IMMULITE 2000 ANTI-TG AB CALIBRATION VERIFICATION MATERIAL, IMMULITE 2000 ANTI-TPO AB CALIBRATION VERIFICATIN MATERIAL
Applicant
Siemens Helathcare Diagnostics, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Sep 20, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The IMMULITE® Anti-TG Ab Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Anti-TG Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert. The IMMULITE® Anti-TPO Ab Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Anti-TPO Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert. The IMMULITE® Thyroglobulin Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Thyroglobulin assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
Device Story
Calibration verification materials (CVM) for IMMULITE 2000 systems; used to verify assay calibration; consists of 4-level lyophilized vials (human or equine serum matrix); used by laboratory technicians; input is CVM sample run on IMMULITE 2000; output is recovered dose value compared against target ranges; assists in ensuring system accuracy and clinical decision-making reliability.
Clinical Evidence
Bench testing only. Stability validated up to 6-9 years. Traceability established to WHO 1st IRP 65/93 (Anti-TG), WHO 1st IRP 66/387 (Anti-TPO), and BCR 457 (Thyroglobulin). Performance validated using 27 replicates across multiple instruments and reagent lots; recovery within ±10-25% of assigned dose.
Technological Characteristics
Lyophilized human or equine serum matrix with preservatives; 4-level concentration; stored at 2-8°C; traceable to international reference materials (WHO/BCR); intended for use on IMMULITE 2000 immunoassay systems.
Indications for Use
Indicated for in vitro diagnostic use to verify calibration of IMMULITE 2000 Anti-TG, Anti-TPO, and Thyroglobulin assays. For prescription use only.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
K133122 — IMMULITE 2000; 3GALLERGY SPECIFIC IGE CONTROL VERIFICATION MATERIAL (CVM), TOTAL IGE CONTROL VERIFICATION MATERIAL (CVM) · Siemens Healthcare Diagnostics, Inc. · Oct 30, 2013
K142811 — IMMULITE 2000 LH Calibration Verification Material (CVM), IMMULITE 2000 Free T3 Calibration Verification Material (CVM), IMMULITE 2000 Gastrin Calibration Verification Material (CVM) · Siemens Healthcare Diagnostics, Inc. · Oct 29, 2014
K140818 — IMMULITE: 2000 FREE T4 CALIBRATION VERIFICATION MATERIAL, 2000 IGF-I CALIBRATION VERIFICATION MATERIAL, 2000 PROLACTIN C · Siemens Healthcare Diagnostics, Inc. · May 1, 2014
K143636 — IMMULITE 2000 Androstenedione Calibration Verification Material, IMMULITE 2000 Troponin I Calibration Verification Material · Siemens Healthcare Diagnostics, Inc. · Feb 25, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92. the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K132391
| 1. Submitter | Siemens Healthcare Diagnostics Inc. | | SEP 2 0 2013 |
|--------------------------|--------------------------------------------------------------------------------|--|--------------|
| Mailing Address: | 511 Benedict Avenue<br>Tarrytown, NY 10591 | | |
| Contact Person: | Garo Mimaryan, MS, RAC<br>Technical Regulatory Affairs Specialist III | | |
| Phone Number: | (914)-524-3270 | | |
| Fax Number: | (914)-524-2101 | | |
| E-mail Address: | garo.mimaryan@siemens.com | | |
| Date Prepared: | September 16, 2013 | | |
| 2. Device Name | | | |
| Proprietary Name: | IMMULITE® 2000 Anti-TG Ab Calibration Verification Material | | |
| Measurand: | Quality Control materials for IMMULITE® 2000 Anti-TG assay | | |
| Type of Test: | Calibration Verification Material (CVM) for IMMULITE® 2000<br>Anti-TG Ab assay | | |
| Regulation Section: | 21 CFR 862.1660, Quality Control Material | | |
| Classification: | Class I Reserved | | |
| Products Code: | JJX – Single (Specified) Analyte Controls (Assayed and<br>Unassayed) | | |
| Panel: | Clinical Chemistry (75) | | |
| 3. Predicate Device Name | Elecsys Anti-TG CalCheck | | |
| Predicate 510(k) No: | K020369 | | |
| 4. Device Description: | The Calibration Verification Material (CVM) contains one set of | | |
matrix with preservatives. LTGCVM2, LTGCVM3 and LTGCVM4 contain low, intermediate and high levels of Anti-TG Ab respectively, in human serum/ buffer matrix matrix with preservatives.
> See Indications for Use Statement below The IMMULITE® Anti-TG Ab Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Anti-TG Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
four vials, 2 mL each. LTGCVM i contains human serum/ buffer
For prescription use only
IMMULITE® 2000 Systems
Special Conditions for Use Statement(s): Special Instrument Requirements:
5. Intended Use:
Indication for Use:
{1}------------------------------------------------
### 6. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Anti-TG Ab Calibration Verification Material (CVM) is substantially equivalent to the predicate device, Elecsys Anti-TG CalCheck, as summarized in Table 1.
Table 1: Substantial Equivalence Comparison
| SIMILARITIES | | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Anti-TG Ab CVM | Elecsys Anti-Tg CalCheck |
| Intended<br>Use | The IMMULITE® Anti-TG Ab<br>Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in<br>the verification of calibration of the<br>IMMULITE Anti-TG Ab assay on the<br>IMMULITE 2000 systems as indicated<br>in the CVM Package Insert. | The Elecsys Anti-Tg CalCheck is for<br>use in the verification of the<br>calibration established by the Elecsys<br>Anti-Tg reagent on the indicated<br>Elecsys and cobas e immunoassay<br>analyzers. |
| Analyte | Anti-TG Ab | Anti-Tg |
| Form | Lyophilized | Lyophilized |
| Matrix | Human Serum | Human Serum |
| stability | Stable until the expiration date when<br>stored refrigerated. | Stable until the expiration date when<br>stored refrigerated. |
| Storage | 2-8°C | 2-8°C |
| | DIFFERENCES | |
|--------|------------------------------|--------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Anti-TG Ab CVM | Elecsys Anti-Tg CalCheck |
| Levels | 4 | 3 |
| Use | Single Use Only | Not For Single Use |
#### 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Anti-TG Ab Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The IMMULITE® 2000 Anti-TG Ab Calibration Verification Materials (CVMs) are stable up to 7 years when stored refrigerated at 2-8°C prior to opening.
{2}------------------------------------------------
#### Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
#### Table 2: Stability Protocol Summary
| CVM Level | Time-Points | | | |
|-----------|-------------|-----|-----|-----|
| | (Days) | | | |
| LTGCVM1 | 1 | 182 | 365 | 548 |
| LTGCVM2 | 1 | 182 | 365 | 548 |
| LTGCVM3 | 1 | 182 | 365 | 548 |
| LTGCVM4 | 1 | 182 | 365 | 548 |
#### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE 2000 Anti-TG Ab Calibration Verification Material (CVM) consists of 2 parts. Part 1 consists of the Guideline Acceptance Criteria which requires the dose value stability CVM to fall between ±15% of the assigned dose. Part 2 consists of the Review Limits Acceptance Criteria which requires the dose value of the controls to be within 2SD of the control target value generated from the stability calibrator curve. If the result is not within the acceptable dose range of ±15%, then additional data review is performed using the part 2 acceptance criterion.
#### Traceability:
The IMMULITE® 2000 Anti-TG Ab CVMs are traceable to WHO 1st IRP 65/93. The CVMs are manufactured using qualified materials and measurement procedures.
#### Value Assignment:
The IMMULITE Anti-TG Ab (Anti-Thyroglobulin antibody) CVMs are 4 level materials which are a subset of 8 level Anti-TG Ab calibrators, Calibrators are not commercialized but are used internally during manufacture and release testing of Anti-TG Ab reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The assigned reference calibrators are prepared using Anti-TG antibody stock and are traceable to WHO 1ª IRP 65/93. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Two levels of commercially available controls, and 30 patient samples (10 spiked normal patients samples, two patient samples and 18 spiked samples) are used to validate CVM value assignments.
The CVMs are manufactured using qualified materials and measurement procedures. The CVMs were tested on 27 replicates in total comprised of nine runs and three replicates per run on seven IMMULITE 2000 systems and five different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
{3}------------------------------------------------
#### Expected Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run. 5 different reagent kit lots and 7 different instruments were used to gain 27 replicates. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected values are provided in the IMMULITE® 2000 CVM Calibration verification Material lot-specific value card. The expected assay range is 20 -3000 IU/mL. The target values in Table 3 can be considered as guidelines.
| Table 3: | | Target Values |
|----------|--|---------------|
|----------|--|---------------|
| Analyte target<br>levels | Level | Target Mean<br>(IU/mL) | SD | Guideline ±2SD Range<br>(IU/mL) |
|--------------------------|----------|------------------------|--------|---------------------------------|
| | 1 | 0.00 | - | 0.00 ≤20.00 |
| | 2 | 73.5 | - | - |
| | 50% CVM1 | 36.8 | 2.75 | 31.3 42.3 |
| | 50% CVM2 | | | |
| | 3 | 452 | 34 | 384 520 |
| | 4 | 3130 | 234.75 | 2661 3600 |
| Assay Range | | 20-3000 IU/mL | | |
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material .
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
#### Conclusion: 9.
The IMMULITE® 2000 Anti-TG Ab (Anti-Thyroglobulin antibody) Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Elecsys Anti-Tg CalCheck. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device. Elecsys Anti-TG CalCheck, The IMMULITE® 2000 Anti-TG Ab Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{4}------------------------------------------------
## 510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K132391
- 1. Submitter Mailing Address:
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
- 2. Device Name Proprietary Name: Measurand: Type of Test:
Regulation Section: Classification: Products Code:
Panel:
- 3. Predicate Device Name Predicate 510(k) No: 4. Device Description:
5. Intended Use:
Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com September 16, 2013
IMMULITE® 2000 Anti-TPO Ab Calibration Verification Material Quality Control materials for IMMULITE® 2000 Anti-TPO assay Calibration Verification Material (CVM) for IMMULITE® 2000 Anti-TPO Ab assay 21 CFR 862.1660. Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)
Elecsys Anti-TPO CalCheck K001014 The Calibration Verification Material (CVM) contains one set of four vials, 2 mL each. LTOCVM1 contains human serum/ buffer matrix with preservatives. LTOCVM2, LTOCVM3 and LTOCVM4 contain low, intermediate and high levels of Anti-TPO Ab respectively, in human serum/ buffer matrix matrix with preservatives.
See Indications for Use Statement below The IMMULITE® Anti-TPO Ab Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Anti-TPO Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
For prescription use only
IMMULITE® 2000 Systems
{5}------------------------------------------------
#### A comparison of the device features, intended use, and other 6. Technological Characteristics information demonstrates that the IMMULITE® 2000 Anti-TPO` and Substantial Equivalence Ab Calibration Verification Material (CVM) is substantially Comparison with Predicate: equivalent to the predicate device, Elecsys Anti-TPO CalCheck, as summarized in Table 1.
| SIMILARITIES | | |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Anti-TPO Ab CVM | Elecsys Anti-TPO CalCheck |
| Intended<br>Use | The IMMULITE® Anti-TPO Ab<br>Calibration Verification Material (CVM)<br>is for in vitro diagnostic use in the<br>verification of calibration of the<br>IMMULITE Anti-TPO Ab assay on the<br>IMMULITE 2000 systems as indicated<br>in the CVM Package Insert. | The Elecsys Anti-TPO CalCheck is<br>for use in the verification of the<br>calibration established by the Elecsys<br>Anti-TPO reagent on the indicated<br>Elecsys and cobas e immunoassay<br>analyzers. |
| Analyte | Anti-TPO Ab | Anti-TPO |
| Form | Lyophilized | Lyophilized |
| Matrix | Human serum | Human serum |
| Stability | Stable until the expiration date when<br>stored refrigerated. | Stable until the expiration date when<br>stored refrigerated. |
| Storage | 2-8°C | 2-8°C |
| | Table 1: Substantial Equivalence Comparison | |
|--|----------------------------------------------|--|
| | | |
| | DIFFERENCES | |
|--------|-------------------------------|---------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Anti-TPO Ab CVM | Elecsys Anti-TPO CalCheck |
| Levels | 4 | 3 |
| Use | Single Use Only | Not For Single Use |
#### 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Anti-TPO Ab Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The IMMULITE® 2000 Anti-TPO Ab Calibration Verification Materials (CVMs) are stable up to 9 years when stored refrigerated at 2-8℃ prior to opening.
### Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
{6}------------------------------------------------
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LTOCVM1 | 1 | 2555 | 2920 | 3285 |
| LTOCVM2 | 1 | 2555 | 2920 | 3285 |
| LTOCVM3 | 1 | 2555 | 2920 | 3285 |
| LTOCVM4 | 1 | 2555 | 2920 | 3285 |
### Table 2: Stability Protocol Summary
#### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE 2000 Anti-TPO Ab Calibration Verification Material (CVM) consists of 2 parts. Part 1 consists of the Guideline Acceptance Criteria which requires the dose value stability CVM to fall between ±25% of the assigned dose for CVM level 2 and ±15% of assigned dose for CVM levels 3 and 4. Part 2 consists of the Review Limits Acceptance Criteria which requires the dose value of the controls to be within 2SD of the control target value generated from the stability calibrator curve. If the result is not within the acceptable dose range of £25% for level 2 and ±15% for levels 3 and 4, then additional data review is performed using the part 2 acceptance criterion.
#### Traceability:
The IMMULITE Anti-TPO Ab CVMs are traceable to WHO 15 IRP 66/387. The CVMs are manufactured using qualified materials and measurement procedures.
#### Value Assignment:
Anti-TPO Ab (Anti-Thyroid Peroxidase Antibody) CVMs are 4 level materials which are a subset of 7 level Anti-TPO Ab calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Anti-TPO Ab reagents and two point adjustors. The assigned reference calibrators are prepared using Anti-TPO antibody stock and are traceable to WHO Ist IRP 66/387. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Two levels of commercially available controls, and 30 patient samples and 25 spiked samples) are used to validate CVM value assignments.
The CVMs are manufactured using qualified materials and measurement procedures. The CVMs were tested on 27 replicates in total comprised of nine runs and three replicates per run on five IMMULITE 2000 systems and four different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
{7}------------------------------------------------
#### Expected Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run. 4 different reagent kit lots and 5 different instruments were used to gain 27 replicates. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected values are provided in the IMMULITE® 2000 CVM Calibration verification Material lot-specific value card. The expected assay range is 10-1000 IU/mL .. The target values in Table 3 can be considered as guidelines.
Table 3: Target Values
| Analyte target<br>levels | Level | Target Mean<br>(IU/mL) | SD | Guideline ±2SD Range<br>(IU/mL) | |
|--------------------------|-----------------|------------------------|-------|---------------------------------|--|
| | 1 | 0.00 | - | ≤10.00 | |
| | 2 | 34.1 | 3.4 | 27.3 40.9 | |
| | 3 | 281 | 21 | 239 323 | |
| | 4 | 950 | 71.25 | 808 1093 | |
| Assay Range | 10 - 1000 IU/mL | | | | |
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators
- Guidance for Industry and FDA Staff Assayed and Unassaved Quality Control Material .
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
#### 9. Conclusion:
Anti-TPO Ab (Anti-Thyroid Peroxidase Antibody) Calibration Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Elecsys Anti-TPO CalCheck. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device. Elecsys Anti-TPO CalCheck. The IMMULITE® 2000 Anti-TPO Ab Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{8}------------------------------------------------
### 510(k) Summary
According to the requirements of 21 CFR 807.92, the following information provides Introduction: sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K132391
- 1. Submitter Mailing Address:
511 Benedict Avenue Tarrytown, NY 10591
Contact Person:
Phone Number: Fax Number: E-mail Address: Date Prepared:
2. Device Name Proprietary Name: Measurand: Type of Test:
> Regulation Section: Classification: Products Code: Panel:
3. Predicate Device Name Predicate 510(k) No:
4. Device Description:
- 5. Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
Garo Mimaryan, MS, RAC Technical Regulatory Affairs Specialist III (914)-524-3270 (914)-524-2101 garo.mimaryan@siemens.com September 16, 2013
Siemens Healthcare Diagnostics Inc.
IMMULITE® 2000 Thyroglobulin Calibration Verification Material Quality Control materials for IMMULITE® 2000 Thyroglobulin assay Calibration Verification Material (CVM) for IMMULITE® 2000 Thyroglobulin assay 21 CFR 862.1660. Quality Control Material
Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)
ADVIA Centaur® Intact PTH Master Curve Material (MCM) K020217
The Calibration Verification Material (CVM) contains one set of four vials, 2 mL each. LTYCVM1 in horse serum matrix with preservatives. LTYCVM2. LTYCVM3 and LTYCVM4 contain low, intermediate and high levels of Thyroglobulin respectively, in horse serum matrix with preservatives.
See Indications for Use Statement below The IMMULITE® Thyroglobulin Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Thyroglobulin assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
#### For prescription use only
IMMULITE® 2000 Systems
{9}------------------------------------------------
- A comparison of the device features, intended use, and other 6. Technological Characteristics and Substantial Equivalence information demonstrates that the IMMULITE® 2000 Thyroglobulin Calibration Verification Material (CVM) is Comparison with Predicate: substantially equivalent to the predicate device, ADVIA Centaur® Intact PTH Master Curve Material, as summarized in Table 1.
### Table 1: Substantial Equivalence Comparison
| SIMILARITIES | | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE® 2000 Thyroglobulin<br>Calibration Verification Material | ADVIA Centaur® Intact PTH<br>Master Curve Material |
| Intended<br>Use | The IMMULITE® Thyroglobulin<br>Calibration Verification Material (CVM) is<br>for in vitro diagnostic use in the verification<br>of calibration of the IMMULITE<br>Thyroglobulin assay on the IMMULITE<br>2000 systems as indicated in the CVM<br>Package Insert. | For in vitro diagnostic use for<br>evaluating the ADVIA Centaur®<br>Intact PTH assays. This material is<br>intended to be run singly as<br>unknown samples after a two-point<br>calibration has been performed on<br>the system. |
| Form | Lyophilized | Lyophilized |
| Matrix | Equine Serum | Equine Serum |
| Use | Single Use Only | Single Use Only |
| | DIFFERENCES | |
|-----------|---------------------------------------------------------------------------|----------------------------------------------------------|
| | Candidate Device | Predicate Device |
| | IMMULITE 2000 Thyroglobulin<br>Calibration Verification Material<br>(CVM) | ADVIA Centaur® Intact PTH<br>Master Curve Material (MCM) |
| Analyte | Thyroglobulin | Intact PTH |
| Levels | 4 | 7 |
| Stability | Stable until the expiration date when<br>stored refrigerated. | Stable until the expiration date<br>when stored frozen. |
| Storage | 2-8°C | -20°C |
#### 7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
#### Stability Summary:
The stability study was conducted to validate shelf life claim for the IMMULITE 2000 Thyroglobulin Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The IMMULITE® 2000 Thyroglobulin Calibration Verification Materials (CVMs) are stable up to 6 years when stored refrigerated at 2-8℃ prior to opening.
{10}------------------------------------------------
#### Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) and the dose value determined from the reference calibrator curve and is summarized in Table 2.
#### Table 2: Stability Protocol Summary
| CVM Level | Time-Points<br>(Days) | | | |
|-----------|-----------------------|------|------|------|
| LTYCVM1 | 1 | 1825 | 2190 | 3285 |
| LTYCVM2 | 1 | 1825 | 2190 | 3285 |
| LTYCVM3 | 1 | 1825 | 2190 | 3285 |
| LTYCVM4 | 1 | 1825 | 2190 | 3285 |
#### Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE 2000 Anti-TG Ab Calibration Verification Material (CVM) consists of 2 parts. Part 1 consists of the Guideline Acceptance Criteria which requires the dose value stability CVM to fall between ±10%. Part 2 consists of the Review Limits Acceptance Criteria which requires the dose value of the controls to be within 2SD of the control target value generated from the stability calibrator curve. If the result is not within the acceptable dose range of ±10%, then additional data review is performed using the part 2 acceptance criterion.
#### Traceability:
The IMMULITE Thyroglobulin CVMs are traceable to Certified Reference Material from human thyroglobulin (BCR 457) in terms of the Institute of Reference Materials and Measurements of the European Commission (formerly the Community Bureau of Reference. The CVMs are manufactured using qualified materials and measurement procedures.
#### Value Assignment:
Thyroglobulin CVMs are 4 level materials which are a subset of 9 level Thyroglobulin calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Thyroglobulin reagents and two point adjustors. The IMMULITE Thyroglobulin CVMs are traceable to Certified Reference Material from human thyroglobulin (BCR 457) in terms of the Institute of Reference Materials and Measurements of the European Commission (formerly the Community Bureau of Reference. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Three levels of commercially available controls, and 30 patient samples (5 normal patients samples and 25 spiked samples) are used to validate CVM value assignments.
{11}------------------------------------------------
The CVMs are manufactured using qualified materials and measurement procedures. The CVMs were tested on 27 replicates in total comprised of nine runs and three replicates per run on four IMMULITE 2000 systems and four different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. The controls must fall within their target ranges
#### Expected Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run. 4 different reagent kit lots and 4 different instruments were used to gain 27 replicates. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected values are provided in the IMMULITE® 2000 CVM Calibration verification Material lot-specific value card. The expected assay range is 02-300 NG/mL. The target values in Table 3 can be considered as guidelines.
| Analyte target<br>levels | Level | Target Mean<br>(ng/mL) | SD | Guideline ±2SD Range<br>(ng/mL) | | |
|--------------------------|----------------|------------------------|------|---------------------------------|-------|--|
| | 1 | 0.00 | | 0.00 | ≤20.0 | |
| | 2 | 9.10 | 0.5 | 8.10 | 10.1 | |
| | 3 | 36.2 | 2.15 | 31.9 | 40.5 | |
| | 4 | 430 | - | - | - | |
| | 30% CVM1 | | | | | |
| | 70% CVM4 | 301 | 21 | 259 | 343 | |
| Assay Range | 0.2 -300 ng/mL | | | | | |
Table 3: Target Values
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic ● Calibrators
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material .
#### Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
{12}------------------------------------------------
## siemens
#### 9. Conclusion:
Thyroglobulin Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed ADVIA Centaur® Intact PTH Master Curve Material. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, ADVIA Centaur® Intact PTH Master Curve Material, The IMMULITE® 2000 Thyroglobulin Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{13}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/13/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 20, 2013
SIEMENS HEALTHCARE DIAGNOSTICS, INC. C/O MR. GARO MIMARYAN TECHNICAL REGULATORY AFFAIRS SPECIALIST III 511 BENEDICT AVENUE TARRYTOWN. NY 10591-5097
Re: K132391
Trade/Device Name: IMMULITE® 2000 Anti-TG Ab Calibration Verification Material; IMMULITE® 2000 Anti-TPO Ab Calibration Verification Material; IMMULITE® 2000 Thyroglobulin Calibration Verification Material; Regulation Number: 21 CFR 862.1660
Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I Reserved Product Code: JJX Dated: July 31, 2013 Received: August 1. 2013
Dear Mr. Mimaryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{14}------------------------------------------------
Page 2-Mr. Mimaryan
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Maria M. Chan -S
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration -
## Indications for Use
510(k) Number (if known) K 13239 l :
#### Device Name
IMMULITE® 2000 Anti-TG Ab Calibration Verification Material
#### Indications for Use (Describe)
The IMMULITE® Anti-TG Ab Calibration Material (CVM) is for in vitro diagnostic use in the verification of callbration of the IMMULITE Anti-TG Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Maria M. Chan -S
Image /page/15/Picture/18 description: The image shows a logo with the letters FDA. The letters are stylized and appear to be outlined. The logo is simple and monochromatic.
FORM FDA 3881 (9/13)
### Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
PSC Publishing Services (301) 343-6740 EF
{16}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K 132391
#### Device Name
IMMULITE® 2000 Thyroglobulin Calibration Verification Material
#### Indications for Use (Describe)
The IMMULITE® Thyroglobulin Calibration Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Thyroglobulin assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Maria M. Chan -
Image /page/16/Picture/18 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is in black and white and features the letters "FDA" in a stylized font. The letters are bold and have a slightly rounded appearance. The logo is simple and recognizable, and it is often used on products that have been approved by the FDA.
FORM FDA 3881 (6/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{17}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132391
#### Device Name
IMMULITE® 2000 Anti-TPO Ab Calibration Verification Material
#### Indications for Use (Describe)
The IMMULITE® Anti-TPO Ab Calibration Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Anti-TPO Ab assay on the IMMULITE 2000 systems as indicated in the CVM Package Insert.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Maria M. Chan -S
FDA
FORM FDA 3881 (6/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.