K132946 · Teleflexmedical, Inc. · CCK · Jun 4, 2014 · Anesthesiology
Device Facts
Record ID
K132946
Device Name
SOFTECH PLUS ETC02 CANNULA
Applicant
Teleflexmedical, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Jun 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Device Story
SOFTECH Plus ETCO2 Cannula is a non-sterile, single-patient-use nasal cannula; functions as an adjunct to oxygen therapy. Device delivers low-flow oxygen through one nasal prong while simultaneously sampling exhaled gas from the other prong. Used in clinical settings for patients requiring oxygen therapy and ventilatory rate monitoring. Healthcare providers connect the device to oxygen supply and gas monitoring equipment; output allows clinicians to assess patient oxygenation and respiratory status. Benefits include simultaneous oxygen delivery and CO2 monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ETCO2 performance testing under simulated conditions at various oxygen flow rates, biocompatibility testing (cytotoxicity, sensitization, irritation, genotoxicity, implantation), and DEHP testing.
Technological Characteristics
Materials: Polyvinylchloride (PVC), Polypropylene, Polycarbonate. Non-DEHP PVC used. Split nasal prong design. Connectivity: Luer lock connectors for gas sampling. Single-patient use, non-sterile. No software or electronic components.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring oxygen therapy to improve blood oxygen levels and monitoring of expired gas to determine ventilatory rate.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
Additional Information Response
## 510(k) SUMMARY
### A. Name, Address, Phone and Fax Number of Applicant
JUN 0 4 2014
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8050 919-433-4996 Fax:
### B. Contact Person
Angela Bouse Senior Regulatory Affairs Specialist
### C. Date Prepared
June 3, 2014
### D. Device Name
| Trade Name: | SOFTECH® Plus ETCO2 Cannula |
|--------------------------|----------------------------------------------|
| Classification Name: | Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase |
| Product Code: | CCK |
| Additional Product Code: | CAT |
| Regulation Number: | 868.1400 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
### E. Predicate Device
This submission demonstrates substantial equivalence to the predicate device Double Lumen Oxygen Delivery Gas Sampling Nasal Cannula, Hudson RCI Bi-Flo -K961150.
## F. Device Description
The SOFTECH Plus ETCO2 Cannula is a non-sterile disposable, single patient use device that acts as an adjunct to oxygen therapy with its primary function providing a means to deliver low flow oxygen, while sampling part of the patients exhaled gas. The SOFTECH Plus ETCO2 Cannula has a split nare blank with oxygen delivery through one nasal prongs while allowing sampling of the patient's exhaled gas from the corresponding nasal prongs.
{1}------------------------------------------------
## G. Indications for Use
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Patient Population: Infant, Pediatric, Adult
# H. Technological Characteristics and Material Comparison to the predicate
The proposed SOFTECH Plus ETCO2 Cannula is substantially equivalent to the predicate device listed above in that the indications for use, the intended use, and fundamental scientific technology remain unchanged. Table 1 summarizes the differences between the proposed and predicate devices.
| Features | Predicate<br>K961150<br>Double Lumen Oxygen<br>Delivery Gas Sampling<br>Nasal Cannula, Hudson<br>RCI Bi-Flo | Proposed<br>SOFTECH® Plus ETCO2 Cannula | Performance<br>Testing |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Nasal Prongs<br>(Nares) | Dual-channel nasal prong<br>that allows oxygen delivery<br>and gas sampling from both<br>nares | Split nasal prong that directs oxygen<br>flow into one nare and samples<br>expired gas from the other nare | ETCO2<br>Performance<br>Testing with<br>Simultaneous<br>Oxygen Delivery |
| Oxygen supply<br>tubing and CO2<br>sampling line<br>length | Oxygen supply tubing: 7 ft<br>CO2 sampling line: 7 in | Oxygen supply tubing: 7 ft and 14 ft<br>CO2 sampling line: 2 in, 7 ft and 14 ft | ETCO2<br>Performance<br>Testing with<br>Simultaneous<br>Oxygen Delivery |
| Materials | | | |
| Nares | Polyvinylchloride | Polyvinylchloride | Biocompatibility<br>Testing: |
| Bolo Slide | Polycarbonate or<br>Polyvinylchloride | Polyvinylchloride | Testing:<br>Cytotoxicity, |
| Tubing Clip | None | Polypropylene | Sensitization, |
| Lariat Tubing | Polyvinylchloride | Polyvinylchloride | Irritation, |
| Oxygen Supply<br>Tubing | Polyvinylchloride | Polyvinylchloride | Genotoxicity,<br>Implantation |
| CO2 Sampling<br>Line | Polyvinylchloride | Polyvinylchloride | |
| Oxygen<br>Connector | Polyvinylchloride | Polyvinylchloride | |
| Tubing Connector | Polyvinylchloride | Polyvinylchloride | |
| Male and Female<br>Luer Lock<br>Connectors | Acrylic | Polycarbonate | |
| 22mm Oxygen | None | Polypropylene | |
| Table 1 - Differences Between the Proposed and Predicate Devices<br> |
|----------------------------------------------------------------------|
|----------------------------------------------------------------------|
Teleflex Medical, Inc.
{2}------------------------------------------------
Additional Information Response
## K132946 – SOFTECH Plus ETCO2 Cannula Tab 006 - Section 011 - 510(k) Summary
| Tubing Adaptor | | | |
|----------------|---------------|---------------|--------------|
| Adhesive | Cyclohexanone | Cyclohexanone | |
| PVC | PVC with DEHP | Non-DEHP PVC | DEHP Testing |
## I. Performance Data
The proposed device was tested for ETCO2 sampling at various oxygen flow rates, biocompatibility, and DEHP testing. The test results demonstrate that the device is substantially equivalent to the predicate device.
The tests performed are summarized in Table 2 below.
| | 114 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 | Table 2 - Performance Testing Summary | |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|--|
| | | | |
| General Description |
|--------------------------------------------------------------------------------------------------------------------------------|
| ETCO2 performance in simulated conditions at various oxygen flow rates |
| Biocompatibility testing as per ISO 10993-1: cytotoxicity, sensitization, irritation,<br>genotoxicity and implantation testing |
| DEHP Testing |
## J. Conclusion
The device data and test results demonstrate that the device is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo features a stylized human figure with three arms reaching upwards, symbolizing health, hope, and well-being. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
#### June 4, 2014
Teleflex Medical, Inc. Angela Bouse Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K132946
Trade/Device Name: SOFTECH® Plus ETCO2 Cannula Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK, CAT Dated: May 02, 2014 Received: May 05, 2014
Dear Ms. Bouse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of general voluless provincitive, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Angela Bouse
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MarySuBunner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132946
Device Name
Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula
Indications for Use (Describe)
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Patient Population: Infant, Pediatric, Adult
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
Paul Career For Comment FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR SECTION CONSE
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Todd D. Courtney -S
2014.06.03 16:33:20 -0400
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# ·DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.