The MiNS grasping instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp, hold, and manipulate other soft internal tissues as well as items such as hernia mesh.
Device Story
MiNS Needlescopic Resposable Laparoscopic device system consists of a reusable autoclavable handle and a sterile, disposable, single-use needle/cannula shaft housing a retractable grasper (NLU). The device is used by surgeons during general laparoscopic procedures. The instrument shaft is introduced percutaneously to the surgical site without a traditional insertion conduit. Once at the site, the working portion is deployed to approximate, grasp, and manipulate soft tissue or hernia mesh. The system provides a minimally invasive means to access internal anatomy. The device benefits patients by allowing for needlescopic access, potentially reducing trauma compared to larger diameter instruments.
Clinical Evidence
Bench testing only. The device underwent mechanical and attribute evaluations. Biocompatibility was established based on identical materials to the predicate device (K070686), which previously passed ISO 10993 testing.
Technological Characteristics
Materials: Medical grade stainless steel (17-7 SS, 300 SS, 475 SS) and polyester. Dimensions: 2.4mm ± 0.1mm diameter, 300mm length. Principle: Manual mechanical grasping/manipulation. Sterilization: Handle is autoclavable (moist heat, ISO 17665); disposable NLU is gamma-irradiated (ISO 11137). Connectivity: None. Software: None.
Indications for Use
Indicated for use in minimally invasive laparoscopic procedures to penetrate soft tissue, grasp, hold, and manipulate internal tissues and hernia mesh.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Mini Lap Technologies Inc. The logo includes the words "MINI LAP" in a stylized font, along with a graphic of an atom. The text "Mini Lap Technologies Inc." is printed to the right of the logo. The text "K132232" is printed below the company name.
510(k) Summary (as specified by 21 CFR 807.92) Prepared July 11, 2013. Modified December 3, 2013
DEC 04 2013
Device Name: MiNS Needlescopic Resposable Laparoscopic device system
#### Intended Use
The MiNS grasping instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp, hold, and manipulate other soft internal tissues as well as items such as hernia mesh.
#### No Previous NSE Decision
This device has not been the subject of a previous NSE decision nor has there been other submissions or communication to the FDA regarding this device ..
#### Establishment Registration Number
Regulatory Contact Allan Alward 145 Palisade Street Dobbs Ferry, NY 10522
#### Sponsor/Manufacturer
Mini Lap Technologies Inc. 145 Palisade Street Dobbs Ferry, NY 10522 Contact: Dr. S. Ravikumar Tel: 914.591.8400
FDA Establishment Registration Number is 3007123990
#### Device Trade or Proprietary Names
The device trade names are: MiNS Needlescopic Resposable Laparoscopic device system
#### Device Common, Usual or Classification Names
Laparoscopic Instruments, Graspers, Cannula, Trocar, Manual Surgical Instruments.
#### Classification Panel
Classification of this device falls under the responsibility of the General & Plastic Surgery panel.
#### Class
Class 2 device under the following product codes/regulations:
OCW, 21 CFR 876.1500, Endoscope and accessories ■
Page 1 of 5
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Mini Lap Technologies Inc. The logo includes the text "MINI LAP" in a stylized font, with a graphic of an atom-like structure. To the right of the graphic is the text "Mini Lap Technologies Inc." in a bold, sans-serif font.
## Applicable Standards
## Compliance with Section 514 of the Food, Drug and Cosmetic Act
None. Section 514 has not established performance standards for this device.
#### Device Description
#### Summary of the function of the device and its major components:
The MiNS devices are a family of disposable and re-useable devices. Consisting of an integrated, insulated needle/cannula shaft that houses a retractable grasper (NLU) which are inserted into a reusable Handle. The disposable NLU's when inserted into a reusable Handles are used for mobilization and manipulation of soft tissue during general laparoscopic procedures. The system has the ability to directly penetrate soft tissue to access certain areas of the human anatomy without the need for a traditional insertion conduit. The shaft of the instrument can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate, grasp and manipulate soft tissue.
#### Patient Contact Materials
The device is composed of biocompatible materials that have been used in medical devices for many years. The patient contact materials are biocompatible and are identical to the predicate Minilap device (K070686) which passed testing to ISO 10993.
## Cleaning, Disinfection, Sterilization and Pyrogenicity
The MiNS Needlescopic Resposable Laparoscopic device NLU's are a family of sterile disposable single patient packaged in a pet blister tray and sealed with a tyvek lid for the sterile barrier use ends that attaches to an autoclavable reusable handle.
The MiNS Needlescopic Resposable Laparoscopic device Handle is a reusable autoclavable portion of the MiNS system which has separate cleaning and sterilization techniques located in the information booklet and is packed as a non sterile device in a separate package..
The Handle portion is designed for moist heat sterilization, validated in production to an SAL of 10th per ISO 17665:2006 utilizing a half cvcle overkill method to meet the requirements of ISO17665-1:2006, Steam sterilization of medical devices.
The disposable NLU's are sterilized using gamma radiation, validated in production to a SAL of 10° per ISO 11137:2006. Method 1. in order to reduce gamma exposure/aging. When applicable, the VDmax provisions of 1137:2006 will be used to substantiate the standard 25-to-40 kGy dose.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Mini Lap Technologies Inc. The logo includes the text "Mini Lap Technologies Inc." in a bold, sans-serif font. To the left of the text is a stylized graphic that includes the words "MINI" and "LAP" with a circular design.
## Software
The subject device does not include software.
## 510(k) Summary of Safety and Effectiveness
We have compared various characteristics of the MiNS Needlescopic Resposable device system to existing technologies that would have an adverse affect upon the safety and efficacy of the system and have found that the Mini Lap MiNS Needlescopic Resposable device system does not present any new risks to the patient and therefore, we believe that it is substantially equivalent to other devices and technologies cleared by the FDA.
## Technological Characteristics
Mini Lap Technologies, Inc. believes that the subject device is substantially equivalent to other devices that have previously received FDA 510(k) clearance including the predicate devices.
## Predicate Device
The following devices have been identified as predicate devices:
- · Mini Lap Instruments K070686
- · Snowden-Pencer-Resposable Laparoscopic Scissors- K113407
- · Aesculap Needlescopic Instrument System- K982623
## Predicate Device Comparison
The Mini Lap MiNS Needlescopic Resposable device systems are, in principal and function, identical to existing technologies. A variety of reusable laparoscopic instruments are already widely manufactured and used in surgical procedures and have many similarities to the MiNS Needlescopic Resposable device svstem.
Discussion on Comparison:
## Mini Lap Instruments
The Mini Lap Instruments (K070686) were included as part of the comparison because the needle design and jaw design dimensions as well as the patient contact materials are identical to the subject device along with sterilization method.
- The main differences between the Mini Lap Instruments and the MiNS Needlescopic Resposable ● device system are the a) handle design which is reusable , b) the grasper shaft assembly attaches to a reusable handle .
Snowden-Pencer Resposable Laparoscopic Scissors (K113407) were included as predicates because they are a family of products with a reusable handle and sterile disposable tips indicated for mechanically cutting soft tissue in laparoscopic procedures.
- The two sets of devices share similar handle design, composed of Medical Grade stainless steel and . plastic components.
- The main difference between the MiNS Needlescopic Resposable device system and the Snowden-. Pencer Resposable Laparoscopic device is the outer diameter of the subject device is 2.4mm and the predicate is a 5mm.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Mini Lap Technologies Inc. The logo includes the text "MINI LAP" in a stylized font, accompanied by a geometric design resembling intersecting ellipses. To the right of the logo is the company name, "Mini Lap Technologies Inc.", written in a bold, sans-serif font.
A esculap Needlescopic Instrument System (K 982623) were included as predicates because they are a family of products with a reusable handles and attachable Shafts/tips indicated for use in endoscopic and laparoscopic procedures.
- The main difference between the MiNS Needlescopic Resposable device system and the Aesculap . Needlescopic Instrument System is the Aesculap system is reusable.
In the following section, we compare various characteristics of the MiNS Needlescopic Resposable device system to existing technologies. The Mini Lap MiNS Needlescopic Resposable device system does not present any new risks to the patient and therefore, we believe that it is substantially equivalent to other devices and technologies cleared by the FDA.
| Characteristic | Mini Lap<br>Technologies<br>MiniLap Resposable<br>Devices | Aesculap<br>Needlescopic<br>instruments | Mini Lap<br>Technologies<br>MINI LAP<br>Instruments | Snowden-Pencer<br>Resposable<br>Laparoscopic<br>Scissors |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K132232 | K982623 | K070686 | K113407 |
| Use | The MiNS grasping<br>instruments are a family<br>of minimally invasive<br>devices with the means<br>to penetrate soft tissue<br>to access certain areas<br>of the human anatomy.<br>The devices are used<br>to grasp, hold, and<br>manipulate other soft<br>internal tissues as well<br>as items such as hernia<br>mesh. | The Aesculap<br>Needlescopic<br>instrument set is<br>indicated for use in<br>adult and pediatric<br>diagnostics and<br>therapeutic general<br>endoscopy and<br>laparoscopy surgery. | The Minilap<br>instruments are a<br>family of<br>minimally invasive<br>devices with the<br>means to<br>penetrate soft<br>tissue to access<br>certain areas of the<br>human anatomy.<br>The devices are<br>used to grasp,<br>hold, and<br>manipulate other<br>soft internal tissues<br>as well as items<br>such as hernia<br>mesh. | The Snowden<br>Pencer<br>Laparoscopic<br>Ergonomic<br>Resposable Scissors<br>is a monopolar<br>electrosurgical<br>instrument<br>indicated to be used<br>in general<br>laparoscopic and<br>gynecologic<br>procedures to allow<br>high frequency<br>monopolar cutting<br>and coagulation.<br>The Resposable<br>scissors are<br>indicated to<br>mechanical cut<br>tissue and suture. |
| Needle diameter | 2.4mm ± .1mm | 2.5 mm | 2.4mm ± .1mm | 5mm |
| Device length | 300 mm | Various | 300 mm | 24CM,36CM,45CM |
| Material<br>Composition | Medical Grade Stainless<br>Steel<br>17-7 SS, 300 SS, | Medical Grade<br>Stainless Steel | Medical Grade<br>Stainless Steel<br>Polyester | Medical Grade<br>Stainless Steel<br>MT500 Polyolefin |
Predicate Comparison Chart
Page 4 of 5
20
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Mini Lap Technologies Inc. The logo includes the text "MINI" and "LAP" stacked vertically and connected by a stylized atom symbol. To the right of the logo is the text "Mini Lap Technologies Inc."
| | 475SS, Polyester | | | |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------|
| Sterilization | Autoclavable, non<br>sterile, reusable handles<br>with attachable<br>disposable sterile,<br>single use shaft tip<br>combinations. | Non sterile, reusable<br>detachable ends | Sterile, single use | Autoclavable, non<br>sterile reusable<br>handle Single use<br>sterile disposable<br>shaft and scissors<br>tip |
## Table of Contents
A table of content is included at the beginning of this submission. The pages referred to in the contents correspond to the sequentially numbered pages of this premarket notification.
## Truthful and Accurate Statement
A "truthful and accurate" statement regarding all information provided in this premarket notification is present on page .
#### Confidentiality
Mini Lap Technologies Inc. considers certain in this premarket notification to be confidential business information, and has taken measures to protect the release of this information. Mini Lap Technologies Inc. requests the FDA respect the confidentiality of this information to the extent possible under law. We expect the FDA will consult with Mini Lap Technologies Inc. prior to the release of any information in this premarket notification (outside the 510(k) summary) for any reason, including requests under the Freedom of Information Act.
## Performance Data |21 CFR 807.92(b)(1)|
The subject device has been subjected to and passed a variety of bench tests for mechanical and attribute evaluations. Additionally, the device is composed of biocompatible materials with a history of usage in the medical device industry and share common design components with K070686.
## Conclusion 121 CFR 807.92(b)(3)]
We believe the changes are minimal and conclude that the subject device is as safe and effective as the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Arenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 4, 2013
MiniLap Technologies, Inc. Mr. Allan Alward Vice President, Research and Development 145 Palisade Street Dobbs Ferry. New York 10522
Re: K132232
Trade/Device Name: MiNS Needlescopic Resposable device system Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCW Dated: October 29, 2013 Received: November 5, 2013
Dear Mr. Alward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Allan Alward
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Mini Lap Technologies Inc. The logo consists of the words "Mini Lap Technologies Inc." in a simple, sans-serif font. To the left of the text is a stylized graphic of what appears to be a molecule or atom, with the words "MINI LAP" incorporated into the design.
## Indications for Use
510(k) Number: K132232
Device Name: MiNS Needlescopic Resposable device system
Indications for Use:
ﻟﻤﺴﺎﺑﻘﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ
The MiNS grasping instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp, hold and manipulate other soft internal tissues as well as items such as hernia mesh.
X Prescription Use (Part 21 CFR 801 Subpart D)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHIER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Long H. Chen -A
DN CUS, O-U.S. Government ou-HHS.
OU=FDA, Our People, chotong H. Chen-A
0.9.2342.19200300 100 1,1-1300369056
Date, 2013 11 2207 36 160500
for BSA
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132232
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.