K133207 · Spine Wave, Inc. · MQP · Dec 6, 2013 · Orthopedic
Device Facts
Record ID
K133207
Device Name
STAXX(R) SYSTEM
Applicant
Spine Wave, Inc.
Product Code
MQP · Orthopedic
Decision Date
Dec 6, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allograft and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System. The StaXx® XDL System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be used with autograft or allograft and supplemental spinal fixation. The StaXx® IB System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IB System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device. The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IBL System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.
Device Story
StaXx System comprises expandable implants, interlocking wafers, cartridge, and surgical gun. Surgical gun delivers wafers from cartridge to implant to incrementally adjust height. Used in thoracolumbar/lumbar spine procedures; operated by surgeons. Implants made of PEEK-OPTIMA with barium sulfate and tantalum markers for fluoroscopic visualization. System restores vertebral height or facilitates intervertebral fusion. Benefits include height restoration and stabilization of spinal segments.
Clinical Evidence
Bench testing only. Evidence includes simulated use testing, insertion testing, and cadaveric testing to demonstrate performance equivalence to predicate systems.
Technological Characteristics
System components: expandable implants, interlocking wafers, cartridge, surgical gun. Materials: PEEK-OPTIMA, 6% Barium Sulfate, tantalum markers. Mechanical delivery of wafers to form stack. Standalone manual surgical instrument.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine (L2-S1) or thoracolumbar vertebral body replacement (T1-L5) due to tumor or trauma. DDD patients may have up to Grade I spondylolisthesis or retrolisthesis and require 6 months of failed non-operative treatment.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
DEC 0 6 2013
**Spine**WAVE
# 510(k) Summary StaXx® System
## 1. Submitter Information
Submitter: Address:
Spine Wave, Inc. Three Enterprise Drive Suite 210 Shelton, CT 06484 Telephone: 203-712-1847 203-944-9493 Telefax:
Contact:
Joseph Mercado
Date Prepared: December 3, 2013
- 2. Device Information Trade Name:
StaXx® System
Fusion Device
Common Name(s):
Classification Name(s):
Spinal Intervertebral Body Fixation Orthosis; Intervertebral Fusion Device with Bone Graft, Lumbar
Vertebral Body Replacement; Intervertebral Body
Classification(s)/Code(s):
Class II per 21 CFR 888.3060/MQP; Class II (special controls) per 21 CFR 888.3080/MAX
## 3. Purpose of Submission
The purpose of this submission is to gain clearance for a new Surgical Gun to be used with the StaXx® Systems.
### 4. Predicate Device Information
The StaXx® Systems described in this submission are substantially equivalent to the following predicates:
| Predicate Device | Manufacturer | Most recent<br>510(k) No. |
|-------------------|------------------|---------------------------|
| StaXx® XD System | Spine Wave, Inc. | K121889 |
| StaXx® XDL System | Spine Wave, Inc. | K102315 |
| StaXx® IB System | Spine Wave, Inc. | K123461 |
| StaXx® IBL System | Spine Wave, Inc. | K132719 |
{1}------------------------------------------------
## 5. Device Description
The StaXx® Systems are composed of expandable implants and interlocking wafers, which are implanted through a delivery system that includes a Cartridge and a Surgical Gun. The Cartridge is inserted into the Surgical Gun which, when activated, sequentially delivers locking wafers from the Cartridge to the expandable implant. The wafers are designed to be inserted incrementally to form a stack and therefore attain the desired height.
The StaXx® XD and XDL Systems are vertebral body replacement devices intended for use in the thoracolumbar spine (TI-L5). They are composed of wafers that are stacked into an expandable implant to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The implant components are manufactured from PEEK-OPTIMA with 6% Barium Sulfate and contain tantalum markets for additional visualization under fluoroscopy. The StaXx® XD System is to be placed bilaterally while the StaXx® XDL System is to be placed unilaterally and both are to be used with autograft or allograft and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System.
The StaXx® IB and IBL Systems are intervertebral body fusion devices composed of wafers that are stacked into an expandable implant to adjust the height of the implant. The implants are to be used with autogenous bone graft material. The StaXx® IB implant components are manufactured from both PEEK-OPTIMA and PEEK-OPTIMA with 6% Barium Sulfate while the StaXx® IBL implant components are manufactured from PEEK-OPTIMA with 6% Barium Sulfate and tantalum markers.
### 6. Intended Use
The StaXx® Systems have the following indications:
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allografi and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System.
The StaXx® XDL System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed. damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be used with autograft or allograft and supplemental spinal fixation.
{2}------------------------------------------------
The StaXx® IB System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IB System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.
The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IBL System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.
### 7. Comparison of Technological Characteristics
The substantial equivalence of the StaXx® Systems to the predicates is shown by similarity in intended use, indications for use, materials and performance.
### 8. Performance Data
The following non-clinical tests were performed to demonstrate that the subject StaXx® Systems are substantially equivalent to the listed predicate devices:
- Simulated Use Testing .
- . Insertion Testing
- . Cadaveric Testing
### 9. Conclusion
Based on the indications for use, technological characteristics and comparison to the predicates, the subject StaXx® Systems have been shown to be substantially equivalent to the predicate devices identified in this submission, and do not present any new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2013
Spine Wave, Incorporated Mr. Joseph Mercado Regulatory Affairs Specialist 3 Enterprise Drive, Suite 210 Shelton, Connecticut 06484
Re: K133207
Trade/Device Name: StaXx * Surgical Gun Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MOP, MAX Dated: November 5. 2013 Received: November 6, 2013
Dear Mr. Mercado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mr. Joseph Mercado
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiqqins
- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K133207
#### Device Name StaXx(R) Surgical Gun
#### Indications for Use (Describe)
The StaXx® Systems have the following indications:
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (TI-LS) to replace and restore height of a collapsed, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allograft and supplemental fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System:
The StaXx® XDL System is a vertebral body replacement device in the thoracolumbar spine (TI-LS) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof. due to turnor or trauma (i.e., fracture), The system is to be used with autograft or allograft and supplemental spinal fixation.
The StaXx® IB System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IB System is to be used with and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral body fusion device.
The StaXx® IBL System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The StaXx® IBL System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral body fusion device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| | ---------------- | <br>Company Concessful Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children Ch | <br>ANNO FOR A LA B |
|------------------|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| Concurrence of t | Center for I | *or Devices and Radiological Health (CDRH) i | |
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.