K132353 · Inex · QPB · Jul 30, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132353
Device Name
SUCTION LIPOPLASTY ACCESSORIES
Applicant
Inex
Product Code
QPB · General, Plastic Surgery
Decision Date
Jul 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The aspiration and infusion cannulae and needles are indicated for aesthetic body contouring and general tissue aspiration.
Device Story
INEX Cannulae and Needles are manual surgical instruments used for aesthetic body contouring and general tissue aspiration. The device consists of a hollow stainless steel tube with multiple tip configurations, a plastic handle, and a plastic hub. The device connects via tubing to an external aspirator, wall suction, or pump. Operated by surgeons in clinical settings, the device facilitates the removal of fluid, soft tissue, and exudates, or the infiltration of fluids. The surgeon grips the handle to maneuver the cannula/needle, allowing for precise tissue aspiration or infiltration. The device is supplied sterile for single use, eliminating the need for reprocessing.
Clinical Evidence
Bench testing only. Biocompatibility testing was performed in accordance with ISO-10993 standards to demonstrate safety.
Technological Characteristics
Materials: Stainless steel tubes/tips, plastic handles, polypropylene luer hubs. Design: Hollow tube with eyelets for suction/infiltration. Connectivity: Manual connection to external suction/pump via tubing. Sterilization: Supplied sterile for single use. No software or electronic components.
Indications for Use
Indicated for aesthetic body contouring and general tissue aspiration in patients requiring such procedures.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 9, 2021
Inex Raymond Patel Business Manager Za La Gobette 60540 Puiseux Le Hauberger Puiseux Le Hauberger, 60540 France
Re: K132353
Trade/Device Name: Suction Lipoplasty Accessories Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Raymond Patel:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 30, 2014. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 30, 2014
INEX Mr. Raymond Patel Business Manager ZA LA GOBETTE 60540 Puiseux Le Hauberger France
Re: K132353
Trade/Device Name: Cannulae and Needles Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU, GEA Dated: June 27, 2014 Received: June 30, 2014
Dear Mr. Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
K132353
Device Name Cannulae and Needles
Indications for Use (Describe)
The aspiration and infusion cannulae and needles are indicated for aesthetic body contouring and general tissue aspiration.
Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY ﺗﺮ : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
04'00
{5}------------------------------------------------
INEX
K132353
510(k) Summary in accordance with 21CFR 807.92.
Traditional 510(k) Submission by:
JUL 3 0 2014
INEX ZA LA GOBETTE 60540 Puiseux Le Hauberger France laboinex@wanadoo.fr Contact Person: Raymond Patel INEX 33(0)3.44.74.19.95
Date Prepared: July 29, 2014
Establishment Registration Number: 10042355
| Suction Lipoplasty System | 21 CFR 878.5040 (05 Jan 1998) |
|---------------------------|-------------------------------------------------------------|
| | Manual Surgical Instruments for General Use |
| | 21 CFR 878.4800 (1994) |
| Common Usual Name: | Aspiration and infiltration Cannulae/Needles |
| Proprietary Name: | Lipoplasty/Liposuction Aspiration and |
| | Tumescent Infiltration Cannulae/Needles |
| | (Cannulae and Needles) |
| Classification Name: | Class II, 21 CFR 878.5040, Suction Lipoplasty System, Panel |
| | 79, MUU, General and Plastic Surgery |
| Predicate Device: | Suction Cannulae and Needles |
| | Manufacturer: Black & Black Surgical, Inc. |
| | 510(k) Number: K113795 |
| | Substantially Equivalence Date: August 7, 2012 |
Device Description: INEX Cannulae and Needles are used to remove fluid, soft tissue, andCannulae and Needles are used to remove fluid, soft tissue, and exudates and infusion, INEX Device Description: utilizing a hollow stainless steel tube and multiple tips, handle and attachment connectors that are disposable in configuration. They are used during general, plastic, and reconstructive procedures.
Device Description Chart Length Features Packaging Description * Product Code 1 each, 10/box & Cannulae & A - Z letters and Various Straight and . curved 20/box 1 - 0 numbers. Needles
INEX Cannulae and Needles are manufactured of stainless steel tubes, with Plastic Handles and Plastic Hubs.
The aspiration and infusion cannulae and needles are indicated for aesthetic body contouring Indication for Use: and general tissue aspiration.
#### Shared Technical Characteristics with Predicate Device:
The INEX Cannulae and Needles are substantially equivalent in function and intended use to the Black Surgical Cannulae and Needles (K113795). Both devices are used for aesthetic body contouring and general tissue aspiration by means of aspiration and infiltration.
Biocompatibility testing per ISO-10993 demonstrated that the device is biocompatible.
{6}------------------------------------------------
| Device | Design | Material | Function |
|---------------|------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INEX | Handle; Cannula<br>tube; Tip | Plastic Handle;<br>Stainless Steel<br>Cannula Tube;<br>Stainless Steel Tip<br>with eyelets(holes) | The Handle is used to grip (hold) and to<br>connect via tubing to the aspirator, wall<br>suction or pump; the Cannula tube is<br>used to provide length and strength; the<br>Tip is used to provide suction and<br>infiltration through eyelets (holes) in the<br>tip. |
| Black & Black | Handle; Cannula<br>tube; Tip | Aluminum Handle;<br>Stainless Steel<br>Cannula Tube;<br>Stainless Steel Tip<br>with eyelets(holes) | The Handle is used to grip (hold) and to<br>connect via tubing to the aspirator, wall<br>suction or pump; the Cannula tube is<br>used to provide length and strength; the<br>Tip is used to provide suction and<br>infiltration through eyelets (holes) in the<br>tip. |
| INEX | Luer Hub; Needle Tip | Polypropylene Luer hub;<br>Stainless Steel Cannula<br>Tube; Stainless Steel Tip<br>with eyelets(holes) | The Luer hub is used to connect to an<br>aluminum handle via tubing to the<br>aspirator, wall suction, or pump; the<br>Cannula tube is used to provide length and<br>strength; the Tip is used to provide suction<br>and infiltration through eyelets (holes) in<br>the tip. |
#### Summary
The only difference between the Black Cannula and the INEX Cannula is the handle material. The Black and Black has an aluminum handle that can be autoclaved for multiple uses and the INEX Cannula has a plastic single use handle and is supplied sterile for single use. The two handles is the same which is to allow the surgeon to grip and maneuver the cannula and to connect the sterile tubing to an aspirator, wall suction, or pump.
The only difference between the Black needle and the INEX needle is the luer hub material. The Black and Black needle has a stainless steel luer hub that can be autoclaved for multiple uses and the INEX needle has a polypropylene single use luer hub and is supplied sterile for single use. The function of the same which is to allow the surgeon to join the needle to a handle and to connect the sterile tubing to an aspirator, wall suction, or pump.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.