← Product Code [OLP](/productcode/OLP) · K132838

# NUTRA LIGHT BLUE (K132838)

_Nutra Luxe MD, LLC · OLP · May 23, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132838

## Device Facts

- **Applicant:** Nutra Luxe MD, LLC
- **Product Code:** [OLP](/productcode/OLP.md)
- **Decision Date:** May 23, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Nutra Light Blue is a non-invasive LED light device is intended/indicated for over- the-counter use. The device delivers low intensity light emitting diodes (LED's) in the blue region at a wavelength of 410 +/- 5 nanometers, for the treatment of mild to moderate inflammatory acne.

## Device Story

Nutra Light Blue (Model: NL-2213) is a handheld, non-invasive LED light device for home use. It utilizes high spectral purity LEDs to emit blue light at 410 +/- 5 nm wavelength at 75 mW. The device includes a power supply, internal Ni-MH rechargeable battery, and control unit housed in medical-grade injection-molded plastic. The user operates the device by moving it over the skin surface for 3 minutes per target area, following a regimen of 2 treatments per week for 6 weeks. A physical STOP button allows immediate power removal. The device provides uniform, hot-spot-free illumination to treat mild to moderate inflammatory acne. It does not use software.

## Clinical Evidence

An OTC usability study was conducted with 80 participants to evaluate intended user representation, self-selection based on labeling, and the ability to correctly assemble, operate, and care for the device. All study goals were met.

## Technological Characteristics

Medical-grade biocompatible plastics (injection molded). Light source: LEDs, 410 +/- 5 nm wavelength, 75 mW output. Power: Internal Ni-MH rechargeable battery. Standards: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), IEC 60601-2-57 (Non-Laser Light), IEC 62471 (Photobiological Safety), ISO 10993 (Biocompatibility). No software.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Evis MD Platinum Blue ([K082232](/device/K082232.md))
- Tanda Skincare System (0800591)
- Quasar Blue ([K093963](/device/K093963.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the words "K132838 MAY 23 2014 nutra luxe MD NUTRA LIGHT". The words "NUTRA LIGHT" are in a larger font size than the other words. The words "nutra luxe MD" are in a smaller font size and are located above the words "NUTRA LIGHT". The date "MAY 23 2014" is above the words "nutra luxe MD".

# 510K Summary of Safety and Effectiveness for Nutra Light Blue

| 1. | General Information       |                                                                                                                                    |
|----|---------------------------|------------------------------------------------------------------------------------------------------------------------------------|
|    | Submitter:                | NutraLuxe MD, LLC<br>6835 International Center Blvd Suite 4-5<br>Fort Myers, Florida 33912                                         |
|    | Contact Person:           | C/O Jill Creasy<br>RIC - Regulatory & ISO Consulting<br>5575 Santa Rosa Ct.<br>Sparks, NV 89436<br>775-622-9591<br>info@ricreg.com |
|    | Summary Preparation Date: | May 15, 2014                                                                                                                       |
| 2. | Device Name               |                                                                                                                                    |
|    | Device Name:              | Nutra Light Blue (Model: NL-2213)                                                                                                  |
|    | Classification Name:      | Laser Surgical Instrument, for use in General and Plastic<br>Surgery and in Dermatology,                                           |
|    | Regulation Numbers:       | 21 CFR 878.4810                                                                                                                    |
|    | Regulatory Class:         | II                                                                                                                                 |
|    | Product Codes:            | OLP                                                                                                                                |

# 3. Predicate Device

The Nutra Light Blue is substantially equivalent to the Evis MD Platinum Blue (K082232), Tanda Skincare System (0800591) and Quasar Blue (K093963)

{1}------------------------------------------------

#### Device Description 4.

The Nutra Light Blue Model: NL-2213, is deviccs that utilize Light Emitting Diodes to provide LED light to the body. The Nutra Light Blue is a visible light source of high spectral purity. They provide uniform or "hot-spot" free illumination. The output is pre-tuned to one wavelength with a narrow spectral bandwidth. The output wavelength of Blue is 410 +/- 5 at 75 mW. The Nutra Light Blue unit contains the power supplies, internal NI-MH rechargeable battery and the control unit. The enclosure and LED lens are made of medical grade biocompatility plastics via injection molding. It has the capabilities to deliver the light to the skin as the unit is moved over the skin surface. The STOP button directly on the unit allows the user to immediately remove all power to the unit. The Nutra Light Blue does not use any software.

#### 5. Intended Use and Indications:

The Nutra Light Blue is a non-invasive LED light device is intended/indicated for over- the-counter use. The device delivers low intensity light emitting diodes (LED's) in the blue region at a wavelength of 410 +/- 5 nanometers, for the treatment of mild to moderate inflammatory acne.

#### Substantial Equivalency & Comparison of Technological Similarities and Differences 6.

The intended use and technological characteristics of the Nutra Light Blue are equivalent to the listed predicates.

- Has the same intended use as the predicates (i.e., Treatment of mild to moderate acne). 1.
- Has a similar output (i.e., 75 mW/cm- ) as the predicates 2.
- Utilizes a similar wavelength (i.e., 410 nm) as the predicates 3.
- Utilizes a same treatment duration (i.e., 3 minutes per target area) as the predicates 4.
- Utilizes a similar treatment regimen (i.e., 2 treatments per week for 6 weeks) 5.

Any differences between the Nutra Light Blue and the predicates device are not significant to its safety or effectiveness for its intended use.

#### Clinical Performance Data 7.

An OTC Usability Study was conducted with the following four goals in mind 80 participants;

- 1. To attract participants that represented the "intended users" of the device;
- 2. To determine if consumers could correctly self-select using the packaging labeling only
- 3. To test consumer knowledge of the packaging labeling and user manual and actually assembly, operate and care for the device correctly. All four goal of the study were met.

#### 8. Nonclinical Performance Data

The Nutra Light Blue device was tested and complies with Electrical Safety and EMC testing, which include the requirements of IEC 60601-1 "Medical Electrical Equipment Part 1 - General Requirements for Safety" IEC 60601-1-2 "Medical Electrical Equipment Part 1-2, General Requirements for Safety -Collateral Standard Electromagnetic Compatibility Requirements and Tests. IEC 60601-2-57, Medical Electrical Equip. - Part 2-57 Basic Safety and Performance of Non-Laser Light Cosmetic/Aesthetic use. IEC 62471 Photobiological Safety of lamps and lamp systems. In addition, testing and analysis have demonstrated compliance of the plastic within ISO 10993 (Biocompatibility).

#### 9. Regulatory Requirements

NutraLuxe MD manufactures under strict quality assurance guidelines and US FDA Good Manufacturing Practice (GMP)

Nutra Luxe MD is fully compliant with 21 CFR Part 820, US FDA Quality Systems Regulations and risk analysis conforms to ISO 14971.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Nutra Luxe MD Nutra-Light. The words "nutra luxe MD" are in a smaller font size and are located above the words "NUTRA-LIGHT", which are in a larger font size. There is a flower-like symbol above the word "luxe". The logo is surrounded by some decorative lines.

## 10. Substantial Equivalence

Based upon an analysis of the overall performance characteristics the Nutra Light Blue device was found to have the same intended use and indication for use as the predicate devices. The device also has similar technological characteristics to its predicate devices. The minor technological differences between the Nutra Light Blue and predicate devices do not raise any issues of safety or effectiveness.

Therefore, Nutra Luxe MD, Inc. found the Nutra Light Blue device to be substantially equivalent to the legally marketed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird-like figure.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

-

Public Health Service

Food und Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

May 23, 2014

NutraLuxe MD, LLC % Ms. Jill Creasy Regulatory & ISO Consulting 5575 Santa Rosa Court Sparks, Nevada 89436

Re: K132838

Trade/Device Name: Nutra Light Blue Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OLP Dated: May 6, 2014 Received: May 14, 2014

Dear Ms. Creasy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnacunent date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

1f your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Jill Creasy

device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/Medica]Devices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K132838

Device Name Nutra Light Blue

Indications for Use (Describe)

The Nutra Light Blue is a non-invasive LED light device is indicated for over- the -counter use for the treatment of mild to moderate inflammatory acne.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/5/Picture/13 description: The image contains the name "Neil R Ogden -S" and a timestamp. The timestamp reads "2014.05.23 15:30:23 -0410". There is also a logo to the right of the timestamp.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the i me to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K132838](https://fda.innolitics.com/device/K132838)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
