INSUFLOW SYNERGY XL PORT

K132204 · Lexion Medical, LLC · HIF · Aug 22, 2013 · Obstetrics/Gynecology

Device Facts

Record IDK132204
Device NameINSUFLOW SYNERGY XL PORT
ApplicantLexion Medical, LLC
Product CodeHIF · Obstetrics/Gynecology
Decision DateAug 22, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Intended Use

Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.

Device Story

Device is a gas conditioning insufflator system; attaches to CO2 source; warms, humidifies, and filters gas prior to delivery into surgical cavity. System comprises sterile, single-use disposable (filter, heater/humidifier, tubing, path-of-entry access port) and reusable control module. Access ports (10mm/12mm) include duckbill/tool seals; XLR model features additional cannula seal to prevent endoscope fogging. Obturator includes transparent optical window for visualization during insertion. Used in OR by surgeons; provides conditioned gas to maintain surgical cavity; reduces fogging; improves visualization; facilitates instrument entry.

Clinical Evidence

Bench testing only. Performance testing included flow/pressure characterization, gas temperature and humidity output, insertion/removal force testing, and seal leak integrity. XLR model specifically tested to confirm reduction in endoscope fogging and improved visualization.

Technological Characteristics

Gas conditioning system with inline filter, heater/humidifier, and tubing. Access ports available in 10mm and 12mm configurations. XLR model includes endoscopic cannula seal. Obturator features transparent optical window. System utilizes reusable control module for safety/heating circuits. Single-lumen gas delivery. Sterilized, single-use disposable components.

Indications for Use

Indicated for patients undergoing thoracic, abdominal, or gynecologic minimally invasive endoscopic surgical procedures requiring CO2 insufflation and instrument access.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 5.0 510(k) Summary Insuflow SynergyTM XL and XLR Port | Date Prepared: | July 12, 2013; updated August 19, 2013 | | |----------------|------------------------------------------------------------------------------------------------------------------------------------|--------------| | Submitter: | Lexion Medical, LLC<br>545 Atwater Circle<br>St. Paul, MN 55103<br>Telephone: 651-855-1447<br>Fax: 651-636-1671 | | | Contact: | Mr. Bernard Horwath<br>Regulatory Affairs Consultant<br>4486 Timberline Ct<br>Vadnais Heights, MN 55127<br>Telephone: 651-231-1761 | AUG 2 2 2013 | #### Proprietary Name Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® SynergyTM Modified Device: XLR Port (10mm and 12mm with endoscopic cannula seal) Common/Usual Name: Gas Conditioner Insufflator Device with integral path of entry device Classification Name: Class II per regulations 884.1730, Product Code: HIF Class II per regulations 876.1500, Product Code: GCJ ### Establishment Registration Number: 2135348 #### Predicate Devices: LEXION Medical believes that Insuflow Synergy™ XL and XLR Port devices are substantially equivalent to the following devices: - Insuflow Synergy™ Port, KI20640 . - EndoPath Xcel Trocar, K032676 ### Description: The Insuflow® Synergy™ XL and XLR Port are a gas conditioning device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow Synergy™ XL and XLR Port consist of a sterile, disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the system. The technology for gas conditioning and path of entry are the same as the predicate marketed device. Regulated CO2 gas flows into the Insuflow Synergy™ XL and XLR Port, through the inline filter, continues along the tubing to enter the path of entry access device that contains the {1}------------------------------------------------ #### K132204 #### pg 2 of 3 heating element and humidification media, and through the path of entry access device lumen for delivery into the patient's surgical cavity. The integral path of entry access device in the Insuflow Synergy™ XL and XLR Port is designed and constructed similarly to the Insuflow® Synergy™ Port but comes in larger sizes just as the predicate EndoPath trocar device. The Insuflow® Synergy™ XL Port and the Insuflow® Synergy™ XLR Port have access devices in 10 mm and 12 mm configurations with a single-lumen working channel configuration for conditioned gas delivery. The access device has a working channel with duckbill and tool seals for instrument entry into the surgical cavity and delivers conditioned insufflation gas through the working channel. The Insuflow® Synergy™ XLR Port has an endoscope compatible configuration that has an additional seal at the bottom of the main cannula, which serves as a means to seal off around the endoscope, forcing all gas flow out of the peripheral gas exhaust ports. This minor change is intended to reduce fogging and improve visualization. The obturator for the Insuflow Synergy™ XL and XLR Port has a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue layers during insertion. #### Indications for Use: The Insuflow® Synergy™ XL Port and the Insuflow® Synergy™ XLR Port have the same intended use as the Insuflow® Synergy™ Port predicate device. Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity. ### Technological characteristics, comparison to predicate device. The Insuflow Synergy™ XL and XLR Port utilize the same gas conditioning technology, materials and path of entry access port technology as the predicate Insuflow® Synergy™ Port cleared under 510(k) K120640. Both devices perform a similar gas conditioning function with minor design configuration changes due to the larger trocar. The Insuflow Synergy™ XL and XLR Port path of entry access devices are technically equivalent to the predicate Insuflow Synergy™ Port cleared under 510(k) K120640 but are configured for 10 mm and 12 mm surgical tools where the predicate has a 5 mm configuration. The Insuflow® Svnergy™ Port is configured in 5 mm single and dual lumen configurations, while the Insuflow® Synergy™ XL and XLR Port have only a single lumen configuration, similar to the EndoPath Trocar which also is available in sizes up to 12 mm. Just as with the predicate EndoPath trocar, the obturator for the Insuffow Synergy™ XL and XLR Port has a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue lavers during insertion. The Insuflow Synergy™ XLR Port devices are endoscope compatible, utilizing a seal at the bottom of the access device cannula that prevents conditioned gas from fogging the endoscope lens. {2}------------------------------------------------ K132204 pg 3 of 3 #### Discussion of performance testing. Extensive performance testing has been conducted to assure that the Insuflow® Synergy™ XL and XLR Port perform in accordance with its specifications and applicable standards. Flow/pressure performance, gas temperature and humidity characterization, insertion/removal testing, and seal leak integrity testing were successfully completed. In addition, for the Insuflow® Synergy™ XLR Port endoscope compatible models, a test was successfully conducted to confirm that the cannula seal configuration does reduce scope fogging and improve visualization. #### Conclusion: Based on a comparison to the predicate devices and information provided, it is the conclusion of LEXION Medical that Insuflow Synergy™ XL Port and the Insuflow® Synergy™ XLR Port devices are substantially equivalent to devices already on the market being used for these applications (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 22, 2013 Lexion Medical, LLC % Bernard Horwath Regulatory Consultant HRG 4486 Timberline Court Vadnais Heights, MN 55127 Re: K132204 Trade/Device Name: Insuflow® Synergy" XL and Insuflow® Synergy" XLR Port Regulation Number: 21 CFR 870.1730 Regulatory Class: Class II Product Code: HIF, GCJ Dated: July 12, 2013 Received: July 23, 2013 Dear Bernard Horwath, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {4}------------------------------------------------ Page 2 - Bernard Horwath (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### 510(k) Number (if known): K132204 ## Insuflow® Synergy™ XL Port and Insuflow® Synergy™ XLR Port : . #### Indications for Use: Insuflow® Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Office of Device Evaluation (ODE) # Herbert P. Lerner -S (Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K132204 510(k) Number _
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...