← Product Code [KCT](/productcode/KCT) · K133015

# INSTRU-SAFE INSTRUMENT PROTECTION SYSTEM (K133015)

_Summit Medical, Inc. · KCT · Mar 13, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K133015

## Device Facts

- **Applicant:** Summit Medical, Inc.
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Mar 13, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility.

## Device Story

Instru-Safe Instrument Protection System consists of specialized cassettes designed to organize and protect surgical instruments during sterilization; used by healthcare providers in clinical settings. Cassettes hold instruments in place during pre-vacuum steam sterilization cycles (270°F/132°C). System requires use with external sterilization wraps or Aesculap rigid containers to maintain sterility post-process. Benefits include organized instrument handling and protection of delicate devices during the sterilization cycle.

## Clinical Evidence

Bench testing only. Validation performed for steam pre-vacuum sterilization cycles at 270°F (132°C) with 4 minutes exposure and 30 minutes dry time. Testing confirmed compatibility with specific Aesculap rigid containers and validated lumen sizes for various cassette models.

## Technological Characteristics

Cassette-based instrument organization system; metallic construction; designed for pre-vacuum steam sterilization (270°F/132°C). Features include specific lumen size accommodations (1mm to 5mm) and instrument capacity limits. System is non-sterile and requires external sterilization packaging (wrap or rigid container).

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Reference Devices

- Aesculap rigid container

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 13, 2014

Summit Medical, Incorporated Ms. Nicole Dove Quality Assurance/Regulatory Affairs Manager 815 Northwest Parkway, Suite 100 St.Paul, MN 55121

Re: K133015

Trade/Device Name: Instru-Safe® Instrument Protection System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: February 12, 2014 Received: February 14, 2014

Dear Ms. Dove:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Dove

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

Sincerely yours.

http://www.fda.gov/MedicalDeviccs/Resourcesfor You/Industry/default.htm.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three black diamonds arranged horizontally.

## Indications for Use Statement

Page of

510(k) number: Unknown, new submission - TBD K133015

Device Name: Instru-Safe® Instrument Protection System

Indications for Use:

Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility.

Autoclave Sterilization Parameter

Cycle: Pre-vacuum Temperature: 270°F (132°C) Exposure Time: 4 minutes Minimum Dry Time: 30 minutes

| Summit Cassette Model | Aesculap Container Model |
|-----------------------|--------------------------|
| IN-8823-AE            | *JN444                   |
| IN-2880               | *JK444                   |
| IN-6105               | *JN742                   |

*Validated by Summit Medical for use in steam prevacuum sterilizers ONLY operating at 270°F (132°C) for 4 minutes exposure time. Consult container instructions to ensure that contents do not exceed the sterilization containers intended load claims.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F 2014.03.12 15

Prescription Use: (Per 21 CFR 801.109) OR

Over-The-Counter x

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, bold letters, with three black diamonds arranged horizontally above the word. Below the word "SUMMIT" is the word "medical" in smaller letters.

## Indications for Use Statement

Page of

| Summit<br>Cassette Model | Minimum<br>Inside<br>Diameter | Maximum<br>Length | Number of<br>Lumens |
|--------------------------|-------------------------------|-------------------|---------------------|
| IN-8823-CF               | 3mm                           | 400mm             | 4                   |
| IN-8823-CF               | 3mm                           | 200mm             | 2                   |
| IN-2880                  | 1mm                           | 76mm              | 2                   |
| IN-2880                  | 3mm                           | 177mm             | 1                   |
| IN-6105                  | 5mm                           | 241mm             | 1                   |
| IN-2681                  | 1mm                           | 65mm              | 1                   |
| IN-2681                  | 3mm                           | 200mm             | 1                   |
| IN-7823                  | 1mm                           | 400mm             | 17                  |

Lumen size of instrumentation validated includes:

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three black diamonds, arranged horizontally.

:

## Indications for Use Statement

| Attachment 1 - Device Models |                          |                                              |
|------------------------------|--------------------------|----------------------------------------------|
| Part Number                  | Maximum # of Instruments | Estimated Weight - Tray w/ instruments (lbs) |
| IN-2681                      | 13                       | 0.64                                         |
| IN-2682                      | 13                       | 0.89                                         |
| IN-2683                      | 13                       | 1.39                                         |
| IN-2840                      | 36                       | 8.75                                         |
| IN-2842                      | 24                       | 6.5                                          |
| IN-2843                      | 36                       | 8.75                                         |
| IN-2880                      | 56                       | 12.1                                         |
| IN-2900                      | 22                       | 4.18                                         |
| IN-2950                      | 12                       | 4.1                                          |
| IN-3030                      | 34                       | 9.5                                          |
| IN-4000                      | 20                       | 3.2                                          |
| IN-4003                      | 30                       | 3.25                                         |
| IN-4010                      | 10                       | 1.98                                         |
| IN-5401                      | 6                        | 2                                            |
| IN-5401-02                   | 2                        | 1                                            |
| IN-5401-03                   | 2                        | 1                                            |
| IN-5401-08                   | 8                        | 3.2                                          |
| IN-5401-12                   | 12                       | 3.25                                         |
| IN-6103                      | 2                        | 2.15                                         |
| IN-6105                      | 2                        | 2.15                                         |
| IN-6110                      | 3                        | 2.15                                         |
| IN-6203                      | 2                        | 2.75                                         |
| IN-6205                      | 2                        | 2.75                                         |
| IN-6210                      | 2                        | 2.75                                         |
| IN-6240                      | 2                        | 2.75                                         |
| IN-6303                      | 2                        | 3.28                                         |
| IN-6305                      | 2                        | 3.28                                         |
| IN-6310                      | 2                        | 3.28                                         |
| IN-6403                      | 2                        | 3.28                                         |
| IN-6405                      | 2                        | 3.28                                         |
| IN-6410                      | 2                        | 3.28                                         |
| IN-7012                      | 1                        | 1.07                                         |
| IN-7032                      | 2                        | 1.1                                          |
| IN-7150                      | 8                        | 1.9                                          |
| IN-7153                      | 6                        | 1.7                                          |
| IN-7223                      | 10                       | 9.2                                          |
| IN-7224                      | 15                       | 7.2                                          |
| IN-7225                      | 10                       | 9.5                                          |
| IN-7723                      | 15                       | 7.18                                         |
| IN-7724                      | 15                       | 7.2                                          |
| IN-7725                      | 10                       | 9.5                                          |
| IN-7823                      | 25                       | 4                                            |
| IN-7940                      | 20                       | 13.25                                        |
| IN-8240                      | 20                       | 13.5                                         |
| IN-8420                      | 20                       | 13.5                                         |
| IN-8610                      | 2                        | 6.65                                         |
| IN-8612                      | 2                        | 6.8                                          |
| IN-8613                      | 2                        | 6.1                                          |
| IN-8615                      | 2                        | 5.8                                          |
| IN-8616                      | 2                        | 5.8                                          |
| IN-8620                      | 3                        | 7.2                                          |
| IN-8621                      | 4                        | 7.18                                         |
| IN-8622                      | 4                        | 7.18                                         |
| IN-8630                      | 3                        | 6.5                                          |
| IN-8632                      | 3                        | 6.45                                         |
| IN-8633                      | 3                        | 6.8                                          |
| IN-8640                      | 4                        | 5.35                                         |
| IN-8642                      | 4                        | 5.35                                         |
| IN-8643                      | 5                        | 5.35                                         |
| IN-8645                      | 4                        | 5.35                                         |
| IN-8660                      | 4                        | 5.35                                         |
| IN-8662                      | 4                        | 5.35                                         |
| IN-8663                      | 4                        | 5.35                                         |
| IN-8810                      | 20                       | 13.5                                         |
| IN-8820                      | 15                       | 8.75                                         |
| IN-8823                      | 45                       | 14                                           |
| IN-8830                      | 15                       | 8.75                                         |
| IN-8833                      | 45                       | 14                                           |
| IN-8840                      | 20                       | 13.75                                        |
| IN-8850                      | 15                       | 8.75                                         |
| IN-8853                      | 45                       | 14                                           |
| IN-8860                      | 15                       | 8.75                                         |
| IN-8862                      | 30                       | 10.5                                         |
| IN-8863                      | 45                       | 14                                           |
| IN-8880                      | 2                        | 3.28                                         |
| IN-8882                      | 16                       | 12.1                                         |
| IN-8883                      | 2                        | 3.28                                         |
| IN-8884                      | 4                        | 5.35                                         |
| IN-8885                      | 1                        | 2.25                                         |
| IN-8886                      | 6                        | 12.1                                         |
| IN-8889                      | 6                        | 12.1                                         |
| IN-8891-S                    | 1                        | 2                                            |
| IN-8891-SI-12-S              | 1                        | 2                                            |
| IN-8891-SI-85-S              | 1                        | 2                                            |
| IN-8892-01                   | 12                       | 12.1                                         |
| IN-8893                      | 9                        | 7.5                                          |
| IN-8894                      | 5                        | 16.1                                         |
| IN-8897                      | 8                        | 6                                            |
| IN-8898                      | 10                       | 10.25                                        |
| IN-8899                      | 7                        | 6.5                                          |
| IN-8901                      | 1                        | 2.25                                         |
| IN-8902                      | 22                       | 17                                           |
| IN-8903                      | 15                       | 13.25                                        |
| IN-8904                      | 22                       | 17                                           |
| IN-8907                      | 7                        | 12.5                                         |
| IN-8945                      | 2                        | 5.18                                         |
| IN-8946                      | 9                        | 6.1                                          |
| IN-8980-01                   | 20                       | 9.5                                          |
| IN-8982-01                   | 17                       | 9.5                                          |
| IN-8983-01                   | 16                       | 9.5                                          |
| IN-8984-01                   | 15                       | 9.5                                          |
| IN-8986-S                    | 2                        | 6.5                                          |
| IN-8987-S                    | 2                        | 6.5                                          |
| IN-8988-S                    | 2                        | 6                                            |
| IN-8989-S                    | 2                        | 6                                            |
| IN-9999-160                  | 6                        | 12.1                                         |
| IN-9999-162-S                | 2                        | 5.8                                          |
| IN-9999-168-S                | 2                        | 5.8                                          |
| IN-9999-172-S                | 2                        | 5.8                                          |
| IN-9999-178-S                | 2                        | 5.8                                          |

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA phone: 651.789.3939 | fax: 651.789.3979 www.summitmedicalusa.com

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "SUMMIT" are three black diamonds, arranged horizontally.

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Summit Medical. The logo consists of the word "SUMMIT" in large, bold, sans-serif font, with the word "medical" in a smaller font size underneath. Above the word "SUMMIT" are three diamond shapes, arranged horizontally. The logo is simple and modern, and the use of the diamond shapes suggests a sense of precision and accuracy.

Summit Medical, Inc. | 815 Northwest Parkway, Suite 100 | St. Paul, MN 55121 USA

---

**Source:** [https://fda.innolitics.com/device/K133015](https://fda.innolitics.com/device/K133015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
