Clinical Investigation of the moorLDLS-BI for Burn Healing Potential Assessment; Multi-centre clinical investigation
Patients with burn wounds (48 hours to 5 days post-burn); Sample Size: 596 burns cases for 204 burns patients; Number of Sites: 5 burns centres (2 UK, 1 USA, 1 Belgium, 1 Australia)
moorLDI2-BI burns imager
Accuracy of burn wound healing potential prediction compared to healing records
Indications for Use
The moorLDLS-BI laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%. The device is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device.
Device Story
Imaging device using laser Doppler technique to quantify blood flow in burn-damaged skin; aids clinicians in predicting healing potential. Input: laser light projected onto tissue; linear detector array samples reflected light as line sweeps across tissue. Output: color-coded blood flow image; integrated CCD camera records color photograph for spatial correlation. Used in hospital settings by burns surgeons. Device provides visual map of perfusion; helps clinicians decide between conservative management or surgery. Benefits: rapid scanning minimizes scan time and patient discomfort compared to raster-scanning predicates. Interpretation requires consideration of confounding factors like infection, tattoos, or movement.
Clinical Evidence
Multi-center clinical investigation (5 centers: UK, USA, Belgium, Australia) on 596 burn cases (204 patients). Primary endpoint: accuracy of healing potential prediction compared to healing records. Results: 94.2% overall accuracy. Agreement with predicate: 95.4% (HP<14), 94.2% (HP14-21), 98.5% (HP>21). Non-inferiority analysis showed subject device unlikely to perform >1.7% worse than predicate.
Technological Characteristics
Laser Doppler imaging system; uses low-power laser line and linear detector array. Class II device. Electrical safety, laser radiation safety, and EMC compliance. Software-controlled image processing. Standalone imaging unit.
Indications for Use
Indicated for patients with burn wounds (up to 30% Total Body Surface Area) at 48 hours to 5 days post-burn. Used by burns surgeons in hospital environments to assess blood flow in debrided burn wounds to aid in predicting healing potential.
Regulatory Classification
Identification
An extravascular blood flow probe is an extravascular ultrasonic or electromagnetic probe used in conjunction with a blood flowmeter to measure blood flow in a chamber or vessel.
{0}------------------------------------------------
struments laser Doppler blood flow assessment
#### moorLDLS-Bl 510(k) Summary ર
K132163 : Page 1 of 3
| Submitter: | Moor Instruments Ltd | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Address: | Millwey<br>Axminster, Devon<br>EX13 5HU, United Kingdom | JAN 29 2014 |
| Telephone: | (+44) 1297 35715 | |
| Fax: | (+44) 1297 35716 | |
| Contact: | Xiabing Huang | |
| Contact title: | Managing Director | |
| E-mail: | xhuang@moor.co.uk | |
| Date: | July 1, 2013 | |
| Model Name: | moorLDLS-BI Laser Doppler Burns Imager | |
| Model Number: | moorLDLS-BI | |
| Common Name: | Laser Doppler Burns Imager | |
| Classification Name: | Extravascular blood flow probe, DPT, 21 CFR 870.2120<br>Laser surgical instrument for use in general and plastic surgery<br>and in dermatology, GEX, 21 CFR 878.4810 | |
| Regulatory Status: | Class II | |
| Establishment Reg No: | 8043564 | |
| Type of 510(k): | Traditional | |
| Reason for submission: | New Device | |
| Predicate Device: | moorLDI2-BI Laser Doppler Burns Imager | |
5-1
510(k) Number K060976
Moor Instruments Ltd Millwey Axminster Devon EX13 5HU UK tel +14 (0)1297 35715 fax +44 (0)1297 35716 · amail sales@moor.co.uk website www.moor.co.uk Company Registered in England No. 2200367 VAT Ringisualion No. 08490987008
ﺑ
Image /page/0/Picture/7 description: The image contains two logos. The logo on the left has a check mark inside of a circle with the text "ISO 9001:2000 SGS" underneath. The logo on the right has a check mark with a crown above it and the text "UKAS Management Systems 005" underneath. Both logos appear to be quality assurance or certification marks.
Image /page/0/Picture/8 description: The image shows two semi-circular shapes, one above the other. Both shapes have a similar pattern of lines radiating outwards from a central point. The lines appear to be slightly curved and unevenly spaced, giving the impression of a natural or organic structure.
.
{1}------------------------------------------------
#### K132163 Description of the Device
Page 2 of 3
The moor DLS-BI laser Doppler burns imager is an imaging device to aid the clinician to judge the healing potential of bums and the need for surgery.
It uses the laser Doppler imaging technique to quantify the blood flow in an area of skin damaged by a burn. The device uses a line of laser light projected onto the tissue and a linear detector arrav that sample from the line as it is swept across the tissue to rapidly build up a colour coded image of blood flow in the burn area and the surrounding normal skin for healing potential prediction. In addition, a CCD camera is integrated into the scanner unit for recording a colour photograph at the time of scanning, corresponding closely with the blood flow image in size and aspect.
# Intended Use
The moorLDLS-BI laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%.
The device is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device.
## Substantial Equivalence
## Technological Characteristics
The operation and design of the moorLDLS-BI laser Doppler burns imager and the predicate device moorLDI2-B1 are very similar. Both devices rely on the same physical principle, i.e. the laser Doppler imaging technology, to measure the tissue blood perfusion for healing potential prediction of burn wounds. The moorLDLS-Bl uses the same dedicate colour palette as the predicate device to predict three types of burn wound healing potentials, i.e. HP14: healing in <14 days, HP14-21: healing in 14-21 days and HP>21: healing in > 21 days.
The main difference between two devices is the scanning method. The predicate device moorLDI2-BI scans a low power laser beam (single laser spot) over the tissue surface in a raster pattern to produce a two dimensional colour coded blood perfusion image, while the moorLDLS-BI laser Doppler burns imager uses a low power laser line to sweep across the tissue for rapid scanning which could be important in clinical environment to minimize scan time and patient discomfort.
#### Indication for Use
Both devices have the same intended use, i.e. assessing the blood flow in debrided burn wounds of the skin and to be used as an aid to burn wound management for patients with burn wounds.
Both devices are intended to be used within the hospital environments by burns surgeons for the same patient population, i.e. patients with burn wounds at 48 hours to 5 days post-burn.
#### Performance Data
The moorLDLS-BI burns imager has been designed and tested for compliance with the standards for electrical safety, laser radiation safety, electromagnetic compatibility and programmable medical device.
5-2
{2}------------------------------------------------
K132163
Page 3 of 3
To establish the substantial equivalent between the predicate device for the effectiveness of burns assessment, two types of performance bench testing and a clinical investigation have been carried out. The bench testing includes a flow model experiment and image scan of normal skin tissue using both devices. The results show that the moorLDLS-Bl and the predicate device have a good correlation for tissue blood flow measurement.
A multi-centre clinical investigation entitled 'Clinical Investigation of the moorLDLS-Bl for Burn Healing Potential Assessment' was conducted in five burns centres (2 in the UK, 1 in the USA, 1 in the Belgium, 1 in the Australia).
The objectives of this clinical investigation were
To assess the performance and accuracy of a new moorLDLS-BI line scan burns 1. imager for prediction of burn wound healing potential;
- 2. To compare the performance of the moorLDLS-Bl with the predicate device moorLDI2-BI burns imager.
The clinical investigation of the performance and accuracy for burns assessment using both devices has been performed on 596 burns cases for 204 burns patients. An overall accuracy of 94.2% was found for moorLDLS-Bl when compared with healing records. The agreement between the moorLDLS-Bl and moorLDI2-Bl as the non-reference stand was 95.4% for HP14, 94.2% for HP14-21 and 98.5% for HP>21.
A specific analysis of non-inferiority was performed and found that the moorLDLS-Bl is unlikely to perform more than 1.7% worse than the predicate device moorLDI2-Bl. In fact, the moorLDLS-Bl performed slightly better than the moorLD12-Bl for these data, so there may well be no drop in accuracy at all.
The laser Doppler blood flow images of burns and/or their interpretation can be adversely influenced by a number of factors: these must be considered during image interpretation. These are referred to as patient and technical related confounding factors such as un-debrided skin, infection, medication/sickness, Tattoos, edge effect, high reflection, and movement artefact;
## Conclusions
Based on the indications for use, design, technology, performance and functional testing, and clinical evaluation, it can be concluded that the moorLDLS-Bl laser Doppler burns imager is substantial equivalent to the predicate device moorLD12-B1 as an aid to assessing healing potential in burn damaged skin in terms of effectiveness and safety.
5-3
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Moor Instruments Ltd. Xiabing Huang Managing Director Millwey, Axminster Devon, EX13 5HU UNITED KINGDOM
January 29, 2014
Re: K132163
Trade/Device Name: MoorLDLS-BI laser-doppler burns imager Regulation Number: 21 CFR 870.2120 Regulation Name: Extravascular blood flow probe Regulatory Class: Class II Product Code: DPT, GEX Dated: December 12, 2013 Received: December 26, 2013
Dear Dr. Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 – Xiabing Huang
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4 Indications for Use Statement
510(k) Number:_K132163
Device Name: moorLDLS-81 Laser Doppler Burns Imager
Indications For Use:
.
::
The moorl.DLS-81 laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%.
Prescription Use: Yes
(Part 21 CFR 601 SubpartD)
AND/OR
Over-The-Counter User No (21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R Ogden: Neil R Ogden - System - 05'00'
(Division Slgn-off) for BSA
Division of Surgical Devices
510(k) Number_K132163
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.