K132348 · Tepha, Inc. · NWJ · Sep 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132348
Device Name
TEPHAFLEX BRAIDED SUTURE
Applicant
Tepha, Inc.
Product Code
NWJ · General, Plastic Surgery
Decision Date
Sep 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4494
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TephaFLEX braided sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
Device Story
TephaFLEX Braided Suture is a sterile, absorbable surgical suture composed of poly-4-hydroxybutyrate (P4HB). The device features a multifilament inner core covered by a braided multifilament sheath. It is available in dyed (D&C Violet No. 2) or undyed forms, with or without attached needles. Used by surgeons for soft tissue approximation and ligation, the suture provides extended wound support. The device is absorbed in vivo over time, with strength retention profiles comparable to predicate absorbable sutures. Clinical benefit includes secure tissue approximation during the healing period followed by gradual absorption, reducing the need for suture removal. The device is provided sterile for clinical use.
Clinical Evidence
No human clinical trials were conducted. Evidence consists of bench testing and in vivo animal studies. Bench testing confirmed compliance with USP 33 standards for diameter, knot pull tensile strength, and needle attachment. Biocompatibility testing (ISO 10993) included cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, genotoxicity, and subcutaneous implantation in rabbits (12-78 weeks), showing non-toxic responses. In vivo rabbit studies demonstrated tensile strength retention and mass loss rates equivalent to predicate devices.
Indicated for general soft tissue approximation and/or ligation. Contraindicated for cardiovascular, neurological, microsurgery, or ophthalmic procedures.
Regulatory Classification
Identification
An absorbable poly(hydroxybutyrate) surgical suture is an absorbable surgical suture made of material isolated from prokaryotic cells produced by recombinant deoxyribonucleic acid (DNA) technology. The device is intended for use in general soft tissue approximation and ligation.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Absorbable Poly(hydroxybutyrate) Surgical Suture Produced by Recombinant DNA Technology.” For the availability of this guidance document see § 878.1(e).
{1}------------------------------------------------
K132348 page 1/4
Image /page/0/Picture/1 description: The image shows the logo for TEPHA Medical Devices. The logo features a stylized black graphic element above the text. The text "TEPHA" is in large, bold, sans-serif font, with "MEDICAL DEVICES" in a smaller font size below it.
99 Hayden Avenue
Suite 360 Lexington, MA 02421 tel: (781) 357-1700 fax: (781) 357-1701
#### Section X 510(k) Summary
(Prepared on July 26, 2013)
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Tepha, Inc. is submitting the following summary of information respecting safety and effectiveness:
| Trade Name: | TephaFLEX Braided Suture |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Tepha, Inc.<br>99 Hayden Avenue, Suite 360<br>Lexington, MA 02421 |
| Contact Person: | Mary P. LeGraw, V.P., Regulatory Affairs |
| Telephone: | 781-357-1709 |
| Fax: | 781-357-1701 |
| Email: | legraw@tepha.com |
| Device Classification Name: | CFR §878.4494 - Product Code: NWJ<br>Absorbable Poly(hydroxybutyrate) Surgical Suture Produced by<br>Recombinant DNA Technology |
| Classification: | According to Section 13 of the Federal Food, Drug and Cosmetic<br>Act, the device classification is Class II, Performance Standards. |
| Predicate Devices: | Tepha, Inc., TephaFLEX Absorbable Suture - K052225, K082178<br>Tornier, Inc., BioFiber Suture – K122487, K130422<br>Ethicon, Inc., Perma-Hand Silk Braided Suture - K930360<br>U.S. Surgical (Covidien), Dexon S Suture - K972566<br>Ethicon, Inc., Vicryl Suture - K022269<br>Please see attached Substantial Equivalence table comparing<br>the TephaFLEX Braided suture to the predicate devices. |
| Device Description: | TephaFLEX suture is a sterile, braided, surgical suture<br>constructed of poly-4-hydroxybutyrate (P4HB). The suture<br>consists of an inner core of multifilament fibers covered by a<br>braided sheath made of multifilament fibers. The suture is<br>provided dyed (D&C Violet No. 2) or un-dyed and is offered in a<br>variety of cut lengths, with or without needles attached. |
| Indications for Use: | The TephaFLEX sutures are indicated for use in general soft<br>tissue approximation and/or ligation, but not for use in<br>cardiovascular or neurological tissues, microsurgery, or<br>ophthalmic surgery. |
| Safety and Performance: | All testing was performed in compliance with the FDA Guidance:<br>Class II Special Controls Guidance: Absorbable<br>Poly(hydroxybutyrate) Surgical Suture Produced by Recombinant<br>DNA Technology. |
SEP 0 5 2013309
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for TEPHA MEDICAL DEVICES. The logo features a stylized swoosh above the word "TEPHA" in large, bold, sans-serif font. Below "TEPHA" are the words "MEDICAL DEVICES" in a smaller, sans-serif font. The logo is black and white.
K132348 page 2/4
99 Hayden Avenue Suite 360 Lexington, MA 02421 tel: (781) 357-1700 fax: (78) 357-1701
Physical/mechanical testing performed on the TephaFLEX suture verified conformance to USP 33 for absorbable surgical sutures, including <861> Suture Diameter, <881> Knot Pull Tensile Strength, and <871> Needle Attachment Strength.
Testing was also performed for conformance to ISO-10993 for biocompatibility. Tepha performed Cytotoxicity, Intracutaneous Irritation, Sensitization, Acute Systemic Toxicity, Pyrogenicity, Genotoxicity, and 12, 26, 52 and 78 week Subcutaneous Implantation studies in rabbits. All testing yielded a non-toxic response.
In vivo implantation studies were conducted in rabbits to demonstrate rates of tensile strength and mass loss. Results show the in vivo strength retention of the TephaFLEX braided suture to be equivalent to the published strength retention of the predicate devices with the same clinical indications over the critical healing period. Therefore, the TephaFLEX braided suture is equivalent to the predicate device in regard to its strength retention profile.
Conclusion:
Based on the indications for use, technological characteristics, and the results of safety and performance testing described above, the TephaFLEX braided suture has been shown to be substantially equivalent to predicate devices used for the same clinical indications under the Federal Food, Drug and Cosmetic Act.
310
{3}------------------------------------------------
| Characteristic | Tepha, Inc.<br>TephaFLEX Braided Suture | Tepha, Inc.<br>TephaFLEX Absorbable Suture<br>K052225, K081099, K082178 | Ethicon, Inc., Perma-Hand Silk<br>Suture<br>K930360 | Covidien/US Surgical<br>Dexon Braided Suture<br>K972566 |
|----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Indicated for use in soft<br>tissue approximation<br>and/or ligation but not for<br>use in cardiovascular or<br>neurological surgery,<br>microsurgery or<br>ophthalmic surgery. | Indicated for use in soft tissue<br>approximation and/or ligation but not<br>for use in cardiovascular or<br>neurological surgery, microsurgery<br>or ophthalmic surgery.<br>These sutures are particularly useful<br>where the combination of an<br>absorbable suture and extended<br>wound support (up to six months) is<br>desirable. | Perma-Hand sutures are indicated<br>for use in general soft tissue<br>approximation and/or ligation,<br>including use in cardiovascular,<br>ophthalmic and neurological<br>procedures. | Braided Dexon S sutures are<br>indicated for use as absorbable<br>sutures in general soft tissue<br>approximation and/or ligation,<br>including use in ophthalmic<br>procedures, but not for use in<br>cardiovascular or neural tissue. |
| Material | Poly-4-hydroxybutyrate<br>(P4HB) | Poly-4-hydroxybutyrate (P4HB) | Organic fibroin protein derived<br>from domesticated species<br>Bombyx mor (B. mori). Dyed<br>black and coated with a special<br>wax mixture | Composed of the homopolymer<br>of glycolic acid. Coated with<br>Polycaprolate, a co-polymer of<br>glycoloide and epsilon-<br>caprolactone. |
| Dyed, Undyed<br>Filament Type | Dyed and Un-dyed<br>Multifilament | Dyed and Un-dyed<br>Monofilament | Dyed<br>Multifilament | Dyed & Un-dyed<br>Multifilament |
| Size | 2 (various lengths)<br>with or w/out needles<br>attached | 5-0 through 2 (various lengths)<br>with or w/out needies attached | 9-0 through 5 (various lengths)<br>with or w/out needles attached | 8-0 through 2 (various lengths) |
| Suture Diameter<br>Knot Pull Strength<br>Needle Attachment<br>Strength<br>Absorption Profile | All characteristics meet<br>USP Requirements,<br>except for diameter.<br>Approximate % Strength<br>remaining (BSR) | All characteristics meet USP<br>Requirements, except for diameter.<br>Approximate % Strength remaining<br>(BSR): | All characteristics meet USP<br>Requirements, except for<br>diameter.<br>From IFU | All characteristics meet USP<br>requirements.<br>Approximate % strength<br>remaining (BSR): |
| | Size 2<br>2 - 4 weeks:<br>8 - 12 weeks:<br>26 weeks: | Sizes: 5-0, 4-0 3-0, 2-0 0, 1, 2<br>4 weeks: ~60% ~60% ~90%<br>8 weeks: ~30% ~50% ~80%<br>12 weeks: ~20% ~40% ~60%<br>26 weeks: ~0% ~30% ~30% | While silk sutures are not<br>absorbed, per se, progressive<br>degradation of the proteinaceous<br>silk fiber in vivo may result in<br>gradual loss of all of the suture's<br>strength over time. | 2 weeks (6-0 and larger): ~65%<br>3 weeks (6-0 and larger): ~35%<br>2 weeks (7-0 and smaller): ~55%<br>3 weeks (7-0 and smaller): ~20% |
| | Absorption essentially<br>complete within 12-18<br>months | Absorption essentially complete<br>within 12 - 24 months | From Ethicon Wound Closure<br>Manual: | Absorption is essentially<br>complete between 60 and 90<br>days |
| Packaging | Foil packaging with<br>removable Tyvek header | Foil packaging with removable<br>Tyvek header | Tyvek with foil inner pack | Foil package / Tyvek |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
| Characteristic<br>Indications for Use | Table 10 (continued)<br>Ethicon - Vicryl Braided Suture - K022269<br>Coated Vicryl suture is indicated for use in general soft tissue<br>approximation and/or ligation, including use in ophthalmic<br>procedures, but not for usue in cardiovascular and neurological<br>tissues | | Tornier - BioFiber Suture -K122487, K130422<br>Indicated for use in soft tissue approximation and/or ligation but not for<br>use in cardiovascular or neurological surgery, microsurgery or<br>ophthalmic surgery. | |
| Material | Co-polymer of 90% glycolide and 10% L-lactide. Coating of<br>equal parts of co-polymer of glycolide and lactide with calcium<br>stearate | | Poly-4-hydroxybutyrate (P4HB) | |
| Dyed, Undyed | Dyed & Un-dyed | | Dyed & Un-dyed | |
| Filament Typo | Multifilament | | Multifilament | |
| Size | 6-0 though 3 (various lengths) with or without needles attached | | 2 (various lengths), with or w/out needles attached | |
| Suture Diametor<br>Knot Pull Strength<br>Needle Attachment<br>Strength<br>Absorption Profile | All characteristics meet USP requirements, except for diameter | | All characteristics meet USP Requirements, except for diameter | |
| | Approximate % Strength remaining (BSR):<br>2 weeks:<br>~75%<br>3 weeks (6-0 and larger):<br>~50%<br>3 weeks (7-0 and larger):<br>~40%<br>4 weeks (6-0 and larger):<br>~25%<br>Absorption is essentially complete between 56 and 70 days | | Approximate % Strength remaining (BSR)<br>2 - 4 weeks: ~60%<br>8-12 weeks: ~40%<br>26 weeks: ~20%<br>Absorption essentially complete within 18-24 months | |
| Packaging<br>Sterilization | Tyvek with foil inner pack<br>Ethylene Oxide (EO) | | Tyvek with foil inner pack<br>Ethylene Oxide (EQ) | |
Table 10 Table 10 Table 10
到
新闻
- ---
1 .
K132348 page 3/4
{4}------------------------------------------------
.
.
:
K132348 page 4/4
.
312
名田
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 5, 2013
Tepha, Incorporated Ms. Marv P. LeGraw Vice President, Regulatory Affairs 99 Hayden Avenue. Suite 360 Lexington. Massachusetts 02421
Re: K132348
Trade/Device Name: Tephal-LEX Braided Suture Regulation Number: 21 CFR 878.4494 Regulation Name: Absorbable poly(hydroxybutyrate) surgical suture produced by recombinant DNA technology Regulatory Class: Class II Product Code: NWJ Dated: July 26, 2013 Received: July 29, 2013
Dear Ms. LeGraw:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean. that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Mary P. LeGraw
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda,gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known): Not Assigned
Device Name: TephaFLEX Braided Suture
## Indications for Use:
TephaFLEX braided sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
Prescription Use: X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
/C
## David Krause -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132348
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.