← Product Code [DXT](/productcode/DXT) · K133189

# DUAL SHOT ALPHA 7 (K133189)

_Nemoto Kyorindo Co., Ltd. · DXT · Jun 30, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133189

## Device Facts

- **Applicant:** Nemoto Kyorindo Co., Ltd.
- **Product Code:** [DXT](/productcode/DXT.md)
- **Decision Date:** Jun 30, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1650
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The contrast delivery system Dual Shot Alpha 7 is an intravascular injection system intended for the administration of ionic and non-ionic contrast media and saline used in conjunction with computed X-ray tomography (CT).

## Device Story

Dual Shot Alpha 7 is an electromechanical angiographic injector for CT imaging; delivers contrast media and saline via dual-syringe powerhead. System comprises powerhead, console with TFT LCD touch interface, and hand switch. Operator sets injection parameters (pressure, volume, flow rate, time) on console; powerhead uses two electromechanical actuators to drive syringes. Features include body weight-based protocols, multi-phase injection (up to 5 phases), and interphase delay. Device integrates with CT scanners via NCOM for synchronized injection/scan timing. Used in clinical radiology settings by trained professionals. Output includes automated, controlled delivery of contrast/saline to patient vascular system; provides real-time pressure monitoring and injection history. Benefits include precise, reproducible contrast delivery, enhancing diagnostic image quality while maintaining safety through software-based stop mechanisms and pressure limits.

## Clinical Evidence

Bench testing only. No clinical data provided. Device performance verified through conformance to EN 60601-1 (safety), EN 60601-1-2 (EMC), EN 60601-1-6 (usability), and EN 62304 (software life cycle).

## Technological Characteristics

Electromechanical dual-syringe injector; motor-driven actuators. Console features TFT LCD touch panel. Connectivity includes CT scanner interlock (NCOM). Software-controlled injection parameters (pressure 10-300 PSI, flow 0.1-10 mL/sec). Power supply integrated with universal voltage input. Complies with EN 60601-1, EN 60601-1-2, EN 62304, and EN 62366.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Dual Shot Alpha ([K062168](/device/K062168.md), [K071691](/device/K071691.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| Nemoto Kyorindo Co., Ltd.  |
|----------------------------|
| 2-12-4 Aoki, Kawaguchi-shi |
| Saitama, 332-0031          |

# JUN 3 0 2014 K133189

ry

Special 510k Section 011 Page 1 of 4

# Special 510(k) SUMMARY CONTRAST DELIVERY SYSTEM - Dual Shot Alpha 7

DATE SUMMARY PREPARED: June 19, 2013

OFFICIAL CONTACT. Ken Wakabayashi Vice Director, Overseas Department Nemoto Kyorindo CO., LTD. 2-12-4 Aoki Kawaguchi-shi Saitama 332-0031, Japan TEL : +81 48 250 3255 FAX : +81 48 250 3256

CLASSIFICATION NAME(S): Angiographic Injector with Syringe

DEVICE CLASSIFICATION: Class II

COMMON NAME: Powered Injector with Syringe PROPRIETARY NAME: Contrast Delivery System -- Dual Shot Alpha 7

PREDICATE DEVICE:

Dual Shot Alpha- CONTRAST DELIVERY SYSTEM from NEMOTO KYORINDO K062168 and K071691

INTENDED USE:

The contrast delivery system Dual Shot Alpha 7 is an intravascular injection system intended for the administration of ionic and non-ionic contrast media and saline used in conjunction with computed X-ray tomography (CT).

## DESCRIPTION OF DEVICE:

The contrast delivery system ~ Dual Shot Alpha 7 is an angiographic injector that is used in conjunction with Computed Tomography, and is intended for use by doctors, radiology technologists and other licensed medical practitioners. This device is designed to be used in conjunction with various injection methods of contrast media and saline, and utilized with multi-slice CT scanners.

The Dual Shot Alpha 7 has two driving mechanisms to deliver contrast media and/or saline, each side A and B respectively are capable of utilizing either a 100mL or 200mL size syringe. Syringes are connected to the patient via an intravascular catheter. The Dual Shot Alpha 7 consists following components;

{1}------------------------------------------------

| Nemoto Kyorindo Co., Ltd.  | Special 510k |
|----------------------------|--------------|
| 2-12-4 Aoki, Kawaguchi-shi | Section 011  |
| Saitama, 332-0031          | Page 2 of 4  |

#### Iniector Powerhead

The injector Powerhead is intended to provide accurate, automatic delivery of contrast media by two electromechanically driven actuators. Sterile empty-syringes or pre-filled syringes can be set onto the Injector head. For the sterile emplysyringes, the injector head provides a Quick Return function to fill the syringe with contrast media or saline. The Powerhead is composed of operation indicator LED, switches for manual operation, start-switch to start infusion and stop-switch to stop infusion. Injector head detects injection conditions and transmits the Injector head status information to the console.

#### Console

The new style Console, now has an integrated power supply with universal voltage input, was designed to take up very little space on a control desk, and consists of TFT LCD display and touch panel interface to set/display various injections set by an operator. The operator can set injection pressure, time, volume, flow rate and a patient examination region of interest. The Dual Shot Alpha 7 also includes as a standard item the Body Weight Protocol from our previous Dual Shot Alpha platform.

- Options .
## Hand Switch

The Hand Switch is connected to the Console and consists of start and stop switches to start/stop injection according to the protocols already set on the console. The Hand Switch provides LED style display to display scan time and injection state,

## SUBSTANTIAL EQUIVALENCE:

The contrast delivery system Dual Shot Alpha 7 maintains the same intended use as the predicate device. It is intended for the specific purpose of injecting contrast media and saline solution into a patient's vascular system to obtain diagnostic images in X-ray computed tomography (i.e. "CT").

The contrast delivery system Dual Shot Alpha 7 consists of two main components the Powerhead and Console unlike the predicate device that included an Injector head, a Console, and a Power Supply, Both the Dual Shot Alpha 7 and predicate device consist of the same or substantially equivalent materials and technology. They are motor driven, electromechanical devices that are controlled by software.

Below is a table that compares the predicate device (Dual Shot Alpha) to the proposed modified Dual Shot Alpha 7,

{2}------------------------------------------------

# Nemoto Kyorindo Co., Ltd.
2-12-4 Aoki. Kavaguchi shi
Saitama. 332-0031

.

Special 510k Section 011
· Page 3 of 4

| Feature                                | Proposed Device:                                                                                                                                                                                                                             | Predicate Device:                                                 |  |      |  |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|--|------|--|
|                                        | Contrast Delivery System -<br>Dual Shot Alpha 7                                                                                                                                                                                              | Contrast Delivery System -<br>Dual Shot Alpha K062168 and K071691 |  |      |  |
| Intended Use                           | The contrast delivery system Dual Shot<br>Alpha 7 is an intravascular injection<br>system intended for the administration of<br>ionic and non-ionic contrast media and<br>saline used in conjunction with computed<br>X-ray tomography (CT). | Same                                                              |  |      |  |
| Single or Dual Syringe<br>System       | Dual syringe model                                                                                                                                                                                                                           | Same                                                              |  |      |  |
| Information Display                    | Color LCD                                                                                                                                                                                                                                    | Same                                                              |  |      |  |
| Programming Keys                       | Non-dedicated keys - software determined                                                                                                                                                                                                     | Same                                                              |  |      |  |
| Touch Screen                           | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Multi-Phase                            | 1-5 Phases per injection                                                                                                                                                                                                                     | Same                                                              |  |      |  |
| Interphase Delay                       | 0-300 seconds, 1 sec. increments                                                                                                                                                                                                             | Same                                                              |  |      |  |
| Rise Time                              | Off = 0 seconds, On = 2 seconds                                                                                                                                                                                                              | Same                                                              |  |      |  |
| Arming Modes                           | Single                                                                                                                                                                                                                                       | Same                                                              |  |      |  |
| Protocol Storage<br>Capability         | 80 protocols                                                                                                                                                                                                                                 | 125 protocols                                                     |  |      |  |
| Hold Capability                        | Until user operate.                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Scan Delay                             | 1 - 300 seconds                                                                                                                                                                                                                              | Same                                                              |  |      |  |
| Variable Jog Speeds                    | 0.5ml/sec, 1.5ml/sec, 8.0ml/sec                                                                                                                                                                                                              | Same                                                              |  |      |  |
| Body weight Setting                    | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Body weight                            | 10-200kg 1kg increment<br>10-440lb 1lb increment                                                                                                                                                                                             | Same                                                              |  |      |  |
| Contrast media                         | 300, 320, 350, 370, 400<br>or 3 preset parameter                                                                                                                                                                                             | Same                                                              |  |      |  |
| Iodine volume                          | 10-1000 mg/kg<br>1mg/kg increment                                                                                                                                                                                                            | Same                                                              |  |      |  |
| Time                                   | 0:01-5:00 or 3 preset parameter                                                                                                                                                                                                              | Same                                                              |  |      |  |
| Volume Ratio                           | 5.95-95.5% 1% increment                                                                                                                                                                                                                      | Same                                                              |  |      |  |
| Injection Results History              | 100                                                                                                                                                                                                                                          | 40                                                                |  |      |  |
| Timing Bolus Duration<br>Mode          | Yes                                                                                                                                                                                                                                          | No                                                                |  |      |  |
| Safety Stop Mechanism                  | Multi layered software stops with backup<br>monitoring.                                                                                                                                                                                      | Same                                                              |  |      |  |
| Syringe System                         | A-head: 200mL or 100mL syringe<br>B-head: 200mL or 100mL syringe                                                                                                                                                                             | A-head: 200mL or 100mL syringe<br>B-head: 100ml syringe           |  |      |  |
| Programmed Volume                      | A or B-head: 1 to 200ml. or 1 to 100ml.<br>(depending on syringe size)                                                                                                                                                                       | Same                                                              |  |      |  |
| Volume Remaining<br>Readout            |                                                                                                                                                                                                                                              | Graphic and numeric on LCD                                        |  | Same |  |
| Flow Rate                              | 0.1mL/sec to 10mL/sec                                                                                                                                                                                                                        | Same                                                              |  |      |  |
| Programmable Pressure Limit            | Settable from 10 to 300PSI                                                                                                                                                                                                                   | Same                                                              |  |      |  |
| Pause                                  | Programmable –<br>1 sec to 300 seconds in 1sec increments                                                                                                                                                                                    | Same                                                              |  |      |  |
| Quick Return                           | Fill rate 0.5, 1.5 and 8.0 mL/sec                                                                                                                                                                                                            | Same function different name Auto-Return                          |  |      |  |
| Quick Purge                            | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Remote Start Switch                    | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Pressure Graph                         | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Syringe Sensing                        | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Test Injection                         | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Syringe Heat Maintainer                | Yes                                                                                                                                                                                                                                          | Same                                                              |  |      |  |
| Power Saver Mode                       | Yes                                                                                                                                                                                                                                          | No                                                                |  |      |  |
| Integrated Power Supply                | Yes                                                                                                                                                                                                                                          | No                                                                |  |      |  |
| Interlocking with CT scanner with NCOM | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Injection Start                        | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Injection Abort                        | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Injection Hold                         | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Powerhead Indicator lights             | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Power-on Self Test                     | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |
| Selectable Injection Patterns          | Yes                                                                                                                                                                                                                                          | Yes                                                               |  |      |  |

.

{3}------------------------------------------------

| Nemoto Kyorindo Co., Ltd.  | Special 510k |
|----------------------------|--------------|
| 2-12-4 Aoki, Kawaguchi shi | Section 011  |
| Saitama, 332-0031          | Page 4 of 4  |

## PERFORMANCE DATA:

The contrast delivery system Dual Shot Alpha 7 has been tested in conformance with the following recognized standards, and is substantially equivalent to the predicate device Dual Shot Alpha:

| EN 60601-1 (2006) | Medical Electrical Equipment, Part 1: General Requirements<br>for basic Safety and essential performance |
|-------------------|----------------------------------------------------------------------------------------------------------|
|-------------------|----------------------------------------------------------------------------------------------------------|

- EN 60601-1-2 (2007) Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
- EN 60601-1-6 (2010) Medical electrical equipment Part 1-6: General requirements for safety and essential performance -Collateral standard. Usability
- EN 62304:2006 Medical Device Software - Software Life Cycle Processes
- EN 62366:2008 Medical Devices - Application of usability engineering to medical devices
- EN 1041:2008 Information supplied by the manufacturer with medical devices
- EN ISO13485:2012 Medical devices - Quality management systems -Requirements for regulatory purposes
- EN ISO14971:2012 Medical devices -- Application of risk management to medical devices

Biocompatibility testing has not been performed because the contrast delivery system Dual Shot Alpha 7 does not include a sterile syringe.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird in flight.

Public Health Service

Food and Drug Administration ! 0903 New Hampshire Avenue Document Contro! Center - WO66-G609 Silver Spring, MD 20993-0002

June 30, 2014

Nemoto Kyorindo Co., Ltd. Ken Wakabayashi Vice Director 2-27-20 Hongo Bunkyo-ku Tokyo 113-0033 Japan

Re: K133189

> Trade/Device Name: Dual Shot Alpha7 Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector with Syringe Regulatory Class: Class II Product Code: DXT Dated: April 4, 2014 Received: June 2, 2014

Dear Ken Wakabayashi,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{5}------------------------------------------------

Page 2 - Ken Wakabayashi

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

M.A. Hilleman

for

Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Section 009, Page 1 of 1

# Indications for Use

510(k) Number (if known): 长133189

Device Name: DUAL SHOT alpha7

Indications for Use:

The DUAL SHOT alpha7 is an intravascular injection system intended for the administration of ionic and non-ionic contrast media and saline used in conjunction with computed X-ray tomography (CT).

Prescription Use O (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Hillemann

---

**Source:** [https://fda.innolitics.com/device/K133189](https://fda.innolitics.com/device/K133189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
