← Product Code [MAX](/productcode/MAX) · K133192

# MECTALIF TIPEEK (K133192)

_Medacta International · MAX · Jan 30, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133192

## Device Facts

- **Applicant:** Medacta International
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jan 30, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.

## Device Story

MectaLIF TiPEEK is an intervertebral body fusion device designed to replace degenerative discs and restore spinal column height. The device consists of PEEK-OPTIMA LT1 implants with a commercially pure titanium (CPTi) coating and tantalum markers. It is available in oblique and posterior configurations. Used in PLIF or TLIF surgical techniques, the device is implanted by a surgeon in an operating room setting. It functions as a structural spacer to be used in conjunction with autogenous bone graft and supplemental posterior fixation (e.g., pedicle screw systems) to facilitate spinal fusion. The device provides mechanical support to the intervertebral space, potentially alleviating discogenic pain and stabilizing the spinal segment.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included static and dynamic axial compression (ASTM F2077), static and dynamic shear compression (ASTM F2077), subsidence (ASTM F2267), wear analysis (ASTM F2077, ASTM F1877), abrasion resistance (ASTM F1978), static tensile strength (ASTM F1147), static shear strength (ASTM F1044), shear fatigue strength (ASTM F1160), and micrographical analysis (ASTM F1854).

## Technological Characteristics

Implant material: PEEK-OPTIMA LT1 (ASTM F 2026) with commercially pure titanium (CPTi) plasma-sprayed coating (ASTM F 1580). Markers: Tantalum (ISO 13782 / ASTM F 560). Form factor: Oblique and posterior interbody fusion device. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- MectaLIF ([K110927](/device/K110927.md))
- MectaLIF Extension ([K131671](/device/K131671.md))
- Calix PC Spinal Implant System ([K112036](/device/K112036.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in a bold, sans-serif font, with two triangles above the "M" that form a stylized mountain range. Below the word "Medacta" is the word "International" in a smaller font, and to the right of "International" is a plus sign.

JAN 30 2014

# 510(k) Summary

Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Mr. Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805) 322-3289 Fax: (805) 437-7553 Email: AGross@medacta.us.com

January 20, 2014 Date Prepared:

# DEVICE INFORMATION

Trade/Proprietary Name: MectaLIF TiPEEK Common Name: Intervertebral Body Fusion Device Classification Name: Intervertebral Fusion Device with Bone Graft, Lumbar 21 CFR 888.3080 Class II Device Product Codes: MAX

Predicate Devices:

| 510(k)  | Product                        | 510(k) Holder         | Clearance<br>Date |
|---------|--------------------------------|-----------------------|-------------------|
| K110927 | MectaLIF                       | Medacta International | 6/13/2011         |
| K131671 | MectaLIF Extension             | Medacta International | 7/05/2013         |
| K112036 | Calix PC Spinal Implant System | X-spine Systems, Inc  | 11/8/2011         |

{1}------------------------------------------------

# Product Description

MectaLIF TiPEEK is characterized by different sizes of PEEK-OPTIMA LT1 (Polyetheretherketone) implants with a Titanium coating and Tantalum Markers. MectaLIF TiPEEK consists of the Oblique Interbody Fusion Device and Posterior Interbody Fusion Device. The Oblique and Posterior Interbody Fusion Devices are used to replace a degenerative disc in order to restore the height of the spinal column structure. MectaLIF TIPEEK can be applied with common surgical techniques such as PLIF (Posterior Lumbar Intervertebral Fusion) and TLIF (Transforaminal Lumbar Intervertebral Fusion). The devices are intended to be used in combination with posterior fixation (e.g. Pedicle Screw System) as well as an autogenous bone graft. The materials of MectaLIF TiPEEK are as follows: Implant: PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) with commercially pure titanium (CPTi, ASTM F 1580) coating and Marker: Tantalum (ISO 13782 / ASTM F 560).

# Indications for Use

The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.

# Comparison to Predicate Devices

MectaLIF TiPEEK is identical to the implants cleared under K110927 and K131671 except for the addition of the commercially pure titanium (CPT). ASTM F 1580) coating. which is the same coating used with K112036 Calix PC. The indications for use of MectaLIF TiPEEK are identical to the previously cleared predicate devices. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The substantial equivalence of the MectaLIF TiPEEK implants are supported by the performance testing, materials information, and data analysis provided within this Premarket Notification.

{2}------------------------------------------------

## Performance Testing

Performance testing was conducted in accordance with FDA Guidance - Class II Special Controls Guidance Document: Intervertebral Body Fusion Device as well as in accordance with FDA Guidance - Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements. Additional wear analysis for particulates was conducted and it was determined that MectaLIF TiPEEK passed all requirements.

MectaLIF TiPEEK was tested using the worst-case device for each of the following tests:

Static and dynamic axial compression - ASTM F2077 Static and dynamic shear compression - ASTM F2077 Subsidence - ASTM F 2267 Wear Analysis - ASTM F2077, ASTM F1877 Abrasion resistance - ASTM F1978 Static Tensile strength - ASTM F1147 Static Shear strength - ASTM F1044 Shear fatigue strength - ASTM F1160 Micrographical Analysis - ASTM F1854

Conclusion:

Based on the above information, the MectaLIF TIPEEK can be considered as substantially equivalent to its predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with stylized wings and tail feathers.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 30, 2014

Medacta International Mr. Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 1556 West Carroll Avenue Chicago, Illinois 60607

Re: K133192

Trade/Device Name: MectaLIF TiPEEK Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: October 18, 2013 Received: November 14, 2013

Dear Mr. Gross:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

## Page 2 - Mr. Adam Gross

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Vincen EDDevlin -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K133192

# Device Name

MectaLIF TiPEEK

## Indications for Use (Describe)

The Mecal.[F implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with desc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment with the devices.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement on last page.

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

・・・・・ : . . . . . . FOR FDA USE ONLY . . . . . . . . . .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# EDmitrie

{6}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K133192](https://fda.innolitics.com/device/K133192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
