K132583 · Vidacare Corporation · FMI · Feb 11, 2014 · General Hospital
Device Facts
Record ID
K132583
Device Name
EZ-IO INTRAOSSEOUS INFUSION SYSTEM
Applicant
Vidacare Corporation
Product Code
FMI · General Hospital
Decision Date
Feb 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K132583 · Feb 11, 2014
EZ-IO INTRAOSSEOUS INFUSION SYSTEM
Vidacare Corporation
Published clinical literature (Frascone et al., 2009) reporting on routine clinical use by EMS providers
The sponsor utilized findings from a published clinical study of routine EMS practice to justify expanding the weight range indication for the 25 mm needle set, citing observations regarding soft tissue depth as the limiting factor for successful insertion.
EMS clinical practice; Pediatric intraosseous access; Label expansion; Clinical literature
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Use of the pediatric EZ-IO needle by emergency medical services providers (Frascone et al., 2009); Clinical study of routine clinical practice
Pediatric patients requiring intraosseous access
Not applicable for this study
Successful insertion and limiting factors (soft tissue depth)
Indications for Use
The EZ-IO Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Device Story
The EZ-IO Intraosseous Infusion System facilitates vascular access via the bone marrow. The system consists of a needle set (stylet and catheter) and a reusable cordless driver or manual handle. Clinicians identify anatomical landmarks, clean the site, and use the driver or handle to advance the needle through the bone cortex into the medullary space. A tactile 'pop' or 'give' indicates successful entry. The driver is detached, the stylet is removed, and the remaining Luer-lock catheter allows for the administration of fluids and medications. The device is used in clinical settings by healthcare providers. This submission expands the weight range for the 25 mm needle set from 40 kg or over to 3 kg or over to accommodate anatomical variability in soft tissue depth, allowing clinicians to select the appropriate needle length based on patient anatomy rather than weight alone.
Clinical Evidence
No new clinical trials were conducted. Evidence supporting the label change includes cadaveric pediatric studies and previously published clinical literature (Frascone et al., 2009) evaluating the use of pediatric EZ-IO needles by EMS providers. These studies demonstrate that the 25 mm needle length is safe and often necessary for patients under 40 kg due to variations in soft tissue depth over the insertion site.
Technological Characteristics
Needle sets are single-use, composed of 304 stainless steel with polycarbonate hubs. Available in 15 mm, 25 mm, and 45 mm lengths. Features depth markers on the catheter. Connects to a reusable cordless driver or manual handle via a magnetic disc. Driver is battery-powered. No changes to materials, design, or sterilization compared to predicates.
Indications for Use
Indicated for adult and pediatric patients requiring intraosseous access in the proximal tibia, distal tibia, or humeral head when intravenous access is difficult or impossible in emergent, urgent, or medically necessary cases.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
K132583- [Traditional 510(k)] EZ-IO Intraosseous Infusion System by Vidacare Corporation
October 30, 2013
# 005_510(k) Summary (Revised)
[807.92(c)]
vidacare
4350 Lockhill Selma Road Shavano Park, TX 78249+2095 210-375-8500 www.vidacare.com
#### SUMMARY (Revised)
Submitter's name [807.92(a)(1)]: Vidacare Corporation
Address:
4350 Lockhill Selma Road
210-375-8500
Shavano Park, TX 78249-2095
Phone:
210-375-8537 Fax number:
Name of contact person: Larry Miller, MD, CMO/ Diana Montez, BSN, RN
Date Summary was prepared: October 30, 2013
[807.92(a)(2)]: EZ-IO Intraosseous Infusion System Trade Name of the devices: Intraosseous Infusion System Common or usual name: Hypodermic single lumen needle Classification name:
The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:
| 510(k) number | Trade or Proprietary or<br>Model Name | Manufacturer |
|---------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K091140 | EZ-MIO Distal Tibia,<br>EZ-IO Distal Tibia; EZ-IO<br>(formerly Vidaport)<br>Intraosseous Infusion System;<br>EZ-IO Humeral Head | Vidacare Corporation |
| K101026 | Powered PD IO Infusion<br>System; EZ-IO Humeral<br>Head; Powered PD-IO | Vidacare Corporation |
{1}------------------------------------------------
| 510(k) number | Trade or Proprietary or<br>Model Name | Manufacturer |
|--------------------|-------------------------------------------------------------|------------------|
| K981853<br>K062940 | Bone Injection Gun™ Adult<br>Proximal Tibia<br>Humeral Head | Waismed |
| K915409<br>K913258 | Cook Intraosseous Infusion<br>Needles | Cook Medical Inc |
## Device Description [807.92(a)(4)]:
The EZ-IO Intraosseous Infusion System including previously cleared K091140 EZ-MIO Distal Tibia, EZ-10 Distal Tibia; EZ-10 (formerly Vidaport) Intraosseous Infusion System ; EZ-IO Humeral Head and K101026 Powered PD IO Infusion System; EZ-IO Humeral Head; Powered PD-IO is designed to allow the user to insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate intraosseous infusion of desired fluids for vascular access. All system materials are medical grade. Needle sets are single-use and composed of 304 stainless steel with polycarbonate hubs and available in 15 mm (3-39 kg); 25 mm (40 kg or over) and 45 mm (40 kg or over). Black lines on the needle set catheter serve as depth markers. The needle sets connect to the driver/drill/ manual handle shaft with a magnetic disc. The reusable cordless driver/drill is powered by lithium batteries with a battery-power indicator light. The EZ-MIO consists of a manual handle device which is primarily used by the armed forces.
Clinicians locate anatomical landmarks and clean the insertion site. Using the cordless driver or manual handle with needle set attached the needle set is pressed through the soft tissue to the outer cortex of the bone. Depth markers on the catheter must be visible prior to powering driver or manually inserting the needle set to ensure placement within the medullary space. If using the powered driver clinicians then squeeze the driver trigger and apply moderate, steady pressure. Trigger is released when a sudden "give" or "pop" is felt, which indicates entry into medullary space; the catheter will not always be inserted to the hub. If using the manual handle, clinicians apply downward pressure and rotation through the bone until a change in pressure is felt as a "give" or "pop" or desired depth is reached. For both methods after insertion of the needle set, the driver unit is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the catheter by turning the stylet hub counter clockwise leaving a standard Luer lock catheter securely seated in the bone. The catheter Luer lock permits attachment of standard syringes and IV tubing for administration of medications and fluids.
{2}------------------------------------------------
This submission requests a label change only to the 25 mm needle set from "40 kg or over" to" 3 kg or over" utilizing the same technique and devices previously cleared via 510(k)s K091140 and K101026.
### Proposed Expanded Weight Range for the 25 mm needle set:
Vidacare is requesting a label change only to the 25 mm needle set from "40 kg or over" to "3 kg or over" to provide clinicians with an option to choose the correct needle length not only based on weight but also variability in each patient's anatomy. Due to differences in the amount of soft tissue overlying the target anatomical area, many patients less than 40 kg require additional needle length the 25 mm offers over the 15 mm needle length. This has been noted by authors in a clinical study which stated the limiting factor for successful insertions is the amount of tissue covering the approved insertion site.1 Cadaveric pediatric studies with representative photographs demonstrate the 25 mm length to be safe and often necessary in patients less than 40 kg (see RTA response section J number 36 Performance Data included with this response package). There would be no change in the Directions for Use as insertion technique is the same.
In consideration of our cited predicate similarities, performance data, observations noted in the cited clinical studies and Vidacare's experience with the EZ-IO Intraosseous Infusion System we conclude that expanding the weight range for the 25 mm needle set is a safe and correct modification. This modification will allow clinicians to choose the correct needle length based on the best anatomical site option, soft tissue depth over the insertion site with a wider weight range that more accurately suits patients' anatomy and needs, with a weight range of 3 kg or over .
1. Frascone RJ, Jensen J, Wewerka SS, Salzman JG. Use of the pediatric EZ-IO needle by emergency medical services providers. Pediatric Emergency Care 2009;25(5):329-32.
## [807.92(a)(5)]:
## (Proposed)Indications for Use:
The EZ-10 Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
The above proposed Indications For Use provide the same indication as the predicate devices but is presented in a condensed version.
{3}------------------------------------------------
# [807.92(a)(6)]:
# Summary of the technological characteristics of our device compared to the predicate devices:
The labeling change requires no new technology to facilitate the safe application of the product. There have been no changes to the design or components of the Vidacare devices cleared under K091140 and K101026 and therefore, the comparison of technological characteristics listed below are identical:
- Target Population .
- Driver Design Features .
- . Needle Design
- Technique .
- Anatomical Sites .
- Sterility .
- Biocompatibility .
- Where Device is Used (types of facilities/environments) .
- Types of Clinicians (that would use device) .
For predicates K981853, K062940 The Bone Injection Gun and K915409, K913258 Cook Intraosseous Infusion Needles the following characteristics are similar or identical:
- . Target Population
- Needle Design .
- Technique: .
- . For insertion -Cook and EZ-MIO devices
- For locating correct sites; validating insertion; indications and . contraindications
- Anatomical sites .
- . Sterility
- Biocompatibility .
- Where Device is Used (Types of facilities/environments) .
- Types of Clinicians (that would use device) .
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 11, 2014
Vidacare Corporation Dr. Larry Miller CMO 4350 Lockhill Selma Road Shavano Park, TX 78249
Re: K132583
Trade/Device Name: EZ-IO Intraosseous Infusion System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 19, 2013 Received: August 21, 2013
Dear Dr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. Miller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/7 description: The image shows the text "Kwame Ulmer-s" in a stylized font. The text is arranged in two lines, with "Kwame" on the top line and "Ulmer-s" on the bottom line. There is a logo to the right of the text, which appears to be a circle with a design inside and a triangle with a design inside.
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Premarket Notification [Special 510(k)] Device Modification: Vidacare Corporation August 12, 2013
#### 6. Indications for Use Statement
Indications for Use
510(k) Number (if known): _K132583
#### Device Name:
The EZ-IO Intraosseous Infusion System provides intraosseous access in the proximal tibia, distal tibia and humeral head (proximal humerus) of adult and pediatric patients when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Prescription Use X ___________________________________________________________________________________________________________________________________________________________
Over-The-Counter Use -------
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
FDA
Digitally signed by Richard C. Chapman Date: 2014.02.10 16:20:00 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.