Bistos' BT-500 is a Infant Incubator and intended to keep Premature Neonate infants in a warm environment in the hood for Neonatal hypothermia, Observation and Examination in newborn nurseries. BT-500 provides Heat and Air in a controlled manner to neonates or premature infants who are unable to thermo- requlate on their own physiology. BT-500 is not intended for transport.
Device Story
Infant incubator providing controlled thermal environment; hood, control box, LCD monitor, main body. Inputs: skin temperature probe, humidity/temperature sensors. Operation: heater and circulation system maintain set skin temperature and air conditions. Outputs: real-time monitoring and graphing of humidity, air temperature, skin temperature, weight, and SpO2. Used in newborn nurseries by clinical staff. Benefits: prevents neonatal hypothermia; supports infants unable to thermoregulate. Output used by clinicians to monitor infant status and adjust environmental parameters.
Clinical Evidence
No clinical data. Bench testing performed for safety and performance in accordance with FDA guidance for neonatal incubators. Biocompatibility testing conducted per ISO 10993 (parts 4, 5, 10).
Technological Characteristics
Class II neonatal incubator. Components: hood, control box, LCD monitor. Sensors: humidity, air temperature, skin temperature probe. Energy: electric heater. Connectivity: LCD display for monitoring/graphing. Biocompatibility: materials identical to predicate (K102710), tested per ISO 10993.
Indications for Use
Indicated for premature or neonate infants unable to thermoregulate; used for warming, observation, and examination in newborn nurseries. Not for transport.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}------------------------------------------------
#### BT-500 Infant Incubator
#### Summary of safety and effectiveness K132543
In accordance with section 513(1) of the SMDA as defined in 21CFR part 807.92 This summary is submitted to obtain Pre-Market 510(K) notification.
#### 1. Submitter, Manufacturer
Prepared on Sept 20,2013
Bistos Co,. Ltd. Reg Number 3006179052 7th fl, WooRim Lions Valley 5, 144-3 SangDae Won, JungWon Gu, SeoungNam, GyeongGi Do, Republic of Korea T: 31 750 0340 F: 31 750 0344
2. Preparer / Contact person. Mr. Young Chi, President. Bio-Med USA Inc. 27 New England Drive, Ramsey, NJ 07446. USA T: 973 278 5222 F: 201 934 6030 e mail: biomedusa@msn.com
#### 3. Name of Device.
> : BT-500 Infant Incubator Trade Name Common or usual name : Neonatal Incubator Regulation number/class : 880.5400 / II Product Code : FMZ Classification panel : General Hospital
4. Substantial Equivalence
K102710 Atom Infant incubator model 102, Atom Medical Inc
Bistos BT-500 infant incubator is substantially equivalent in Intended Use, Design, Function, Performing, Direction to use, software, Technology etc to Predicate device.
#### 5. Device Description.
BT-500 Infant Incubator consists of Hood, Control Box, LCD external monitor and main body with stand ( optional), This device incorporates two sensors to control Humidity, and Temperature inside the Hood and monitors the Air Conditions, and controls Circulation with proper temperature.
Also Heater is equipped inside of the hood to control the Skin Temperature of Infant at a fixed level as measured by the skin probe and showing, measuring and graphing the humidity, Air Temperature, Skin Temperature, Weight and SpO2
This Incubator is provided with function to control the infant's skin temperature.
{1}------------------------------------------------
# 6. Intended use.
Bistos' BT-500 is a Infant Incubator and intended to keep Premature Neonate infants in a warm environment in the hood for Neonatal hypothermia, Observation and Examination in newborn nurseries.
BT-500 provides Heat and Air in a controlled manner to neonates or premature infants who are unable to thermo- requlate on their own physiology.
BT-500 is not intended for transport.
# 7 Labeling
Back label, Market promoting leaflet designed by labeling requirement regulation under (21CFR part 801) is on Technical Construction File.
#### 8.Biocompatibility test.
BT-500 used same material of patient contacted part as already cleared Neo-Servo under K102710, and done Biocompatibility test by NAMSA ( North American Science ) by FDA guidance Blue Book Memo G95-1 use of ISO 10993 Biological Evaluation of Medical Device part 4, 5, 10.
# 9. Test Data
BT-500 Infant Incubator done various Performing, Safety test voluntary and accordance with the guidance for Industry and FDA Staff -Neonatal Transport Incubator Premarket notifications., and all test report attached
#### 10, Conclusion.
BT-500 Infant Incubator is substantially equivalent in Intended use Design, Function, Performing, direction to use, software, Technology/Principle of operation as already cleared predicate device K102710 model 102.
So, BT-500 Infant Incubator has no new issues in safety and effectiveness. End of Summarv
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 3, 2014
Bistos Company, Limited C/O Mr. Young Chi Bio-Med USA, Incorporated 27 New England Drive Ramsey, New Jersey 07446
Re: K132543
Trade/Device Name: BT-500 Infant Incubator Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal incubator Regulatory Class: Class II Product Code: FMZ Dated: June 24, 2014 Received: June 26, 2014
Dear Mr. Chi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Chi
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mary S. Runner -S
Erin Keith, M.S. Division Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132543
Device Name BT-500 Infant Incubator
Indications for Use (Describe)
Bisto's BT-500 is a Infant Incubator and intended to keep Premature in a warm environment in the hood for Neonatal hypothermia, Observation and Examination in newborn nurseries.
BT-500 provides Heat and Air in a controlled manner to neonates or premature infants who are unable to themn-regulate on their own physiology.
BT-500 is not intended for transport.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
: . .
Digitally signed by Richard C. Chapman -S Date: 2014.07.03 11:54:25 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, Including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.