K132587 · Ned Display Solutions , Ltd. · LLZ · Aug 29, 2013 · Radiology
Device Facts
Record ID
K132587
Device Name
MD242C2 24 DIAGNOSTIC IMAGING LCD MONITOR
Applicant
Ned Display Solutions , Ltd.
Product Code
LLZ · Radiology
Decision Date
Aug 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The MD242C2 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers. MD242C2 cannot be used for a life-support system. This device must not be used in digital mammography. This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
Device Story
MD242C2 is a 24.1-inch color TFT LCD monitor providing 2.3-megapixel (1920x1200) resolution for medical imaging. It receives digital image data from NEC-approved display controllers via DVI, DisplayPort, or HDMI inputs. The device transforms these inputs into high-resolution visual output for clinical review in environments such as CT, MRI, HIS, and PACS. Operated by trained physicians, the monitor assists in diagnostic decision-making by providing calibrated grayscale and color rendering. It is designed for exclusive interconnection with IEC60601-1-1 certified equipment. The device benefits patients by providing precise, DICOM-compliant visualization of diagnostic images.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical safety (AAMI/ES 60601-1), electromagnetic compatibility (EN60601-1-2), and DICOM conformance testing.
Technological Characteristics
24.1" TFT Color LCD monitor; 1920x1200 resolution; 10-bit color depth (1,073,741,824 colors); 1000:1 contrast ratio. Inputs: DVI, DisplayPort, HDMI. Software-based luminance calibration. Compliant with DICOM Part 14. Certified to ANSI/AAMI ES 60601-1:2005 and EN60601-1-2.
Indications for Use
Indicated for displaying and viewing digital medical images for diagnosis by trained physicians. Contraindicated for use in life-support systems and digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K132587
Page 1 of 4
# 510(K) Summary of Safety and Effectiveness
As required by 807.92
#### DEVICE ESTABLISHMENT AND CONTACT PERSON l.
Mr. Satoru Kotani Manager NEC Display Solutions Ltd. 4-28, Mita 1-chome, Minato-ku, Tokyo, Japan Ph: +81-465-85-2384 Fax: +81-465-85-2393
### 2. COMPANY REISTRATION NUMBER 3003623028
#### DATE SUMMARY PREPARED 3.
12 October 2012
#### 4. DEVICE NAME
·
| Trade Name: | MD242C2 24.1" Diagnostic Imaging LCD monitor |
|----------------------|------------------------------------------------------------------|
| Model Name: | MD242C2 |
| Common Name: | Color LCD Monitor, Color Diagnostic Display, etc. |
| Classification Name: | System, Image Processing, Radiological (CLASS II CF<br>892.2050) |
#### 4. PREDICATE DEVICE
L217TG TFT Color LCD Monitor by NEC Display Solutions Ltd. (K083907)
### DEVICE DESCRIPTION 5.
Medical Display, MD242C2 is a 24.1" Color LCD monitor that displays image for medical use, It provides 2.3 mega pixel (1920*1200) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.
{1}------------------------------------------------
#### DEVICE OF INTENDED USE 6.
The MD242C2 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.
To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers.
MD242C2 cannot be used for a life-support system.
This device must not be used in digital mammography.
This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
## 7. SE Comparison Table:
| Items | L217TG | MD242C2 |
|---------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| 510(k) Number | K083907 | |
| Panel Size and Type | 21.3" TFT Color LCD Monitor | 24.1" TFT Color LCD Monitor |
| Pixel Pitch | 0.270 mm x 0.270 mm | 0.270 mm x 0.270mm |
| Display Color | 16,777,216 | 1,073,741,824 |
| Viewing Angles (°) | H:176, V:176 | H:178, V:178 |
| Scanning Frequency (H, V) | 31.5-91.1kHz, 50-85 Hz | 31.5-93.8, 118.4kHz, 50-85 Hz |
| Native Resolutions | 1600X1200 | 1920X1200 |
| Brightness | 400 cd/m² calibrated,<br>850 cd/m² Max. | 180 cd/m² calibrated,<br>350 cd/m² Max. |
| Contrast Ratio | 1050 : 1 (typical) | 1000 : 1 (typical) |
| DOT Clock | 162 MHz | 202.5 MHz (Max) (Analog)<br>162 MHz (Max) (Digital) |
| Input Signals | Three connectors: one D-sub<br>analog VGA: and two DVI-I<br>(VGA analog or digital) | Three connectors: one DVI port,<br>one Display port, one HDMI<br>port |
| Input Terminals | DVI-D, D-sub | DVI-D, Display port, HDMI |
| | | port |
| USB (option) / Standard | No | No |
| Active Display Size (H x | Landscape: 432mmX324mm | Landscape: 518.4mmX324mm |
| V) | Portrait: 324X432mm | Portrait: 324X518.4mm |
| Viewable Image Size | 540 mm (diagonal) | 540 mm (diagonal) |
| Luminance Calibration | Software | Software |
| Default Gamma | DICOM part 14 + off, 1.8,2.0,2.2<br>1.8.2.0,2.2 | DICOM part 14 |
| | user | |
| Power | AC100-240V, 50/60Hz | AC100-240V, 50/60Hz |
| Power Consumption | 100W (Max) | 38.4W (Max) |
| Power Save Mode | <2W | <2W |
| | | |
| Dimensions | W: | W: |
| (W x H x D) | Landscape: 467.8mm | Landscape: 556.8mm |
| | Portrait:361.6 mm | Portrait: 362.4 mm |
| | H: | H: |
| | Landscape: 434.3-584.3mm | Landscape: 378 - 528mm |
| | Portrait: 487.4-637.4mm | Portrait: 572.4-625.2mm |
| | D: 306 mm | D: 227.6 mm |
| NET Weight | 10.7 kg | 10.2 kg |
| Intended of use | Displaying and viewing of digital | Displaying and viewing of |
| | images for diagnosis by trained | digital images for diagnosis by |
| | physicians | trained physicians |
| | This device can not use for a life | This device can not use for a life |
| | support system. | support system. |
| | This device must not be use in | This device must not be use in |
| | digital mammography. | digital mammography. |
| | This device is designed for | This device is designed for |
| | exclusive interconnection with | exclusive interconnection with |
| | IEC60601-1 certified equipment | IEC60601-1-1 certified |
| | | equipment |
| Certifications & | CE ITE/Medical Device Directive, | CE ITE/Medical Device |
| Standards | UL/cUL (UL60601-1, CSA C22.2 | Directive, UL/cUL |
| | No.601-1), FCC Class B, | (ANSI/AAMI ES |
| | EN60601-1-2, DIN V 6868-57, | 60601-1:2005), FCC Class B, |
| | DICOM | EN60601-1-2, DIN V 6868-57, |
| | | DICOM |
### Comparison tables between MD242C2 & L217TG
{2}------------------------------------------------
. .
{3}------------------------------------------------
K132587
Page 4 of 4
### CONCLUSION
These two devices have the same target population of trained practitioner in hospital; it shares the same design, same performance and is the same in radiation safety (EN60601-1-2), mechanical safety, electrical safety (AAMI/ES 60601-1) human factors and DICOM conformance. It use similar material, and have same compatibility with environment and other device. Comparison table of the principal characteristics of two devices is shown in the Section 3, table 3.3. These two devices also have the same intended use; Therefore we concluded that it is substantially equivalent to L217TG by NEC Display Solutions Ltd. (K083907)
{4}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
NEC Display Solutions Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Service LLC. 1394 25th Street NW BUFFALO MN 55313
Re: K132587
Trade/Device Name: MD242C2 24" Diagnostic Imaging LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 15, 2013 Received: August 16, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Costinent for (100) that to nevice, subject to the general controls provisions of the Act. The r ou may, mereleve, manns of the Act include requirements for annual registration, listing of general volurols pro Prefering practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quand provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132587
MD242C2 24" Diagnostice Imaging LCD Monitor Device Name:
Indications for Use:
The MD242C2 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers. MD242C2 cannot be used for a life-support system.
This device must not be used in digital mammography. This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF .NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh.7)
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.