← Product Code [HSB](/productcode/HSB) · K132428

# IDEAL (K132428)

_Jiangsu Ideal Medical Science & Technology Co., LT · HSB · Dec 20, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132428

## Device Facts

- **Applicant:** Jiangsu Ideal Medical Science & Technology Co., LT
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Dec 20, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Ideal™ Intramedullary Nail System is intended to provide temporary fracture fixation and stabilization of the tibia. (Simple, compound first- and second-degree tibial shaft fractures, pseudarthrosis and delayed union).

## Device Story

Ideal™ Intramedullary Nail System provides temporary fixation and stabilization for tibial fractures. System comprises intramedullary nail, locking screws, and end cap. Nail inserted into tibia; locking screws passed through proximal and distal holes to prevent rotation and axial compression; end cap threaded into nail to allow intraoperative length adjustment and prevent tissue ingrowth. Available in various lengths and diameters. Used by orthopedic surgeons in clinical/surgical settings. Benefits include stabilization of tibial shaft fractures, pseudarthrosis, and delayed union.

## Clinical Evidence

Bench testing only. Testing included static bending, static torsion, dynamic bending (ASTM F1264-03), and pull-out strength (ASTM F543-07) to verify design specifications.

## Technological Characteristics

Material: Ti-6Al-4V alloy (ASTM F136). Components: Intramedullary nail, locking screws, end cap. Mechanical fixation via intramedullary insertion and screw locking. No software or energy source.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Intramedullary Nail System ([K121312](/device/K121312.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### Section 5 of Traditional 510(K) Submission:

#### 510 (K) Summary

## This 510(K) Summary of safety and effectiveness information is being submitted in accordance with requirement of 21 CFR807.92

### The assigned 5 10(k) Number: K132428

1. Date of Submission: Sept. 11.2013

2. Submitter / 510(K)Holder

Jiangsu Ideal Medical Science & Technology Co., Ltd. East Area, Jinfeng Industry Park Zhangjiagang Citv Jiangsu Province, China 215625

Contact Person: Mr. Hui Gang Cai Tel: (86) 0512-58550488 Fax: (86) 0512-58550988 E-mail: caihuig@126.com

#### 3. Proposed Device Name

Trade name: Ideal™ Common name: Intramedullary Nail System Classification Name: Intramedullary fixation rod Device Class: Class II Classification Panel: Orthopedic Panel Product Code: HSB Regulation Number: 21 CFR 888.3020

#### 4. Predicate Devices

510 (k) Number: K121312 Product Name: Intramedullary Nail System Submitter: Weigao Orthopaedic Device Co., Ltd.

#### 5. Device Description

The Ideal™ Intramedullary Nail System is intended to provide temporary fracture and stabilization of the tibia. (Simple, compound first- and second-degree tibial shaft fractures, pseudarthrosis and delayed union). It consists of intramedullary nail. locking screw and end cap.

# DEC 2 0 2013

{1}------------------------------------------------

The Intramedullary nail is available in a variety of lengths and diameters to meet assorted anatomical needs. Each of the nails is secured by a series of screws that pass through holes manufactured into the proximal and distal sections of each nail. The locking screw passes through the holes at the proximal and distal sections of intramedullary nail for preventing rotation and axial compression. The end cap which screws into the threaded end of the intramedullary nail provides intraoperative lengths adjustment and prevents tissues growth into nail threads.

All implants of intramedullary nail system are manufactured from Ti-6A1-4V allov that meets the requirements of ASTM F-136. The materials are widely used in the industry with well-known biocompatibility. No new materials are used in the development of this implant.

#### 6. Indication for Use/Intended Use

The Ideal" Intramedullary Nail System is intended to provide temporary fracture fixation and stabilization of the tibia. (Simple, compound first- and second-degree tibial shaft fractures, pseudarthrosis and delaved union).

#### 7. Non-Clinical Testing

Bench tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that proposed device complies with the following standards:

ASTM F1264-03 (Reapproved 2007). Standard Specification and Test Methods for Intramedullary Fixation Devices, including the following items:

- * Static bending test
- * Static torsion test
- * Dynamic bending test

ASTM F 543-07. Standard Specification and Test Methods for Metallic Medical Bone Screws- pullout strength, including the following item:

* Pull out test

#### 8. Substantially Equivalent Conclusion

The Ideal™ Intramedullary Nail System has same intended use and similar technological characteristics as the predicate device. The proposed device, the Ideal 131 Intramedullary Nail System, is deternined to be Substantially Equivalent (SE) to the predicate device. K121312 Intramedullary Nailing System, in respect to safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2013

Jiangsu Ideal Medical Science & Technology Company, Limited % Ms. Bo Gong Shanghai Yarui Consultant Company, Limited Room 503, Building 8, 600 Liu Zhou Road Shanghai, 200233 China

Re: K132428

Trade/Device Name: Ideal™ Intramedullary Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Codes: HSB Dated: September 19, 2013 Received: September 30, 2013

Dear Ms. Gong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Ms. Bo Gong

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4 of Traditional 510(K) Submission:

# Indications for Use

510(k) Number: K132428

Device Name: Ideal™ Intramedullary Nail System

Indications For Use:

The Ideal™ Intramedullary Nail System is intended to provide temporary fracture fixation and stabilization of the tibia. (Simple, compound first- and second-degree tibial shaft fractures, pseudarthrosis and delayed union).

Prescription Use_ X AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

**Elizabeth L. Frank -S**

Division of Orthopedic Devices

ldeal™, Jiangsu Ideal Medical

---

**Source:** [https://fda.innolitics.com/device/K132428](https://fda.innolitics.com/device/K132428)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com