The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
Device Story
SFI Bar® Implant Abutments serve as connection interfaces between dental implants and removable overdentures. The system consists of an abutment screwed into an endosseous dental implant, which then attaches to a stress-free bar to support the prosthesis. Used in dental clinics by practitioners to restore chewing function in edentulous patients. The device provides a stable prosthetic framework. Benefit includes retention and support for removable dental prostheses.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via tolerance analysis of platforms and functional/torque testing. Acceptance criteria met.
Technological Characteristics
Material: Wrought Titanium 6 Aluminum-4 Vanadium ELI Alloy (ASTM F136). Device is a mechanical dental abutment. Non-sterile. No software or energy source.
Indications for Use
Indicated for partially or totally edentulous patients requiring removable dental prostheses to restore chewing function. Used with dental implants as a prosthetic framework.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
SFI Bar® Implant Abutments for 7 Platforms (K130183)
{1}------------------------------------------------
K13 27164
Fr: 2 5 2014
Sterngold Dental, LLC Abbreviated 510(k) Premarket Notification
September 6, 2013 SFI Bar® Implant Abutments for 9 Implant Platforms
# 510(k) Summary
| Sponsor: | Sterngold Dental, LLC<br>23 Frank Mossberg Drive<br>Attleboro, MA 02703 |
|----------------------|----------------------------------------------------------------------------------------------------------------|
| Contact: | Maria Rao, QA/RA Director<br>Ph: 508-226-5660 ext 1206 |
| Trade Name: | SFI Bar® Implant Abutments |
| Common Name: | Implant Abutments |
| Classification Name: | Endosseous Dental Implant Abutment |
| Classification: | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II |
| Product Code: | NHA (21CFR 872.3630) |
Legally Marketed Device to which Equivalence is claimed (Predicate Devices): Predicate Device(s): K130183, K102382.
| K130183 | SFI Bar® Implant Abutments for 7 Platforms |
|---------|------------------------------------------------------------------|
| K102382 | SFI Bar® Implant Adapter Straumann, SFI Bar® Implant Adapter Neo |
#### Description of Device:
Device Description: The SFI Bar® Implant Abutments provide the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI Bar® Implant Abutments consist of an abutment, which is attached to a stress free bar for the fixation of removable overdentures. The implant abutment is screwed into the dental abutment.
The implant abutments fit Straumann Bone Level RP, Straumann Bone Level NP, Ankylos, Nobel 5mm Trilobe, Nobel Conical RP, Nobel Conical NP, Astra 4.5/5.0, Astra 3.5/4.0, Zimmer TSV 5.7mm.
### See table 1 for platform compatibility.
There are nine (9) different platforms and each platform is compatible with one or more implant types. The platforms of the abutments are [BD], [BE], [AE], [AN], [AY], [AY], [AJ], [AK], [BF]. Table 1 indicates which implants are compatible with these platforms. The difference between each platform is the internal connection with the specific implant.
The devices are supplied non-sterile, therefore there is no shelf life.
{2}------------------------------------------------
# Abutment Insertion
Choose the abutment with the proper cuff height that fits on the existing implant. Abutment platforms should be 1 to 2 mm above the gingival level and approximately parallel to the occlusal plane. However, to allow the subsequent placement of the sections of SFI-Bar on top of these abutments so that the bars are approximately parallel to the occlusal plane, it may be necessary to choose some abutments with different heights. Abutments are screwed into each implant.
### Intended Use of the Device:
The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
Model Implant Brand Straumann Bone Level RC, Blue Sky Bio Square Taper RC Straumann Straumann Bone Level NC, Blue Sky Bio Square Taper NC Straumann Ankylos Ankvlos Nobel Replace WP, Nobel Replace Select WP, NobelSpeedy Replace WP, Nobel Biocare Implant Direct 5.0 RePlant, BlueSky Bio 5.0 Trilobe Nobel Biocare Nobel Conical Connection RP Nobel Conical Connection NP, Blue Sky Bio Max Nobel Biocare Astra 4.5 / 5.0, Blue Skv Bio Conus 12 4.5 / 5.0 Astra Dental Astra 3.5 / 4.0. Blue Sky Bio Conus 12 3.5 / 4.0 Astra Dental Zimmer TSV 5.7mm, Implant Direct Legacy 5.7, BioHorizon 5.7 Zimmer Dental
The SFI-Bar® Implant Abutments are compatible with the following implant systems:
# Summary Technological Characteristics:
The proposed implant abutments are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.
The material of the implant abutments conform to ASTM F136, Wrought Titanium 6 Aluminum-4 Vanadium ELI Alloy.
# Comparison/Compatibility
# Substantial Equivalence:
The proposed implant abutments are substantially equivalent to the currently marketed predicate devices. The intended use, basic design, fundamental operating principles and manufacturing procedures are the same as the predicate devices.
To ensure compatibility the following process was carried out: The implant abutments were designed and developed, and manufactured according to manufacturer's
specifications and controlled procedures. A validation protocol was done in accordance with Design Control requirements per FDA CFR820.30.
Table 2 summarizes the substantial equivalence comparison to the predicate devices.
{3}------------------------------------------------
#### Performance Data:
Application and functional testing have been conducted to evaluate the performance characteristics of the SFI Bar® Implant Abutments. The test methods used were the same as in predicate devices. Testing has shown that the SFI-Bar® Implant Abutments included in this application are equivalent in performance characteristics to the predicate SFI Bar®. The acceptance criteria were met.
# Summary of Testing to Demonstrate Safety and Effectiveness / Conclusion:
Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. The non-clinical testing consisted of tolerance analysis of platforms -to-identify-worst-case-test-samples ... Fatigue testing .was-not-done as the basic design is the same as the predicate devices. The evaluation was based on FDA guidance "Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments." Torque tests, application and functional tests have been carried out.
The summary of technological characteristics as well as application and functional testing indicate that the device is safe and effective for its intended use and performs as well or better than the predicate devices.
13.4
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
#### February 25, 2014
Sterngold Dental LLC Ms. Maria Rao Quality Assurance Director 23 Frank Mossberg Drive Attleboro, MA 02703
Re: K132814
Trade/Device Name: SFI Bar Implant Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 21, 2014 Received: January 22, 2014
Dear Ms. Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Ms. Cheng
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/8 description: The image shows the name "Kwame O. Ulmer" in bold, black font. Below the name, there is a faded, patterned design that is difficult to discern. The letter "S" is located to the left of the design, also in bold, black font.
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Sterngold Dental, LLC Abbreviated 510(k) Premarket Notification
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.. . . . . .
September 6, 2013 SFI Bar® Implant Abutments for 9 Implant Platforms
# INDICATIONS FOR USE STATEMENT
# SECTION 2
:
.. . . . . . . . . .
and the country of the country of the country of the country
2. |
..
- •
,
・
: . · ﺑ . . .
:
.
·
: .
············ :
. . .
{7}------------------------------------------------
510(k) Number (if known): K132814
#### Device Name: SFI Bar® Abutments for 9 Implant Platforms
# Indications for Use:
The SFI-Bar® Implant Abutments are indicated to be used with dental implants as a prosthetic framework to support and /or retain removable dental prostheses in the treatment of partially or totally edentulous patients to restore chewing function. The abutment screws are intended to secure the bar to the endosseous implants.
The SFI-Bar® Implant Abutments are compatible with the following implant systems:
| Implant Brand | Model |
|---------------|---------------------------------------------------------------------------------------------------------------------------|
| Straumann | Straumann Bone Level RC, Blue Sky Bio Square Taper RC |
| Straumann | Straumann Bone Level NC, Blue Sky Bio Square Taper NC |
| Ankylos | Ankylos |
| Nobel Biocare | Nobel Replace WP, Nobel Replace Select WP, NobelSpeedy Replace WP,<br>Implant Direct 5.0 RePlant, BlueSky Bio 5.0 Trilobe |
| Nobel Biocare | Nobel Conical Connection RP |
| Nobel Biocare | Nobel Conical Connection NP, Blue Sky Bio Max |
| Astra Dental | Astra 4.5 / 5.0, Blue Sky Bio Conus 12 4.5 / 5.0 |
| Astra Dental | Astra 3.5 / 4.0, Blue Sky Bio Conus 12 3.5 / 4.0 |
| Zimmer Dental | Zimmer TSV 5.7mm, Implant Direct Legacy 5.7, BioHorizon 5.7 |
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Mary S. Runner - S 2014.02.24 0.28.46 05.00 Over-the -Counter Use AND/OR Prescription Use × (21 CFR 807 Subpart D) (Part 21 CFR 801 Subparts D) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. . ...
• • •
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.