The intended use is to acquire, record, and store one or more days of multi-lead ECG data for patients undergoing continuous ambulatory electrocardiography. The 300-2 performs no cardiac analysis by itself and is intended to be used only with the DMS Premier II Holter system (K062088), as is the predicate 300-2 device.
Device Story
The 300-2W is a battery-powered, ambulatory Holter ECG recorder worn by patients for 24+ hours. It captures multi-lead ECG signals via electrode cables, storing data on internal memory chips. The device features an event button and auxiliary outputs (USB and WiFi) for real-time monitoring of electrode quality or operational status. After recording, data is transferred via USB to a PC running the DMS Premier II Holter software. The physician reviews the generated report to diagnose cardiac conditions. The device is used in clinical settings and operated by medical professionals. Modifications from the predicate 300-2 include increased memory capacity and the addition of a WiFi auxiliary output. The device does not perform onboard analysis; it serves as a data acquisition and storage tool.
Clinical Evidence
Bench testing only. The device underwent electromagnetic compatibility and electrical safety testing per EN 60601-1-2:2007, including electrostatic discharge, radiated emissions, RF electromagnetic field immunity, and power frequency magnetic field tests. No clinical data was required or provided.
Technological Characteristics
Multi-channel ambulatory ECG recorder; 3-lead configuration; 8-bit resolution; 0.05-100 Hz bandwidth; >60 dB common mode rejection. Powered by AA alkaline batteries. Housed in plastic casing (mobile phone size). Connectivity: USB for data transfer, WiFi auxiliary output for real-time monitoring. Standards: IEC 60601-1, ISO 13485, EN 60601-1-2:2007, IEC 60601-2-47.
Indications for Use
Indicated for use in a clinical setting by qualified medical professionals for recording multi-lead ECG data of patients for a minimum of 24 hours. Used for evaluation of patients with symptoms suggesting arrhythmias, patients with pacemakers, heart rate changes, QRS interval changes, and response to drug therapy. Not a life-supporting system; not connected to AC power.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
K133014 b 1/a
# Special 510(k) Notification: Device Summary
#### 300-2W Holter ECG Recorder
#### Date: September 16, 2013
#### Submitter:
William Parsons, Official Correspondent Diagnostic Monitoring Software 290 Kingsbury # 3 P.O. Box 3109 Stateline, NV 89449 Tel: 775 589 6049 . Fax: 714 953 0260 e-mail: holterdms@gmail.com Contact: William Parsons Submitted Device: 300-2W Predicate Device: 300-2 Common Name: Holter ECG Recorder Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: II (two) Product Code: MWJ
Establishment Registration Number: 2028190 Owner/Operator Number: 9003252 Payment Identification Number:
The elements of this Special 510(k) Summary include the following:
- 1. Legally 510(k) marketed device to which Substantial Equivalence is claimed.
- 2. Description.
- ന് Comparison to the Sponsor's Predicate Device.
- 4. Intended Uses and Indications for Use of the Modified Device.
- 5. Technology.
- 6. Design Control Activities Summary.
- 7. Compliance with Performance Standards.
- 8. Labeling,
{1}------------------------------------------------
9. Hazard Risk Analysis.
10. Pictures.
# Legally 510(k) marketed device to which Substantial Equivalence is claimed:
In October 2006, the FDA issued its 510(k) number K062959 for the DMS Holter Recorder, Model 300-2, which is the predicate device for this submission. The Regulation Number was 21 CFR 870 2800, and the Regulation Name was Ambulatory Electrocardiograph without analysis. The Regulatory Class was Class II, and the Product Code was 74 MWJ.
The subject of this Device Modification submission is the DMS Holter Recorder, Model 300-2W. The 300-2W is a minor modification to the predicate 300-2 Holter Recorder. The modifications are (1) increasing the size of the memory storage without affecting the characteristics of the stored ECG signal, and (2) including an additional auxiliary ECG output for the traditional realtime ECG viewing of the operational status of a Holter Recorder.
The history of the predicate 300-2 device is that this 300-2W is its first modification for a submission of a Special 510(k) submission.
# Principle of Device Operation:
The principle of operation of a Holter ECG recorder is that a battery powered device is worn by a patient for the purpose of a 24-hour or longer ambulatory ECG recording on either a hospital in-patient or out-patient. After the Holter ECG recording is completed, the ECG data is transferred from the Holter ECG recorder to a PC that has a Holter ECG processing software program. From the PC, a Holter ECG report is generated for a qualified physician to review. Only a licensed physician can order the Holter ECG test.
# Description:
The 300-2W is a Holter Recorder that records the ECG signal in the exact same manner as the predicate 300-2 device. The 300-2W transfers the Holter ECG from its memory in the exact same manner as the predicate 300-2. The 300-2W uses the same USB cable as is used by the predicate 300-2 to transfer the Holter file from the Holter Recorder to the Holter PC. The Holter PC program processes the 300-2W Holter ECG recording with the exact same software program as is used by the 300-2 predicate device. All of the Holter ECG screen displays and print pages are exactly the same when using either the 300-2W device or the 300-2 predicate device. The Holter report from which the physician makes a diagnosis is the exact same format when using either the 300-2W device or the 300-2 predicate device. The 300-2W uses the same firmware as the predicate 300-2 to record the Holter ECG data and to transfer the Holter recording to the Holter PC. The ECG signal in the 300-2W is the same ECG signal as the
{2}------------------------------------------------
predicate 300-2. The Intended Uses of the 300-2W and the predicate 300-2 are the same. The Indications for Use of the 300-2W and the predicate 300-2 are the same. The Design Controls and Performance Standards for the 300-2W and the predicate 300-2 are from the same source processes. The manufacturing process has ISO 13485 certification.
The 300-2W and the predicate 300-2 are multi-day Holter recorders. Both the 300-2W and the predicate 300-2 use the same alkaline or lithium batteries that are off-the-shelf AA batteries. Both the 300-2W and the predicate 300-2 use off-the-shelf commercial memory chips. The 300-2W uses a larger capacity memory chip, and both recorders operate with the memory chip in the same manner. The larger commercial memory chips are now larger capacity and cost less money than 300-2 Holter ECG recorder of 2006. All Holter ECG recorders have auxiliary real-time ECG outputs. These auxiliary outputs have no effect on the safety, performance, Intended Use, or Indications for Use with either the 300-2W or the predicate 300-2. Both the 300-2W and the predicate 300-2 have two (2) auxiliary outputs. The traditional uses of these auxiliary outputs is to view the quality of the electrode application to the patient prior to leaving the doctor's office, or to see the operational ECG of the Holter recording. The 300-2W and the predicate 300-2 have the same USB auxiliary output. The 300-2W has a WiFi auxiliary output and the predicate 300-2 has a transtelephonic auxiliary output.
Both the 300-2W and the predicate 300-2 have successfully passed TUV Rheinland CE Test Reports for Holter Recorders. Both the 300-2W and the predicate 300-2 use the same ECG electrode cables. Both Holter recorders provide the physician with the same Holter ECG results. See Sections 12 and 13 for the TUV CE Test Reports.
The enclosed TUV CE Test Report 15030071 001 shows that the 300-2W Holter Recorder passed all the test specifications for EN 60601=1-2:2007; including Enclosure Electrostatic Discharge, Radiated Emission on the Frequency Range above 30 MHz, RF electromagnetic field immunity test, and Power frequency magnetic field.
A photograph of the 300-2W is listed as an attachment in the Pictures section 15.
The physical properties of both the 300-2W and the predicate 300-2 are the same. The Holter circuit board is housed in a plastic type casing. The devices are similar in size to a mobile phone. There is an input for connecting an ECG cable, and an output for transferring the ECG that is stored on a memory chip on the circuit board. It is not physically possible to connect the ECG Cable input and the ECG data transfer output at the same time. As in all Holter ECG recorders there are auxiliary outputs to view the ECG in a real-time ECG mode. The physical properties of the 300-2W have been thoroughly tested for safety and performance as shown in the attached TUV Test Reports.
{3}------------------------------------------------
The comparison chart for the Device Modification and the Predicate Device is shown on the following page.
# Comparison to the Sponsor's Predicate Device:
.
The DMS 300-2W is substantially equivalent (SE) to the below predicate device, (300-2).
| Specifications | 300-2W | 300-2 |
|-----------------------|----------------------|----------------------|
| Predicate Device | No | Yes |
| Owner | DMS | DMS |
| 510(k) number | | K062959 |
| ECG Leads | 3-Lead ECG | 3-Lead ECG |
| Resolution | 8-bit | 8-bit |
| Recording Duration | 30-day | 30-day |
| Bandwidth | 0.05 to 100 Hz | 0.05 to 100 Hz |
| Common Mode Rej | >60 db | >60 db |
| Power Source | AA Alkaline | AAA Alkaline |
| Average Current Drain | 3 mA | 3 mA |
| Event ECG Button | Yes | Yes |
| Operating Temperature | 0 to 60 C | 0 to 60 C |
| 1 MV Input = | 1-CM Square Wave | 1-CM Square Wave |
| Dimensions | 4.90 x 2.75 x .94 in | 4.90 x 2.75 x .94 in |
| Weight | 3 oz. w/o battery | 3 oz. w/o battery |
| Processing System | Premier II Holter | Premier II Holter |
| ECG Data Transfer | PC's hard disk | PC's hard disk |
{4}------------------------------------------------
# Intended Uses and Indications for Use of the Modified Device:
The "Intended Uses" of the modified 300-2W is exactly the same as the predicate device (model 300-2). The intended use is to acquire, record, and store one or more davs of multi-lead ECG data for patients undergoing continuous ambulatory electrocardiography. The 300-2 performs no cardiac analysis by itself and is intended to be used only with the DMS Premier II Holter system (K062088), as is the predicate 300-2 device.
The "Indications for Use" of the modified DMS 300-2W recorder is indicated for use in a clinical setting by qualified medical professionals only for recording multi-lead ECG data of patients during a minimum ambulatory time period of 24-hours. It is not a life supporting system, and is not connected to an AC power source. Ambulatory 24-hour electrocardiography is used for the below indications:
- . Evaluation of patients with symptoms suggesting arrhythmias.
- Evaluation of patients with pacemakers. .
- Evaluation of patients heart rate changes.
- . Evaluation of patients QRS interval changes.
- · Evaluation of patients response to drug therapy treatment.
The "Indications for Use" for the modified 300-2W is exactly the same as the predicate 300-2 device.
# Technology:
The modified 300-2W and the predicate 300-2 are both multi-channel ambulatory electrocardiograph devices that have the same intended uses and indications for use, that function in the same manner, produce the same ECG results, and are both processed by the same DMS Premier II Holter system. The Premier II Holter was granted 510(k) marketing approval in 2006 with K062088. Both devices use the same technological characteristics to record the Holter ECG.
### Design Control Activities Summary:
Both the 300-2W and the predicate 300-2 Holter recorders use the same Design Control requirements that are an integral part of the manufacturing Quality System Regulation (QSR). The QSR describes the good manufacturing practices (GMP) requirements, and have been CE certified by TUV with the ISO 13485 certification. A copy of such ISO 13485 certification is attached (Section 16) as a document. Since this is a Special 510(k) submission, we hereby declare Conformance to Design Controls, and such is also attached in Section 8 as a document. The original 510(k) number for the predicate 300-2 device is K062959.
{5}------------------------------------------------
# K133014 b 6/9
Design controls 21 CFR 820.30 were used during the development of both the 300-2W device, and its predicate 300-2 device. Such design controls are available for FDA on-site inspections.
# Compliance with Performance Standards:
Design Control procedures within the Quality Management System use standards IEC 60601-1, ISO 9001, ISO 13485, EN 60601-1-2:2007, IEC 60601-2-47, 21 CFR 820.30, and MDD 93/42/EEC. These standards are listed in the Standards Data Report for 510(k)s in Section 14.
### Labeling:
A picture of the 300-2W labeling is enclosed in Section 15. The label identifies the device as the 300-2W. The Holter ECG test can only be prescribed by a licensed physician. Labeling changes:
- � Does the change affect the Indications for Use? No
- Is it a change in warnings or precautions? . No
- Does the change add a contraindication? No
- Does the change delete a contraindication? No
- . ls the labeling being revised for clarity to insure safer or more effective use? No
# CDRH Premarket Review Submission Cover Sheet:
See Section 2
510(k) Cover Letter:
See Section 3
Indications for Use Statement: See Section 4
Premarket Notification 510(k) Statement per 21 CFR 807.93: See Section 6
Truthful and Accuracy Statement per 21 CFR 807.87(k): See Section 7
Declaration of Conformity for Design Controls: See Section 8
{6}------------------------------------------------
Hazard Risk Analysis: See Section 9
Statement of Legal Rights to Distribute Submitted Device: See Section 10
Screening Checklist for Premarket Notification 510(k): See Section 11
Electromagnetic Compatibility and Electrical Safety (YUV): See Sections 12 and 13
Standards Data Report for 510(k):
See Section 14
Pictures: See Section 15
300-2W Operator's Manual See Section 16
Special 510(k) Criteria:
- . This 510(k) is submitted for the purpose of modifying a legally marketed predicate device, and is submitted by the same holder of the 510(k) of the predicate device.
- The "Indications for Use" of the submitted device are unchanged from the predicate . device.
- . The scientific technology of the submitted device is unchanged from the legally marketed predicate device.
- . The submission includes only summary level information.
The submission contains an "Indications for Use" statement (Section 4). .
The submission contains a 510(k) Summary (this Section 5).
The submission contains a 510(k) Summary Checklist (see section11).
The submission includes a Truthful and Accurate Statement (see section 7).
The submission references the numbers of the applicable national or international standards.
The submission relied upon published guidance documentations by the FDA.
{7}------------------------------------------------
The submitter believes that this submission is properly descriptive of the submitted device, and that the device description and labeling are consistent.
The submission's Principle of Operation is described in the 510(k) Summary section.
The description of the submitted device (300-2W) is in the Description section.
The Description section identifies the device modifications and the reasons for the changes.
The history of change from the predicate device to the modified device is described in the second page of Section 5 of this submission that was identified as "Legally 510(k) marketed device to which Substantial Equivalence is claimed."
The submitter included pictures and descriptions, and submitter believes that no engineering drawings are applicable; other than those shown in the Sections 12 & 13 TUV CE Test Reports. The submitted device is not intended to be marketed with multiple components. The 300-2W Holter recorder has its ECG data transferred to a PC with the Premier II Holter program, for which 510(k) number K062088 was provided by the FDA to Diagnostic Monitoring Software in August 2006. The predicate 300-2 device uses the same Premier II Holter program.
The submitter has included the predicate device's 510(k) number (K062959), the trade name (Holter ECG Recorder), and the model number (300-2W).
The identified predicate device is consistent throughout the submission; such as, being the Substantially Equivalent predicate device.
The submission identifies the Indications for Use and the Technology of the predicate 300-2 and the modified 300-2W as the same.
The submission describes why any differences between the submitted device and the predicate device do not constitute (a) a new intended use, (b) affect safety or effectiveness, or (c) raise different questions of safety and effectiveness.
The Risk Analysis methodology as applied to the modified submission device is described in the Hazard Risk Analysis (Section 9).
Based on the Hazard Risk Analysis, the attached TUV CE Test Reports describe the verification and validation requirements that were tested and passed, and such resulted from the various components of the ISO 13485 Quality Management System of the manufacturer.
20
{8}------------------------------------------------
This submission includes the Declaration of Conformity with Design Controls for (1) statement that all verification and validation activities were performed by designated individuals and the results were as expected, (2) statement that manufacturing facility complies with design controls as specified by 21 CFR 820.30, and (3) statement is signed by the responsible person.
This submission includes labeling and operator's manual, which includes a description of the device, its intended use, and directions for use.
This submission includes picture of labeling changes that result from device modification; such as, CE identification and model number.
The sections on Description and Intended Uses describe that the Intended Use of the modified device has not changed as a result of the modification. Also, as described in the Label process of the operator's manual the Intended Use has not changed as a result of the modification. Also, a separate signed statement on the Indications for Use is attached in Section 4.
Cordially,
William Parsons Official Correspondent Diagnostic Monitoring Software
{9}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. In the center of the seal is a stylized eagle-like symbol with three curved lines representing its wings. The seal appears to be a logo or emblem associated with the U.S. Department of Health and Human Services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(14 Silver Spring, MD 20993-002
May 19, 2014
Diagnostic Monitoring Software c/o Mr. William Parsons Official Correspondent 290 Kingsbury Grade # 3 Stateline, NV 89449 US
Re: K133014
> Trade/Device Name: Dms 300-2W Holter ECG Recorder Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory Electrocardiograph Regulatory Class: Class II Product Code: MWJ Dated: April 22, 2014 Received: April 23, 2014
Dear Mr. William Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
#### Page 2 - Mr. William Parsons
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### K133014
#### Indications for Use
510(k) Number (if known):
Device Name: DMS Holter ECG Recorder - Model 300-2W
Indications for Use: The "Indications for Use" of the modified 300-2W recorder are indicated for use in a clinical setting, by qualified medical professionals only, for recording multi-lead ECG data of patients during a minimum ambulatory time period of 24-hours. It is not a life support system, and is not connected to an AC power source. The "Indications for Use" of the modified 300-2W are exactly the same as the predicate devices (DMS 300-2 and 300-7). Ambulatory multi-day electrocardiography is used for the below indications:
- Evaluation of patients with symptoms suggesting arrhythmias .
- Evaluation of patients with pacemakers .
- Evaluation of patient heart rate changes .
- Evaluation of patient QRS interval changes .
- -Evaluation of patient response to drug therapy treatment
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
FDA Date: 2014.05.19
358 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.