K133086 · Zimmer Spine, Inc. · NKB · Nov 27, 2013 · Orthopedic
Device Facts
Record ID
K133086
Device Name
TITLE 2 POLYAXIAL SPINAL SYSTEM
Applicant
Zimmer Spine, Inc.
Product Code
NKB · Orthopedic
Decision Date
Nov 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TiTLE 2 Polyaxial Spinal System is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the thoracic, lumbar and sacral spine. The TiTLE 2 Polyaxial Spinal System is a Pedicle Screw System intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis). The TiTLE 2 Polyaxial Spinal System is also indicated for pedicle screw fixation for severe spondylolisthesis (grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1. In addition, TiTLE 2 Polyaxial System, when not used with pedicle screws, is indicated for sacral screw fixation from T1 to the illium sacrum. The non-pedicle screw indications are spondylolisthesis, degeneralive disc disease, (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis, and kyphosis), tumor, fracture, and previous failed fusion surgery. The TiTLE 2 Polyaxial System can also be linked to the Minit® Posterior Cervical and Upper Thoracic Fixation System.
Device Story
TiTLE® 2 Polyaxial Spinal System; rod and top-loading screw fixation system; temporary construct for spinal stabilization during posterior fusion procedures; implants include screws, rods, and connection components; surgical placement via open or endoscopic techniques; system removed after solid fusion; links to Minit® Posterior Cervical and Upper Thoracic Fixation System; aids surgical correction of spinal instabilities/deformities; benefits patient by providing immobilization to assist normal healing.
Clinical Evidence
Bench testing only. Performed static axial compression, static torsion, axial compression fatigue (ASTM F1717-12), and static/dynamic gripping/flexion-extension testing (ASTM F1798-97). Cadaver lab testing conducted to evaluate human factors regarding instrument and labeling design changes. Biocompatibility and sterilization/dry time testing confirmed equivalence.
Technological Characteristics
Rod and top-loading screw fixation system. Implants: Titanium alloy (Ti-6AI-4V ELI) per ASTM F136. Instruments: Stainless Steel per ASTM A564 and ASTM F899. Non-sterile, single-use. Polyaxial design.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Treats degenerative disc disease, degenerative spondylolisthesis with neurologic impairment, severe spondylolisthesis (grades 3-4 at L5-S1), fractures, dislocations, scoliosis, kyphosis, lordosis, spinal tumors, and pseudoarthrosis. Applicable to thoracic, lumbar, and sacral spine segments.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Minit® Posterior Cervical and Upper Thoracic Fixation System
Submission Summary (Full Text)
{0}------------------------------------------------
## Zimmer Spine - 510(k) - TiTLE® 2 Polyaxial Spinal System - Appendix A
Image /page/0/Picture/2 description: This document is a 510(k) Summary for the TITLE® 2 Polyaxial Spinal System. It includes the date of summary preparation as September 27, 2013, and the submitter as Zimmer Spine, Inc. located in Minneapolis, MN. The document also lists the establishment registration number, primary and secondary company contacts, trade name, device name, device classification as Class III, product codes, regulation numbers, and a description of the regulation.
legally marketed predicate devices:
| TITLE® 2 Polyaxial Spinal System<br>Predicate Device Name | Submission ID<br>Number | Clearance Date |
|-----------------------------------------------------------|-------------------------|-------------------|
| TITLE® 2 Polyaxial Spinal System | K073510 | February 11, 2008 |
| TITLE® 2 Polyaxial Spinal System | K072840 | December 7, 2007 |
| TITLE® 2 Polyaxial Spinal System | K070367 | April 12, 2007 |
| TITLE® 2 Polyaxial Spinal System | K060990 | April 26, 2006 |
| TITLE® 2 Polyaxial Spinal System | K041808 | October 1, 2004 |
{1}------------------------------------------------
## General Device Description:
The TiTLE® 2 Polyaxial Spinal System Implants are intended to be used as a temporary construct that assists in normal healing and are not intended to replace normal body structures. The system is intended to stabilize the spinal operative site during posterior fusion procedures, attaching to the spine by means of screws joined with spinal rods and should be removed after fusion.
The TiTLE® 2 Polyaxial Spinal System is designed to aid in the surgical correction of several types of spinal conditions, as stated in the section below. The TiTLE® 2 Polyaxial Spinal System consists of screws, rods, connection components that are intended to be used as a temporary construct that assists in normal healing and are not intended to replace normal body structures. The system is intended to stabilize the spinal operative site during posterior fusion procedures, attaching to the spine by means of screws joined with spinal rods and should be removed after fusion. The polyaxial spinal system is a rod and top-loading screw fixation system. This system can be placed surgically using either open or endoscopic surgical techniques. The system is provided non-sterile, for single use only.
The TiTLE® 2 Polyaxial Spinal System can also be linked to the Minit® Posterior Cervical and Upper Thoracic Fixation System.
#### Indications for Use:
The TiTLE 2 Polyaxial Spinal System is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). Levels of fixation are for the Thoracic, Lumbar and Sacral spine.
The TiTLE 2 Polyaxial Spinal System is a Pedicle Screw System intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The TiTLE 2 Polyaxial Spinal System is also indicated for pedicle screw fixation for severe spondylolisthesis (grades 3 and 4) at L5-S1, in skefetally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1.
In addition, TiTLE 2 Polyaxial System, when not used with pedicle screws, is indicated for sacral screw fixation from T1 to the illum sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease, (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, and kyphosis), tumor, fracture, and previous failed fusion surgery.
The TiTLE 2 Polyaxial System can also be linked to the Minit® Posterior Cervical and Upper Thoracic Fixation System.
### Summary of Technological Characteristics:
The subject TiTLE® 2 Polyaxial Spinal System shares the same technological characteristics as its predicate device, TiTLE® 2 Polyaxial Spinal System. The characteristics include the same design, same materials, same range of sizes and, substantially equivalent performance characteristics and intended use.
The subject and predicate TiTLE® 2 Polyaxial Spinal System both consist of screws, rods, connection components and the instruments necessary to implant the spinal system. All implant components are made from medical grade titanium alloy (Ti-6AI-4V ELI) that meets ASTM F136. The delivery instruments are manufactured from Stainless Steel that meets ASTM A564 and ASTM F899.
{2}------------------------------------------------
The TiTLE® 2 Polyaxial Spinal System is designed to aid in the surgical correction of several types of spinal conditions, as stated in the section above. The TiTLE® 2 Polyaxial Spinal System consists of screws, rods, connection components that are intended to be used as a temporary construct that assists in normal healing and are not intended to replace normal body structures. The system is intended to stabilize the spinal operative site during posterior fusion procedures, attaching to the spine by means of screws joined with spinal rods and should be removed after fusion. The oolvaxial system is a rod and top-loading screw fixation system. This system can be placed surgically using either open or endoscopic surgical techniques. The subject and predicate systems are provided non-sterile, are for single use only.
## Summary of Performance Testing:
The TiTLE® 2 Polyaxial Spinal System is substantially equivalent to the predicate devices in design, materials, function and intended use.
The performance testing included components of the subject TiTLE® 2 Polyaxial Spinal System, which were reviewed and tested appropriately for design verification, design validation, biocompatibility and sterilization. The test results conclude the subject TiTLE® 2 Polyaxial Spinal System to be substantially equivalent to its predicate device, TiTLE® 2 Polyaxial Spinal System.
- Bench testing (Static Axial Compression (Bending), Static Torsion, and Axial . Compression (Bending) Fatique per ASTM F1717-12; and Static Axial Gripping Capacity, Static Torsion Gripping Capacity, Static Flexion-Extension Moment, and Dynamic Flexion-Extension Moment per ASTM F1798-97 (Reapproved 2008)) for implants, screws, rods, and connection component, confirmed the product performance of the subject TiTLE® 2 Polvaxial Spinal System is suitable for its intended use.
- Cadaver lab testing of the subject TiTLE® 2 Polyaxial Spinal System to evaluate human . factors regarding the combination of instrument design changes and labeling design changes, as well as interaction with implants to confirm the substantial equivalence of the changes compared to the identified predicate devices.
- . Biocompatibility testing ensured the subject TiTLE® 2 Polyaxial Spinal System materials are biocompatible after manufacturing based on the minor design changes made in comparison to the predicate devices.
- Sterilization, and Dry Time testing ensured the subject TiTLE® 2 Polyaxial System . steam sterilization, and dry time requirements and instructions are substantially equivalent to the predicate devices.
### Substantial Equivalence:
Zimmer Spine considers the subject TiTLE® 2 Polyaxial Spinal System product performance to be substantially equivalent to its predicate device. TiTLE® 2 Polyaxial Spinal System because there are no changes to the product performance specifications or device functional scientific technology.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission of promoting health, well-being, and human services. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Arenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 27, 2013
Zimmer Spine, Incorporated Ms. Michelle Lenz Regulatory Affairs Specialist 7375 Bush Lake Road Minneapolis, Minnesota 55439
Re: K133086
Trade/Device Name: TiTLE® 2 Polyaxial Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB. MNH. MNI, KWP Dated: September 27, 2013 Received: September 30. 2013
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse cvents) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Ms. Michelle Lenz
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Zimmer Spine - 510(k) -- TiTLE® 2 Polyaxial Spinal System -- Section 4. Indications for Use Statement
Indications for Use
K133086 510(k) Number (if known):
Device Name:
TiTLE® 2 Polyaxial Spinal System
Indications for Use:
The TiTLE 2 Polyaxial Spinal System is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic . studies). Levels of fixation are for the thoracic, lumbar and sacral spine.
The TiTLE 2 Polyaxial Spinal System is a Pedicle Screw System intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).
The TiTLE 2 Polyaxial Spinal System is also indicated for pedicle screw fixation for severe spondylolisthesis (grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1.
In addition, TiTLE 2 Polyaxial System, when not used with pedicle screws, is indicated for sacral screw fixation from T1 to the illium sacrum. The non-pedicle screw indications are spondylolisthesis, degeneralive disc disease, (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis, and kyphosis), tumor, fracture, and previous failed fusion surgery.
The TiTLE 2 Polyaxial System can also be linked to the Minit® Posterior Cervical and Upper Thoracic Fixation System.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)
Over-the Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K133086
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.