EQUASHIELD CLOSED SYSTEM DRUG TRANSFER DEVICE (CSTD)

K132899 · Equashield Medical , Ltd. · ONB · May 13, 2014 · General Hospital

Device Facts

Record IDK132899
Device NameEQUASHIELD CLOSED SYSTEM DRUG TRANSFER DEVICE (CSTD)
ApplicantEquashield Medical , Ltd.
Product CodeONB · General Hospital
Decision DateMay 13, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

Closed System drug Transfer Device (CSTD) for preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress.

Device Story

Equashield is a multi-component Closed System drug Transfer Device (CSTD) used in hospitals/clinics by pharmacists and healthcare professionals. System components include a piston syringe unit, vial adaptor, infusion bag spike adaptors, luer lock adaptors, and protective plugs. Operation relies on a double-membrane method; when components join, membranes press together and are pierced by needles to create a leak-proof, drug-residual-free connection. The syringe unit features an airtight, sealed barrel isolating the plunger rod and interior; an integrated closed pressure equalization chamber containing sterile air prevents aerosol/vapor escape. The system requires no priming and is not a gravity-feed system. Accessories include Y-site tubing and secondary tubing with drip chambers. By mechanically prohibiting environmental contaminant transfer and drug/vapor escape, the device minimizes exposure to hazardous drugs and prevents microbial ingress. The system is single-use, non-pyrogenic, and provided sterile.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing and performance data demonstrating compliance with recognized consensus standards, including ISO 10993 for biological evaluation (cytotoxicity, blood interaction), ISO 594 for luer fittings, and ISO 23908 for sharps injury protection. The system was substantiated to prevent microbial ingress for up to 7 days and supports multiple component reconnections up to 10 times.

Technological Characteristics

Multi-component system; components include piston syringe, vial/spike/luer adaptors, and protective plugs. Materials evaluated per ISO 10993-1, -4, -5. Features double-membrane (septum) connection mechanism; integrated closed pressure equalization chamber; needle-stick safe design. Sterile, single-use, EO sterilized (SAL 10^-6). No DEHP plasticizer in tubing. Complies with ISO 594-1/2 (luer tapers), ISO 7886-1 (syringes), ISO 8536-4 (infusion equipment), and ISO 23908 (sharps protection).

Indications for Use

Indicated for use by licensed pharmacists and healthcare professionals for the preparation, compounding, and administration of drugs, including antineoplastic and hazardous drugs, in hospital and clinical settings.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 | Israel www.equashieid.com #### 510(k) Summary K132899 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Date Prepared: May 12, 2014 Applicant Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 , Israel Contact Person Raymond Kelly Licensale Inc. 57 Lazy Brook Rd Monroe, CT 06468 USA Phone: (203) 880-4091 Alternate Contact Dorit Valal Projects Manager Equashield Medical Ltd. POB 12 , Tefen Industrial Park Migdal Tefen 24959 . Israel Phone: +972-49873737 (ex. 114) Cell: +972-549952810 Device Information Trade Name: Equashield Closed System drug Transfer Device (CSTD) Model Numbers: Generations 1 & 2: (VA-x, SU-x, SA-x, LL-x, MC-x, FC-x, PP-x) Classification Name: Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System Review Panel: General Hospital Product Code: ONB Device Class: Class II Regulation: 21 C.F.R. §880.5440 ### Predicate Device Information K101940 Equashield Closed System, cleared November 19, 2010 K 130197 PhaSeal Closed System, cleared February 27, 2013 {1}------------------------------------------------ ### Intended Use/Indications for Use Closed System drug Transfer Device (CSTD) for safe preparation, reconstitution, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress. ### Principle of Operation Equashield (first and second generation) is a device which has been substantiated to prevent microbial ingress up to a time period of 7 days and has been substantiated for multiple reconnections of components up to 10 times. Equashield is a Closed System drug Transfer Device (CSTD) which does not require priming, does not include an Insertion Tube, and is not a "Gravity Feed" System. Equashield does contain accessories which contain tubing, one accessory uses "drip therapy" but no tubing contains plasticizer DEHP. Equashield is a single use, nontoxic, non-pyrogenic fluid path system which is provided sterile as long as the package is intact, undamaged, and protective caps are secure. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress. System contains filters and needles using a needle stick safe design, ### Device Description The Equashield system consists of a piston syringe (Syringe Unit), an adaptor to the medication vial (Vial Adaptor), an adaptor for the infusion bag for injection (Spike Adaptor 1), an adaptor for the infusion bag for withdrawal (Spike Adaptor W), an adaptor for female luer lock ports (Luer Lock Adaptor), nursing pair connectors for male and female luer lock ports (Female LL Connector and Male LL Connector), and a protective plug (Protective Plug). The Equashield closed system consists of 8 components that are dedicated to each other to create a system. The components are not intended to be used individually. Vial Adaptor is 48 x 31mm, 23mm at distal end. Plug is 51 x 17mm large. Female luer lock is 50 x 23 mm with a 5.2mm ID. Male luer lock is 42 x 21mm with a 15mm depth. Spike Adaptor is 75 x 42 mm and the height is 22mm. ### Accessories Description Y-Site Tubing - this accessory consists of a Luer Lock Adaptor permanently attached to a Y-Site. Short pieces of tubing extend from each of the other two Y-Site ports, creating one tubing line. One tube is terminated with a female luer lock and the other is terminated with a male luer lock. The Y-Site Tubing Accessory is intended to be attached in series to an infusion line and the Equashield Luer Lock Adaptor enables safe connection and injection with the Equashield Syringe Unit or Female Connector. Priming volume: 1.9ml, length 25cm, Y-angle 30 degrees. {2}------------------------------------------------ Spike Adaptor Secondary Tubing - consists of a Spike Adaptor-1 for injection with a drip chamber at its outlet. The drip chamber with the attached tubing, the roller clamp and the male luer lock at the tubing end, all constitute standard secondary infusion tubing sets. The integration of the standard infusion tubing set with the Equashield Spike Adaptor provides a closed system protection during injection of drugs into the infusion bag and during their administration. It reduces the assembly steps and the risk of exposure to drugs associated with such assembly procedures. Length is 100 cm, priming volume 7.5ml. ### Performance No performance standards or special controls have been developed under Section 514 of the Federal Food, Drug, and Cosmetic Act ("FD&C Act") for Closed System Transfer Devices (CSTD). However, the Equashield System Generation 2, its accessories, and its components follow FDA recognized consensus standards: - . Recognition Number 6-11; ISO 594-1:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements - . Recognition Number 6-129: ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings - . Recognition Number 6-170: ISO 7886-1:1993, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use - Recognition Number 6-245: ISO 8536-4:2010, Infusion equipment for medical use Part . 4: Systems for single use, gravity feed - Recognition Number 2-156: ISO 10993-1:2009 Biological evaluation of medical devices . -- Part 1: Evaluation and testing within a risk management process - ISO 10993-4:2002, Biological evaluation of medical devices -- Part 4: Selection of tests . for interactions with blood - Recognition Number 2-153: ISO 10993-5:2009, Biological evaluation of medical devices . -- Part 5: Tests for In Vitro cytotoxicity - . Recognition Number 5-40: ISO 14971:2007, Medical devices - Application of risk management to medical devices - . ISO 22413:2010, Transfer sets for pharmaceutical preparations -- Requirements and test methods - Recognition Number 6-273: ISO 23908:2011, Sharps injury protection Requirements . and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling ### Substantial Equivalence Equivalence was determined using a side by side tabular comparison between the predicate and proposed devices which included: Features, Intended Use, Labeling, Materials, Specifications, Performance Data, and Technological Aspects {3}------------------------------------------------ | Predicate Device (K13019 | Same | system and the escape of drum vapor contractions outside of drug system thereby minimizing ndividual and environments<br>System is an airtight and leak-<br>sposure to drug vapor, aeroso and spills. The system also prevents microbial ingress.<br>vironmental contaminants | Class II | 888.544( | ONB | |---------------------------|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------------|------------| | Predicate Device (K101940 | Same | nacists or other healthcar<br>fessionals to prepare drugs<br>ned system to be used | Class II | 888.5440 | r H | | Proposed Device | Same | | | 888.544( | ONB | | | tended Us | dications for<br>Use | assificatio | Regulation Number | roduct Cod | SE Table of Attributes Page 32 {4}------------------------------------------------ | | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) | |---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Vial Adaptor | Vial Adaptor | Vial Adaptor ("Protector") | | | Syringe Unit | Syringe Unit | Syringe connector ("Injector"<br>with female luer lock) | | | Spike Adaptor I- for injection | Spike Adaptor I- for injection | Spike Adaptor ("Infusion<br>Adaptor") | | | Spike Adaptor W- for withdrawal | Spike Adaptor W- for withdrawal | | | | Luer Lock Adaptor | Luer Lock Adaptor | Male "Connector Luer Lock" | | System<br>components | Male Luer Lock Connector | Male Luer Lock Connector | Male "Connector Luer Lock" | | | Female Luer Lock Connector | Female Luer Lock Connector | Female luer lock connector<br>("Injector") | | | Protective Plug | Protective Plug | NA | | | Y-Site Tubing<br>(Accessory) | NA | Luer Lock Adaptor with Y-site<br>tubing ("Y-Site Connector") | | | Secondary Tubing<br>(Accessory) | NA | Spike adaptor with secondary<br>tubing ("Secondary Set with Drip<br>Chamber") | | Characteristics | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug | Closed System used for<br>antineoplastic and hazardous drug<br>reconstitution, transfer and<br>administration, in order to prevent<br>contamination of the surrounding<br>environment and of the drug | | Principles of<br>Operation | Multi-component system.<br>Components are intended to be used<br>as a system. | Multi-component system.<br>Components are intended to be used<br>as a system. | Multi-component system.<br>Components are intended to be<br>used as a system. | | Technological<br>Characteristics | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) | | | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system. | A leak-proof connector with a<br>single use syringe permanently<br>attached to it as part of the system. | A leak-proof connector<br>("Injector") that needs to be<br>connected to an off the shelf<br>single use syringe. | | | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization. | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two membranes<br>are pressed together and then<br>pierced by needles. System has<br>integrated closed pressure<br>equalization. | All system components are sealed<br>with resealing membranes<br>(Septum). When components are<br>joined together the two<br>membranes are pressed together<br>and then pierced by needles.<br>System has integrated closed<br>pressure equalization. | | | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel. | The system syringe is closed from<br>all sides. The syringe barrel is<br>sealed airtight also at its rear end,<br>thereby isolating the plunger rod<br>and the interior of the barrel. | NA | | | Proposed Device | Predicate Device (K101940) | Predicate Device (K130197) | | Secondary Tubing<br>("Secondary Set with Drip Chamber") | Spike Adaptor with drip chamber and secondary tubing attached. | NA | Spike Adaptor ("infusion adaptor") with drip chamber and secondary tubing attached. | | System | A fully encapsulated Syringe Unit is the active transfer device of this closed system.<br><br>The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.<br><br>For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | A fully encapsulated Syringe Unit is the active transfer device of this closed system.<br><br>The syringe barrel is sealed airtight also at its rear end, thereby isolating the plunger rod and the interior of the barrel. A leak-proof connector is permanently welded to the syringe. The closed pressure equalization system with a chamber containing sterile air is built-in the Syringe Unit and makes the system airtight consequently containing all aerosols, particles and vapors.<br><br>For transfer of fluids the Syringe Unit connects to the passive Vial Adaptors, infusion bag and infusion tubing adaptors of the system, using the double membrane method to create a leak-proof and drug residual-free connection. | The system uses a double membrane to create a leak-proof connection. The closed expansion chamber makes the system airtight consequently containing all aerosols, particles and vapors. The expansion chamber also equalizes the pressure within the system. It can handle most types of IV sets and syringes to enable a totally safe and convenient administration of drugs. | | Device Type | Rx/Single Use | Rx/Single Use | Rx/Single Use | | Target Users | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals | Licensed Pharmacists/Health Care Professionals | | Environment | Hospitals and clinics | Hospitals and clinics | Hospitals and clinics | | Sterilization | EO/SAL 10-6 | EO/SAL 10-6 | EO/SAL 10-6 | {5}------------------------------------------------ ... . Page 34 the country of the county of and the comments of the country of {6}------------------------------------------------ : : : . ﺮ . Page 35 {7}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ### May 13, 2014 Equashield Medical Ltd. c/o Raymond J Kelly IV Arazy Group 57 Lazy Brook Rd MONROE, CT 06468 Re: K132899 Trade/Device Name: Equashield Closed System drug Transfer Device (CSTD) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: II Product Code: ONB Dated: April 14, 2014 Received: April 17, 2014 Dear Mr. Kelly, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {8}------------------------------------------------ Page 2 - Mr. Kelly forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. · Sincerely yours, ## Mary S. Runner -S Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K 132899 #### Device Name Equashield Closed System drug Transfer Device (CSTD) ### Indications for Use (Describe) Closed System drug Transfer Device (CSTD) for preparation, compounding and administration of drugs, including antineoplastic and hazardous drugs. This closed system mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills and also prevents microbial ingress. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) . . . . ### Digitally signed by Richard C. Chapman Date: 2014.05.13 14:21:18 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...