RENASYS EZ MAX NEGATIVE PRESSURE WOUND THERAPY DEVICE
K132446 · Smith & Nephew, Inc. · OMP · Oct 23, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132446
Device Name
RENASYS EZ MAX NEGATIVE PRESSURE WOUND THERAPY DEVICE
Applicant
Smith & Nephew, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Oct 23, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials. Examples of appropriate wound types include: Chronic, Acute, Traumatic, Sub-acute and dehisced wounds, Ulcers (such as pressure or diabetic), Partial-thickness burns, Flaps and grafts. The RENASYS EZ MAX Professional Healthcare Facility model (66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
Device Story
RENASYS EZ MAX is a powered suction pump for negative pressure wound therapy (NPWT). It applies continuous or intermittent negative pressure to wounds to remove exudates, irrigation fluids, and infectious materials into a disposable canister. The device is used in acute care and professional healthcare environments, operated by qualified healthcare professionals. It connects to wound dressing kits via tubing. The pump operates on mains power or internal battery. Healthcare providers use the device to manage wound healing; the output (fluid removal) facilitates the healing process. The device includes mechanical features for mounting to IV poles or bed rails and includes alarm systems for safety.
Clinical Evidence
Bench testing only. No clinical data provided. Testing verified pumping capacity, pressure settings (40-200mmHg), operating modes, interface compatibility with existing dressing kits, and mechanical integrity (mounting/impact resistance). Electrical safety and electromagnetic compatibility were verified against IEC 60601-1 (3rd Ed), IEC 60601-1-2, IEC 60601-1-8, and IEC 60601-1-6 standards.
Technological Characteristics
Powered suction pump; continuous/intermittent negative pressure (40-200mmHg). Mains or battery-powered. Compatible with existing RENASYS dressing kits. Standards: IEC 60601-1 (3rd Ed), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 60601-1-6/IEC 62366 (usability). Mechanical mounting for IV poles and bed rails.
Indications for Use
Indicated for patients requiring negative pressure wound therapy for wound healing via fluid removal. Applicable to chronic, acute, traumatic, sub-acute, dehisced wounds, ulcers (pressure/diabetic), partial-thickness burns, flaps, and grafts. Intended for use in professional healthcare settings under supervision.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K132466
2 We are smith&nephew
#### Wound Management 727 392-1261
に
Smith & Nephew, Inc. F 727 392-6914 or 727 392-0797 970 Lake Carillon Drive Customer Care Center: 1 800 876-1261 Suite 110 www.smith-nephew.com St. Petersburg, FL 33716
## 510(k) Summary
**OCT 23 2013**
General Information Submitters Name/Address:
Smith & Nephew, Inc. 970 Lake Carillon Drive Suite 110 St. Petersburg, FL 33716
3006760724 Establishment Registration Number: Contact Person: Laura Reynolds Director Regulatory Affairs Phone Number: (727) 329-7702 Date Prepared: September 20, 2013 Device Description RENASYS™ EZ MAX Negative Pressure Wound Trade Name: Therapy Device Generic/Common Name: Powered Suction Pump Classification Name: Powered Suction Pump; 21 CFR 878.4780 Product Code: OMP
Predicate Device Information
| 510(k) # | Device | Clearance Date |
|----------|-----------------|----------------|
| K102001 | RENASYS EZ PLUS | 08/06/2010 |
#### Device Description
The RENASYS EZ MAX Negative Pressure Wound Therapy NPWT device is a lightweight, suction device intended for wound management via application of continuous or intermittent negative pressure wound therapy to the wound for removal of fluids, including wound exudates, irrigation fluids, and infectious materials. The pump is connected to the wound dressing via a tube connected to a disposable canister. The device provides negative pressure wound therapy to the wound at a range of pressure settings and removes exudates from the wound site to the disposable canister. The device can operate either by a mains power supply or internal battery. The technological characteristics of the new device have not changed. The primary changes made are in relation to compliance with the IEC 60601-1 electrical safety standards. The predicate device complies with the IEC 60601-1 200 edition series while the RENASYS EZ MAX NPWT complies with the IEC 60601-1 3rd Edition series of electrical safety standards
RENASYS EZ MAX NPWT is compatible with existing Smith & Nephew wound dressing kits currently on the market.
{1}------------------------------------------------
## Indications for Use
RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials
Examples of appropriate wound types include:
- . Chronic
- Acute .
- Traumatic .
- Sub-acute and dehisced wounds .
- Ulcers (such as pressure or diabetic) .
- Partial-thickness burns .
- Flaps and grafts .
The RENASYS EZ MAX Professional Healthcare Facility model (66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
## Non-Clinical Tests (Bench)
Testing has been conducted to verify the modifications to the RENASYS EZ MAX device meet design specifications and demonstrate substantial equivalence to the predicate device.
The list below summarizes the bench testing undertaken and successfully completed for the RENASYS EZ MAX NPWT device:
- Pumping capacity is equivalent to the predicate device. .
- Device provides negative pressure at individual pressure settings of 40, 50, 60, . 70, 80, 90, 100, 120, 140, 160, 180 and 200mmHg, identical to the predicate device.
- . The device delivers negative pressure in a continuous and intermittent operating mode identical to the predicate device.
- . The device including canisters interface with existing RENASYS NPWT wound dressing kits.
- Device pressure setting remains unchanged when the device is locked and . actuation of the Therapy Knob is attempted
- . Device has a protection feature/mechanism to maintain the intended functional state (Continuous ON, OFF, Intermittent ON).
- Device attaches to IV poles ranging in diameter from 1/2 inches. .
- . Device bed rail clamp attaches to a bed rail (or equivalent) up to 3 inches in diameter.
- . Device canister bracket withstands side impact load simulating a worst case failure mode for clinical use.
- . Device complies with the following standards:
- IEC 60601-1-2:2007(3rd edition) Medical electrical equipment Part 1-2: o General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility -Requirements and tests, Interpretation Sheet
- IEC 60601-1:2005 (3rd edition). Medical Electrical Equipment Part 1: O
{2}------------------------------------------------
General Requirements for Safety,
- ANSI/AAMI ES60601-1:2005 Version (R2012) Medical Electrical o Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-8:2006 (2nd edition) Medical electrical equipment Part 1o 8: General requirements for basic safety and essential performance -Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical
- IEC 60601-1-6:2010 (3rd Edition) Medical electrical equipment Part 1-0 6: General requirements for basic safety and essential performance -Collateral standard. Usability
- IEC 62366:2007 (1st edition) Medical devices Application of usability o engineering to medical devices. (General)
### Summary of Safety and Effectiveness
In establishing substantial equivalence to the currently marketed predicate device, Smith & Nephew, Inc. evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing and electrical safety testing has been successfully completed to demonstrate that the RENASYS EZ MAX is substantially equivalent to the predicate device and is safe and effective for the intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Incorporated Ms. Laura Reynolds Director of Regulatory Affairs 970 Lake Carillon Drive, Suite 110 Saint Petersburg, Florida 33716
October 23, 2013
Re: K132446
Trade/Device Name: RENASYS" EZ MAX Negative Pressure Wound Therapy Device Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: September 20, 2013 Received: September 23, 2013
Dear Ms. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Laura Reynolds
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mark Nielkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K132446
Device Name: RENASYS™ EZ MAX Negative Pressure Wound Therapy Device
Indications For Use:
RENASYS EZ MAX is indicated for patients who would benefit from a suction device (negative pressure wound therapy), as it may promote wound healing via removal of fluids, including irrigation and body fluids, wound exudates and infectious materials.
Examples of appropriate wound types include:
- Chronic .
- . Acute
- Traumatic .
- Sub-acute and dehisced wounds .
- Ulcers (such as pressure or diabetic) ●
- . Partial-thickness burns
- Flaps and grafts .
The RENASYS EZ MAX Professional Healthcare Facility model (66801309) is intended for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/16 description: The image shows the name "Jiyoung Dang -S" in a simple, sans-serif font. The text is horizontally aligned and appears to be digitally rendered. The letters are black against a white background, providing a clear contrast.
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.