← Product Code [DPT](/productcode/DPT) · K132163

# MOORLDLS-BI LASER DOPPLER BURNS IMAGER (K132163)

_Moor Instruments, Ltd. · DPT · Jan 29, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132163

## Device Facts

- **Applicant:** Moor Instruments, Ltd.
- **Product Code:** [DPT](/productcode/DPT.md)
- **Decision Date:** Jan 29, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K132163 · Jan 29, 2014 | MOORLDLS-BI LASER DOPPLER BURNS IMAGER | Moor Instruments, Ltd. | Clinical healing records (routine clinical outcomes) | The clinical investigation compared the device's burn healing potential predictions against actual patient healing records to determine accuracy. | burn healing potential; clinical healing records; accuracy assessment |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical Investigation of the moorLDLS-BI for Burn Healing Potential Assessment; Multi-centre clinical investigation | Patients with burn wounds (48 hours to 5 days post-burn); Sample Size: 596 burns cases for 204 burns patients; Number of Sites: 5 burns centres (2 UK, 1 USA, 1 Belgium, 1 Australia) | moorLDI2-BI burns imager | Accuracy of burn wound healing potential prediction compared to healing records |

## Indications for Use

The moorLDLS-BI laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%. The device is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device.

## Device Story

Imaging device using laser Doppler technique to quantify blood flow in burn-damaged skin; aids clinicians in predicting healing potential. Input: laser light projected onto tissue; linear detector array samples reflected light as line sweeps across tissue. Output: color-coded blood flow image; integrated CCD camera records color photograph for spatial correlation. Used in hospital settings by burns surgeons. Device provides visual map of perfusion; helps clinicians decide between conservative management or surgery. Benefits: rapid scanning minimizes scan time and patient discomfort compared to raster-scanning predicates. Interpretation requires consideration of confounding factors like infection, tattoos, or movement.

## Clinical Evidence

Multi-center clinical investigation (5 centers: UK, USA, Belgium, Australia) on 596 burn cases (204 patients). Primary endpoint: accuracy of healing potential prediction compared to healing records. Results: 94.2% overall accuracy. Agreement with predicate: 95.4% (HP<14), 94.2% (HP14-21), 98.5% (HP>21). Non-inferiority analysis showed subject device unlikely to perform >1.7% worse than predicate.

## Technological Characteristics

Laser Doppler imaging system; uses low-power laser line and linear detector array. Class II device. Electrical safety, laser radiation safety, and EMC compliance. Software-controlled image processing. Standalone imaging unit.

## Regulatory Identification

An extravascular blood flow probe is an extravascular ultrasonic or electromagnetic probe used in conjunction with a blood flowmeter to measure blood flow in a chamber or vessel.

## Predicate Devices

- moorLDI2-BI Laser Doppler Burns Imager ([K060976](/device/K060976.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

struments laser Doppler blood flow assessment

#### moorLDLS-Bl 510(k) Summary ર

K132163 : Page 1 of 3

| Submitter:             | Moor Instruments Ltd                                                                                                                                                 |             |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address:               | Millwey<br>Axminster, Devon<br>EX13 5HU, United Kingdom                                                                                                              | JAN 29 2014 |
| Telephone:             | (+44) 1297 35715                                                                                                                                                     |             |
| Fax:                   | (+44) 1297 35716                                                                                                                                                     |             |
| Contact:               | Xiabing Huang                                                                                                                                                        |             |
| Contact title:         | Managing Director                                                                                                                                                    |             |
| E-mail:                | xhuang@moor.co.uk                                                                                                                                                    |             |
| Date:                  | July 1, 2013                                                                                                                                                         |             |
| Model Name:            | moorLDLS-BI Laser Doppler Burns Imager                                                                                                                               |             |
| Model Number:          | moorLDLS-BI                                                                                                                                                          |             |
| Common Name:           | Laser Doppler Burns Imager                                                                                                                                           |             |
| Classification Name:   | Extravascular blood flow probe, DPT, 21 CFR 870.2120<br>Laser surgical instrument for use in general and plastic surgery<br>and in dermatology, GEX, 21 CFR 878.4810 |             |
| Regulatory Status:     | Class II                                                                                                                                                             |             |
| Establishment Reg No:  | 8043564                                                                                                                                                              |             |
| Type of 510(k):        | Traditional                                                                                                                                                          |             |
| Reason for submission: | New Device                                                                                                                                                           |             |
| Predicate Device:      | moorLDI2-BI Laser Doppler Burns Imager                                                                                                                               |             |

5-1

510(k) Number K060976

Moor Instruments Ltd Millwey Axminster Devon EX13 5HU UK tel +14 (0)1297 35715 fax +44 (0)1297 35716 · amail sales@moor.co.uk website www.moor.co.uk Company Registered in England No. 2200367 VAT Ringisualion No. 08490987008

ﺑ

Image /page/0/Picture/7 description: The image contains two logos. The logo on the left has a check mark inside of a circle with the text "ISO 9001:2000 SGS" underneath. The logo on the right has a check mark with a crown above it and the text "UKAS Management Systems 005" underneath. Both logos appear to be quality assurance or certification marks.

Image /page/0/Picture/8 description: The image shows two semi-circular shapes, one above the other. Both shapes have a similar pattern of lines radiating outwards from a central point. The lines appear to be slightly curved and unevenly spaced, giving the impression of a natural or organic structure.

.

{1}------------------------------------------------

#### K132163 Description of the Device

Page 2 of 3

The moor DLS-BI laser Doppler burns imager is an imaging device to aid the clinician to judge the healing potential of bums and the need for surgery.

It uses the laser Doppler imaging technique to quantify the blood flow in an area of skin damaged by a burn. The device uses a line of laser light projected onto the tissue and a linear detector arrav that sample from the line as it is swept across the tissue to rapidly build up a colour coded image of blood flow in the burn area and the surrounding normal skin for healing potential prediction. In addition, a CCD camera is integrated into the scanner unit for recording a colour photograph at the time of scanning, corresponding closely with the blood flow image in size and aspect.

# Intended Use

The moorLDLS-BI laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%.

The device is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device.

## Substantial Equivalence

## Technological Characteristics

The operation and design of the moorLDLS-BI laser Doppler burns imager and the predicate device moorLDI2-B1 are very similar. Both devices rely on the same physical principle, i.e. the laser Doppler imaging technology, to measure the tissue blood perfusion for healing potential prediction of burn wounds. The moorLDLS-Bl uses the same dedicate colour palette as the predicate device to predict three types of burn wound healing potentials, i.e. HP14: healing in <14 days, HP14-21: healing in 14-21 days and HP>21: healing in > 21 days.

The main difference between two devices is the scanning method. The predicate device moorLDI2-BI scans a low power laser beam (single laser spot) over the tissue surface in a raster pattern to produce a two dimensional colour coded blood perfusion image, while the moorLDLS-BI laser Doppler burns imager uses a low power laser line to sweep across the tissue for rapid scanning which could be important in clinical environment to minimize scan time and patient discomfort.

#### Indication for Use

Both devices have the same intended use, i.e. assessing the blood flow in debrided burn wounds of the skin and to be used as an aid to burn wound management for patients with burn wounds.

Both devices are intended to be used within the hospital environments by burns surgeons for the same patient population, i.e. patients with burn wounds at 48 hours to 5 days post-burn.

#### Performance Data

The moorLDLS-BI burns imager has been designed and tested for compliance with the standards for electrical safety, laser radiation safety, electromagnetic compatibility and programmable medical device.

5-2

{2}------------------------------------------------

K132163

Page 3 of 3

To establish the substantial equivalent between the predicate device for the effectiveness of burns assessment, two types of performance bench testing and a clinical investigation have been carried out. The bench testing includes a flow model experiment and image scan of normal skin tissue using both devices. The results show that the moorLDLS-Bl and the predicate device have a good correlation for tissue blood flow measurement.

A multi-centre clinical investigation entitled 'Clinical Investigation of the moorLDLS-Bl for Burn Healing Potential Assessment' was conducted in five burns centres (2 in the UK, 1 in the USA, 1 in the Belgium, 1 in the Australia).

The objectives of this clinical investigation were

To assess the performance and accuracy of a new moorLDLS-BI line scan burns 1. imager for prediction of burn wound healing potential;

- 2. To compare the performance of the moorLDLS-Bl with the predicate device moorLDI2-BI burns imager.
The clinical investigation of the performance and accuracy for burns assessment using both devices has been performed on 596 burns cases for 204 burns patients. An overall accuracy of 94.2% was found for moorLDLS-Bl when compared with healing records. The agreement between the moorLDLS-Bl and moorLDI2-Bl as the non-reference stand was 95.4% for HP14, 94.2% for HP14-21 and 98.5% for HP>21.

A specific analysis of non-inferiority was performed and found that the moorLDLS-Bl is unlikely to perform more than 1.7% worse than the predicate device moorLDI2-Bl. In fact, the moorLDLS-Bl performed slightly better than the moorLD12-Bl for these data, so there may well be no drop in accuracy at all.

The laser Doppler blood flow images of burns and/or their interpretation can be adversely influenced by a number of factors: these must be considered during image interpretation. These are referred to as patient and technical related confounding factors such as un-debrided skin, infection, medication/sickness, Tattoos, edge effect, high reflection, and movement artefact;

## Conclusions

Based on the indications for use, design, technology, performance and functional testing, and clinical evaluation, it can be concluded that the moorLDLS-Bl laser Doppler burns imager is substantial equivalent to the predicate device moorLD12-B1 as an aid to assessing healing potential in burn damaged skin in terms of effectiveness and safety.

5-3

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Moor Instruments Ltd. Xiabing Huang Managing Director Millwey, Axminster Devon, EX13 5HU UNITED KINGDOM

January 29, 2014

Re: K132163

Trade/Device Name: MoorLDLS-BI laser-doppler burns imager Regulation Number: 21 CFR 870.2120 Regulation Name: Extravascular blood flow probe Regulatory Class: Class II Product Code: DPT, GEX Dated: December 12, 2013 Received: December 26, 2013

Dear Dr. Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 – Xiabing Huang

CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Joshua C. Nipper -S

For

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# 4 Indications for Use Statement

510(k) Number:_K132163

Device Name: moorLDLS-81 Laser Doppler Burns Imager

Indications For Use:

.

::

The moorl.DLS-81 laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%.

Prescription Use: Yes
(Part 21 CFR 601 SubpartD)

AND/OR

Over-The-Counter User No (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden: Neil R Ogden - System - 05'00'

(Division Slgn-off) for BSA

Division of Surgical Devices

510(k) Number_K132163

---

**Source:** [https://fda.innolitics.com/device/K132163](https://fda.innolitics.com/device/K132163)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
