← Product Code [IPF](/productcode/IPF) · K132284

# VECTRA NEO CLINICAL THERAPY SYSTEM (K132284)

_Djo, LLC · IPF · Apr 10, 2014 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132284

## Device Facts

- **Applicant:** Djo, LLC
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** Apr 10, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K132284 · Apr 10, 2014 | VECTRA NEO CLINICAL THERAPY SYSTEM | Djo, LLC | ISO 10993-1:2009 Annex C literature review; Historical biocompatibility data from predicate devices (K062354) | A literature review was conducted to evaluate the biocompatibility of new patient-contacting materials (specifically for the patient remote) by comparing them to materials used in previously cleared predicate devices. | Biocompatibility; Literature Review; ISO 10993-1 |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Biocompatibility evaluation of patient remote; Literature review and comparative analysis; Follow-up/Duration: Not applicable; Study Period: Not applicable | Not applicable; Sample Size: Not applicable; Number of Sites: Not applicable | Predicate device (K062354) remote | Biocompatibility of patient-contacting materials |

## Indications for Use

Vectra Neo Clinical Therapy System is indicated for: For VMS-(Pulsed Mode, Burst Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS) - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy - Increasing local blood circulation - Muscle re-education - Maintaining or increasing range of motion - Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), Symmetrical Biphasic (TENS), and HAN - Symptomatic relief or management of chronic, intractable pain - Post-traumatic acute pain - Post-surgical acute pain For DC Continuous Mode - Relaxation of muscle spasm For FES - Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait For EMG triggered Stim - Stroke rehab by muscle re-education - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy - Increase local blood circulation - Muscle re-education - Maintaining or increasing range of motion For EMG - To determine the activation timing of muscles for: - Retraining of muscle activation - Coordinating of muscle activation - An indication of the force produced by muscle for control and maintenance of muscle contractions - Relaxation muscle training - Muscle re-education For Ultrasound - Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as: - Relief of pain, muscle spasms and joint contractures - Relief of pain, muscle spasms and joint contractures that may be associated with: - Adhesive capsulitis - Bursitis with slight calcification - Myositis - Soft tissue injuries - Shortened tendons due to past injuries and scar tissues - Relief of sub-chronic and chronic pain and joint contractures resulting from: - Capsular tightness - Capsular scarring For Infrared Lamp (laser) - To provide topical heating for the following: - Temporary increase in local blood circulation - Temporary relief of minor muscle and joint aches, pains and stiffness - Relaxation of muscles - Temporary relief of muscle spasms - Temporary relief of minor pain and stiffness associated with arthritis

## Device Story

Vectra Neo Clinical Therapy System is a modular, integrated electrotherapy device for clinical use by healthcare professionals. It combines muscle stimulation, ultrasound, laser, and sEMG biofeedback modalities. Inputs include user-selected therapy parameters via touchscreen and sEMG signals from patient electrodes. The device processes these inputs to deliver various waveforms (e.g., Interferential, TENS, Russian, VMS), ultrasound energy (1/3.3 MHz), and laser therapy. Outputs include therapeutic electrical stimulation, deep heat, and visual/stored biofeedback data. Clinicians use the system to manage pain, muscle spasms, and joint contractures, and to facilitate stroke rehabilitation and muscle re-education. The device features a cart with storage and leadwire management. Clinical decision-making is supported by on-screen protocol guidance and patient data management. Benefits include consolidated treatment options, improved workflow, and targeted therapeutic outcomes for musculoskeletal and neurological conditions.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench performance testing, including verification and validation of software, and a summative usability study. The device was tested for compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-5 (ultrasound), IEC 60601-2-10 (stimulators), IEC 60601-2-22 (laser), IEC 60601-2-57 (light source), and IEC 60825-1 (laser safety). Biocompatibility of patient-contacting materials was evaluated per ISO 10993-1:2009.

## Technological Characteristics

Modular electrotherapy system. Modalities: 12 waveforms (Interferential, TENS, Russian, etc.), dual-frequency ultrasound (1/3.3 MHz), laser therapy, and 2-channel sEMG. Materials: Hydrogel, cellulose, polyester, aluminum, chrome, copper-nickel, polycarbonate, nylon, ABS, silicone. Connectivity: USB for data storage. Interface: Touchscreen. Standards: IEC 60601-1, 60601-1-2, 60601-2-5, 60601-2-10, 60601-2-22, 60601-2-57, IEC 60825-1. Sterilization: N/A (non-sterile).

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- Vectra Genisys ([K062354](/device/K062354.md))
- Vectra Genisys Laser System (Intelect XT Laser System) ([K040662](/device/K040662.md))

## Reference Devices

- Kaoosta ([K874469](/device/K874469.md))
- Kaoosta ([K852267](/device/K852267.md))
- Kaoosta ([K970426](/device/K970426.md))
- [K955246](/device/K955246.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "DJO" is in a larger font than the word "GLOBAL", and there is a trademark symbol next to the "O" in "DJO". The logo is black and white.

#### 510(k) Summary

### APR 1 0 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is: K132284

Submitted by:

DIO, LLC 1430 Decision Street Vista, CA 92081

Contact Person:

Lorri Trotter Regulatory Affairs Manager Phone: 760-734-3049 Fax: 760-734-5544 Email: lorri.trotter@djoglobal.com

Vectra Neo Clinical Therapy System

Vectra Neo Clinical Therapy System

Date Summary Prepared:

July 19, 2013

Trade Name:

Common/Usual Name:

Classification Name:

21 CFR 890.5500 Infrared lamp 21 CFR 890.5850 Powered muscle stimulator 21 CFR 890.5860 Ultrasound and Muscle stimulator 21 CFR 882.5890 TENS for pain relief 21 CFR 882.5050 Biofeedback device 21 CFR 882.5810 External functional neuromuscular stimulator Interferential current stimulator Unclassified

#### Product Code:

- ILY Infrared lamp
- IPF Powered muscle stimulator
- IMG Ultrasound and muscle stimulator
- Transcutaneous electrical nerve stimulator for pain relief GZJ
- HCC Biofeedback device
- GZI External functional neuromuscular stimulator
- LIH Interferential current therapy

{1}------------------------------------------------

Traditional 510(k): New Device Vectra Neo Clinical Therapy System

Image /page/1/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The globe is black and white and has a curved line running through it. The text "DJO" is in a bold, sans-serif font, and the text "GLOBAL" is in a smaller, sans-serif font below it.

| Regulatory Class:    | Class II *                                                          |
|----------------------|---------------------------------------------------------------------|
| Panels:              | 89 – Physical Medicine<br>84 – Neurology                            |
| Predicate Device(s): | Vectra Genisys Laser System (Intelect XT Laser System)<br>(K040662) |

#### Device Description:

The Vectra Neo Clinical Therapy System is an electrotherapy product offering clinicians a modular design of muscle stimulation, ultrasound, laser and biofeedback modalities in one combination device. The Vectra Neo Clinical Therapy System is designed to give the most treatment options in one compact and integrated package.

Vectra Genisys (K062354)

Clinicians have a variety of choices to best suit the needs of their individual practice. Below is an overview of the system choices/features.

The electrotherapy mode offers one of the largest selections of multiple waveforms cleared to market by FDA. The numeric pain scales can be recorded with the patient data management system. The therapy system cart provides three storage drawers to conveniently house clinical essentials. The leadwire management system allows for easy, quick access in an uncluttered arrangement.

The electrotherapy module offers 12 waveforms: Interferential, Premodulated (IFS), Asymmetrical Biphasic (TENS), Microcurrent, VMS, VMS Burst, VMS FR, Russian, High Voltage Pulsed Current, Symmetrical Biphasic (TENS), Direct Current, and HAN (TENS). The Vectra Neo Clinical Therapy System allows to sequence these waveforms for ease of use.

The Neo ultrasound module is dual frequency (1 MHz and 3.3 MHz) with selectable duty cycles of 10%, 20%, 50% and 100%, low BNR (5.0:1) {FDA method of measurement), pulse frequency 100Hz and selectable pulse durations of 1 mSec, ±20%; 2 mSec, ±20%; 5 mSec, ±20%.

| Model<br>Number | Description                       | BNR<br>(FDA<br>method) | Standard<br>Deviation<br>(BNR) | ERA<br>(cm²) | ERA Limits<br>(cm²) |     | Beam<br>Type<br>(Pattern) | Frequency<br>(MHz) | Number<br>of<br>Crystals | Max.<br>Output<br>Power<br>(W) |
|-----------------|-----------------------------------|------------------------|--------------------------------|--------------|---------------------|-----|---------------------------|--------------------|--------------------------|--------------------------------|
| 27333           | 1 cm²<br>ultrasound<br>applicator | 5:1                    | 1                              | 0.9          | 1                   | 0.4 | Collimating               | 3.3                | 1                        | 1.5                            |
| 27334           | 2 cm²<br>ultrasound<br>applicator | 5:1                    | 1                              | 1.8          | 2                   | 1.4 | Collimating               | 1.0 and<br>3.3     | 1                        | 3.0                            |

The following ultrasound applicators are provided with the Vectra Neo Clinical Therapy System:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The globe is black and white and appears to be divided into sections. The text "DJO" is in a bold, sans-serif font, and the text "GLOBAL" is in a smaller, sans-serif font below it.

| Model<br>Number | Description                        | BNR<br>(FDA<br>method) | Standard<br>Deviation<br>(BNR) | ERA<br>(cm²) | ERA Limits<br>(cm²) |     | Beam<br>Type<br>(Pattern) | Frequency<br>(MHz) | Number<br>of<br>Crystals | Max.<br>Output<br>Power<br>(W) |
|-----------------|------------------------------------|------------------------|--------------------------------|--------------|---------------------|-----|---------------------------|--------------------|--------------------------|--------------------------------|
|                 |                                    |                        |                                |              | High                | Low |                           |                    |                          |                                |
| 27335           | 5 cm²<br>ultrasound<br>applicator  | 5:1                    | 1                              | 4.0          | 5                   | 3   | Collimating               | 1.0 and<br>3.3     | 1                        | 6.0                            |
| 27336           | 10 cm²<br>ultrasound<br>applicator | 5:1                    | 1                              | 8.5          | 10                  | 7   | Collimating               | 1.0 and<br>3.3     | 1                        | 15.0                           |

The laser therapy module is indicated for the temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness, temporary relaxation of muscles, temporary relief of muscle spasms, temporary relief of minor pain and stiffness associated with arthritis.

The sEMG biofeedback module provides two channels of surface EMG. Feedback can be stored onto a USB Thumb Drive. The sEMG features a clinician chosen trigger point that activates therapeutic stimulation. The sEMG feature is often used to treat stroke patients and for muscle re-education.

Online guided assistance is provided through Clinical Protocol Setup and On-Screen Indications to help guide therapy selections for electrotherapy waveform rationale, parameter selections, electrode placement images, laser and ultrasound applicator recommendations.

Combination electrotherapy is used for the management of pain and muscle spasm. All functions of 1 or 3.3 MHz Ultrasound can be combined with Interferential, Premodulated (IFS), Asymmetrical Biphasic (TENS), VMS, VMS Burst, VMS FR and High Voltage Pulsed Current.

#### Intended Use:

Vectra Neo Clinical Therapy System is indicated for:

For VMS-(Pulsed Mode, Burst Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS)

- . Relaxation of muscle spasms
- . Prevention or retardation of disuse atrophy
- Increasing local blood circulation .
- . Muscle re-education
- . Maintaining or increasing range of motion
- . lmmediate postsurgical stimulation of calf muscles to prevent venous thrombosis

Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), Symmetrical Biphasic (TENS), and HAN

- Symptomatic relief or management of chronic, intractable pain .

{3}------------------------------------------------

- Post-traumatic acute pain .
- . Post-surgical acute pain

#### For DC Continuous Mode

- Relaxation of muscle spasm �
#### For FES

- . Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait
#### For EMG triggered Stim

- . Stroke rehab by muscle re-education
- . Relaxation of muscle spasms
- . Prevention or retardation of disuse atrophy
- . Increase local blood circulation
- . Muscle re-education
- . Maintaining or increasing range of motion

#### For EMG

- To determine the activation timing of muscles for: .
	- Retraining of muscle activation -
	- Coordinating of muscle activation -
- . An indication of the force produced by muscle for control and maintenance of muscle contractions ·
	- Relaxation muscle training ।
	- Muscle re-education -

#### For Ultrasound

.

Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as:

- . Relief of pain, muscle spasms and joint contractures
	- Relief of pain, muscle spasms and joint contractures that may be associated with :
		- -Adhesive capsulitis
		- Bursitis with slight calcification l
		- -Myositis
		- Soft tissue injuries ।
		- Shortened tendons due to past injuries and scar tissues -
- Relief of sub-chronic and chronic pain and joint contractures resulting from: .
	- Capsular tightness -
	- Capsular scarring -

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "DJO" is in a larger font size than the word "GLOBAL". The logo is black and white.

#### For Infrared Lamp (laser)

To provide topical heating for the following:

- Temporary increase in local blood circulation .
- Temporary relief of minor muscle and joint aches, pains and stiffness
- . Relaxation of muscles
- . Temporary relief of muscle spasms
- . Temporary relief of minor pain and stiffness associated with arthritis

#### Indication for Use Comparison to Predicate Devices:

- Vectra Genisys (K062354) primary predicate .
The Vectra Neo Clinical Therapy System has the same Indications for Use as the Vectra Genisys for the following types of treatment:

- o VMS-(Pulsed Mode, Burst Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS)
- Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, o Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), and Symmetrical Biphasic (TENS) and HAN
- For FES 0
- For DC Continuous Mode O
- For EMG ಂ
- For EMG triggered Stim ಂ
- For Ultrasound ਼
- Vectra Genisys Laser System (Intelect XT Laser System) (K040662) secondary predicate .

The Vectra Neo Clinical Therapy System has the same Indications for Use as the Vectra Genisys Laser System (Intelect XT Laser System)

- For Infrared Lamp (laser) o
#### Technological Comparison to Predicate Devices:

The Vectra Neo Clinical Therapy System is technologically equivalent to Vectra Genisys (K062354) and Vectra Genisys Laser System (Intelect XT Laser System) (K040662) with respect to technical and performance characteristics.

The Vectra Neo Clinical Therapy System has the same indications for use and the same fundamental technology as the two predicate devices. It is intended to be used by the same target population and in the same clinical environments. The Vectra Neo Clinical Therapy System simply combines the functionality of the two predicates into one modular, adaptable device and has been upgraded with more modern components and has been engineered to meet FDA-recognized international consensus standards.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "DJO" is in a larger font size than the word "GLOBAL", and there is a trademark symbol after the "O" in "DJO". The logo is black and white.

Some of the minor differences include: the user interface has been simplified from having 16 buttons to 2 buttons and a touchscreen, additional FDA recognized consensus standards have been conformed to, and some outputs have been modified slightly. These differences are minimal and do not affect the device's performance or efficacy. With respect to safety, these minor differences could only improve device safety. Additionally, bench performance testing has demonstrated that the Vectra Neo Clinical Therapy System has the same waveforms and other outputs as the predicate devices and has passed all verification and validation activities.

All differences noted above do not raise new issues of safety and effectiveness.

#### Performance Testing:

- The device's software has been validated in accordance with the requirements set forth in . the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). The software validation tests demonstrated that the software version meets its design requirements.
- . A Usability Study was conducted to validate the usability of the Vectra Neo Clinical Therapy System. The results of the Summative Validation support the instructions for successfully using the device as intended. The result of the Human Factors and Usability Study substantiates the acceptability of the risks identified during the risk assessment activities.
- . The Vectra Neo Clinical Therapy System was tested and found to comply with the following standards:
	- . IEC 60601-1 for basic safety and essential performance
	- . IEC 60601-1-2 for electromagnetic compatibility
	- . IEC 60601-2-5 for safety of ultrasonic physiotherapy equipment
	- . IEC 60601-2-10 for performance of nerve and muscle stimulators
	- 트 IEC 60601-2-22 for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
	- . IEC 60601-2-57 for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use (Radiology)
	- . IEC 60825-1 for safety of laser products

#### Biocompatibility:

The Vectra Neo Clinical Therapy System has no direct contact with the patient. The table below identifies accessories that have direct contact with the patient at the treatment site, the material in contact with the patient, reference to the predicateand if applicable, reference devices. For new accessories, evaluation for biocompatibility is in accordance ISO 10993-1:2009 – Annex C, Suggested Procedure for Literature Review. The conclusion of these evaluations , as well as reference to the predicate and any associated reference devices, indicate that all materials selected are considered safe and effective for the intended use.

{6}------------------------------------------------

.

Image /page/6/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "DJO" is in a larger, bolder font than the word "GLOBAL". The globe is made up of several curved lines and shapes.

| Accessory              | Patient Contacting<br>Material                       | Predicate                           | Reference<br>Device                      | Biocompatibility Route                                                                                                                                                                                       |
|------------------------|------------------------------------------------------|-------------------------------------|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrodes             | Hydrogel                                             | Identical to predicate<br>(K062354) | Kaoosta<br>K874469<br>K852267<br>K970426 | n/a                                                                                                                                                                                                          |
| Ultrasound Gel         | Conductive gel                                       | Identical to predicate<br>(K062354) | K955246                                  | n/a                                                                                                                                                                                                          |
| Hivolt Probe           | Cellulose Pad<br>Polyester                           | ldentical to predicate<br>(K062354) | n/a                                      | n/a                                                                                                                                                                                                          |
| Ultrasound Applicators | No direct patient<br>contact<br>(see Ultrasound Gel) | Identical to predicate<br>(K062354) | n/a                                      | n/a                                                                                                                                                                                                          |
| Laser Applicators      | Aluminum<br>Chrome<br>Copper-Nickel<br>Polycarbonate | Identical to predicate<br>(K040662) | n/a                                      | n/a                                                                                                                                                                                                          |
| Laser Safety Glasses   | Nylon                                                | Identical to predicate<br>(K040662) | n/a                                      | n/a                                                                                                                                                                                                          |
| Patient Remote         | Polycarbonate/ABS<br>Silicone                        | NEW                                 | n/a                                      | ISO 10993-1:2009 - Annex C,<br>Suggested Procedure for<br>Literature Review, which included<br>a comparison to the same patient<br>contacting materials used for the<br>predicate device (K062354)<br>remote |

#### Conclusion:

:

:

Based on the performance testing and the supporting documentation, it can be concluded that the Vectra Neo Clinical Therapy System is safe, effective and substantially equivalent to the predicate devices.

:

.

.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic. The text is in all capital letters and is smaller than the central graphic.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 10, 2014

DIO, LLC Lorri Trotter Regulatory Affairs Manager 1430 Decision Street Vista, CA 92081

Re: K132284

Trade/Device Name: Vectra Neo Clinical Therapy System Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF; IMG; GZJ; HCC; GZJ; LIH; ILY Dated: March 6, 2014 Received: March 7, 2014

Dear Ms. Trotter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{8}------------------------------------------------

Page 2 - Ms. Lorri Trotter

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Felipe Aquel -S

Carlos Peña, PhD, MS for

Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

#### Indications for Use

#### 510(k) Number (if known) K132284

Device Name Vectra Neo Clinical Therapy System

Indications for Use (Describe) Vectra Neo Clinical Therapy System is indicated for:

For VMS-(Pulsed Mode, Burst Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS)

- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion
- · Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis

Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), Symmetrical Biphasic (TENS), and HANS

· Symptomatic relief or management of chronic, intractable pain

- · Post-traumatic acute pain
- · Post-surgical acute pain

For DC Continuous Mode

- · Relaxation of muscle spasm
For FES

• Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of thus improve the patient's gait

For EMG triggered Stim

- · Stroke rehab by muscle re-education
- · Relaxation of muscle spasms
- · Prevention or retardation of disuse atrophy
- · Increase local blood circulation
- · Muscle re-education
- · Maintaining or increasing range of motion

#### For EMG

· To determine the activation timing of muscles for:

- Retraining of muscle activation -
- Coordinating of muscle activation ﮯ
- An indication of the force produced by muscle for control and maintenance of muscle contractions
	- Relaxation muscle training •
	- Muscle re-education -

#### For Ultrasound

1 of Unlabours
Application of therapeutic deep heat for the treatment of selected sub-chronic medical conditions such as: · Relief of pain, muscle spasms and joint contractures

See PRA Statement below.

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

{10}------------------------------------------------

| • Relief of pain, muscle spasms and joint contractures that may be associated with : |  |
|--------------------------------------------------------------------------------------|--|
| - Adhesive capsulitis                                                                |  |
| - Bursitis with slight calcification                                                 |  |
| - Myositis                                                                           |  |
| - Soft tissue injuries                                                               |  |
| - Shortened tendons due to past injuries and scar tissues                            |  |
| • Relief of sub-chronic and chronic pain and joint contractures resulting from:      |  |
| - Capsular tightness                                                                 |  |
| - Capsular scarring                                                                  |  |

For Infrared Lamp (laser)

To provide topical heating for the following:

| • Temporary increase in local blood circulation                          |  |
|--------------------------------------------------------------------------|--|
| • Temporary relief of minor muscle and joint aches, pains and stiffness  |  |
| • Relaxation of muscles                                                  |  |
| • Temporary relief of muscle spasms                                      |  |
| • Temporary relief of minor pain and stiffness associated with arthritis |  |

Type of Use (Select one or both, as applicable)

| <span style="text-align:left;"> <span style="float:left;">☑</span>                 Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span style="text-align:left;"> <span style="float:left;">☐</span>                 Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Felipe Aguel -S Date: 2014.04.10 15:34:06
Felipe Aguel -S Dato0'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K132284](https://fda.innolitics.com/device/K132284)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
