← Product Code [IYO](/productcode/IYO) · K132228

# UGEO PT60A DIAGNOSTIC ULTRASOUND SYSTEM (K132228)

_Samsung Medison Co., Ltd. · IYO · Aug 20, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132228

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYO](/productcode/IYO.md)
- **Decision Date:** Aug 20, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal, Abdominal, Small Organs, Adult Cephalic, Muscular-Skeletal (Conventional, Superficial), Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

## Device Story

UGEO PT60A is a general-purpose, hand-held, software-controlled diagnostic ultrasound system. It acquires ultrasound data via transducers and displays it in B-mode, M-mode, Color/Power Doppler, PW Spectral Doppler, and Harmonic imaging. Used by healthcare professionals in clinical settings to visualize anatomical structures and fluid flow. Features include measurement/analysis packages, Quick Scan (Q Scan), Spatial Compound Imaging, Dynamic MR Plus (SMDR), Auto IMT, and Needle Mate (Advanced Needle Visualization). System displays real-time mechanical and thermal indices. Output assists clinicians in diagnosis and biopsy guidance. Benefits include portable diagnostic imaging and procedural guidance.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including acoustic output measurements, biocompatibility, and electrical/mechanical safety testing per recognized standards (IEC 60601-1, IEC 60601-2-37, ISO 10993-1).

## Technological Characteristics

Hand-held ultrasound system; linear, curved, and phased array transducers. Modes: B, M, PW, CW, Color/Power Doppler, Harmonic, 3D/4D. Connectivity: Standard ultrasound interfaces. Safety: IEC 60601-1, IEC 60601-2-37, ISO 10993-1. Features: Q Scan, Spatial Compound Imaging, SMDR, Auto IMT, Needle Mate. Software-controlled signal processing.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- UGEO H60 Diagnostic Ultrasound System ([K122583](/device/K122583.md))
- MySono U6 Diagnostic Ultrasound System ([K113381](/device/K113381.md))
- SonoSite Maxx Series: M turbo ([K101757](/device/K101757.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K132228

510(k) Premarket Notification

UGEO PT60A Diagnostic Ultrasound System

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807. Subpart E, Section 807.92.

## 1. Submitter's Information: 21 CFR 807.92(a)(1)

SAMSUNGMEDISON CO., LTD. 42. Teheran-ro 108-gil, Gangnam-gu. Seoul, Korea

Contact Person: Kyeong-Mi. Park Regulatory Affairs Manager

Telephone: 82.2.2194.1373 Facsimile: 82.2.554.7567

Data Prepared: May 24, 2013

**AUG** 20 2013

.

### 2. Name of the device:

| Common/Usual Name:                           |           |              |
|----------------------------------------------|-----------|--------------|
| Diagnostic Ultrasound System and Accessories |           |              |
| Proprietary Name:                            |           |              |
| UGEO PT60A Diagnostic Ultrasound System      |           |              |
| Classification Names:                        | FR Number | Product Code |
| Ultrasonic Pulsed Doppler Imaging System     | 892.1550  | IYN          |
| Ultrasound Pulsed Echo Imaging System        | 892.1560  | IYO          |
| Diagnostic Ultrasound Transducer             | 892.1570  | ITX          |

## 3. Identification of the predicate or legally marketed device:

- UGEO H60 Diagnostic Ultrasound System(K 122583) -
- MySono U6 Diagnostic Ultrasound System(K113381) -
- SonoSite Maxx Series: M turbo (K 101757) .

{1}------------------------------------------------

5 10(k) Premarket Notification

-

### 4. Device Description:

The UGEO PT60A is a general purpose, hand-held, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B mode, M mode, Color Doppler imaging, Power Doppler imaging, PWSpectral Doppler mode, Harmonic imagingor as a combination of these modes. The UGEO PT60A also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The UGEO PT60A has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

#### 5. Intended Uses:

The UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analy sis of the human body.

The clinical applications include:Fetal, Abdominal, Small Organs, Adult Cephalic, Muscular-Skeletal (Conventional, Superficial), Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

#### 6. Technological Characteristics:

The UGEO PT60Ais substantially equivalent with respect to safety, effectiveness, and functionality to the UGEO H60 Diagnostic Ultrasound System (K122583) and MySono U6 Diagnostic Ultrasound System (K113381).

It is substantially equivalent with respect to safety, effectiveness, and functionality to the Advanced Needle Visualization of SonoSite Maxx Series; M Turbo (K101757) in regards to the device with Needle Mate.

All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow. and calculations.

These are described in detail in the technological characteristics comparison table as below.

510(k) Summary / Statement Certification

ATTACHEMENTI

{2}------------------------------------------------

# 510(t) Pemarket Norification

|                                   | The subject device |                       | The predicate devices  |                          |
|-----------------------------------|--------------------|-----------------------|------------------------|--------------------------|
| Feature / Characteristics         | UGEO PT60A         | UGEO H60<br>(K122583) | MySono U6<br>(K113381) | Maxx Series<br>(K101757) |
| Indication for Use                |                    |                       |                        |                          |
| - Fetal                           | √                  | √                     | √                      | √                        |
| - Abdominal                       | √                  | √                     | √                      | √                        |
| - Pediatric                       | √                  | √                     | √                      | √                        |
| - Small Organ                     | √                  | √                     | √                      | √                        |
| - Neonatal Cephalic               |                    | √                     | √                      | √                        |
| - Adult Cephalic                  | √                  | √                     | √                      | √                        |
| - Trans-rectal                    |                    | √                     | √                      | √                        |
| - Trans-vaginal                   |                    | √                     | √                      | √                        |
| - Musculo-skeletal (Conventional) | √                  | √                     | √                      | √                        |
| - Musculo-skeletal (Superficial)  | √                  | √                     | √                      | √                        |
| - Cardiac Adult                   | √                  | √                     | √                      | √                        |
| - Cardiac Pediatric               | √                  | √                     | √                      | √                        |
| - Peripheral vessel               | √                  | √                     | √                      | √                        |
| Scanhead Types                    |                    |                       |                        |                          |
| - Linear Array                    | √                  | √                     | √                      | √                        |
| - Curved Linear Array             |                    | √                     | √                      | √                        |
| - Endocavity                      |                    | √                     | √                      | √                        |
| - Phased Array                    | √                  | √                     | √                      | √                        |
| - Static Probes                   |                    |                       |                        | √                        |
| Scanhead Frequency                |                    |                       |                        |                          |
| 1.0 - 20.0 MHz                    |                    | √                     | √                      | √                        |
| Modes of Operation                |                    |                       |                        |                          |
| - B-mode                          | √                  | √                     | √                      | √                        |
| - M-mode                          | √                  | √                     | √                      | √                        |
| - Pulsed wave (PW) Doppler        | √                  | √                     | √                      | √                        |
| - Continuous wave (CW) Doppler    | √                  | √                     | √                      | √                        |
| - Color Doppler                   | √                  | √                     | √                      | √                        |
| - Power Amplitude Doppler         | √                  | √                     | √                      | √                        |
| - Tissue Harmonic Imaging         | √                  | √                     | √                      | √                        |
| - 3D/4D imaging mode              | √                  | √                     | √                      | √                        |
| - Combined modes                  | √                  | √                     | √                      | √                        |
| Safety & EMC Compliance           |                    |                       |                        |                          |
| - IEC60601-1                      | √                  | √                     | √                      | √                        |
| - UL 60601-1                      |                    |                       |                        |                          |
| - CSA C22.2 No.601.1              |                    |                       |                        |                          |
| - IEC 60601-2-37                  | √                  | √                     | √                      | √                        |
| - IEC 60601-1-2                   | √                  | √                     | √                      | √                        |
| Acoustic Output Display Standard  | √                  | √                     | √                      | √                        |
| Track 3                           |                    |                       |                        |                          |
| Patient Contact Materials         |                    |                       |                        |                          |
| Tested to ISO 10993-1             | √                  | √                     | √                      | √                        |
| Functionality                     |                    |                       |                        |                          |
| - Quick Scan (Q Scan)             | √                  | √                     | √                      | √                        |
| - Spatial Compound Imaging        | √                  | √                     | √                      |                          |
| - SMDR (Dynamic MR Plus)          | √                  | √                     | √                      |                          |

<Technological Characteristics Comparison Table>

:

{3}------------------------------------------------

#### 510(k) Premarket Notification

UGEO PT60A Diagnostic Ultrasound System

| Feature / Characteristics        | The subject device<br>UGEO PT60A | The predicate devices  |                          |
|----------------------------------|----------------------------------|------------------------|--------------------------|
|                                  | UGEO H60<br>(K122583)            | MySono U6<br>(K113381) | Maxx Series<br>(K101757) |
| - Auto IMT                       | ✓                                |                        | ✓                        |
| - Needle Mate                    | ✓                                |                        | ✓1)                      |
| 1) Advanced Needle Visualization |                                  |                        |                          |

1) Advanced Needle

7. A brief discussion of the bench and non-clinical tests conducted on the subject device The device has been evaluated for acoustic output. biocompatibility effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and has been found to conform to applicable medical device safety standards.

The UGEO PT60A and its application comply with voluntary standards as below:

- UL 60601-1. Safety requirements for Medical Equipment

- CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
- IEC60601-2-37, Diagnostic Ultrasound Safety Standards

- EN/IEC60601-1,Safety requirements for Medical Equipment

- EN/IEC60601-1-2,EMC requirements for Medical Equipment

- EN/IEC60601-1-2,EMC requirements for Medical Equipment
 NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices
- on Diagnostic Ultrasound Equipment

- ISO10993-1. Biocompatibility
- ISO10993-1. Biocompatibility
- ISO14971, Application of risk management to medical devices

#### Summary of Clinical Tests:

Not applicable. The subject of this submission, UGEO PT60A, did not require clinical studies to support substantial equivalence.

#### 8. Conclusion

Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design. development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO.. LTD. considers the UGEO PT60Ato be as safe, as effective, and performanceis substantially equivalent to thepredicate devices.

#### END of 510(K) Summary

510(k) Summary / Statement Certification

ATTACHEMENT!

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is a stylized symbol that resembles a caduceus, which is a traditional symbol of medicine.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 20, 2013

SAMSUNG MEDISON CO., LTD C/O MARK JOB RESPONSIBLE THIRD PARTY OFFICIAL REGULATORY TECHNOLOGY SERVICES LLC 1394 25TH STREET NW BUFFALO MN 55313

Re: K132228

Trade/Device Name: UGEO PT60A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 12, 2013 Received: August 13, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the UGEO PT60A Diagnostic Ultrasound System, as described in your premarket notification:

## Transducer Model Number

| C2-8  | LN5-12 | PN2-4 |
|-------|--------|-------|
| C2-5. | LS6-15 |       |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), . it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.lda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

![](https://i.imgur.com/99999999999999999999999999999999.png)

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## SECTION 1.3 INDICATIONS FOR USE

510(k) Number (if known): _K132228

Device Name: UGEO PT60A Diagnostic Ultrasound System

Indications for Use:

The UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal, Abdominal, Small Organ, Adult Cephalic, Muscular-Skeletal (Conventional, Superficial), Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use . (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Indications for Use

Section 1.3. page 1 of 7

{7}------------------------------------------------

510(k) No.: Device Name: UGEO PT60A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (* includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|----------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                  | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                    |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           | N                                                  | N | N   | N   | N                 | Note 1               | Notes 2, 7       |
|                           | Abdominal (See Note 8)       | N                                                  | N | N   | N   | N                 | Note 1               | Notes 2, 7       |
|                           | Intra-operative (See Note 6) |                                                    |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                    |   |     |     |                   |                      |                  |
| Fetal Imaging             | Laparoscopic                 |                                                    |   |     |     |                   |                      |                  |
| &<br>Other                | Pediatric                    |                                                    |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     | N                                                  | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 9  |
|                           | Neonatal Cephalic            |                                                    |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               | N                                                  | N | N   | N   | N                 | Note 1               | Note 7           |
|                           | Trans-rectal                 |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                    |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     | N                                                  | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 9  |
|                           | Musculo-skel. (Superfic.)    | N                                                  | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 9  |
|                           | Intra-luminal                |                                                    |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                | N                                                  | N | N   | N   | N                 | Note 1               | Note 7           |
| Cardiac                   | Cardiac Pediatric            | N                                                  | N | N   | N   | N                 | Note 1               | Note 7           |
|                           | Trans-esophageal (Cardiac)   |                                                    |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |
| Peripheral                | Peripheral vessel            | N                                                  | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 9  |
| Vessel                    | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B+M, B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstetrics and Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

#### Indications for Use

{8}------------------------------------------------

## 510(k) No.: Device Name: C2-5 for use with UGEO PT60A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7       |
|                           | Abdominal (See Note 8)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7       |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication: P= previously cleared by FDA KI13381: E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B+M. B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstetrics and Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

#### Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

{9}------------------------------------------------

## 510(k) No.: Device Name: C2-8 for use with UGEO PT60A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7       |
|                           | Abdominal                    | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7       |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K113381: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstetrics and Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

#### Indications for Use

{10}------------------------------------------------

## 510(k) No.: Device Name: LN5-12 for use with UGEO PT60A

| Clinical Application      |                              |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |
|---------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks 1 & III) | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |   |                                                   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           |   |                                                   |     |     |                   |                      |                  |
|                           | Abdominal                    |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6) |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |   |                                                   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |   |                                                   |     |     |                   |                      |                  |
|                           | Pediatric                    |   |                                                   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     | P | P                                                 | P   |     | P                 | Note 1               | Notes 2,5,6,9    |
|                           | Neonatal Cephalic            |   |                                                   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-urethral               |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |   |                                                   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     | P | P                                                 | P   |     | P                 | Note 1               | Notes 2,5,6,9    |
|                           | Musculo-skel. (Superfic.)    | P | P                                                 | P   |     | P                 | Note 1               | Notes 2.5.6.9    |
|                           | Intra-luminal                |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |
|                           | Cardiac Adult                |   |                                                   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric            |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            | P | P                                                 | P   |     | P                 | Note 1               | Notes 2,5,6,9    |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstettics and Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

{11}------------------------------------------------

510(k) No.:

#### ·Device Name: LS6-15 for use with UGEO PT60A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal (See Note 3)           |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal                    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     | N                                                 | N | N   |     | N                 | Note 1               | Note 9           |
|                           | Musculo-skel. (Superfic.)    | N                                                 | N | N   |     | N                 | Note 1               | Note 9           |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            | N                                                 | N | N   |     | N                 | Note 1               | Note 9           |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication: P= previously cleared by FDA: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstetrics and Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

{12}------------------------------------------------

## 510(k) No.: Device Name: PN2-4 for use with UGEO PT60A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (* includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|----------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                  | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                    |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           |                                                    |   |     |     |                   |                      |                  |
|                           | Abdominal                    | N                                                  | N | N   |     | N                 | Note 1               | Note 7           |
|                           | Intra-operative (See Note 6) |                                                    |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                    |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                    |   |     |     |                   |                      |                  |
|                           | Pediatric                    |                                                    |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |                                                    |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                    |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               | N                                                  | N | N   |     | N                 | Note 1               | Note 7           |
|                           | Trans-rectal                 |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                    |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                    |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                    |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                    |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                    |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                | N                                                  | N | N   |     | N                 | Note 1               | Note 7           |
| Cardiac                   | Cardiac Pediatric            | N                                                  | N | N   |     | N                 | Note 1               | Note 7           |
|                           | Trans-esophageal (Cardiac)   |                                                    |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            |                                                    |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                    |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+PD+PW

Note 2: Includes imaging for guidance of biopsy

Note 3: Obstetrics and Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: Includes Gynecology

Note 9: Spatial Compound Imaging

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Prescription Use (Per 21 CFR 801.109)

Smh7)

Section 1.3, page 7 of 7

Indications for Use

510(k): K132228

---

**Source:** [https://fda.innolitics.com/device/K132228](https://fda.innolitics.com/device/K132228)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
