COMPUTED RADIOGRAPHY SCANNER

K133120 · 3Disc Americas · MQB · Jan 15, 2014 · Radiology

Device Facts

Record IDK133120
Device NameCOMPUTED RADIOGRAPHY SCANNER
Applicant3Disc Americas
Product CodeMQB · Radiology
Decision DateJan 15, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2

Indications for Use

This device is a Computed Radiography System and intended for use in producing digital X-Ray images for general radiography purposes. It comprises of scanner, cassette with reusable phosphor storage plate (IP) and workstation software. It scans X-Ray exposed image plate and produces X-Ray image in digital form. Then, digital image is transferred to workstation for further processing and routing. This device is intended to be operated in a radiological environment by qualified staff. This device is not intended for the acquisition of mammographic image data.

Device Story

FireCR Spark is a computed radiography (CR) scanner; utilizes reusable X-ray storage phosphor plates (IP) to capture latent X-ray images. Device scans exposed IP using laser; releases latent image as light; photomultiplier tube (PMT) collects and converts light into digital image data. System performs signal processing including digital filtering, gain/offset correction, and flat fielding. Output viewed on workstation; stored locally, printed, or sent to PACS. Operated by qualified staff in radiological environments. Modular design improves reliability; multi-layer fiber bundle enhances DQE and MTF performance. Provides digital diagnostic images for general radiography; facilitates clinical decision-making through rapid, high-resolution image acquisition.

Clinical Evidence

Clinical study conducted to evaluate image quality; radiologists compared images produced by FireCR Spark against the predicate device. Results indicated equivalent image quality. Bench testing performed per FDA guidance for solid-state X-ray imaging devices, including electrical, mechanical, and environmental safety (IEC 60601-1) and EMC (IEC 60601-1-2).

Technological Characteristics

Computed radiography scanner; utilizes reusable phosphor storage plates. Features high-sensitivity photomultiplier tube (PMT) and multi-layer fiber bundle. Resolution: 100µm and 200µm. Connectivity: DICOM 3.0 compliant. Power: 100-250VAC, 50/60Hz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62220-1.

Indications for Use

Indicated for general radiography imaging in patients requiring diagnostic X-ray exams. Not indicated for mammography. Intended for use by qualified staff in radiological environments.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Page 1 of 4 JAN 15 2014 ## 510(k) Summary K133120 ## This 510(k) summary information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date prepared: December 12, 2013 1. Company and Correspondent making the submission: Name - 3D Imaging & Simulations Corp. Address - 815. Tamnip-Dong, Yuseong-Gu, Daejeon, Korea Telephone - +82-42-931-2100 Fax - +82-42-931-2299 Contact - Jiin Jung / Vice President E-mail - jiinjung@3-disc.com - 2. Device : | Trade/proprietary name: | FireCR Spark | |-------------------------|------------------------------| | Common Name: | Computed Radiography Scanner | | Classification Name: | Stationary x-ray system | - 3. Predicate Device : | Manufacturer: | 3D Imaging & Simulations Corp. | |----------------|----------------------------------------| | Device: | FireCR | | 510(k) Number: | K102619 (Decision Date – Mar. 1. 2011) | ## 4. Classifications Names & Citations : 21CFR 892.1680, MQB - Stationary x-ray system, Class 2 ## 5. Description : - 5.1 General The FireCR Spark is Computed Radiography System which produces the X-ray diagnostic image in digital format instead of using traditional screens and film. This device utilizes reusable X-ray storage phosphor plate (IP) that is sensitive to X-ray and stores latent image when it is exposed to X-ray. After X-ray exposure to the X-ray storage phosphor plate, X-ray storage phosphor plate is scanned by means of laser in the device. Latent image in the X-ray storage phosphor plate is released in a form of light by laser scanning. Then the light is collect and converted into a form of digital image. The signal processing is made to the digital image data such as the digital filtering, the gain & offset correction and flat fielding. The image can then be viewed on a computer workstation, adjusted if necessary, then stored locally, sent to an archive, printed or sent to PACS system. After acquisition of latent image from the X-ray storage phosphor plate, it is erased thoroughly to be reused. #### 5.2 Main Features Modern Scanning Mechanism FireCR Spark adopts an updated scanning mechanism constructed in a compact and rigid structure. High Throughput: Its unique scanning mechanism enables to improve the efficiency and high throughout. Scanning Resolution: User selectable resolution of 100um and 200um allows user to make diagnosis on variable purposes. Detector High sensitivity photomultiplier tube supplied in the FireCR Spark delivers high gain, wide dynamic range and high speed response for radiographic imaging. Acquisition and Diagnostic Software: QuantorMed Plus Acquisition and Diagnostic Software's accurate and rapid data processing make the scanner powerful. ### 5.3 Product features - Photomultiplier Tube (PMT) - 35cm x 43cm, 24cm x 30cm, 18cm x 24cm imaging area. - Wide dynamic range with 16-bit digitization {2}------------------------------------------------ - Image process parameters are selectable according to the body part to make best images for diagnosis - DICOM3.0 standard compliance - User Selectable Scanning Resolution : 100µm and 200µm - 6. Indications for use : This device is a Computed Radiography System and intended for use in producing digital X-Ray images for general radiography purposes. It comprises of scanner, cassette with reusable phosphor storage plate (IP) and workstation software. It scans X-Ray exposed image plate and produces X-Ray image in digital form. Then, digital image is transferred to workstation for further processing and routing. This device is intended to be operated in a radiological environment by qualified staff. This device is not intended for the acquisition of mammographic image data. ## 7. Comparison with predicate device : 3D Imaging & Simulations Corp. believes that the FireCR Spark is substantially equivalent to FireCR. | | | FireCR<br>3D Imaging & Simulations Corp. | FireCR Spark<br>3D Imaging & Simulations Corp. | |-----------------------------|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) number | | K102619 | K133120 | | Intended Use | | The FireCR imaging system is<br>indicated for capturing, digitization<br>and processing of general<br>radiography images stored in<br>imaging plate recording media. | The FireCR Spark imaging system is<br>indicated for capturing, digitization and<br>processing of general radiography<br>images stored in imaging plate recording<br>media. | | Physical<br>Characteristics | Overall<br>Dimensions | Reader<br>464 x 703 x 117mm | Reader<br>445 x 795 x 134mm | | | Imaging Area | 14" x 17" (35cm x 43cm)<br>10" x 12" (25cm x 30cm) | 35cm x 43cm SAME<br>24cm x 30cm (SLIGHT DIFFERENCE)<br>18cm x 24cm (NEW SIZE) | | | Effective Pixel<br>Pitch | 100µm, 200µm | 100μm, 200μm | | | Spatial<br>Resolution | 3.7lp/mm @ 100um | 3.7lp/mm @ 100um | | | Image Matrix<br>(Pixel) | 35cm x 43cm<br>3500 x 4300 @ 100um<br>1750 x 2150 @ 200um<br>25cm x 30cm<br>2500 x 3000 @ 100um<br>1250 x 1500 @ 200um | 35cm x 43cm SAME:<br>3500 x 4300 @ 100um SAME<br>1750 x 2150 @ 200um SAME<br>24cm x 30cm (SLIGHT DIFFERENCE):<br>2400x3000 @ 100um (SLIGHT<br>DIFFERENCE)<br>1200 x 1500 @ 200um (SLIGHT<br>DIFFERENCE)<br>18cm x 24cm (NEW SIZE):<br>1800 x 2400 @ 100um<br>900 x 1200@ 200um | | | Weight | 30kg | 21.5kg | | | Imaging Device | High Sensitivity Photo Multiplier<br>Tube (s-PMT) | High Sensitivity Photo Multiplier Tube (s-<br>PMT) | {3}------------------------------------------------ | | | FireCR<br>3D Imaging & Simulations Corp. | FireCR Spark<br>3D Imaging & Simulations Corp. | |--------------------------------|------------------------|----------------------------------------------------------|----------------------------------------------------------| | Photo | | Image: FireCR | Image: FireCR Spark | | Operational<br>Characteristics | Operating<br>Condition | Temperature : 0-40°C<br>Humidity: 15%-95% RH | Temperature : 0-40°C<br>Humidity: 15%-95% RH | | | Power<br>Requirements | 100-250VAC +/- 10%,<br>50/60Hz | 100-250VAC +/- 10%,<br>50/60Hz | | | Methods of<br>Exposure | Register Patient → X-ray Exposure | Register Patient → X-ray Exposure | | | X-ray<br>Absorber | Imaging plate | Imaging plate | | Functional<br>Characteristics | Output Data | Dicom3.0 Compatible | Dicom3.0 Compatible | | | DQE | 23.5% @ 0.5 lp/mm | 25% @ 0.5 lp/mm | | | MTF | 79% @ 0.5 lp/mm | 80% @ 0.5 lp/mm | | | Defect<br>Compensation | By Calibration | By Calibration | | | Dynamic<br>Range | 16bit | 16 bit | | | Image<br>Processing | Image processing parameter is<br>selectable by body part | Image processing parameter is<br>selectable by body part | | DICOM Compatibility | | DICOM 3.0 Compliant | DICOM 3.0 Compliant | | Standards | | IEC 60601-1; IEC 60601-1-2<br>IEC 62220-1 | SAME | The FireCR Spark's imaging principle, physical characteristics, target population and intended use are the same as those of FireCR. However, the differences in the design are follows: - The technical specification (including DQE, MTF), mechanical structure and physical appearance of the FireCR Spark is a little different from the FireCR. - The testing of the FireCR Spark demonstrates that the performance is substantially equivalent to the predicate devices cited above. In clinical considerations, - The rating was considered equivalent by radiologists. - As a result of Clinical Study, FireCR Spark is considered that Image quality is equivalent to the Predicate Device. Summary of comparison: The FireCR Spark described in this 510(k) has the same intended use and similar technical characteristics to the FireCR. The similarities and differences between these systems are described in the table shown above. The similarities are as follows: Intended use with FireCR; Capturing image, Image Processing, and DICOM compatible features with FireCR {4}------------------------------------------------ K133120 Page 4 of 4 X-ray exposing technique with FireCR: Effective Pixel Pitch with FireCR: Spatial resolution with FireCR: X-ray absorber material with FireCR; Image making process with FireCR; Energy Sources, Source to skin distance with FireCR; Workstation and operating system with FireCR #### Differences are as follows: (1) Difference in Mechanical Structure and Physical Appearance (Now the structure is more modular) Mechanical structure and physical appearance of FireCR Spark is different from those of FireCR. (2) Difference in DQE and MTF: FireCR Spark shows slightly better DQE and MTF compared to the FireCR. This is due to the uses of a multi layer fiber bundle. Image /page/3/Figure/4 description: The image shows a close-up of a layered material. The top layer appears to be composed of many small, vertical fibers or strands. Below this is a solid, dark layer, and at the bottom, there is a row of small, white, spherical objects embedded within another dark layer. Single Layer Fiber Bundle Image /page/3/Picture/6 description: The image is a close-up shot of a dark, possibly black, surface. A thin, white line runs horizontally across the lower portion of the image, creating a stark contrast against the dark background. The texture of the surface appears smooth, with subtle variations in tone that suggest a slight sheen or reflection. The composition is simple, focusing on the interplay between the dark surface and the bright, linear element. Multi Layer Fiber Bundle (3) A new panel size is now available, allowing the user greater imaging flexibility (see table above). In summary, The FireCR Spark does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate device. ## 8. Safety, EMC, Biocompatibility and Performance Data : Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2(2001). Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory. #### 9. Conclusions : In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification 3D Imaging & Simulations Corp. concludes that the Computed Radiography Reader System (FireCR Spark) is safe and effective and substantially equivalent to predicate device as described herein. These modifications pose no risk to safety and effectiveness because they enhance reliability (modular construction) and they provide slightly better performance characteristics (multi layer fiber bundle) rendering the modifications substantially equivalent to our predicate device. There is also an added imaging plate size which gives the user a greater choice in their imaging needs (18cm x 24cm NEW SIZE, 1800 x 2400 @ 100um and 900 x 1200@ 200um. {5}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image contains the logo for the U.S. Department of Health and Human Services. The logo features a stylized emblem with an abstract human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 15, 2014 3DISC Americas % Mr. Daniel Kamm. P.E. Kamm & Associates 8870 Ravello Court NAPLES FL 34114 Re: K133120 Trade/Device Name: Fire CR Flash Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: December 12, 2013 Received: December 16, 2013 ## Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ Page 2-Mr. Kamm : :. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Smith Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health for Enclosure {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): K133120 Device Name: Computed Radiography Scanner / FireCR Flash Indications for Use: 1. 1. 1. 1. 1. : : : : : ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ This device is a Computed Radiography System and intended for use in producing digital X-Ray images for general radiography purposes. It comprises of scanner, cassette with reusable phosphor storage plate (IP) and workstation software. It scans X-Ray exposed image plate and produces X-Ray image in digital form. Then, digital image is transferred to workstation for further processing and routing. This device is intended to be operated in a radiological environment by qualified staff. This device is not intended for the acquisition of mammographic image data. Prescription Use X AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Image /page/6/Picture/12 description: The image contains a stylized signature or logo. The signature is illegible, but it appears to be written in cursive. The signature is surrounded by a geometric pattern of dashed lines, which creates a border around the signature. The overall impression is that of a stylized and abstract design. (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health K133120 510(k) Page 1 of 1 Page 17 of 594
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...