S HEALTH

K132148 · Samsung Electronics Co., Ltd. · DRG · Dec 4, 2013 · Cardiovascular

Device Facts

Record IDK132148
Device NameS HEALTH
ApplicantSamsung Electronics Co., Ltd.
Product CodeDRG · Cardiovascular
Decision DateDec 4, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2910
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The S Health is a mobile application intended for use in home to help people in reviewing and monitoring vital signs such as non-invasive blood pressure, blood glucose, weight and other data from optional add-on devices for effective health management. The user also can share the data via sharing functions (Email and SMS).

Device Story

S Health is mobile application software for health data management. It receives data from optional add-on devices (blood glucose meters, blood pressure monitors, scales) and records user-entered lifestyle data (step count, calories burnt, food intake, comfort levels). The application displays this information in a user-friendly interface for home use by the patient. Users can share recorded data via email or SMS. The device serves as a data management tool to assist users in tracking health metrics and vital signs. It does not perform diagnostic analysis but facilitates effective health management through data review and sharing.

Clinical Evidence

Bench testing demonstrated accurate data transmission from meters to the software and proper memory data rollover functionality. A usability study with lay users reported 100% satisfaction, with all participants rating the program as easy or somewhat easy to use.

Technological Characteristics

Mobile application software. Functions as a data management system for physiological signals (blood pressure, blood glucose, weight). Connectivity via data transmission from external meters. No specific hardware materials or energy sources defined as it is a software-only device.

Indications for Use

Indicated for home use by individuals to review and monitor vital signs, including non-invasive blood pressure, blood glucose, and weight, using data from optional add-on devices for health management.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary 0 This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: #### Applicant / Submitter: 1. Samsung Electronics Co., Ltd. 129, Samsung-ro. Yeongtong-gu. Suwon-si. Republic of Korea 443-742 **DEC 0 4 2013** Tel : +82-31-301-7374 Fax : +82-31-301-1768 #### Submission Correspondent: 2. Priscilla Chung LK Consulting Group USA. Inc. 1515 E. Katella Ave. Unit 2115. Anaheim, CA 92805 Phone: 714-202-5789. Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com #### Device: 3. r | Proprietary Name: | S Health | |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Healthcare Data Management Mobile Application | | Classification Name: | Transmitters And Receivers, Physiological Signal,<br>Radiofrequency<br>System, Measurement, Blood-Pressure, Non-Invasive<br>System, Test, Blood Glucose, Over The Counter<br>Scale, Stand-On, Patient | | Classification: | Class II,<br>21 CFR 870.2910<br>21 CFR 870.1130<br>21 CFR 862.1345<br>21 CFR 880.2700 | | Classification Product Code: | DRG, DXN, NBW, FRI | 1 {1}------------------------------------------------ ## K132148 Page 2 of 3 #### 4. Predicate Device: MedApps 2.0 - Remote Patient Monitoring System (K083862) by MedApps, Inc. Ambio Remote Health Monitoring System (K130676) by Arrayent Health LLC /dba Ambio Health Apps-Health01 (K122098) by Andon Health Co., Ltd. iBGStar Diabetes Manager Application (K130535) by AgaMatrix. Inc. #### ર. Device Description: The S Health is application software for mobile device. This mobile application helps users to care their personal health better by recording and displaying information such as comfort level(temperature and humidity), step count (pedometer), how much user exercises (calories burnt), the food user may consume (calories intake), weight, blood glucose level, and blood pressure in an effective and user friendly interface. #### ે. Intended Use: The S Health is a mobile application intended for use in home to help people in reviewing and monitoring vital signs such as non-invasive blood pressure, blood glucose, weight and other data from optional add-on devices for effective health management. The user also can share the data via sharing functions (Email and SMS). #### 7. Performance Data(Non-Clinical): Samsung Electronics Co, Ltd. conducted bench testing to demonstrate data accuracy transmission for each meter. Memory data rollover was also tested by adding ten (10) data points to meters that were pre-loaded with full memory. Test results show that all the measurement values, data, and time properly downloaded from the meters to the software. Memory data rollover also functioned properly where the new measurement values replaced the oldest glucose values. In addition, a study intended to assess lay users' performance in using the S Health was conducted. Overall, they rated the S Health at 100% for overall program as easy or somewhat easy. There were no users that rated the software program as somewhat difficult or difficult. 100% of them also responded that they are satisfied with the S Health and its manual. The study results supports that the S Health is effective and provides accurate data management system as other predicate devices in the market. #### 8. Substantial Equivalence The predicate devices and S Health are optional software accessories for use with blood glucose, blood pressure, and weight with data management capability. There are some minor differences between the subject device and the predicate device in software operation options, settings and some other features, but it does not constitute a new intended use. Despite the differences, the validation testing results presented in this 510K supports that the S Health is safe and effective as the predicate devices. {2}------------------------------------------------ K132148 Page 3 of 3 . . : . . · . #### 9. Conclusion: . Based on the testing results. Samsung Electronics concludes that the S Health is safe and effective also, substantially equivalent to the predicate devices. 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - . . . : . : : {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. ## December 4, 2013 Samsung Electronic Con. Ltd. % Priscilla Chung Official Correspondent LK Consulting Group 1515 E Katella Ave. Unit 2115 Anaheim, CA 92805 US Re: K132148 Trade/Device Name: S Health Regulation Number: 21 CFR 870.2910 Regulation Name: Healthcare Data Management Mobile Application Regulatory Class: Class II Product Code: DRG, NBW, DXN, FRI Dated: November 4, 2013 Received: November 7, 2013 Dear Priscilla Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ Page 2 - Priscilla Chung or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDeyices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Owen P. Faris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number: K _ Device Name: S Health Indication for use: The S Health is a mobile application intended for use in home to help people in reviewing and monitoring vital signs such as non-invasive blood pressure, blood glucose, weight and other data from optional add-on devices for effective health management. The user also can share the data via sharing functions (Email and SMS). Prescription Use_ (Part 21CFR801 Subpart D) AND/OR Over-The-Counter Use (Part 21CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%