BONDILOXS TOPICAL HEMOSTATIC DRESSING

K132333 · Medtrade Products , Ltd. · QSY · Jun 25, 2014 · SU

Device Facts

Record IDK132333
Device NameBONDILOXS TOPICAL HEMOSTATIC DRESSING
ApplicantMedtrade Products , Ltd.
Product CodeQSY · SU
Decision DateJun 25, 2014
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

Under the supervision of a healthcare professional Bondiloxs Topical Hemostatic Dressing is indicated for use as a temporary topical dressing for bleeding control associated with minor wounds, including control of minor external bleeding and exudate from sutures and/or surgical procedures and for temporary external treatment for controlling moderate to severe bleeding.

Device Story

Bondiloxs Topical Hemostatic Dressing is a sterile, non-woven dressing composed of chitosan fibers and lactic acid; designed for temporary topical bleeding control. Applied directly to wounds with firm pressure; the dressing absorbs exudate and forms a gel-like plug, creating a physical barrier to stop blood flow and promote localized clotting. Used in clinical settings under healthcare professional supervision for minor to severe external bleeding. Available in various sizes up to 15cm x 15cm, including flat or z-folded configurations for application. Benefits include rapid hemostasis and effective management of surgical or traumatic bleeding. Packaged in foil pouches to maintain sterility.

Clinical Evidence

No clinical data required. Evidence consists of bench testing and in-vivo animal studies. Bench testing included absorbency, tensile strength, gelling properties, pack integrity, and sterility. In-vivo studies utilized epigastric artery, liver cruciate, liver dissection, and saphenous femoral punch models to assess hemostatic efficacy across mild, moderate, and major bleeding severities. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Sterilization validated per ISO 11137-1.

Technological Characteristics

Non-woven dressing composed of chitosan fibers and lactic acid. Sterile, single-use. Sterilized via gamma irradiation (ISO 11137) to SAL 10^-6. Biocompatibility per ISO 10993. Dimensions up to 15cm x 15cm. Physical barrier mechanism of action.

Indications for Use

Indicated for patients requiring temporary topical bleeding control for minor wounds, including minor external bleeding and exudate from sutures/surgical procedures, and for temporary external treatment of moderate to severe bleeding. Use under supervision of a healthcare professional.

Regulatory Classification

Identification

To temporarily control bleeding and cover external wounds.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 21, 2023 Medtrade Product Ltd. Claire Ryan Head of Regulatory and Quality Assurance Electra House, Crewe Business Park Crewe, Cheshire CW1 6GL United Kingdom Re: K132333 Trade/Device Name: Bondiloxs Topical Hemostatic Dressing Regulatory Class: Unclassified Product Code: QSY Dear Claire Ryan: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 25, 2014. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov. Sincerely, # Julie A. Morabito -S Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 25. 2014 Medtrade Product Ltd Mr. Jonathan Ranfield Quality & Regulatory Director Electra House, Crewe Business Park Crewc, Cheshire CWI 6GL United Kingdom Re: K132333 Trade/Device Name: Bondiloxs Topical Hemostatic Dressing Regulatory Class: Unclassified Product Code: FRO Dated: March 19, 2014 Received: March 24, 2014 Dear Mr. Ranfield: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications' for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Mr. Jonathan Ranfield forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE K132333 510(k) Number (if known): Device Name: Bondiloxs Topical Hemostatic Dressing Indications for Use: Under the supervision of a healthcare professional Bondiloxs Topical Hemostatic Dressing is indicated for use as a temporary topical dressing for bleeding control associated with minor wounds, including control of minor external bleeding and exudate from sutures and/or surgical procedures and for temporary external treatment for controlling moderate to severe bleeding. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ______________________________________________________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) ## Jiyoung Dang -S ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Medtrade, a company that produces innovative medical products. The logo features the word "Medtrade" in a simple, sans-serif font, with the words "Innovative Medical Products" written in a smaller font below. A curved line wraps around the word "Medtrade", adding a touch of visual interest to the design. ## Section 5 - Traditional 510(k) Notification :- 510(k) Summary This Traditional 510(k) notification is to provide substantial equivalence for Medtrade Products Bondiloxs Topical Hemostatic Dressing, which is substantially equivalent to currently marketed hemostatic dressings intended for temporary external treatment for controlling minor and moderate to severe bleeding. #### 1. Submitted by: Medtrade Product Ltd Electra House, Crewe Business Park Crewe, Cheshire CW1 6GL United Kingdom Establishment Registration Number: 9614493 ## Contact Person:- Mr Jonathan D Ranfield Quality & Regulatory Director Telephone: + 44(0)1270 500019 Fax:- + 44(0)1270 500045 Email: jonathan.ranfield@medtrade.co.uk #### 2. · · Date Summary prepared: 24th" June 2014 #### 3. Device Information :- | Proprietary Name: | Bondiloxs Topical Hemostatic Dressing | |----------------------|---------------------------------------| | Common Name: | Bondiloxs Topical Hemostatic Dressing | | Trade Names: | Not yet defined | | Product Code: | FRO | | Classification Name: | Dressing, Wound, Drug | | Classification: | Unclassified | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Medtrade, a company that specializes in innovative medical products. The logo features the word "Medtrade" in a simple, sans-serif font, with the words "Innovative Medical Products" underneath in a smaller font. A curved line arches over the word "Medtrade", adding a dynamic element to the design. #### 4. Predicate Devices: CELOX Gauze PRO 510(k) # K113560, manufactured by Medtrade Products Ltd. #### 5. Device Description :- Bondiloxs Topical Hemostatic Dressing is a sterile non-woven dressing comprising of chitosan fibres which aids the gelling and absorbency potential of the dressing. Bondiloxs Topical Hemostatic Dressing achieves the principle intended action of hemostasis by the providing a physical barrier to stop bleeding. By applying the Bondiloxs Topical Hemostatic Dressing directly onto a wound and together with firm pressure the gel-like plug on dressing's surface creates a physical barrier which controls blood flow through the dressing to stop bleeding. The dressing promotes localized clotting formation to help stop bleeding. The Bondiloxs Topical Hemostatic Dressing is packed in a foil/foil pouch. The pouch provide an integral barrier that maintains dressing sterility post irradiation vet allows easy opening and aseptic dressing removal by the end user. The Bondiloxs Topical Hemostatic Dressing is available in various sizes ranging up to a maximum of 15cm x 15cm and is available in a flat dressing or z-folded for fast and intuitive application. #### 6. Indication for Use :- Under the supervision of a healthcare professional Bondiloxs Topical Hemostatic Dressing is indicated for use as a temporary topical dressing for bleeding control associated with minor wounds, including control of minor external bleeding and exudate from sutures and/or surqical procedures and for temporary external treatment for controlling moderate to severe bleeding. #### 7. Substantial Equivalence: - The Bondiloxs Topical Hemostatic Dressings has identical indications for use to the legally marketed Medtrade Products CELOX Gauze Pro (510(k) # K113560) Which is a chitosan based dressing. The following tabulations which summarises the description, Indications for use, and general information, and the substantial equivalence information provides a succinct and direct comparison of equivalence between Bondiloxs Topical Hemostatic Dressings and the approved predicate device listed above. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Medtrade, a company that produces innovative medical products. The logo features the word "Medtrade" in a simple, sans-serif font, with the words "Innovative Medical Products" written in a smaller font below it. A black swoosh design is above the company name. The logo is simple and modern, and it conveys a sense of innovation and professionalism. Summary of Comparison of Bondiloxs Topical Hemostatic Dressing and the Predicate Device . | | Bondiloxs Topical Hemostatic<br>Dressings | CELOX Gauze PRO (510(k) #<br>K113560) | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Description | Bondiloxs Topical Hemostatic<br>Dressing is sterile non-woven<br>dressing comprising of chitosan<br>fibres and lactic acid which aids the<br>gelling and absorbency potential of<br>the dressing. | The device consists of a chitosan<br>Haemostatic granules adhered onto<br>a base fabric (non woven gauze)<br>using a hot melt adhesive. | | Physical<br>Composition | Chitosan fibres and lactic acid non-<br>woven, soft, non fibrous absorbent<br>dressing | Chitosan Granules on Sheet | | Indications for<br>Use /RX only) | Under the supervision of a<br>healthcare professional Bondiloxs<br>Topical Hemostatic Dressing is<br>indicated for use as a temporary<br>topical dressing for bleeding control<br>associated with minor wounds,<br>including control of minor external<br>bleeding and exudate from sutures<br>and/or surgical procedures and for<br>temporary external treatment for<br>controlling moderate to severe<br>bleeding. | Under the supervision of a<br>healthcare professional CELOX<br>Gauze PRO for moderate to severe<br>external bleeding wounds (Rx) is<br>intended for temporary external<br>treatment for controlling moderate to<br>severe bleeding. | · : {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Medtrade, a company that specializes in innovative medical products. The logo features the word "Medtrade" in a simple, sans-serif font, with the tagline "Innovative Medical Products" written in a smaller font below it. A curved line extends over the word "Medtrade", adding a modern and dynamic touch to the logo. . . , | | Bondiloxs Topical Hemostatic<br>Dressings | CELOX Gauze PRO (510(k) #<br>K113560) | |------------------------------------------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use /RX only)<br>(Continue) | | Under the supervision of a<br>healthcare professional CELOX<br>Gauze PRO for minor external<br>bleeding from wounds and<br>procedures (Rx) is intended for use<br>as a temporary topical dressing for<br>bleeding control associated with<br>minor wounds, including control of<br>minor external bleeding and exudate<br>from sutures and/or surgical<br>procedures. | 1. Sales of the same . . . . . . . . . . . . . . {8}------------------------------------------------ : : : ! | | Bondiloxs Topical Hemostatic<br>Dressings | CELOX Gauze PRO (510(k) #<br>K113560) | |--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | Product Code | FRO | FRO | | For single use<br>only | Yes | Yes | | Method of<br>sterilisation | Gamma Irradiation in accordance<br>with ISO 11137 | Gamma Irradiation in accordance<br>with ISO 11137 | | Sterility<br>assurance level | 1x10-6 | 1x10-6 | | Biocompatibility<br>testing completed<br>in accordance<br>with ISO 10993 | Yes | Yes | | Mode of Action | Bondiloxs Topical Hemostatic<br>Dressing achieves the principle<br>intended action of hemostasis by<br>the providing a physical barrier to<br>stop bleeding | CELOX Gauze PRO achieves the<br>principle intended action of<br>hemostasis by the providing a<br>physical barrier to stop bleeding. | . . . {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Medtrade, a company that produces innovative medical products. The logo consists of the word "Medtrade" in a simple, sans-serif font, with the tagline "Innovative Medical Products" written in a smaller font below. The word "Medtrade" is enclosed by a curved line above and below the word. #### 8. Manufacturing :- Bondiloxs Topical Hemostatic Dressing is manufactured according to the product specifications and under good manufacturing practices (GMP). A risk analysis has been performed in accordance with BS EN ISO 14971 to identify possible failure mode during manufacturing and design. Manufacturing controls have been developed and implemented to address the identified risk factors based on the criticality of the failure mode. Bondiloxs Topical Hemostatic Dressings meets all the established specifications prior to release to ensure the device is safe, effective and correctly labelled for its intended use. Bondiloxs Topical Hemostatic Dressings are terminally sterilised by gamma irradiation to a sterility assurance level (SAL) of 10°. #### 9. Non-Clinical Performance Data:- Performance data for the Bondiloxs Topical Hemostatic Dressing has been established using in-vivo testing (Hemostatic properties) and bench testing in accordance to relevant standards where applicable (including absorbency, tensile strength, gelling properties, pack integrity (dye penetration and burst test) and sterility testing. The in-vivo studies were designed and conducted to establish the hemostatic efficacy of the Bondiloxs Topical Hemostatic Dressing in different injury types, including: Epigastric artery wound model with both heparinised and nonheparinised conditions. Liver cruciate model (represents mild to moderate bleeding), liver dissection model (to assess hemostatic efficacy for moderate bleeding), Epigastric sever model (to assess hemostatic efficacy for mild topical bleeding) and Saphenous femoral 2.7mm punch (to assess hemostatic efficacy for moderate to major bleeding). The Bondiloxs Topical Hemostatic Dressing is restricted to external, topical (dermal) use and the various animal models were used to assess the ability of the device to control different severities of bleeding. Shelf Life has been determined using stability studies at ambient conditions (real time aging 25°C/60% RH) and in an environment in which the temperature and relative humidity have been controlled at 40°C/75% RH. The biocompatibility of Bondiloxs Topical Hemostatic Dressing has been demonstrated through assessment according to BS EN ISO 10993-1 (Biological Evaluation of Medical Devices) and appropriate testing these include cytotoxicity, sensitization and irritation. The biocompatibility testing for the Bondiloxs Topical Hemostatic Dressing has demonstrated that the device is safe for the indications of use. {10}------------------------------------------------ Sterilisation validation has been performed in compliance with harmonised standards (ISO 11137-1: 2006 - Sterilization of healthcare products - Radiation -Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices). The biocompatibility and performance testing including the in vivo testing for the Bondiloxs Topical Hemostatic Dressing has demonstrated that the device is safe and effective for the indications of use. #### 10. Clinical Performance Data: Based upon the substantial equivalence determination for the predicate device (CELOX Gauze Pro (510(k) # K113560 no clinical data is required for evaluation of the Bondiloxs Topical Hemostatic Dressing. #### 11. Conclusion The indication for use and performance testing for the Bondiloxs Topical Hemostatic Dressing is substantially equivalent to the predicate device; CELOX Gauze Pro (510(k) # K113560) manufactured by Medtrade Products Ltd. The Bondiloxs Topical Hemostatic Dressing is indicated for topical application as for the control of temporary external bleeding associated with minor to severely bleeding wounds. The biocompatibility and performance testing for the Bondiloxs Topical Hemostatic Dressing has demonstrated that the device is as safe and effective as the predicate devices and raises no new issues of safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...