IMMULITE 2000; TOTAL T3 CALIBRATION VERIFICATION MATERIAL, TOTAL T4 CALIBRATION VERIFICATION MATERIAL, TOTAL TBG CALIBRA
K133124 · Siemens Healthcare Diagnostics, Inc. · JJX · Oct 25, 2013 · Clinical Chemistry
Device Facts
Record ID
K133124
Device Name
IMMULITE 2000; TOTAL T3 CALIBRATION VERIFICATION MATERIAL, TOTAL T4 CALIBRATION VERIFICATION MATERIAL, TOTAL TBG CALIBRA
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Oct 25, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The IMMULITE® Total T3 Calibration Verification Material (CVM) is intended for in vitro diagnostic use in the verification of calibration of the IMMULITE Total T3 assay on the IMMULITE 2000 systems. The IMMULITE® Total T4 Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Total T4 assay on the IMMULITE 2000 systems. The IMMULITE® TBG Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE TBG assay on the IMMULITE 2000 systems. The IMMULITE® FSH Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE FSH assay on the IMMULITE 2000 systems. The IMMULITE® Estradiol Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Estradiol assay on the IMMULITE 2000 systems.
Device Story
The IMMULITE 2000 Calibration Verification Materials (CVMs) are multi-level, lyophilized, human or bovine serum-based control materials containing specific analytes (T3, T4, TBG, FSH, or Estradiol). They are used in clinical laboratory settings by technicians to verify the calibration and reportable range of corresponding IMMULITE 2000 assays. The CVMs are processed on the IMMULITE 2000 system as unknown samples. The system measures the analyte concentration, and the resulting values are compared against assigned target ranges to ensure the assay remains within established performance specifications. This verification process helps healthcare providers ensure the accuracy of patient test results, supporting reliable clinical decision-making for thyroid and reproductive hormone assessments.
Clinical Evidence
Bench testing only. Stability studies validated shelf life (3.5 to 5 years at -20°C). Value assignment was validated using commercially available controls and patient samples (n=25-30 per analyte) to confirm recovery within target ranges. No clinical trials were performed.
Technological Characteristics
Lyophilized human or bovine serum matrix with preservatives. Multi-level (4 levels). Storage at -20°C. Single-use. Traceable to internal reference calibrators or WHO standards (e.g., WHO 1st International Standard 88/638 for TBG, WHO 2nd IRP 78/549 for FSH).
Indications for Use
Indicated for in vitro diagnostic use to verify calibration of IMMULITE 2000 Total T3, Total T4, TBG, FSH, and Estradiol assays on IMMULITE 2000 systems. For prescription use only.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
K142811 — IMMULITE 2000 LH Calibration Verification Material (CVM), IMMULITE 2000 Free T3 Calibration Verification Material (CVM), IMMULITE 2000 Gastrin Calibration Verification Material (CVM) · Siemens Healthcare Diagnostics, Inc. · Oct 29, 2014
K140818 — IMMULITE: 2000 FREE T4 CALIBRATION VERIFICATION MATERIAL, 2000 IGF-I CALIBRATION VERIFICATION MATERIAL, 2000 PROLACTIN C · Siemens Healthcare Diagnostics, Inc. · May 1, 2014
K140258 — IMMULITE 2000 DHEA-SO4; THIRD GENERATION TSH AND INTACT PTH CALIBRATION VERIFICATION MATERIAL · Siemens Healthcare Diagnostics, Inc. · Mar 5, 2014
K140541 — IMMULITE 2000: C-PEPTIDE CALIBRATION VERIFICATION MATERIAL, SHBG CALIBRATION VERIFICATION MATERIAL, TOTAL TESTOSTERONE C · Siemens Healthcare Diagnostics, Inc. · Apr 1, 2014
{0}------------------------------------------------
# SIEMENS
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K133124
| 1. Submitter | | | |
|---------------------|-----------------------------------------------------------------------------------|--|--------------|
| Mailing Address: | Siemens Healthcare Diagnostics Inc.<br>511 Benedict Avenue<br>Tarrytown, NY 10591 | | |
| Contact Person: | Garo Mimaryan, MS, RAC<br>Technical Regulatory Affairs Specialist III | | OCT 2 5 2013 |
| Phone Number: | (914)-524-3270 | | |
| Fax Number: | (914)-524-2101 | | |
| E-mail Address: | garo.mimaryan@siemens.com | | |
| Date Prepared: | October 15, 2013 | | |
| 2. Device Name | | | |
| Proprietary Name: | IMMULITE® 2000 Total T3 Calibration Verification Material | | |
| Measurand: | Quality Control materials for IMMULITE® 2000 Total T3 assay | | |
| Type of Test: | Calibration Verification Material (CVM) for IMMULITE® 2000<br>Total T3 assay | | |
| Regulation Section: | 21 CFR 862.1660, Quality Control Material | | |
| Classification: | Class I Reserved | | |
| Products Code: | JJX - Single (Specified) Analyte Controls (Assayed and<br>Unassayed) | | |
| Panel: | Clinical Chemistry (75) | | |
3. Predicate Device Name Predicate 510(k) No:
4. Device Description:
- 5. Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Elecsys T3 CalCheck 5 K ! 1 1 552
The Total T3 Calibration Verification Material (CVM) contains one set of four vials, 3 mL each. LT3CVM1 contains T3-free human serum matrix with preservatives. LT3CVM2, LT3CVM3 and LT3CVM4 contain low, intermediate and high levels of T3 respectively, in human serum matrix with preservatives. CVMs are supplied frozen in lyophilized form.
See Indications for Use Statement below The IMMULITE® Total T3 Calibration Verification Material (CVM): For in vitro diagnostic use in the verification of calibration of the IMMULITE Total T3 assay on the IMMULITE 2000 systems
For prescription use only
IMMULITE® 2000 Systems
{1}------------------------------------------------
- 6. Technological Characteristics A comparison of the device features, intended use, and other and Substantial Equivalence information demonstrates that the IMMULITE® 2000 Total T3 Comparison with Predicate: Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
| Table 1: Substantial Equivalence Comparison |
|---------------------------------------------|
|---------------------------------------------|
| SIMILARITIES | | |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Total T3 CVM | Elecsys T3 CalCheck 5 |
| Intended<br>Use | The IMMULITE® Total T3 Calibration<br>Verification Material (CVM) is for in vitro<br>diagnostic use in the verification of<br>calibration of the IMMULITE Total T3<br>assay on the IMMULITE 2000 systems | The Elecsys T3 CalCheck 5 is an assayed<br>control for use in calibration verification<br>and for use in the verification of the<br>assay range established by the Elecsys<br>T3 reagent on the indicated Elecsys and<br>cobas e immunoassay analyzers. |
| Analyte | T3 | Same |
| Form | Lyophilized | Same |
| Matrix | Human Serum | Same |
| Stability | Stable unopened until the expiration date | Same |
| DIFFERENCES | | |
|-------------|----------------------------|-----------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Total T3 CVM | Elecsys T3 CalCheck 5 |
| Levels | 4 | 5 |
| Storage | -20°C | 2-8°C |
| Use | Single Use Only | Not for Single Use |
## 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Total T3 Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 5 years when stored frozen at -20°C prior to opening.
{2}------------------------------------------------
#### Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
:
#### Table 2: Stability Protocol Summary
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LT3CVM1 | 1 | 1092 | 1456 | 1820 |
| LT3CVM2 | 1 | 1092 | 1456 | 1820 |
| LT3CVM3 | 1 | 1092 | 1456 | 1820 |
| LT3CVM4 | 1 | 1092 | 1456 | 1820 |
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Total T3 Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of the guideline acceptance criteria which require dose value of stability CVM to fall between ±12% of assigned dose for CVM level 2, ±8% of assigned dose for CVM levels 3 and ±15% for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±12% for level 2 and ±8% for levels 3 and ±15% for CVM level 4 then additional data review is conducted using part 2 criteria. The acceptance criteria is summarized in Table 3.
| CVM<br>Level | Assigned Dose<br>(ng/dL) | Guideline Criteria<br>% difference to<br>assigned dose | Acceptable Dose<br>range (ng/dL) | Review Limits |
|--------------|--------------------------|--------------------------------------------------------|----------------------------------|-------------------------------|
| LT3CVM1 | 0.00 | not applicable | ≤40.00 | Controls are |
| LT3CVM2 | 101 | ±12 | 88.88 - 113.12 | within 2SD of |
| LT3CVM3 | 215 | ±8 | 197.8 - 232.2 | target from |
| LT3CVM4 | 610 | ±15 | 518.5 - 701.5 | stability<br>calibrator curve |
## Table 3 Acceptance criteria for stability of IMMULITE 2000 Total T3 CVM
Traceability:
The IMMULITE Total T3 CVMs are traceable to internal assigned reference calibrators prepared using Total T3 antigen stock solution and are traceable to internal material which is gravis protecally prepared.
{3}------------------------------------------------
#### Value Assignment:
The IMMULITE Total T3 CVMs are 4 level materials which are a subset of 6 level Total T3 calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Total T3 reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The assigned reference calibrators are prepared using Total T3 antigen stock and are traceable to internal material which hass been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run ore each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and control using the assigned CVM values. The controls must fall within their target and Two tri-level commercially available controls, and 25 patient samples (5 normal patients samples and 20 patient panel samples) are used to validate CVM value assignments.
## Expected Values/Target Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run, 6 different reagent kit lots and 8 IMMULITE 2000 systems. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The target values are provided in the IMMULITE® 2000 CVM Calibration verification Material lotspecific value card. The expected assay range is 40-600 ng/dL. The target values in Table 4 and be considered as guidelines.
| Analyte target levels | CVM Level | Target Mean (ng/dL) | SD | Guideline ±2SD Range (ng/dL) |
|-----------------------|----------------|---------------------|-------|------------------------------|
| | LT3CVM1 | 0.00 | - | ≤40.00 |
| | LT3CVM2 | 99.0 | 5.975 | 87.1 - 111 |
| | LT3CVM3 | 201 | 14 | 173 - 229 |
| | LT3CVM4 | 603 | 45 | 513 - 693 |
| Assay Range | 40 - 600 ng/dL | | | |
#### Table 4: Analyte Target Range Values
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, than a tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as quidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators .
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material .
## Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
{4}------------------------------------------------
#### 9. Conclusion:
The IMMULITE® 2000 Total T3 Calibration Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently Elecsys T3 CalCheck 5. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence. product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Total T3 Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{5}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K133124
- 1. Submitter Mailing Address:
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown. NY 10591
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
2. Device Name Proprietary Name: Measurand: Type of Test:
> Regulation Section: Classification: Products Code:
Panel:
- 3. Predicate Device Name Predicate 510(k) No:
4. Device Descrintion:
5. Intended Use: Indication for Use:
> Special Conditions for Use Statement(s): Special Instrument Requirements:
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101
garo.mimaryan@siemens.com October 15, 2013
IMMULITE® 2000 Total T4 Calibration Verification Material Quality Control materials for IMMULITE® 2000 Total T4 assay Calibration Verification Material (CVM) for IMMULITE® 2000 Total T4 assav 21 CFR 862.1660. Quality Control Material Class I Reserved JJX – Single (Specified) Analyte Controls (Assayed and Unassaved) Clinical Chemistry (75)
Elecsys T4 CalCheck 5 K112528
IMMULITE® 2000 Total T4 Calibration Verification. Material (CVM) contains one set of four vials, 3 mL each. LT4CVM1 contains T4-free human serum matrix with preservatives. LT4CVM2, LT4CVM3 and LT4CVM4 contain low, intermediate and high levels of T4 respectively, in human serum matrix with preservatives. CVMs are supplied frozen in lyophilized form.
See Indications for Use Statement below The IMMULITE® Total T4 Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Total T4 assay on the IMMULITE 2000 systems
For prescription use only
IMMULITE® 2000 Systems
{6}------------------------------------------------
- 6. Technological Characteristics A comparison of the device features, intended use, and other and Substantial Equivalence information demonstrates that the IMMULITE® 2000 Total T4 Comparison with Predicate: Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
| SIMILARITIES | | |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Total T4 CVM | Elecsys T4 CalCheck 5 |
| Intended<br>Use | The IMMULITE® Total T4 Calibration<br>Verification Material (CVM) is for in vitro<br>diagnostic use in the verification of<br>calibration of the IMMULITE Total T4<br>assay on the IMMULITE 2000 systems | The Elecsys T4 CalCheck 5 is an assayed<br>control for use in calibration verification<br>and for use in the verification of the<br>assay range established by the Elecsys<br>T4 reagent on the indicated<br>Elecsys and cobas e immunoassay<br>analyzers |
| Analyte | T4 | Same |
| Form | Lyophilized | Same |
| Matrix | Human Serum | Same |
| Stability | Stable unopened until the expiration date | Same |
Table 1: Substantial Equivalence Comparison
| DIFFERENCES | | |
|-------------|----------------------------|-----------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Total T4 CVM | Elecsys T4 CalCheck 5 |
| Levels | 4 | 5 |
| Storage | -20°C | 2-8°C |
| Use | Single Use Only | Not for Single Use |
### 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
## Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Total T4 Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 3.5 years when stored frozen at -20°C prior to opening.
{7}------------------------------------------------
#### Stability Protocol Summary:
The CVMs are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from reference calibrator curve and is summarized in Table 2.
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|-----|------|------|
| LT4CVM1 | 1 | 910 | 1092 | 1274 |
| LT4CVM2 | 1 | 910 | 1092 | 1274 |
| LT4CVM3 | 1 | 910 | 1092 | 1274 |
| LT4CVM4 | 1 | 910 | 1092 | 1274 |
#### Table 2: Stability Protocol Summary
#### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Total T4 Calibration Material (CVM) are in 2 parts. Part 1 consists of the guideline acceptance criteria which require dose value of stability CVM to fall between ±10% of assigned dose. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability callbrator curve. If the result is not within acceptable dose range of ±10% then additional data review is conducted using part 2 criteria. The acceptance criteria is summarized in Table 3.
#### Table 3 Acceptance criteria for stability of IMMULITE 2000 Total T4 CVM
| CVM<br>Level | Assigned Dose<br>(µg/dL) | Guideline<br>Criteria %<br>difference to<br>assigned dose | Acceptable Dose<br>range (µg/dL) | Review Limits |
|--------------|--------------------------|-----------------------------------------------------------|----------------------------------|-------------------------------|
| LT4CVM1 | 0.00 | not applicable | ≤1.00 | Controls are |
| LT4CVM2 | 3.82 | ±10 | 3.44 - 4.20 | within 2SD of |
| LT4CVM3 | 15.80 | ±10 | 14.22 - 17.38 | target from |
| LT4CVM4 | 25.00 | ±10 | 22.50 - 27.50 | stability<br>calibrator curve |
#### Traceability:
The IMMULITE Total T4 CVMs are traceable to internal assigned reference calibrators prepared using Total T4 antigen stock solution and are traceable to internal material which is gravimetrically prepared.
{8}------------------------------------------------
#### Value Assignment:
The IMMULITE® 2000 Total T4 CVMs are 4 level materials which are a subset of 6 level Total T4 calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Total T4 reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The assigned reference calibrators are prepared using Total T4 antigen stock and are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The controls must fall within their target ranges. Two tri-level commercially available controls, and 25 patient samples (5 normal patients samples and 20 patient panel samples) are used to validate CVM value assignments. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
## Expected Values/Target Values/Reference Range:
The CVMs are manufactured using qualified materials and measurement procedures. The TOTAL T4 CVMs were tested on 27 replicates in total comprised of 9 runs, 7 IMMULITE 2000 systems and 6 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 1.0 -24 ug/dL. The target values in Table 4 can be considered as guidelines.
| Analyte<br>target levels | CVM Level | Target Mean<br>(µg/dL) | SD | Guideline ±2SD Range<br>(µg/dL) |
|--------------------------|-----------|------------------------|-------|---------------------------------|
| | LT4CVM1 | 0.00 | - | 0.00<br>≤1.00 |
| | LT4CVM2 | 3.87 | 0.33 | 3.21<br>4.53 |
| | LT4CVM3 | 16.0 | 0.95 | 14.1<br>17.9 |
| | LT4CVM4 | 25.0 | 1.375 | 22.3<br>27.8 |
Table 4: Analyte Target Range Levels
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local. acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators .
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material
{9}------------------------------------------------
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
#### 9. Conclusion:
The Total T4 Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Elecsys T4 CalCheck 5. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device. The IMMULITE® 2000 Total T4 Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{10}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K133124
- 1. Submitter
Mailing Address:
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
- 2. Device Name Proprietary Name: Measurand: Type of Test:
Regulation Section: Classification: Products Code: Panel:
- 3. Predicate Device Name Predicate 510(k) No:
4. Device Description:
- 5. Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com October 15. 2013
IMMULITE® 2000 TBG Calibration Verification Material Quality Control materials for IMMULITE® 2000 TBG assay Calibration Verification Material (CVM) for IMMULITE® 2000 TBG assay 21 CFR 862.1660, Quality Control Material Class | Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)
ADVIA Centaur® Intact PTH Master Curve Material K020217
IMMULITE® 2000 TBG Calibration Verification Material (CVM) contains one set of four vials, 2.0 mL each. LTBCVM1 contains a bovine-based matrix with preservatives. LTBCVM2, LTBCVM3 and LTBCVM4 contain low, intermediate and high levels of TBG respectively, in human-based matrix with preservatives. CVMs are supplied frozen in lyophilized form.
## See Indications for Use Statement below
The IMMULITE® TBG Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of the IMMULITE TBG assay on the IMMULITE 2000 systems
For prescription use only
IMMULITE® 2000 Systems
{11}------------------------------------------------
# SIEMENS
- 6. Technological Characteristics A comparison of the device features, intended use, and other and Substantial Equivalence information demonstrates that the IMMULITE® 2000 TBG Comparison with Predicate: Calibration Verification Material (CVM) is substantially equivalent to the predicate device, as summarized in Table 1.
| Table 1: Substantial Equivalence Comparison<br>CARDE IN COLLECTION COLLEGION COLLEGION COLLEGION<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| SIMILARITIES | | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 TBG CVM | ADVIA Centaur Intact PTH MCM |
| Intended<br>Use | The IMMULITE® TBG Calibration<br>Verification Material (CVM) is for in vitro<br>diagnostic use in the verification of<br>calibration of the IMMULITE TBG assay<br>on the IMMULITE 2000 systems | For in vitro diagnostic use for evaluating<br>the ADVIA Centaur® Intact PTH assays.<br>This material is intended to be run singly<br>as unknown samples after a two-point<br>calibration has been performed on the<br>system |
| Form | Lyophilized | Same |
| Stability | Stable unopened until the expiration date | Same |
| Storage | -20°C | Same |
| Use | Single Use Only | Same |
| DIFFERENCES | | |
|-------------|-----------------------|---------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 TBG CVM | ADVIA Centaur Intact PTH Master<br>Curve Material (MCM) |
| Analyte | TBG | Intact PTH |
| Levels | 4 | 7 |
| Matrix | Human Serum | Bovine Serum |
## 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
## Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 TBG Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 5 years when stored frozen at -20℃ prior to opening.
{12}------------------------------------------------
#### Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
#### Table 2: Stability Protocol Summary
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LTBCVM1 | 1 | 1456 | 1638 | 1820 |
| LTBCVM2 | 1 | 1456 | 1638 | 1820 |
| LTBCVM3 | 1 | 1456 | 1638 | 1820 |
| LTBCVM4 | 1 | 1456 | 1638 | 1820 |
#### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE TBG Calibration Material (CVM) are in 2 parts. Part 1 consists of the guideline acceptance criteria which require dose value of stability CVM to fall between ±20% of assigned dose for CVM level 2, ±15% of assigned dose for CVM levels 3 and ±20% for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±20% for level 2 and ±15% for levels 3 and ±20% for CVM level 4 then additional data review is conducted using part 2 criteria. The acceptance criteria is summarized in Table 3.
| CVM<br>Level | Assigned Dose<br>(µg/mL) | Guideline Criteria %<br>difference to assigned<br>dose | Acceptable<br>Dose range<br>(µg/mL) | Review Limits |
|--------------|--------------------------|--------------------------------------------------------|-------------------------------------|----------------------|
| LTBCVM1 | 0.00 | not applicable | ≤3.50 | Controls are within |
| LTBCVM2 | 5.60 | ±20 | 4.48 - 6.72 | 2SD of target from |
| LTBCVM3 | 51.0 | ±15 | 43.35 - 58.65 | stability calibrator |
| LTBCVM4 | 99.5 | ±20 | 79.60-119.40 | curve |
## Table 3 Acceptance criteria for stability of IMMULITE_2000 TBG CVM
#### Traceability:
The IMMULITE TBG CVMs are traceable to WHO 1ST International Standard 88/638. The CVMs are manufactured using qualified materials and measurement procedures.
{13}------------------------------------------------
### Value Assignment:
The IMMULITE® 2000 TBG CVMs are 4 level materials which are a subset of 7 level TBG calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of TBG reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. Two levels of commercially available controls, and 30 patient samples (5 spiked normal patients samples and 25 patient panel samples) are used to validate CVM value assignments. Quality control is performed by calculating the recovery of patient samples. spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
#### Expected Values/Target Values/Reference Range:
The TBG CVMs were tested on 15 replicates in total comprised of 5 runs and 3 replicates per run, 4 IMMULITE 2000 systems and 4 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 3.5-80 ug/mL. The target values in Table 4 can be considered as guidelines.
| Analyte<br>target levels | CVM Level | Target Mean<br>µg/mL | SD | Guideline ±2SD Range<br>(µg/mL) |
|--------------------------|----------------------------------------|----------------------|-------|---------------------------------|
| | LTBCMV1 | 0.00 | - | 0.00<br>≤3.50 |
| | LTBCVM2 | 5.85 | 0.585 | 4.68<br>7.02 |
| | LTBCVM3 | 49.9 | 3.75 | 42.4<br>57.4 |
| | LTBCVM4 | 99.0 | - | - |
| | (85% of LTBCVM4 +<br>15% of LTB1CVM1)* | 84.2 | 8.4 | 67.4<br>101 |
| Assay Range | 3.5 - 80 µg/mL | | | |
#### Table 4: Analyte Target Range Levels
*Note: LTBCVM4 requires dilution to ensure that the target value is within the +10% of the top of the reportable range of the assay.
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision · tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators .
- Guidance for Industry and FDA Staff Assayed and Unassaved Quality Control Material .
{14}------------------------------------------------
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
## 9. Conclusion:
TBG Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed ADVIA Centaur Intact PTH MCM . The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 TBG Calibration Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{15}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K133124
- 1. Submitter Mailing Address:
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
- 2. Device Name Proprietary Name: Measurand: Type of Test:
Regulation Section: Classification: Products Code: Panel:
3. Predicate Device Name Predicate 510(k) No:
4. Device Description:
- 5. Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com October 15, 2013
IMMULITE® 2000 FSH Calibration Verification Material Quality Control materials for IMMULITE® 2000 FSH assay Calibration Verification Material (CVM) for IMMULITE® 2000 FSH assay 21 CFR 862.1660, Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)
Elecsvs FSH CalCheck K964829
IMMULITE® 2000 FSH Calibration Verification Material (CVM) contains one set of four vials, 3 mL each. LFSCVM1 contains a bovine serum with preservatives. LFSCVM2. LFSCVM3 and LFSCVM4, 2 mL each, contain low, intermediate and high levels of human source FSH respectively, in bovine serum matrix with preservatives. CVMs are supplied frozen in lyophilized form.
See Indications for Use Statement below
The IMMULITE® FSH Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of the IMMULITE FSH assay on the IMMULITE 2000 systems
For prescription use only IMMULITE® 2000 Systems
{16}------------------------------------------------
| 6. Technological | A comparison of the device features, intended use, and other |
|-----------------------------|---------------------------------------------------------------|
| Characteristics | information demonstrates that the IMMULITE® 2000 FSH |
| and Substantial Equivalence | Calibration Verification Material (CVM) is substantially |
| Comparison with Predicate: | equivalent to the predicate device, as summarized in Table 1. |
Table 1: Substantial Equivalence Comparison
| SIMILARITIES | | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 FSH CVM | Elecsys FSH CalCheck |
| Intended<br>Use | The IMMULITE® FSH Calibration<br>Verification Material (CVM) is for in<br>vitro diagnostic use in the verification of<br>calibration of the IMMULITE FSH assay<br>on the IMMULITE 2000 systems | For use in the verification of the<br>calibration established by the Elecsys<br>FSH reagent on the indicated Elecsys<br>and cobas e immunoassay analyzers. |
| Analyte | FSH | Same |
| Form | Lyophilized | Same |
| Stability | Stable unopened until the expiration date | Same |
#### DIFFERENCES
| | Candidate Device | Predicate Device |
|---------|-----------------------|----------------------|
| | IMMULITE 2000 FSH CVM | Elecsys FSH CalCheck |
| Levels | 4 | 3 |
| Matrix | Bovine Serum | Human Serum |
| Storage | -20°C | 2-8°C |
| Use | Single Use Only | Not for Single Use |
## 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 TBG Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 5 years when stored frozen at -20°C prior to opening.
-
{17}------------------------------------------------
## Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
## Table 2: Stability Protocol Summary
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LFSCVM1 | 1 | 1460 | 1642 | 1825 |
| LFSCVM2 | 1 | 1460 | 1642 | 1825 |
| LFSCVM3 | 1 | 1460 | 1642 | 1825 |
| LFSCVM4 | 1 | 1460 | 1642 | 1825 |
### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE FSH Calibration Material (CVM) are in 2 parts. Part 1 consists of the Guideline acceptance criteria which require dose value of stability CVM to fall between ±10% of assigned dose. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±10% then additional data review is conducted using part 2 criteria. The acceptance criteria is summarized in Table 3.
## Table 3 Acceptance criteria for stability of IMMULITE 2000 FSH CVM
| CVM Level | Assigned Dose (mIU/mL) | Guideline Criteria % difference to assigned dose | Acceptable dose range (mIU/mL) | Review Limits |
|-----------|------------------------|--------------------------------------------------|----------------------------------|----------------------------|
| LFSCVM1 | 0 | not applicable | $ \le $ 0.10 | |
| LFSCVM2 | 4.64 | $ \pm $ 10 | 4.18 - 5.10 | Controls are within 2SD of |
| LFSCVM3 | 96 | $ \pm $ 10 | 86.40 - 105.60 | target on each |
| LFSCVM4 | 174 | $ \pm $ 10 | 156.60 - 191.40 | curve |
#### Traceability:
The IMMULITE FSH CVMs are traceable to WHO 2nd IRP (78/549, interim replacement code 94/632).
{18}------------------------------------------------
#### Value Assignment:
The IMMULITE® 2000 FSH CVMs are 4 level materials which are a subset of 9 level FSH calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of FSH reagents and two point adjustors. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. Six levels of commercially available controls, and 30 patient samples were used to validate CVM value assignments. Quality control is performed by calculating the recovery of patient samples spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
#### Expected Values/Target Values/Reference Range:
The FSH CVMs were tested on 15 replicates in total comprised of 5 runs and 3 replicates per run. 5 IMMULITE 2000 systems and 3 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 0.1 -170 mlU/mL. The target values in Table 4 can be considered as guidelines.
| Analyte<br>target<br>levels | CVM Level | Target Mean<br>(mIU/mL) | SD | Guideline ±2SD Range<br>(mIU/mL) |
|-----------------------------|------------------|-------------------------|-------|----------------------------------|
| | LFSCVM1 | 0.00 | - | 0.00<br>≤0.10 |
| | LFSCVM2 | 4.59 | 0.23 | 4.13<br>5.05 |
| | LFSCVM3 | 98 | 7.925 | 82.3<br>114 |
| | LFSCVM4 | 174 | 14 | 146<br>202 |
| Assay<br>Range | 0.1 - 170 mIU/mL | | | |
#### Table 4: Analyte Target Range Levels
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assigninent is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents �
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators .
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material .
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
্ৰ
{19}------------------------------------------------
#### ﻢ. Conclusion:
The FSH Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Elecsys FSH CalCheck. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 FSH Calibration Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{20}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
#### The assigned 510(k) Number: K133124
- 1. Submitter Mailing Address:
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
- 2. Device Name Proprietary Name: Measurand: Type of Test:
Regulation Section: Classification: Products Code: Panel:
3. Predicate Device Name Predicate 510(k) No:
- 4. Device Description:
- 5. Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com October 15. 2013
IMMULITE® 2000 Estradio1 Calibration Verification Material Quality Control materials for IMMULITE® 2000 Estradiol assay Calibration Verification Material (CVM) for IMMULITE® 2000 Estradiol assay 21 CFR 862.1660. Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)
ADVIA Centaur® Enhanced Estradiol (eE2) MCM K102904
IMMULITE® 2000 Estradiol Calibration Verification Material (CVM) contains one set of four vials. 2 mL each. LE2CVM1 contains human serum with preservatives. LE2CVM2. LE2CVM3 and LE2CVM4 contain low, intermediate and high levels of human source Estradiol respectively, in human serum matrix with preservatives. CVMs are supplied frozen in lyophilized form.
See Indications for Use Statement below
The IMMULITE® Estradiol Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of the IMMULITE Estradiol assay on the IMMULITE 2000 systems
For prescription use only IMMULITE® 2000 Systems
{21}------------------------------------------------
## 6. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Estradiol Calibration Verification Material (CVM) is substantially equivalent to the predicate device, as summarized in Table 1.
Table 1: Substantial Equivalence Comparison
| SIMILARITIES | | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Estradiol CVM | The ADVIA Centaur® Enhanced<br>Estradiol (eE2) MCM |
| Intended<br>Use | The IMMULITE® Estradiol Calibration<br>Verification Material (CVM) is for in vitro<br>diagnostic use in the verification of<br>calibration of the IMMULITE Estradiol<br>assay on the IMMULITE 2000 systems | The ADVIA Centaur® Enhanced<br>Estradiol (eE2) Master Curve Material is<br>for in vitro diagnostic<br>use in the verification of calibration and<br>reportable range of the ADVIA Centaur<br>Enhanced Estradiol assay. |
| Analyte | Estradiol. | Same |
| Form | Lyophilized | Same |
| Matrix | Human Serum | Same |
| Stability | Stable unopened until the expiration date | Same |
| Use | Single Use Only | Same |
#### DIFFERENCES
| | Candidate Device | Predicate Device |
|---------|-----------------------------|----------------------------------------------------|
| | IMMULITE 2000 Estradiol CVM | The ADVIA Centaur® Enhanced<br>Estradiol (eE2) MCM |
| Levels | 4 | 6 |
| Storage | -20°C | 2-8°C |
## 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Estradiol Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The CVM. The Clibration Verification Materials are stable up to 5 years when stored frozen at -20℃ prior to opening.
{22}------------------------------------------------
## Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
#### Table 2: Stability Protocol Summarv
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LE2CVM1 | 1 | 1460 | 1642 | 1825 |
| LE2CVM2 | 1 | 1460 | 1642 | 1825 |
| LE2CVM3 | 1 | 1460 | 1642 | 1825 |
| LE2CVM4 | 1 | 1460 | 1642 | 1825 |
## Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Estradiol Calibration Verification Material (CVM) are in 2 parts. Part I consists of the guideline acceptance criteria which require dose value of stability CVM to fall between ±14% of assigned dose for CVM level 2, ±10% of assigned dose for CVM levels 3 and ±8% for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±14% for level 2 and ±10% for levels 3 and ±8% for Cr VM level 4 then additional data review is conducted using part 2 criteria. The acceptance criteria is summarized in Table 3.
| CVM level | Assigned<br>Dose<br>(pg/mL) | Guideline Criteria<br>% difference to assigned<br>dose | Acceptable Dose<br>range (pg/mL) | Review Limits |
|-----------|-----------------------------|--------------------------------------------------------|----------------------------------|----------------|
| LE2CVM1 | 0.00 | not applicable | ≤20.00 | Controls are |
| LE2CVM2 | 52 | ±14% | 44.72 - 59.28 | within 2SD of |
| LE2CVM3 | 502 | ±10% | 451.80 - 552.20 | target on each |
| LE2CVM4 | 2323 | ±8% | 2137.16 - 2508.84 | curve |
## Table 3 Acceptance criteria for stability of IMMULITE 2000 Estradiol CVM
#### Traceability:
The IMMULITE Estradiol CVMs are traceable to internal assigned reference calibrators prepared using an Estradiol antigen stock solution and are traceable to internal material which is gravimetrically prepared.
{23}------------------------------------------------
### Value Assignment:
The IMMULITE Estradiol CVMs are 4 level materials which are a subset of 7 level Estradiol calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Estradiol reagents and two point adjustors. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. Four levels of commercially available controls, and 30 patient samples (5 normal patients samples and 25 spiked samples) are used to validate CVM value assignments. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
## Expected Values/Target Values/Reference Range:
The FSH CVMs were tested on 15 replicates in total comprised of 5 runs and 3 replicates per run, 5 IMMULITE 2000 systems and 3 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 20 -2000 pg/mL. The target values in Table 4 can be considered as guidelines.
| Analyte<br>target<br>levels | CVM Level | Target Mean<br>(pg/mL) | SD | Guideline ±2SD Range<br>(pg/mL) |
|-----------------------------|-----------------|------------------------|-------|---------------------------------|
| | LE2CVM1 | 0.00 | - | ≤20.00 |
| | LE2CVM2 | 61.5 | 9.85 | 81.20 |
| | LE2CVM3 | 478 | 33.5 | 545 |
| | LE2CVM4 | 2034 | 142.5 | 2319 |
| Assay<br>Range | 20 - 2000 pg/mL | | | |
### Table 4: Analyte Target Range Levels
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total recision tabulated in the respective assay instructions for use may be considered as one factor when establishing local. acceptable ranges. The values provided above may be considered as guidelines, Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators .
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material .
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
{24}------------------------------------------------
#### 9. Conclusion:
The Estradiol Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent the currently marketed The ADVIA Centaur® Enhanced Estradiol (eE2) MCM. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Estradiol Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{25}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/25/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping waves or curves. The logo is black and white and appears to be a scanned image.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 25, 2013
SIEMENS HEALTHCARE DIAGNOSTICS INC. c/o Garo Mimaryan 511 Benedict Ave. TARRYTOWN NY 10591
Re: K133124
Trade/Device Name: IMMULITE® 2000 Total T3 Calibration Verification Material IMMULITE® 2000 Total T4 Calibration Verification Material IMMULITE® 2000 TBG Calibration Verification Material IMMULITE® 2000 FSH Calibration Verification Material IMMULITE® 2000 Estradiol Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I, reserved Product Code: JJX Dated: September 27, 2013 Received: September 30, 2013
Dear Garo Mimaryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{26}------------------------------------------------
Page 2-Garo Mimaryan
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFF) Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{27}----------------------------------------------…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.