← Product Code [MAX](/productcode/MAX) · K133184

# ARDIS INTERBODY SYSTEM (K133184)

_Zimmer Spine, Inc. · MAX · Jan 30, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133184

## Device Facts

- **Applicant:** Zimmer Spine, Inc.
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jan 30, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Ardis Interbody System is indicated for use with autogenous bone graft as an intervertebral body fusion device at one or two contiguous levels in the lumbosacral region (L2-S1) in the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients with previous non-fusion spinal surgery at involved level may be treated with the device. Patients should be skeletally mature and have had six months of non-operative treatment. The Ardis Interbody System is implanted using a posterior or transforaminal approach and is intended to be used singly or in pairs with supplemental fixation.

## Device Story

Ardis Interbody System is a spinal implant for interbody fusion; hollow PEEK-OPTIMA spacer with textured surfaces and internal cavity for autogenous bone graft; includes tantalum radiographic markers. Used in lumbosacral region (L2-S1) via posterior or transforaminal approach; requires supplemental fixation. System includes single-use implants and reusable stainless steel instruments (rasps, tamps, shavers, trials). Operated by surgeons in clinical/OR settings. Implant provides structural support for fusion; markers allow radiographic visualization. Benefits include stabilization of spinal segments in DDD patients.

## Clinical Evidence

Bench testing only. A simulated surgical evaluation was performed using a cadaver specimen to validate instrument function, performance, and conformance to user needs. Participating healthcare professionals confirmed the system met all established acceptance criteria.

## Technological Characteristics

Implant material: PEEK-OPTIMA (ASTM F2026). Radiographic markers: Tantalum wires/beads (ASTM F560). Instruments: Stainless steel with AlTiN or TiN PVD coating. Design: Hollow spacer with textured convex surfaces. Sterilization: Gamma radiation for implants; non-sterile for instruments (facility-sterilized). Connectivity: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Ardis Spacer ([K073202](/device/K073202.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|                                    | 510(k) SUMMARY                                                                                                                      | JAN 30 2014 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------|
| P. BROKERS<br>spine                | Ardis® Interbody System                                                                                                             |             |
| Date of Summary Preparation:       | October 08, 2013                                                                                                                    |             |
| Submitter:                         | Zimmer Spine, Inc.<br>7375 Bush Lake Road<br>Minneapolis, MN 55439                                                                  |             |
| Establishment Registration Number: | 2184052 (Minneapolis)                                                                                                               |             |
| Company Contact:                   | Michelle Lenz<br>Regulatory Affairs Specialist<br>Email: Michelle Lenz@Zimmer.com<br>Office 952.830.6243<br>Fax 952.837.6843        |             |
| Secondary Contact:                 | Jonathan Gilbert<br>Regulatory Affairs Director<br>Email: Jonathan. Gilbert@Zimmer.com<br>Office: 952.830.6385<br>Fax: 952.837.6985 |             |
| Trade Name:                        | Ardis® Interbody System                                                                                                             |             |
| Device Name (Common Name):         | Interbody Fusion Device                                                                                                             |             |
| Device Classification:             | Class II                                                                                                                            |             |
| Product Code(s):                   | MAX                                                                                                                                 |             |
| Regulation Number:                 | 21 CFR § 888.3080                                                                                                                   |             |
| Regulation Description:            | Intervertebral Body Fusion Device                                                                                                   |             |
| Predicate Device:                  | Ardis Spacer (K073202)                                                                                                              |             |

## General Device Description:

The Ardis® Interbody System is an implant device for interbody fusion of the anterior column of the spine. It is available in various height, width and length options. The device has a textured "tooth" pattern on both the cranial and caudal sides of the device and teardrop holes on the medial and lateral sides of the device. The implant incorporates an internal cavity that allows for the placement of bone graft material.

The system includes the single-use spacer implant, as well as the associated reusable instruments used for site preparation, trialing, and placement/extraction of the device, such as rasps, tamps, shavers, bone funnel, bone tamp, slap-hammer, T-handle and trials.

## Indications for Use:

The Ardis Interbody System is indicated for use with autogenous bone graft as an intervertebral body fusion device at one or two contiguous levels in the lumbosacral region (L2-S1) in the 

{1}------------------------------------------------

treatment of degenerative disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients with previous non-fusion spinal surgery at involved level may be treated with the device. Patients should be skeletally mature and have had six months of non-operative treatment.

The Ardis Interbody System is implanted using a posterior or transforaminal approach and is intended to be used singly or in pairs with supplemental fixation.

### Summary of Technological Characteristics:

Implant Design: The Ardis Interbody System implant is a hollow device with texture on two opposing convex sides, and is offered in various lengths, widths and heights.

Implant Placement: Ardis Interbody System implants are designed to be placed through a posterior or transforaminal approach and to address vertebrae in the lumbosacral region of the spine. The system contains implants of various sizes to accommodate different patient anatomy, and instruments for site preparation, trialing, and placement/extraction.

Materials, Implants and Instruments: Ardis Interbody System implants are made from (polyetheretherketone) PEEK-OPTIMA® (ASTM F2026). As PEEK-OPTIMA is radiolucent, radiographic markers are included in the distal and proximal ends of the PEEK implants. The markers consist of tantalum wires and beads (ASTM F560) that are press-fit into small holes in the implant.

Patient-contacting, reusable instruments for use with the Ardis implants are manufactured from stainless steel and may include an aluminum titanium nitride (AITiN) PVD and/or titanium nitride (TiN) PVD coating.

Sterility, Implants and Instruments: Ardis implants are gamma sterilized and provided to the end user in sterile packaging.

The associated instruments and perforated instrument cases for use with the Ardis implants are supplied non-sterile and must be sterilized by the healthcare facility prior to use.

#### Summary of Performance Testing:

A simulated surgical evaluation was conducted using a cadaver specimen to verify that the design of the Ardis Interbody System and associated instruments meet documented user needs and intended use. All instruments in the system were evaluated during the course of the procedure. The function, performance, and conformance to user needs and intended use were validated based on responses received from the participating healthcare professionals. All responses met the established acceptance criteria.

## Substantial Equivalence:

Zimmer Spine considers the subject Ardis Interbody System product performance to be substantially equivalent to its predicate device, Ardis Spacer (K073202) because there are no changes to the intended use, design, materials or function.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like figure with its wings spread, symbolizing protection and care. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the organization's name and national affiliation. The overall design is simple yet recognizable, representing the department's role in promoting the health and well-being of the nation.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

January 30, 2014

Zimmer Spine, Incorporated Ms. Michelle Lenz Regulatory Affairs Specialist 7375 Bush Lake Road Minneapolis, Minnesota 55439

Re: K133184

Trade/Device Name: Ardis® Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: October 8, 2013 Received: October 17, 2013

Dear Ms. Lenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Ms. Michelle Lenz

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

## Vincen Devlin -S

- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K133184

Device Name Ardis® Interbody System

## Indications for Use (Describe)

The Ardis Interbody System is indicated for use with autogenous bone graft as an intervertebral body fusion device at one or two contiguous levels in the lumbosacral region (1.2-S1) in the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients with previous non-fusion spinal surgery at involved level may be reated with the device. Patients should be skeletally mature and have had six months of non-operative treatment.

The Artis Interbody System is implanted using a posterior or transforminal approach and is intended to be used singly or in pairs with supplemental fixation.

Type of Use (Select one or both, as applicable)

[x] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

,一个 100 000 000 000 FOR FOR FOR FOR FOR FOR FOR FOR FOR F :

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# on E. Dmitrie

FORM FDA 3881 (1/14)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

PSC Poblishing Scrvices (301) 443-6740

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K133184](https://fda.innolitics.com/device/K133184)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
