BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- MEROPENEM (0.125-32 UG/ML)

K132674 · Becton, Dickinson and Company · LON · Dec 27, 2013 · Microbiology

Device Facts

Record IDK132674
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- MEROPENEM (0.125-32 UG/ML)
ApplicantBecton, Dickinson and Company
Product CodeLON · Microbiology
Decision DateDec 27, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

The BD Phoenix™ Automated Microbiology System is intended for in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and Non-Enterobacteriaceae.

Device Story

System performs automated antimicrobial susceptibility testing (AST) using broth microdilution in molded polystyrene panels. Input: pure bacterial culture suspension (0.5 McFarland) in AST broth with redox indicator. Operation: panels incubated at 35°C; instrument monitors growth via redox indicator color change (blue to pink) and turbidity every 20 minutes. Output: MIC values and categorical interpretations (S, I, R). Used in clinical microbiology labs; operated by laboratory technicians. Expert system applies CLSI-derived rules to interpret results. Benefits: rapid (4-16 hours) quantitative susceptibility data to guide antibiotic therapy for infections.

Clinical Evidence

Bench testing only. Performance evaluated using 1089 clinical isolates and 113 challenge isolates. Overall Essential Agreement (EA) 97.8% and Category Agreement (CA) 98.5% compared to CLSI broth microdilution reference method. Major errors 0.3% (n=3), very major errors 0.1% (n=1). Reproducibility >95%. Quality control isolates (E. coli ATCC 25922, P. aeruginosa ATCC 27853) met acceptable ranges.

Technological Characteristics

Broth-based microdilution in molded polystyrene panels (136 wells). Redox indicator (colorimetric oxidation-reduction) for growth detection. Energy source: instrument-based incubation (35°C ± 1°C). Connectivity: barcode scanned, instrument-integrated. Software: rule-based expert system for result interpretation. Sterilization: not specified.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae, Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus, Enterococcus, Streptococcus) from pure culture. Meropenem indicated for E. coli, K. pneumoniae, P. aeruginosa, C. koseri, C. freundii, E. cloacae, K. oxytoca, M. morganii, P. vulgaris, S. marcescens.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132674 # DEC 2 7 2013 #### 510(k) SUMMARY | SUBMITTED BY: | Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4827 Fax: 410-316-4499 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT NAME: | Monica Giguere, RAC Regulatory Affairs Project Manager | | DATE PREPARED: | August 22, 2013 | | DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System - Meropenem (0.125-32 µg/mL) | | DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | DEVICE CLASSIFICATION: | 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System. (Product Code LON) | | PREDICATE DEVICES: | VITEK® System (PMA No. N50510) | | INTENDED USE: | The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae. | ## DEVICE DESCRIPTION: The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components: - BD Phoenix instrument and software. . - BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial . agents for AST determinations. - BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. � - BD Phoenix AST Broth used for performing AST tests only. ● - BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth o determination. {1}------------------------------------------------ The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument. Inoculum for use with the Phoenix system may be prepared either manually or may be automated using the BD Phoenix™ AP System. The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations. The instrument houses the panels where they are continuously incubated at a nominal temperature of 35° ± 1°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible). ## DEVICE COMPARISON: The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance for inoculum prepared manually and inoculum prepared with the BD Phoenix™ AP instrument when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram negative ID/AST or AST only Phoenix panels with this antimicrobial agent. ## SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING: The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", August 28, 2009. Shelf life (stability data) for the drug is being collected and will be maintained on file at BD as indicated in the guidance document. ## Site Reproducibility Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Negative Phoenix panels containing this antimicrobial agent and associated reagents. {2}------------------------------------------------ The results of the study demonstrate that for this antimicrobial agent and the Gram-negative organisms tested there was an overall reproducibility across test sites of greater than 95% (+/- 1 dilution) agreement when compared to the test mode. ## Clinical Studies Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with a Gram Negative Phoenix Panel containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or not susceptible). The following table summarizes the performance for Clinical and Challenge isolates tested in this study. | Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) | |---------------|----------------|--------|--------|--------|--------| | Meropenem | 0.125-32 µg/mL | 1202 | 97.8 | 1202 | 98.5 | ## Performance of BD Phoenix System for Gram-Negative Organisms by Meropenem ## Conclusions Drawn from Substantial Equivalence Studies The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", August 28, 2009. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, with three curved lines representing the wings and a wavy line representing the body. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 27 2013 BECTON DICKINSON AND COMPANY MONICA GIGUERE REGULATORY AFFAIRS PROJECT MANAGER 7 LOVETON CIRCLE - MC 614 SPARKS, MD 21152 USA Re: K132674 Trade/Device Name: BD Phoenix Automated Microbiology System -Meropenem (0.125-32 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON, JWY, LRG, LTT, LTW Dated: November 14, 2013 Received: November 20, 2013 Dear Ms. Giguere: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act).that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Ms. Giguere CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ## Sally A - Dolivat -S Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use ## 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ #### BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Device Name: Meropenem (0.125-32 ug/mL) - Gram-negative ID/AST or AST only Phoenix Panels, ## Indications For Use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus. Enterococcus and Streptococcus. This premarket notification is for the antimicrobial agent meropenem at concentrations of 0.125-32 us/mL to Gram-negative ID/AST or AST only Phoenix panels. Meropenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDAapproved package inserts for this antimicrobial agent. ### Active In Vitro and in Clinical Infections Against: Escherichia coli Klebsiella pneumoniae Pseudomonas aeruginosa Active In Vitro Citrobacter koseri (formerly diversus) Citrobacter freundii Enterobacter cloacae Klebsiella oxytoca Morganella morganii Proteus vulgaris Serratia marcescens Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR) Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...