ONE PLUS IMPLANT SYSTEM

K133050 · Osseofuse, Inc. · DZE · Sep 19, 2014 · Dental

Device Facts

Record IDK133050
Device NameONE PLUS IMPLANT SYSTEM
ApplicantOsseofuse, Inc.
Product CodeDZE · Dental
Decision DateSep 19, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intental or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading.

Device Story

One Plus Implant System consists of endosseous dental implants surgically placed into upper or lower jaw bone; intended to support prosthetic artificial teeth. Device fabricated from Titanium 6AL 4V ELI Gr.23 alloy; features RBM-treated fixture body and TiN-coated abutment. Available in multiple diameters (3.00mm–5.25mm) and lengths (11.5mm–14.5mm). Used by dental clinicians in clinical settings for delayed loading protocols. Provides stable foundation for prosthetic restoration; restores masticatory function in partially edentulous patients.

Clinical Evidence

Bench testing only. Evidence includes sterilization validation per ISO 11137-1/2/3, accelerated shelf-life testing per ASTM F1980-07 and ISO 11607-1, and chemical/SEM analysis to verify absence of residuals post-RBM treatment.

Technological Characteristics

Material: Titanium 6AL 4V ELI Gr.23 alloy. Surface: RBM-treated fixture body, TiN-coated abutment. Form factor: One-piece endosseous dental implant. Sterilization: Gamma radiation. Connectivity: None (mechanical device).

Indications for Use

Indicated for partially edentulous patients requiring missing tooth replacement to restore mastication. 3.0mm diameter implants are indicated for central or lateral mandibular incisors. Intended for single use and delayed loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 19, 2014 OsseoFuse, Inc. C/O Priscilla Chung LK Consulting Group USA, Inc. 2651 East Chapman Avenue, Suite 110 Fullerton, CA 92831 Re: K133050 Trade/Device Name: One Plus Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endossesous Dental Implant Regulatory Class: II Product Code: DZE Dated: August 20, 2014 Received: August 21, 2014 #### Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Mary S. Runner -S Erin I. Keith, M.S Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page 510(k) Number (if known) K133050 Device Name One Plus Implant System Indications for Use (Describe) The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intental or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading. Type of Use (Select one or both, as applicable) × Prescription Use (Part 21 CFR 801 Subpart D) ■ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ # 510(k) Summary (K133050) This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: 09/17/2014 # 1. Applicant / Submitter OsseoFuse, Inc. 5023 North Parkway Calabasas. Calabasas, CA 91302 Tel. 888.446.9995 Fax. 310.356.3183 ## 2. Submission Correspondent LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com ## 3. Device - Trade Name: One Plus Implant System - · Common Name: Dental Implant - Classification Name: Endosseous Dental Implant System - · Product Code: DZE - · Classification regulation: 21CFR872.3640 # 4. Predicate Device: - · MS System by OSSTEM IMPLANT CO., LTD (K083067) - · S-MINI IMPLANT SYSTEM by NEOBIOTECH CO., LTD (K112540) - · OSSEOFUSE DENTAL IMPLANT SYSTEM by Dynamic Innovations Inc. (K110577) - · Spectra System (ScrewDirect Implant) by Implant Direct LLC (K061319) - · Replace One Piece Implant by NOBEL BIOCARE UAS INC. (K023952) - · Lifecore PrimaSolo One-Piece Implant System by Lifecore Biomedical, Inc.(K050506) {4}------------------------------------------------ # 5. Description: The One Plus Implant System is a dental implant system made of Titanium 6AL 4V ELI Gr.23 alloy intended to be surgically placed in the bone of the upper or lower jaw arches. The system is similar to other commercially available products based on the intended use, the technology used, the material composition employed and performance chracteristics, The surface of this system has been treated with R.B.M and the abutment part has TiN coating on it. The One Plus Implant System is available in the following sizes. Sizes: 3.00mm (Dia.) x 11.5/13/14.5 3.75mm (Dia.) x 11.5/13/14.5 4.50mm (Dia.) x 11.5/13/14.5 5.25mm (Dia.) x 11.5/13/14.5 ## 6. Indication for use: The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intended to be used in central or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading. ## 7. Basis for Substantial Equivalence The subject device is substantially equivalent to the noted predicate devices based on tabulated device specifications and properties presented. Based on the comparison analysis, the identical intended use, comparable technological characteristics, and similar general design features, the subject device is substantially equivalent to the predicate devices. There are no significant differences between the One Plus Implant System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to the predicate devices in design, function, material and intended use. The comparison chart can be found on the following pages. {5}------------------------------------------------ #### Comparison Chart | | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(K) Number | K133050 | K083067 | K112540 | K110577 | K061319 | | Device Name | One Plus Implant System | MS System | S-MINI IMPLANT<br>SYSTEM | OSSEOFUSE DENTAL<br>IMPLANT SYSTEM | Spectra System<br>(ScrewDirect Implants) | | Applicant | OsseoFuse Co., Ltd. | OSSTEM Implant Co.,<br>Ltd. | NEOBIOTECH CO., LTD. | Dynamic Innovations Inc. | Implant Direct LLC | | Contract<br>Manufacturer | KJ Meditech Co., Ltd. | - | - | KJ Meditech Co., Ltd. | - | | Indications for<br>Use | The One Plus Implant<br>System is intended to use<br>in the treatment of<br>missing teeth to support<br>prosthetic device, such as<br>artificial teeth, in order to<br>restore mastication in<br>partially edentulous<br>patients. 3.0mm diameter<br>implants are intended to<br>be used in central or<br>lateral mandibular<br>incisors. The One Plus<br>Implant System is<br>intended for single use<br>only. It is for delayed<br>loading. | The MS System<br>(Denture) is intended to<br>be place in the bone of<br>the upper or lower jaw<br>arches to provide support<br>the prosthetic devices to<br>restore the patient's<br>chewing function,<br>including the denture<br>stabilization. MS System<br>(Denture) is intended for<br>single use only.<br>The MS System (Narrow<br>Ridge) is intended to use<br>in the treatment of<br>missing mandibular<br>central and lateral<br>incisors to support<br>prosthetic device, such<br>as artificial teeth, in<br>order to restore chewing<br>function in partially<br>edentulous patients. MS<br>System (Narrow Ridge)<br>is intended for single use<br>only. It is not for | The Cement type is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors or<br>the mandibular central and<br>lateral incisors to serve as<br>temporary support<br>prosthetic devices during<br>the healing phase of<br>permanent endosseous<br>dental implant, such as<br>artificial teeth, in order to<br>restore chewing function in<br>partially edentulous<br>patients. | The OsseoFuse Dental<br>Implant System is intended<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including:<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. This system is<br>dedicated for one and two<br>stage surgical procedures<br>and not dedicated for<br>immediate loading. This<br>system is intended for<br>delayed loading. | The Spectra Dental<br>Implant System consists of<br>one-piece or two-piece<br>implants for single-stage or<br>two-stage surgical<br>procedures that are<br>intended for use in<br>partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. They may be<br>placed in immediate<br>function if initial implant<br>stability can be<br>established.<br>The ScrewDirect 3.0mm<br>implant is indicated for:<br>1. An artificial root<br>structure for single tooth<br>replacement of mandibular<br>central and lateral incisors | | | | immediate loading. | | | and maxillary lateral<br>incisors.<br>2. Multiple tooth<br>replacements or denture<br>stabilization. | | Design | One Piece Implant | One Piece Implant | One Piece Implant | Two Piece Implant | One Piece Implant | | Appearance | Image: One Piece Implant | Image: One Piece Implant | Image: One Piece Implant | Image: Two Piece Implant | Image: One Piece Implant | | Material | Ti 6Al 4V ELI, Gr.23 | Titanium Alloy | Titanium Gr. 4 | Ti 6Al 4V ELI, Gr.23 | Titanium Alloy | | Surface<br>Treatment | •RBM Treatment on the<br>fixture body<br>•TiN coating on the<br>abutment | RBM Treatment on the<br>fixture body | RBM Treatment on the<br>fixture body | •RBM Treatment on the<br>fixture body<br>•TiN coating on the<br>abutment | Roughened - HA blasted | | Implant Sterile | Yes | Yes | Yes | Yes | Yes | | Sterilization<br>Method | Gamma | Gamma | Gamma | Gamma | Gamma | | Implant<br>Diameters | 3.00mm, 3.75mm,<br>4.50mm, 5.25mm | 2.50mm 3.00mm | 2.0mm, 2.5mm, 3.0mm,<br>3.5mm | 3.75mm, 4.1mm, 4.5mm,<br>5.25mm | 3.0mm, 3.7mm, 4.7mm,<br>5.7 mm | | Implant<br>Lengths | 11.5mm, 13mm, 14.5mm | 10.0 mm, 13.0mm,<br>15.0mm | 7.0 - 15.00 mm | 8.5mm - 16.0 mm | 10mm, 13mm, 16mm | | Product Code | DZE | DZE | DZE | DZE, NHA | DZE | {6}------------------------------------------------ {7}------------------------------------------------ | | Predicate Device 5 | Predicate Device 6 | | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | 510(K) Number | K023952 | K050506 | | | Device Name | Replace One Piece Implant | Lifecore PrimaSolo™ One-Piece Implant System | | | Applicant | NOBEL BIOCARE UAS INC. | LIFECORE BIOMEDICAL, INC. | | | Indications for Use | The Replace One Piece Implant is a threaded<br>one-piece implant with an integrated abutment,<br>designed for single-stage surgical procedure<br>and cemented restorations. The Replace One<br>Piece Implant is intended for immediate load<br>on single tooth and multiple tooth applications<br>in good quality bone, to restore chewing<br>function. | Lifecore Biomedical Dental Implant System<br>implants are intended for use in partially or fully<br>edentulous mandibles and maxillae, in support of<br>single or multiple-unit restorations including;<br>cement retained, screw retained, or overdenture<br>restorations, and terminal or intermediate<br>abutment support for fixed bridgework.<br>Specific Intended Uses: The PrimaSolo One-<br>Piece (3.5-5.0mm) Implant is a threaded one-<br>piece implant with integrated abutment designed<br>for single-stage surgical procedure and cemented<br>restorations. The PrimaSolo One-Piece Implant is<br>intended for immediate placement and can be<br>restored with a temporary prosthesis in single<br>tooth and multiple tooth applications with good<br>quality bone.<br>The PrimaSolo One-Piece (3.0mm) Implant is a<br>threaded one-piece implant with an integrated<br>abutment designed for single-stage surgical<br>procedure and is indicated for use in the treatment<br>of missing maxillary lateral incisors or the<br>mandibular central and lateral incisors to support | | | | | prosthetic devices, such as artificial teeth, in<br>order to restore chewing function in partially | | | | | edentulous patients. Mandible central and lateral<br>incisors must be splinted if using two or more | | | | | 3.0mm implants adjacent to one other. | | | Design | One Piece Implant | One Piece Implant | | {8}------------------------------------------------ | Appearance | | | |----------------------|----------------------|-----------------------------------| | Material | Titanium Gr. 4 | Titanium Alloy | | Surface Treatment | | RBM Treatment on the fixture body | | Implant Sterile | Yes | Yes | | Sterilization Method | | | | Implant Diameters | 3.5mm, 4.3mm, 5.0mm, | 3.0mm, 3.5mm, 4.1mm, 5.0mm | | Implant Lengths | 10mm, 13mm, 16mm | 10.0 mm、11.5mm 13.0mm、15.0mm | {9}------------------------------------------------ # 8. Non-Clinical Testing - · Sterilization validating testing has been performed in accordance with ISO 11137-1, ISO 11137-2 and ISO 11137-3. - · Accelerated shelf life testing has been performed in accordance with ASTM1980-07, ISO 11607-1, ISO 11737-1, ISO 11737-2 - Chemical and SEM image analyses have been performed to verify that there is no residual after RBM treatment on the fixture. There might be differences in sterilization parameters, shelf life and manufacturing processes between the subject device and the predicate devices, however, the test results supported that the subject device is substantially equivalent to the predicate devices. # 9. Conclusion The subject device and the predicate devices have intended use and have similar technological characteristics. Overall, the One Plus Implant System has the following similarities to the predicate devices: - · has the same intended use, - · uses the same operating principle, - · incorporates the same basic design, - incorporates the same material and the surface treatment. Based on the similarities, we conclude that the One Plus Implant System is substantially equivalent to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%