures at one level ( C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach and packed with autogenous bone.
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
The implant is designed to accommodate two screws. Two screws should be used to ensure adequate fixation of the implant.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary Crystal®, Mosaic®, and Vertu®
#### 510(k) Number K133218
#### I. SUBMITTER
Spinal Elements, Inc. 3115 Melrose Dr., Suite 200 Carlsbad, CA 92010 P. 760-607-0121 F. 760-607-0125
| Contact Information: | Julie Lamothe |
|----------------------|-----------------------------|
| | Director Regulatory Affairs |
| | Spinal Elements, Inc. |
| | 3115 Melrose Dr., Suite 200 |
| | Carlsbad, CA 92010 |
| | 760-607-1816 |
| | jlamothe@spinalelements.com |
Date Prepared: December 4, 2014
II. DEVICE
Proprietary Name Common Name Device Classification
Device Product Code
Crystal®; Mosaic®; Vertu® Systems Intervertebral Body Fusion Device 21 CFR 888.3080 (Appliance, Fixation Spinal Intervertebral Body) Class II ODP - Crystal OVE - Mosaic and Vertu
#### III. PREDICATE DEVICE
Proposed Regulatory Class
Primary Predicate Name Regulatory Class Submission No Device Code
Predicate Name Regulatory Class Submission No Device Code
Calix PC Class II K112036 ODP
Crystal®; Mosaic®; Vertu® Systems Class II K073351 - K071833 - K122771 ODP - Crystal OVE - Mosaic and Vertu
{6}------------------------------------------------
## IV. DEVICE DESCRIPTION
#### Crystal
Crystal is an intervertebral body fusion device for use in cervical spinal surgery. The device is generally box-shaped with various holes throughout its geometry to allow for the placement of autograft. The exterior of the device has "teeth" or other generally sharp engagement members on the superior and inferior surfaces to help prevent the device from migrating once it is surgically positioned.
#### Mosaic
The Mosaic Spinal Implant System is composed of a device body and fixation screws. The body is a generally box-shaped device with various holes located throughout its geometry for the placement of graft material. Additionally, it has teeth on its superior and inferior external surfaces to keep the device from migrating once placed in its desired location. The box-shaped body has projections (or flanges) that encompass screw holes. Screws pass through screw holes and affix to bone to help prevent implant migration.
#### Vertu
The Vertu Cervical Intervertebral Body Fusion System is composed of an implant body and fixation screws. The implant body is a generally box-shaped device with holes through its body for the placement of graft material. Additionally, it has teeth located on its superior and inferior external surfaces to help keep the device from migrating once placed in its desired location. There are also screw holes located in the implant body. Screws pass through screw holes of the implant body and affix to bone to help prevent implant migration.
#### V. INDICATION FOR USE
#### Crvstal®
The Crystal® device is intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
The Crystal® device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine (i.e., anterior plate systems).
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
#### Mosaic®
The Mosaic® device is an interbody fusion device intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
{7}------------------------------------------------
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
The implant is designed to accommodate two. three, or four screws. The maximum number of screws should be used to ensure adequate fixation of the impant.
#### Vertu®
The Vertu® device is a stand-alone interbody fusion device intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are to be implanted via an open, anterior approach and packed with autogenous bone.
Patients must have undergone a regimen of at least six (6) weeks non-operative treatment prior to being treated with these devices.
The implant is designed to accommodate two screws. Two screws should be used to ensure adequate fixation of the impant.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject Crystal, Mosaic and Vertu devices are identical in indications for use, surgical technique, design features and instrumentation to the predicate devices cleared in K073351, K071833 and K122771, respectively. The difference to the Crystal, Mosaic, and Vertu devices is the addition of the plasma sprayed commercial pure titanium coating on the superior and inferior surfaces of the devices. The performance of the Ti- coating is equivalent to the one of predicate Calix PC cleared in K112036. The addition of this coating does not raise any new issues of safety or effectiveness. In addition to the plasma sprayed commercial pure titanium coating on the Crystal device, line items (sizes) have been added to Crystal. Neither modification raises any new issues of safety or effectiveness.
#### VII. PERFORMANCE DATA
#### Biocompatibility Testing
Biocompatibility testing were performed in accordance with FDA Memorandum #G95-1 and ISO 10993-1 Part 1. The Crystal, Mosaic and Vertu devices are considered permanent implant devices contacting tissue/bone.
The PEEK material biocompatibility relevant to the device contact type is presented in Invibio device master file (MAF 1209). The titanium coating material is commercially pure titanium. The coating is in accordance with ASTMF 1580, which address the biocompatibility of the material. No further testing were required,
#### Electrical safety and electromagnetic compatibility (EMC)
{8}------------------------------------------------
No electrical and electromagnetic compatibility testing were performed.
## Software Verification and Validation Testing
The device does not contain software. Therefore no software verification and validation testing were performed.
## Mechanical testing
Non-clinical testing were used to support the decision of substantial equivalence. Nonclinical testing consisted of the following testing performed in accordance with the FDA guidance Class II Special Controls Guidance Document: Intervertebral Body Fusion Device:
- Dynamic Torsion Testing ASTM F2077-03
- Gravimetric Measurement for Wear Assessment ASTM F2025-06
- . Particle Characterization ASTM F1877-89
- Static Shear ASTM F1044-05
- Static Tension ASTM F1147-05
- . Abrasion Resistance ASTM F1978-00
The wear debris data including number of particulates generated, and particulate size indicates that the device coating performs similarly to predicate Calix PC (K112036).
## Animal Study
No animal studies were performed.
#### Clinical Studies
No clinical studies were performed.
## VIII. CONCLUSIONS
Based on our analysis of the test data, the device performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.