The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy.
Device Story
Gynocular is a handheld, battery-powered colposcope providing magnified visualization (5x, 8x, 12x) of cervical, vaginal, and anogenital tissues. Device uses high-intensity LEDs for illumination and includes a green filter. Operated by physicians or trained colposcopists in clinical settings; can be used handheld or mounted on a standard tripod for hands-free operation. Device does not contact the patient. Output is direct optical visualization of tissue, enabling the clinician to evaluate lesions and identify biopsy sites. Benefits include portability and performance comparable to stationary colposcopes, facilitating diagnostic procedures in various clinical environments.
Clinical Evidence
Clinical evidence includes a randomized, crossover pilot trial of 69 women testing positive for visual inspection with acetic acid. The study compared Swede scores obtained via the Gynocular versus a standard stationary colposcope, correlated with histological diagnosis after biopsy. Results showed 70% overall agreement between the two techniques. The study concluded the Gynocular has a significant level of agreement with stationary colposcopes in assessing cervical lesions.
Technological Characteristics
Handheld colposcope with 300 mm focal distance and 5x, 8x, 12x magnification. Features high-intensity LED illumination, green filter, and rechargeable lithium-ion battery. Includes tripod-mounting capability. Meets IEC 60601-1 and EN 60601-1 safety standards. Includes automatic thermal shut-off mechanism.
Indications for Use
Indicated for magnified visualization of vaginal, cervical, and external genitalia tissues to evaluate and select areas for biopsy in patients requiring colposcopy.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Digital Colposcope & Image Management System (K122973)
{0}------------------------------------------------
GYNIUS AB 510(k) submission Gynoculan Colposcope
## APR 0 8 2014
### 5. 510(k) Summary
April 4th, 2014
(1) Submitter Information Name: Gynius AB Address: Gynius AB Dobelnsgatan 23 11140, Stockholm, Sweden Telephone #: +46 700237670 Email: info@gynocular.com
Contact Person: Matthew Volsky CEO - Gynius AB
(2) Name of Device
Trade Name: Gynocular Common Name: Colposcope Classification: Colposcope, 21 CFR 884.1630, Class II Product Code: HEX
(3) Predicate devices
| Applicant | Medgyn<br>Products,<br>Inc. | Intermed<br>Group Inc. | Medgyn<br>Products,<br>Inc. | Wallach<br>Surgical<br>Devices, Inc. |
|-------------|----------------------------------------------------|------------------------------------|-----------------------------|--------------------------------------|
| Device Name | Digital Colposcope &<br>Image Management<br>System | The Intermed<br>Zoom<br>Colposcope | Gynescope | Wallach Zoom<br>Colopscope |
| Model | AL-106 | IZM-310 | AL 103 | ColpoStar IH |
| K-Number | K122973 | K031629 | K880195 | K853389 |
(4) Device Description
The Gynocular is a hand-held, colposcope, which provides a magnified visualization of the tissues of the cervix, vagina, vulva and anogential area. It is used to evaluate these tissues and select areas for biopsy, without contacting the patient. The user is a medical physician, or a trained colposcopist.
The Gynocular is a colposcope with similar specifications to traditional colposcopes. The Gynocular has a 300 mm focal distance, 3-magnifications: 5x, 8x, and 12x. It is a handheld device that comes with a tripod-mounting clip that screws into a standard tripod enabling the
510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670
1
{1}------------------------------------------------
medical professional to also perform colposcopy hands-free mode for ease of biopsy. The Gynocular uses high intensity LEDs for illumination, has a green filter, and is powered by a rechargeable lithium-ion battery. We also provide a charging base for the Gynocular.
(5) Indications for use (colposcopy):
The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy.
(6) Comparison to Predicate Devices
| Applicant | Medgyn<br>Products, Inc. | Intermed<br>Group Inc. | Medgyn<br>Products,<br>Inc. | Wallach<br>Surgical<br>Devices, Inc. | Gynius AB |
|---------------------|---------------------------------------------------|---------------------------------------|--------------------------------|--------------------------------------|-------------------------------------|
| Device Name | Digital<br>Colposcope<br>&<br>Image<br>Management | The<br>Intermed<br>Zoom<br>Colposcope | Gynescope | WALLACH<br>ZOOM<br>COLPOSCO<br>PE | The<br>Gynocular |
| Model | AL-106 | IZM-310 | AL 103 | ColpoStar IH | VI |
| K-Number | K122973 | K031629 | K880195 | K853389 | n/a |
| Magnification | Optical<br>Display<br>;<br>36x | Standard 10x | 4.3x/7x/11.<br>3x | 12x | 5x/8x/12x |
| Working<br>distance | 200-300 mm | 310 mm | 300 mm | 300 mm | 300 mm |
| Lamp | LEDs | Halogen | Halogen | Halogen | LEDs |
| Filter | Green filter | Green filter | Green filter | Green filter | Green filter |
| Ocular type | Monocular | Binocular | Binocular | Binocular | Monocular |
| Stand | Stationary | Stationary | Stationary | Stationary | Portable/stationary on a<br>tripod. |
| Energy<br>source | Standard AC<br>Power Supply | Standard AC<br>Power<br>Supply | Standard AC<br>Power<br>Supply | Standard AC<br>Power<br>Supply | Lithium-ion<br>Battery |
The Gynocular's performance specifications are substantially equivalent to the listed predicate devices.
#### (7) Performance Data
.
The following testing was conducted to support a substantial equivalence determination.
- . Distortion testing:
- The analysis shows that the Gynocular exhibits a minimal Barrel type geometrical
distortion: 0.47%, 0.75% and 0.81% for 5x, 8x and 12x magnification, respectively. This distortion is at levels of professional camera lenses considered and is considered negligible for a visual examination of tissue. The analysis shows that the Gynocular exhibits a
510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670
2
{2}------------------------------------------------
negligible chromatic distortion: 3.3 um, 5.8 um and 3.3 um for 5x, 8x and 12x magnification, respectively. This chromatic distortion is equal and below the levels of visual acuity of the human eye (5-6 um at the retina).
- . Magnification testing: 4.81x ±0.15,7.74x ±0.15,12.40x ±0.15
- � Optical resolution: 5x = 28.6 cy/mm 8x = 40.5 cv/mm 12x = 56.8cy/mm
Design validation: . Firmware and mechanical tests were conducted to demonstrate that the mechanical functions and optical properties achieve the desired results.
- . Thermal safety:
After 1hr 55 min of continuous use, the Gynocular's outer heat sink reached 39.1 degrees C in an ambient temperature setting of 24 degrees C. The Gynocular has an automatic shut off mechanism if temperatures become unsafe.
- Battery Performance: .
The typical charging time from empty to full charge is 2.5 hours or less. The battery gives approximately 2hrs of illuminator on time (full intensity) from a full charge. The use of life for the battery is 500 full charging cycles.
#### Non-clinical tests
Appropriate risk analysis-driven product testing was conducted to evaluate product safety. The following standards were met:
- · IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)
- · EN 60601-1 2006
#### Clinical tests
Gynius conducted a randomized, crossover, pilot clinical trial for evaluating agreement of diagnosis of cervical lesions by colposcopy using a standard colposcope and the Gynocular. The study population was 69 women that tested positive for visual inspection with acetic acid. Swede scores were given at the time of colposcopy and compared with the final histological diagnosis after directed cervical biopsy. The Swede scores were compared across the two techniques and there was a 70% agreement overall. Conclusions: The study shows that in visual inspection with acetic acid positive women, a battery-driven, pocket-sized colposcope - the Gynocular, has a significant level of agreement with a stationary colposcope in assessing cervical lesions. The study was accepted and published by The Journal of Lower Genital Tract Disease in May 2013.
#### (8) Conclusions
The Gynocular is equivalent in safety and efficacy to the legally marketed predicate devices.
510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, with a double-stranded snake winding around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
April 8, 2014
Gynius AB Matthew Volsky CEO Dobelnsgatan 23 Stockholm 11140 Sweden
Re: K132423
> Trade/Device Name: The Gynocular Regulation Number: 21 CFR§ 884.1630 Regulation Name: Colposcope Regulatory Class: II Product Code: HEX Dated: March 6, 2014 Received: March 10, 2014
Dear Matthew Volsky,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not mislcading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Matthew Volsky
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm.
Sincerely yours,
Herbert P.Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
. GYNIUS AB
510(k) submission Gynocular™ Colposcope
## 4. Indications for Use
510(k) Number: K132423
Device Name: The Gynocular
Indications for Usc:
The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy.
Prescription for Use: X
(Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert P 2014.04.08 13:36:
510(k) Submission
Gynius AB P.O. Box 55503 1 0204, Stockholm Sweden info@gynocular.com T: 011 46 700237670
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.