Quell sound generators are designed for individuals who experience tinnitus. These devices generate a digitally controlled passive analog broad band noise from the circuit. The broad band noise is intended for use with tinnitus masking therapy. These devices are to be dispensed only by a doctor, licensed audiologist, or licensed hearing instrument specialist who are trained in the area of tinnitus treatment. BTE worn sound generators are one of several available sound stimuli means such as; computer generated audio, CD player, iPod, sound pillows, etc. Sound generators are dispensed to an adult population 18 years of age or above with specific instructions for their use.
Device Story
Quell 908TD is a behind-the-ear (BTE) electronic sound generator for tinnitus masking. Device generates digitally controlled, passive analog broadband white noise (750Hz–8KHz bandwidth). Operated by patient via volume control wheel; programmed by hearing healthcare professionals using Hearing Studio software via HI-PRO interface. Output delivered through acoustic soundtip and open dome tip in ear canal. Limits maximum output to 85 dB SPL per OSHA 29CFR 1910.95. Used in conjunction with Tinnitus Retraining Therapy (TRT) to mitigate internal head noise. Benefits include adjustable frequency/intensity to match individual patient needs and open-canal design for comfort.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via probe microphone real-ear measurements and 2cc ANSI test box measurements to confirm white noise characteristics and output safety limits (max 85 dB SPL).
Technological Characteristics
BTE form factor; zinc-air battery (#10) powered. Digital amplifier/speaker circuit generates broadband white noise. Adjustable bandwidth (750Hz-8KHz). Connectivity via HI-PRO interface for programming. Software-controlled frequency response and gain. Open dome tip (6, 8, 10mm). Complies with OSHA 29CFR 1910.95 for sound output limits.
Indications for Use
Indicated for adults 18+ experiencing tinnitus. Used for tinnitus masking therapy. Must be dispensed by trained doctor, audiologist, or hearing instrument specialist.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{0}------------------------------------------------
Page 3
.
Revised: 03/21/2014
.
# 510(k) Summary – Number: K132965
| DATE OF SUBMISSIONS: | September 17, 2013 |
|----------------------------|----------------------------------------------------------------------------------|
| SUBMITTER: | Amplisound Hearing Products & Services<br>594 Putnam Road<br>Danielson, CT 06239 |
| | Phone: (860)779-6500 Fax: (860)779-6501 |
| | CONTACT: |
| TRADE OR PROPRIETARY NAME: | Solace Sound Generators |
| MODEL: | Quell 908TD behind the ear |
| DEVICE COMMON NAMES: | Sound generators, tinnitus makers, noise generators |
| CLASSIFICATION NAME: | 21 CFR 874.3400 (Tinnitus Masker Class II) |
| PRODUCT CODE: | KLW |
| PREDICATE DEVICES: | Resound TSG Module K073636<br>TTC's GHI-T & TN3-T Tinnitus Masker K982451 |
{1}------------------------------------------------
| | Page 4 |
|---------------------------|------------|
| Amplisound 510(k) Summary | 03/04/2014 |
| #K132965 | 09/17/2013 |
#### Device Description:
The Quell 908TD is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. These devices are programmable and offer the ability to adjust the frequency level of the masking sound to fit the individual user's needs. The programming can be done by hearing healthcare professionals using a standard HI-PRO box and Amplisound's Hearing Studio software. The Quell device resembles a miniature behind the ear (BTE) hearing aid. This style device sits on top of and just behind the pinna of the ear and has an acoustic soundtip connected to it. The soundtip is a removable and interchangeable tubing of different lengths, left and right, which carries the sound from the ear. The soundtip connects to the device at one end and bends around the front of the pinna to the other end which is connected to a dome tip which fits inside the ear canal. Quell devices operate on a hearing aid battery and have an easily operated user volume control wheel. The wearer adjusts the sound to a desired loudness level by rolling the volume wheel with a fingertip.
Sound generators can be used for persons who suffer from tinnitus in conjunction with a sound therapy program such as Tinnitus Retraining Therapy (TRT). The sound therapy provides the tinnitus sufferer with an adjustable external sound source to mitigate the presence and annoyance of their personal internal noise.
Device Description - Quell model 908TD
Image /page/1/Figure/5 description: This image shows a hearing aid with labels pointing to its different parts. The labels include connector, processor, sound tube, volume wheel, battery compartment, sound tip, serial number, and retention handle. The hearing aid is a small, beige-colored device with a curved shape. The sound tube is a thin, clear tube that connects the hearing aid to the sound tip.
Image /page/1/Picture/6 description: This image shows a close-up of a human ear. The ear is well-defined, with visible details of the helix, antihelix, and earlobe. The skin appears to be textured, and there are some visible hairs around the ear. The image is in black and white, which emphasizes the contours and shapes of the ear.
Quell 908TD in place on the ear
The Quell 908TD sound generator consists of a miniature digital amplifier and speaker powered by a 1.4 volt battery packaged inside a small hearing aid shaped housing and which generates a digitally adjustable broad band white noise. White noise is defined as noise containing many frequencies at similar intensities. The bandwidth is adjustable from the broadest range of 750Hz to 8KHz, to the narrowest of 1250Hz to 3850Hz. The broad band noise is delivered through a small tubing connected to the housing which curves around the pinna and then is placed inside the ear canal. This small tubing is called a 'soundtip' which comes in left and right sides in short, medium and long lengths. There is a small plastic locking handle which helps retain the device in the ear, and a replaceable open dome tip in 6, 8, 10mm sizes to help the tip be comfortable and well seated in the ear canal.
{2}------------------------------------------------
| Amplisound 510(k) Summary | 03/04/2014 |
|---------------------------|------------|
| #K132965 | 09/17/2013 |
#### Indications for Use:
Quell sound generators are designed for individuals who experience tinnitus. These devices generate a digitally controlled passive analog broad band noise from the circuit. The broad band noise is intended for use with tinnitus masking therapy. These devices are to be dispensed only by a doctor, licensed hearing instrument specialist who are trained in the area of tinnitus treatment. BTE worn sound generators are one of several available sound stimuli means such as; computer generated audio, CD player, iPod, sound pillows, etc. Sound generators are dispensed to an adult population 18 years of age or above with specific instructions for their use.
#### Risks and Warnings for Safe Use:
Because the maximum sound output level of the device is limited to safe levels determined by OSHA, there is no threat to the user with respect to damage to the ear caused by intensity of sound and noise exposure.
#### Rationale of the design:
Research on sound generators concludes that the primary characteristics of the design must include the following: a smooth broadband white noise response with minimal peaks, high frequency emphasis, a very gradual volume control taper with the ability to lower the sound to a zero dB noise floor, a reliable circuit that does not produce extraneous noises to the wearer, a comfortable housing design that is easy to wear and maintains a secure fit with an open canal.
{3}------------------------------------------------
| | Page 6 | |
|---------------------------|--------|------------|
| Amplisound 510(k) Summary | | 03/04/2014 |
| #K132965 | | 09/17/2013 |
#### Software/Programming
When needed, control of the sound emanating from the device is done with the use of programming software called "Hearing Studio". The Quell device is connected to a computer using a cable with flex connector and HiPro interface. The adjustments include tone of any combination of 12 frequency bands and an overall gain slider. There is also an adjustment of the low battery warning signal from 500, 1000, 1500, 2000Hz at a level of 60, 66, 72, 78dB. There is the option of turning this warning off.
Image /page/3/Figure/3 description: The image shows a screenshot of a software interface with several labeled components. The labels point to the frequency band slider, overall gain, and low battery warning. The software appears to be related to audio processing or analysis, with graphs displaying frequency response and controls for adjusting various parameters.
{4}------------------------------------------------
| Amplisound 510(k) Summary | 03/04/2014 |
|---------------------------|------------|
| #K132965 | 09/17/2013 |
## Comparison Information to Predicate Devices
| | Quell 908TD | Resound TSG Module | GHI -T & TN3 Masker |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Sound Generator | Combination Device | Sound Generator |
| Models | BTE | BTE | BTE & ITE |
| Maximum output level | 85 વેક ટેન | 89 db SPL | 94 dB SPL |
| Battery drain | 1.0 mA | 1.2 mA | .4 - .5 mA |
| Peak output frequency | 3KHz | 3KHz | 2700, 3600, 6000Hz |
| Energy source | Zinc-Air battery #10 | Zinc-Air battery #312 | Zinc-Air battery #10,<br>312,13 |
| Intended use | Ear level worn sound<br>generator or masking<br>device for tinnitus. | Same | Same |
| Indications for use | Quell sound generators are<br>designed for individuals<br>who experience tinnitus.<br>These devices generate a<br>digitally controlled passive<br>analog broad band white<br>noise. The broad band<br>noise is intended for use<br>with tinnitus masking<br>therapy. These devices are<br>to be dispensed only by a<br>doctor, licensed audiologist<br>or licensed hearing<br>instrument specialist who<br>are trained in the area of<br>tinnitus treatment. | This tinnitus sound<br>generator module is a tool<br>to generate sounds to be<br>used in a tinnitus<br>management program to<br>relieve patients suffering<br>from tinnitus. The target<br>population is primarily the<br>adult population over 18<br>years of age. This product<br>may also be used with<br>children 5 years of age and<br>older. | These devices are<br>intended for the adult<br>population suffering from<br>a chronic persistent<br>ringing in the<br>ears(tinnitus), who do not<br>need or desire<br>amplification. The<br>products may be used for<br>masking tinnitus as part<br>of a tinnitus retraining<br>therapy(TRT) protocol<br>and should be utilized<br>only in consultation with<br>a licensed hearing<br>healthcare professional<br>who is trained in<br>subsequent rehabilitation<br>therapy, or a qualified<br>audiologist. |
| Warnings | Maximum daily usage<br>should be determined by<br>the hearing healthcare<br>professional. | same | same |
Page 7
.
{5}------------------------------------------------
Amplisound 510(k) Summary 03/04/2014 #KJ32962 09/17/2013
The Amplisound Quell 908TD is comparable to two devices presently on the market - the ReSound TSG Module, which should be noted is a combination hearing aid/sound generator; and TTC's GHI-T and TN3 sound generator device. While it is not a hearing aid, Quell 908TD is similar to the predicate devices in that they are all ear level worn instruments that produce therapeutic sound for tinnitus sufference between the Quell 908TD device and the predicate devices has primarily to do with the physical shape and style of the housing and the type of user volume control. The Quell 908TD has a scroll volume wheel, while the Resound has a toggle volume control and the GHI has a rotational volume control. All three devices are designed to minimize the amount of contact and occlusion of the ear canal , therefore keeping the canal open which is most desirable for sound therapy masking applications. By performing bench testing measurements with probe microphone on actual ears and 2cc ANSI tests in a test box, the sound output has been designed to mimic the predicate device's approved white noise characteristic which is a smooth response (void of any peaks) with less emphasis in the low frequencies and greater emphasis in the high frequencies. The Quell 908TD response is similar in frequency response to the predicate devices yet with lower maximum output to maintain safety and effectiveness. The Quell 908TD device is designed to be worn without any frequency adjustments for most wearers, however, the tone of the therapeutic sound is adjustable by the hearing care professional using the Hearing Studio software which employs band gain slider controls and overall gain control to adjust the frequency response of the sound as needed for individual patient needs. This software adjustment is similar to the Resound TSG module in that the low and high frequencies can be adjusted independently to 'tune' the masking sound if the user prefers a different tone. Output of the Quell 908TD device is limited within the fitting software to a maximum sound pressure level of 85 dB to meet the OSHA 29CFR 1910.95 standard. These settings include limits for the maximum user volume control which is set by the hearing healthcare provider and cannot exceed 85dBspl.
Similar to the predicate devices, Quell 908TD sound generators are programmed, when needed, using the industry standard HiPro box and programming cables. The Quell 908TD device utilizes a standard micro behind the ear instrument housing and is powered by a size 10 zinc-air battery. Non clinical tests using probe microphone real-ear and 2cc test box measurements demonstrate that the Quell 908TD performs similarly to the comparative devices on the market designed for use with tinnitus treatment. Quell miniature BTE sound generator devices are similar in look, function and sound quality to the GHI and Resound ear level tinnitus devices already approved for the market and in use and by tinnitus sufferers.
Page 8
{6}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Image /page/6/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
March 25, 2014
Amplisound . Inc. c/o Mr. Ralph Campagna President 594 Putnam Road Danielson, CT 06239
Re: K132965
Trade/Device Name: Solace Sound Generators Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: December 27, 2013 Received: December 30, 2013
Dear Mr. Campagna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Ralph Campagna
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Amplisound, Inc. Abbreviated 510(k) Submission Solace Sound Generators
# INDICATIONS FOR USE
510(k) Number: K132965
Device Name: Solace Sound Generators
Indications for Use:
Quell sound generators are designed for individuals who experience tinnitus. These devices generate a digitally controlled passive analog broad band noise from the circuit. The broad band noise is intended for use with tinnitus masking therapy. These devices are to be dispensed only by a doctor, licensed audiologist, or licensed hearing instrument specialist who are trained in the area of tinnitus treatment. BTE worn sound generators are one of several available sound stimuli means such as; computer generated audio, CD player, iPod, sound pillows, etc. Sound generators are dispensed to an adult population 18 years of age or above with specific instructions for their use.
OR
Prescription Use: ____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 901 Subpart D)
Over the Counter Use: ________________ (21 CFR 801 Subpart C)
(Please do not write below this line - continue on another page if needed)
Concurrence of CDHR, Office of Device Evaluation (ODE)
Vasant Dasika -S 2014.03.19 23:06:17 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.