LIAISON 25 TOTAL-D

K132492 · DiaSorin, Inc. · MRG · Sep 5, 2013 · Clinical Chemistry

Device Facts

Record IDK132492
Device NameLIAISON 25 TOTAL-D
ApplicantDiaSorin, Inc.
Product CodeMRG · Clinical Chemistry
Decision DateSep 5, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1825
Device ClassClass 2

Intended Use

The LIAISON® 25 TOTAL-D assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of 25-hydroxyvitamin D and other hydroxylated vitamin D metabolites in human serum, EDTA and Lithium heparin plasma. The LIAISON® 25 TOTAL-D assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The DiaSorin LIAISON® 25 TOTAL-D is intended to be used on the LIAISON® XL Analyzer. The DiaSorin LIAISON® 25 TOTAL-D Control Set is intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® 25 TOTAL-D assay. The DiaSorin LIAISON® 25 TOTAL-D Calibration Verifiers are assayed quality control materials intended for use in the quantitative verification of calibration and reportable range of the LIAISON® 25 TOTAL-D assay when performed on the LIAISON® XL Analyzer.

Device Story

The LIAISON® 25 TOTAL-D is a chemiluminescent immunoassay (CLIA) for the quantitative measurement of 25-hydroxyvitamin D in human serum or plasma. The device uses paramagnetic particles coated with goat anti-25 OH Vitamin D antibody. The assay follows a competitive principle: 25 OH Vitamin D from the sample competes with a vitamin D-isoluminol conjugate for antibody binding sites. After incubation and washing, starter reagents initiate a flash chemiluminescent reaction. A photomultiplier measures the light signal (RLU), which is inversely proportional to the analyte concentration. The device is intended for use on the LIAISON® XL Analyzer in clinical laboratory settings. Healthcare providers use the quantitative output to assess vitamin D sufficiency in adult patients. The system includes controls and calibration verifiers to ensure accuracy and precision.

Clinical Evidence

Bench testing only. Method comparison study (n=403) against predicate showed high correlation (r=0.990, slope 0.99). Matrix comparison (serum, SST, EDTA, Li-heparin) confirmed equivalence. Precision/reproducibility study (n=80 replicates per sample) showed total CVs between 5.0% and 13.6%. LoB/LoD/LoQ were determined to be <2.03 ng/mL, 2.91 ng/mL, and 4.0 ng/mL respectively. Recovery studies averaged 96%. Interference testing showed no significant bias (<10%) for common endogenous substances.

Technological Characteristics

Chemiluminescent immunoassay (CLIA) using paramagnetic particles coated with goat anti-25 OH Vitamin D antibody. Reagents include isoluminol-conjugated vitamin D. Operates on LIAISON® XL Analyzer. Measuring range 4-150 ng/mL. Storage 2-8°C. Automated sample processing.

Indications for Use

Indicated for the quantitative determination of 25-hydroxyvitamin D and other hydroxylated vitamin D metabolites in human serum, EDTA, and Lithium heparin plasma to aid in the assessment of vitamin D sufficiency in adults.

Regulatory Classification

Identification

A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.

Special Controls

*Classification.* Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and (2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132492 #### 510(k) SUMMARY #### SUBMITTED BY: Mari Meyer Director, Regulatory Affairs DiaSorin Inc. 1951 Northwestern Avenue P.O. Box 285 Stillwater, MN 55082-0285 Phone (651) 439-9710 Fax (651) 351-5669 Email: mari.meyer@diasorin.com SEP 0 5 2013 DATE PREPARED: PURPOSE OF SUBMISSION: NAME OF DEVICE: Trade Name: LIAISON® 25 TOTAL-D LIAISON® 25 TOTAL-D Control Set LIAISON® 25 TOTAL-D Calibration Verifiers Common Names/Descriptions: Classification Names: Vitamin D August 28, 2013 New Device Vitamin D Test System: Class II 21 CFR 862.1825; Clinical Chemistry (75) Quality Control Material: Class I, reserved 21 CFR 862.1660; Clinical Chemistry (75) Product Code: PREDICATE DEVICE: MRG, JJX LIAISON® 25 OH Vitamin D TOTAL K112725 LIAISON® 25 OH Vitamin D Control K112725 LIAISON® 25 OH Vitamin D Calibration Verifier K090104 #### DEVICE DESCRIPTION: The LIAISON® 25 TOTAL-D consists of one Reagent Intergral with calibrators, which consists of: | Magnetic Particles | Magnetic particles coated with goat antibody against 25 | |--------------------|---------------------------------------------------------| | (2.4 mL) | OH Vitamin D, protein, phosphate buffer, < 0.1% sodium | | | azide. | | Assay Buffer | Buffer with 7.5% ethanol, surfactants and 0.2% ProClin® | | (28.0 mL) | as a preservative. | | Conjugate | 25 OH Vitamin D conjugated to an isoluminol derivative. | | (4.5 mL) | in phosphate buffer with10% ethanol, EDTA and 0.1% | {1}------------------------------------------------ | benzoic acid as a preservative. | | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Calibrator 1<br>(1.2 mL) | Human serum, BSA, <0.1% sodium azide and 25 OH Vitamin D. The calibrator concentrations (ng/mL) are referenced to standard preparations containing highly purified 25 OH Vitamin D. | | Calibrator 2<br>(1.2 mL) | Human serum, BSA, <0.1% sodium azide and 25 OH Vitamin D. The calibrator concentrations (ng/mL) are referenced to standard preparations containing highly purified 25 OH Vitamin D. | ProClin® 300 is a registered trademark of Rohm and Haas Co. The LIAISON® 25 TOTAL-D control consists of 2 levels of human serum, BSA, <0.1% sodium azide and 25 OH Vitamin D. The LIAISON® 25 TOTAL-D calibration verifier consists of 4 levels of human serum, BSA. <0.1% sodium azide and 25 OH Vitamin D. #### INTENDED USE: The LIAISON® 25 TOTAL-D assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of 25-hydroxyvitamin D and other hydroxylated vitamin D metabolites in human serum, EDTA and Lithium heparin plasma. The LIAISON® 25 TOTAL-D assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The DiaSorin LIAISON® 25 TOTAL-D is intended to be used on the LIAISON® XL Analyzer. The DiaSorin LIAISON® 25 TOTAL-D Control Set is intended for use as assaved quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® 25 TOTAL-D assay. The DiaSorin LIAISON® 25 TOTAL-D Calibration Verifiers are assayed quality control materials intended for use in the quantitative verification of calibration and reportable range of the LIAISON® 25 TOTAL-D assay when performed on the LIAISON® XL Analyzer. #### TEST PRINCIPLE: The DiaSorin LIAISON® 25 TOTAL-D assay is a direct competitive chemiluminescence immunoassay (CLA) for quantitative determination of 25 OH vitamin D in human serum or plasma. During the first incubation, 25 OH vitamin D is dissociated from its binding protein and binds to the specific antibody on the solid phases. After 10 minutes the tracer, (vitamin D linked to an isoluminol derivative) is added. After a second 10 minute incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added to initiate a flash chemiluminescent reaction. The light signal is measured by a photomultiplier as relative light units (RLU) and is inversely proportional to the concentration of 25 OH vitamin D present in calibrators, controls, or samples. {2}------------------------------------------------ # COMPARISON TO PREDICATE DEVICE: The DiaSorin LIAISON 25 TOTAL-D assay is substantially equivalent in principle and performance to the LIAISON® 25 OH Vitamin D TOTAL assay (K112725) which was FDA cleared January 20, 2012. | Table 1: Summary of Device Similarities - LIAISON® 25 TOTAL-D | | | | |---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--| | Characteristic | New Device<br>LIAISON® 25 TOTAL-D | Predicate Device<br>LIAISON® 25 OH Vitamin D<br>TOTAL assay (K112725) | | | Intended Use | In-vitro assay for the<br>quantitative determination<br>of 25-hydroxyvitamin D and<br>other hydroxylated vitamin<br>D metabolites. | In-vitro assay for the<br>quantitative determination of<br>25-hydroxyvitamin D and<br>other hydroxylated vitamin D<br>metabolites. | | | Indications for Use | To be used as an aid in the<br>assessment of vitamin D<br>sufficiency in adults. | Same | | | Measured Analyte | 25-hydroxyvitamin D | Same | | | Assay Type | Chemiluminescent<br>Immunoassay | Same | | | Test principle | Competitive<br>chemiluminescent<br>Immunoassay | Same | | | Solid Support | Paramagnetic particles<br>coated with goat antibody<br>against 25 OH Vitamin D | Same | | | Reagent Integral<br>Storage | On-board or in<br>refrigerator@ 2-8°C | Same | | | Sample<br>Handling/Processing | Automated | Same | | | Unit of Measure | ng/mL | Same | | | Calibration | Two-point calibrator<br>verification of stored master<br>curve.<br>Included with the kit. | Same | | | Measuring range | 4 - 150 ng/mL | Same | | | Conjugate Antibody | 25 OH Vitamin D<br>conjugated to an isoluminol<br>derivative | Same | | | Controls | 2 Levels | Same | | | Open storage<br>2-8°C | 4 weeks | Same | | | Table 2 : Summary of Device Differences - LIAISON® 25 TOTAL-D | | | | | Characteristic | New Device<br>LIAISON® 25 TOTAL-D | Predicate Device<br>LIAISON® 25 OH Vitamin D<br>TOTAL assay (k112725) | | | Sample Matrix | Human serum, EDTA and<br>Li heparin plasma | Human serum | | | Sample size | 15 uL | 25 uL | | | Calibration Matrix | Human serum with BSA<br>and, sodium azide | Human serum with buffer<br>salts and sodium azide. | | {3}------------------------------------------------ The DiaSorin LIAISON 25 TOTAL-D Control is substantially equivalent in principle and performance to the LIAISON® 25 OH Vitamin D TOTAL Control (K112725) which was FDA cleared January 20, 2012. | Table 3: Summary of Device Similairites and Difference LIAISON® 25 TOTAL-<br>D Control Set | | | | |--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--| | Characteristic | LIAISON® 25 TOTAL-D<br>Control | LIAISON® 25 OH Vitamin D<br>TOTAL Control (K112725) | | | Intended Use | Intended for use as<br>assayed quality control<br>samples to monitor the<br>accuracy and precision of<br>the assay. | Same | | | Storage | Store at 2-8°C until ready<br>to use | Same | | | Matrix | Human serum with BSA<br>and, sodium azide | Human serum with buffer<br>salts and sodium azide. | | | Levels | 2 levels: low and high | Same | | The DiaSorin LIAISON® 25 TOTAL-D Calibration Verifiers are substantially equivalent in principle and performance to the LIAISON® 25 OH Vitamin D TOTAL Calibration Verifiers (K090104) which were FDA cleared September 9, 2009. {4}------------------------------------------------ # Table 4: Summary of Device Similarities and Differences LIAISON® 25 TOTAL-D Calibration Verifiers | Characteristic | LIAISON® 25 TOTAL-D<br>Calibration Verifiers | LIAISON® 25 OH Vitamin D<br>TOTAL Calibration<br>Verifiers (K090104) | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | Intended Use | Assayed quality control<br>materials for in vitro<br>diagnostic use in the<br>quantification verification of<br>calibration and reportable<br>range of the DiaSorin<br>LIAISON® 25 OH Vitamin D<br>TOTAL Assay. | Same | | Product Storage | 2 to 8°C | Same | | Matrix | Human serum with BSA<br>and, sodium azide | Human serum with buffer<br>salts and sodium azide. | | Levels | Four | Same | | Volume | 5.0 mLs | Same | ## PERFORMANCE DATA: #### Method Comparison: A method comparison study was performed on 403 serum samples following CLSI EP9-A2. The samples were tested in singlicate by both the LIAISON® 25 TOTAL-D assay and the predicate device. Testing for the LIAISON® 25 TOTAL-D assay and the predicate assay was performed at DiaSorin according to the respective Instructions for Use. All specimens were stored at -20°C or below until tested. Samples tested in the method comparison included de-identified residual serum samples spanning the assay range which were obtained from a sample procurement organization. The samples are excess samples drawn from non-selected U.S. and/or European patients with sufficient volume to perform all testing required. Twenty-three (23) of the 403 samples were spiked or diluted samples in order to achieve sample values that reach the upper and/or lower limits of the assay range. Three hundred ninety-one (391) of the 403 samples tested were analyzed. Twelve (12) samples read below the measuring range of the LIAISON® 25 TOTAL-D assay (<4.0 ng/mL) and therefore were not included in the analysis. Individual 25 OH Vitamin D results were plotted. Passing and Bablok regression analyses were performed for all samples across the measuring range of the assays. {5}------------------------------------------------ #### Results The range of the results observed with the LIAISON® 25 TOTAL-D assay ranged from 4.18 - 135 ng/mL. Passing & Bablok regression analysis was applied to these samples and is summarized in the following table. | N | Slope<br>(95%CI) | Intercept<br>(95% CI) | Correlation<br>Coefficient | |-----|---------------------|-----------------------|----------------------------| | 391 | 0.99<br>(0.97-1.01) | -0.22<br>(-0.51-0.08) | 0.990 | #### Table 5: Method Comparison Passing & Bablok Regression Data #### Matrix comparison: #### Serum, SST Serum, Lithium Heparin, EDTA Plasma Sixty four (64) matched patient sets of serum. lithium heparin plasma and EDTA plasma samples spanning the measuring range of the assay (4.0-139.3 ng/mL) were tested to determine if serum, SST serum, lithium heparin plasma and EDTA plasma samples provide equivalent results on the LIAISON® 25 TOTAL-D assay. #### Results A Passing-Bablok method comparison was used for analysis comparing each sample fype separately to the corresponding serum samples, with the specification of a 0.90 to 1.10 slope to serum considered acceptable. | | Bias | 95% CI | | |-------------|--------|---------|-----------| | y-intercept | 0.4098 | -0.1694 | to 1.1652 | | Slope | 0.9942 | 0.9785 | to 1.0113 | #### SST Serum Compared to Serum #### EDTA Plasma Compared to Serum | | Bias | 95% CI | | |-------------|---------|---------|------------| | y-intercept | -0.5785 | -1.1028 | to -0.0976 | | Slope | 1.0092 | 0.9916 | to 1.0253 | #### Lithium Heparin Plasma Compared to Serum | | Bias | 95% Cl | | | |-------------|---------|---------|-----------|--| | v-intercept | -0.0211 | -0.5554 | to 0.4683 | | | Slope | 1.0007 | 0.9869 | to 1.0180 | | #### Conclusion The study shows equivalent testing results from SST serum. EDTA plasma or lithium heparin plasma to serum. #### Reference Range/Expected Values: A reference range study was performed in accordance with CLSI Guideline C28-A2. Participant enrollment took place at four sites in the contiguous US, representing a wide distribution of sun strength: northern, southern, and central US. The collection was conducted on adults (ages 21 - 90 years) in summer and winter, including light and {6}------------------------------------------------ · dark skin individuals, and males and females. Three hundred and ninety five (395) samples met the inclusion/exclusion criteria and were tested with the LIAISON® 25 TOTAL-D Assav. The determination of apparently healthy for this study was based on the following definition: - · Normal serum levels of: - o Total calcium - o Intact parathyroid hormone (PTH) - o Thyroid stimulating hormone (TSH) - No personal history of kidney, gastrointestinal or liver disease . - No personal history of parathyroid, thyroid, or chronic disease (defined by . severity of condition which is linked to treatment and frequency as stated on subject's CRF) - No personal history of seizures . - No bariatric surgery . - No family history of parathyroid or calcium regulatory disease . Other criteria were: at least 50% of subjects NOT taking ANY Vitamin D supplementation, and of the less than 50% of subject taking Vitamin D supplementation, level of supplementation must be <2000 IU/dav. Subjects were not currently taking any medications known to affect absorption (drugs that inhibit cholesterol absorption) or increase catabolism, such as anticonvulsants, glucocorticoids, HAART (AIDS treatment) and anti-reiection medications - 1. Current use of dietary or alternative therapies containing high concentrations of Vitamin D (e.g. adults ≥ 2.000 IU / day) - 2. More than 50% of subjects are taking Vitamin D supplementation - 3. Family history of parathyroid or calcium requlatory disease - 4. Personal history of the following diseases: kidney, gastrointestinal, liver, thyroid, or parathvroid - 5. Personal history of seizures - 6. Bariatric surgerv - 7. Pregnancy or lactation The Instructions for Use contain the following table and statement: Reference Range | Population<br>(395) | Median<br>25 OH Vitamin D | Observed Reference Ranges<br>Central 95% Interval | |---------------------|---------------------------|---------------------------------------------------| | United States | 22.9 ng/mL | 8.8 ng/mL – 54.2 ng/mL | Consider these limits as guidelines only. It is important for each laboratory to establish its own reference range, representative of its typical population. {7}------------------------------------------------ #### Reproducibility/Precision: A twenty day reproducibility/precision study was performed at DiaSorin Inc consisting of a coded panel comprised of 6 frozen serum samples. The coded panel contained samples for each level of low, medium and high 25 OH Vitamin D medical decision concentrations. The LIAISON® 25 TOTAL-D controls (2 levels) were also tested in the study. The CLSI document EP5-A2 was followed in the preparation of the testing protocol. The precision panel samples and kit controls were tested on the LIAISON® 25 TOTAL-D assay on 1 Reagent Integral lot at one site in two replicates per run, 2 runs per day for 20 operating days on 1 analyzer for a total of 80 replicate results per sample. The mean, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens. #### Results The 20 day results obtained at DiaSorin Inc. are summarized in Table 5 as sample overall mean 25 OH Vitamin D concentration in ng/mL, computed SDs and %CVs for within run and total within lot. | Sample<br>ID | N | mean<br>conc<br>(ng/mL) | Within run | | Total | | |--------------|----|-------------------------|------------|------|-------|-------| | KC 1 | 80 | 15.4 | 0.68 | 4.4% | 1.78 | 11.6% | | KC 2 | 80 | 52.8 | 1.57 | 3.0% | 3.29 | 6.2% | | Sample 1 | 80 | 11.2 | 0.43 | 3.8% | 1.52 | 13.6% | | Sample 2 | 80 | 17.7 | 0.56 | 3.1% | 1.82 | 10.3% | | Sample 3 | 80 | 28.2 | 0.80 | 2.9% | 2.36 | 8.4% | | Sample 4 | 80 | 34.3 | 1.03 | 3.0% | 2.96 | 8.6% | | Sample 5 | 80 | 65.1 | 1.35 | 2.1% | 3.97 | 6.1% | | Sample 6 | 80 | 113.5 | 2.27 | 2.0% | 5.70 | 5.0% | Table 7: DiaSorin Inc. - 20 dav Reproducibility/Precision #### Dilution Linearity: A lineanty study was performed using samples slightly above the upper measuring limit of the LIAISON® 25 TOTAL-D Assay, following CLSI EP6-A. Separate serum, SST (serum separator tubes), EDTA plasma and lithium heparin plasma pools that met this criterion were diluted with the assay's Specimen Diluent to yield multiple dilutions with Vitamin D levels spanning the assay's full measuring range of 4 - 150 ng/ml. Each dilution was analyzed in replicates of 4 with one lot of the modified Vitamin D Assay on one LIAISON® XL Analyzer. Expected concentrations for the dilutions were determined by (Vitamin D value of high pool) x (dilution factor). {8}------------------------------------------------ #### Results The results for each sample were analyzed by linear regression (slope and, intercept) of observed 25 OH D concentrations versus expected 25 OH D concentrations in ng/mL. The resulting equation for each sample type is: Serum: y (observed) = 1.025x (expected) - 0.6094 SST serum: y (observed) = 1.0064x (expected) - 0.3887 EDTA plasma: y (observed) = 1.0092x (expected) - 0.4958 Lithium heparin plasma: y (observed) = 1.0421x (expected) - 0.8295 A polynomial regression fit of the serum data yielded statistically insignificant second and third order terms. These results support the claimed measuring range of 4 to 150 ng/mL. #### LoB/LoD/LoQ: The study was performed based upon CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline June · 2012- Second Edition. For LoB, 6 blank sample aliquots were tested in the LIAISON® 25 TOTAL-D assay on two analyzers, with two reagent lots and two technicians over six runs over three days yielding 60 concentration results. For LoD, four (4) samples in the range of the mean LoB to 4 times the mean LoB were tested with the LIAISON® TOTAL-D on one LIAISON® XL Analyzer with two reagent . lots and two technicians over six runs and three days (2replicates/sample/run) vielding a minimum of 96 concentration results. For LoQ, eight (8) samples (4 of which were used in the LoD determination) in the range of approximately 1.5 ng/mL to approximately 5.3 ng/mL were tested in the LIAISON® 25 TOTAL-D assay on two LIAISON® XL Analyzer with two reagent lots and two technicians over six runs and three days (2 replicates/sample/run), yielding 192 concentration results to define LoQ. To verify the LoQ, the same 8 samples were run on 1 LIAISON® XL Analyzer with a third reagent lot and two technicians over four runs (3 replicates/sample/run) yielding 96 concentration results to verify LoQ. #### Results: The limits are reported in the following table: | LoB | LoD | LoQ | |--------------|------------|-----------| | < 2.03 ng/mL | 2.91 ng/mL | 4.0 ng/mL | #### Recovery: Testing was performed to determine trueness or percent recovery on the LIAISON® 25 TOTAL-D. Five (5) high concentration serum samples (spiked or endogenous) and five (5) low concentration serum samples were analyzed on the LIAISON® 25 TOTAL-D assay. {9}------------------------------------------------ Recovery samples were then prepared by mixing defined ratios of the high and low samples and analyzing these in the LIAISON® 25 TOTAL-D assay. The obtained results were determined through calculation of % recovery for each dilution point: % Recovery = (Observed conc. / Expected conc.)* 100 Expected conc. = conc. of High sample x % ratio used + Low sample x % ratio used Each neat serum along with each blend was analyzed. The observed values were then compared to the expected values, based on the neat serum concentrations to determine the % recovery. The mean recovery is 96%. | Serum Samples | Defined<br>Concentration | Expected<br>ng/mL | Observed<br>ng/mL | %<br>Recovery | |---------------------------|--------------------------|-------------------|-------------------|---------------| | Serum High Sample 1 (HS1) | 105.0 | | | | | 2 HS1 : 1LS1 | | 73.5 | 72.3 | 98% | | 1 HS1 : 1LS1 | | 57.2 | 54.5 | 95% | | 1 HS1 : 2LS1 | | 41.0 | 38.3 | 93% | | Serum Low Sample 1 (LS1) | 9.5 | | | | | Serum High Sample 2 (HS2) | 95.5 | | | | | 2 HS2 : 1LS2 | | 67.5 | 63.4 | 94% | | 1 HS2 : 1LS2 | | 53.1 | 49.0 | 92% | | 1 HS2 : 2LS2 | | 38.6 | 35.9 | 93% | | Serum Low Sample 2 (LS2) | 10.6 | | | | | Serum High Sample 3 (HS3) | 101.0 | | | | | 2 HS3 : 1LS3 | | 70.8 | 68.7 | 97% | | 1 HS3 : 1LS3 | | 55.3 | 52.4 | 95% | | 1 HS3 : 2LS3 | | 39.7 | 37.3 | 94% | | Serum Low Sample 3 (LS3) | 9.5 | | | | | Serum High Sample 4 (HS4) | 58.5 | | | | | 2 HS4 : 1LS4 | | 44.4 | 43.7 | 98% | | 1 HS4 : 1LS4 | | 37.1 | 36.4 | 98% | | 1 HS4 : 2LS4 | | 29.8 | 29.3 | 98% | | Serum Low Sample 4 (LS4) | 15.7 | | | | | Serum High Sample 5 (HS5) | 56.9 | | | | | 2 HS5 : 1LS5 | | 42.2 | 42.2 | 100% | | 1 HS5 : 1LS5 | | 34.7 | 33.9 | 98% | | 1 HS5 : 2LS5 | | 27.1 | 26.9 | 99% | | Serum Low Sample 5 (LS5) | 12.4 | | | | | | | Mean Recovery | | 96% | {10}------------------------------------------------ #### Interfering Substances: Vitamin D samples containing concentrations of 30 ng/mL and 60 ng/mL were spiked with hemoglobin, trialycerides, cholesterol, and bilirubin, uric acid, albumin.lgG, biotin, ascorbic acid, metoprolol, propanolol hydrochloride, furosemide, hydrochlorothiazide, paricalcital and doxercalciferol. Replicate samples (n=6 for each test and control sample) were tested and results were compared to control samples without the potentially interfering substances added. The highest concentrations at which no interference, or minimal interference (defined by the sponsor as ≤ 10% bias relative to the control) are shown below. The testing was based on CLSI EP7-A2. | Specimens That Are | Demonstrate < 10% change in Results Up To | |--------------------|-------------------------------------------| | Hemolyzed | 200 mg/dL of hemoglobin | | Lipemic | 638 mg/dL of triglycerides | | Icteric | 40 mg/dL of conjugated bilirubin | | Icteric | 40 mg/dL of unconjugated bilirubin | ### Specimens That Contain Cholesterol Uric acid Total Protein Biotin Ascorbic Acid Metoprolol Propanolol hydrochloride Furosemide Hydrochlorothiazide Paricalcital Doxercalciferol Demonstrate < 10% change in Results Up To 350 mg/dL 20 mg/dL 12 g/dL 0.1 mg/dL 6 mg/dL 1.2 mg/dL 0.23 mg/dL 6 mg/dL 0.6 mg/dL 0.012 ua/mL #### Cross-reactivity: Cross-reactivity studies were performed as described in CLSI EP7-A2 using three serum pools at 25 OH D concentration levels of 15 ng/mL, 50 ng/mL, and 75 ng/mL. Data on the cross-reactivity of the antiserum used in this assay were obtained by spiking up to 100 ng/mL of the potential cross-reactant and assaying. The crossreactivity of each compound, normalized to 25 OH Vitamin D3, is listed below. 0.012 ug/mL | Steroid | % Cross<br>reactivity | |------------------------|-----------------------| | 25 OH Vitamin D3 | 100% | | 25 OH Vitamin D2 | 93% | | Vitamin D3 | 3.6% | | Vitamin D2 | 1.9% | | 1,25 (OH)2 Vitamin D3 | 17.1% | | 1,25 (OH)2 Vitamin D2 | 27.1% | | 3-epi-25 OH Vitamin D3 | 1.9% | {11}------------------------------------------------ #### Stability: Shelf life stability studies were performed with controls and calibration verifiers and demonstrated that they are stable until the expiration date shown on the product labeling when stored as instructed. Controls and calibration verifiers are stable until the expiration date printed on the label when stored as directed. Once opened, controls and calibration verifiers are stable for 4 weeks when properly stored at 2-8℃ between uses. Calibration curve stability and reagent open vial stability were performed and demonstrated that the calibration curve is stable for 7 days and open reagent vials are stable for 4 weeks when stored on board or at 2-8°C. #### Traceability The LIAISON® 25 TOTAL-D Calibrators. Controls and Calibration Verifiers are traceable to UV spectrophotometric analysis of an in-house standard preparation. The standard material used is a commercially available 25-Hydroxyvitamin D. The stock solution of 25 0H Vitamin Dis then used to make calibrators by diluting into human serum. #### Value assignment: Concentrations of calibrators controls and calibration verifiers are assigned through an internal procedure. Master calibrators are prepared from a stock solution made from reference material whose concentration is determined spectrophotometrically. The master calibrators are then used to assign values to the kit calibrators, controls and calibration verifiers using a minimum of 3 LIAISON XL analyzers with at least 2 reagent lots. Each lot of controls and calibration verifiers are tested over several runs and the mean results are used to determine the target values. 25 TOTAL-D Calibrators Level 1= 3.2 -4.8 ng/mL · Level 2= 70.0 – 80.0 ng/mL 25 TOTAL-D Controls have the following target ranges: Level 1= 12.0 - 18.0 ng/mL Level 2= 40.0 - 60.0 ng/mL 25 TOTAL-D Calibration Verifiers have the following target ranges: Level 1= 8.5 - 11.5 ng/mL Level 2= 27.0 - 33.0 ng/mL Level 3= 59.0 - 71.0 ng/mL Level 4= 108 - 122 ng/mL #### CONCLUSION: The material submitted in this premarket notification is complete and supports the basis for substantial equivalence to the LIAISON® 25 OH Vitamin D TOTAL assay (K112725). {12}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wavy lines representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Control - WO66-G609 Silver Spring, MD 20993-0002 September 5, 2013 DiaSorin Inc. C/O Mari Mever P. O. Box 285 1951 Northwestern Avenuc STILLWATER MN 55082-0285 Re: k132492 Trade Name: LIAISON® 25 TOTAL-D LIAISON® 25 TOTAL-D Control Set LIAISON® 25 TOTAL-D Calibration Verifiers Regulation Number: 21 CFR \$862.1825 Regulation Name: Vitamin D Test System Regulatory Class: Class II Product Codes: MRG, JJX Dated: July 29, 2013 Received: August 9, 2013 Dear Ms. Meyer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {13}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHQ0ffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincercly vours. # Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {14}------------------------------------------------ # Indication for Use 510(k) Number (if known): k132492 LIAISON® 25 TOTAL-D, LIAISON® 25 TOTAL-D Control Set and Device Name: LIAISON® 25 TOTAL-D Calibration Verifiers > The LIAISON® 25 TOTAL-D assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of 25-hydroxyvitamin D and other hydroxylated vitamin D metabolites in human serum. EDTA and Lithium heparin plasma. The LIAISON® 25 TOTAL-D assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The DiaSorin LIAISON® 25 TOTAL-D is intended to be used on the LIAISON® XL Analyzer. The DiaSorin LIAISON® 25 TOTAL-D Control Set is intended for use as assayed quality control samples to monitor the accuracy and precision of the DiaSorin LIAISON® 25 TOTAL-D assay. The DiaSorin LIAISON® 25 TOTAL-D Calibration Verifiers are assayed quality control materials intended for use in the quantitative verification of calibration and reportable range of the LIAISON® 25 TOTAL-D assay when performed on the LIAISON® XL Analyzer. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _ (21 CFR Part 801 Subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Yung.W.Chan -S Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) kl32492
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...