← Product Code [LNH](/productcode/LNH) · K132119

# MAGNETOM PRISMA, MAGNETOM PRISMA FIT (K132119)

_Siemens Medical Solutions USA, Inc. · LNH · Nov 22, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132119

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Nov 22, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MAGNETOM Prisma and the MAGNETOM Prisma™ systems are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM Prisma and the MAGNETOM Prisma™ MR systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

## Device Story

MAGNETOM Prisma/Prisma fit are 3T superconducting whole-body MR scanners; designed for increased patient comfort, ergonomics, and workflow efficiency. Systems acquire MR signals to produce transverse, sagittal, coronal, and oblique cross-sectional images, spectroscopic images, and spectra. Operated by trained professionals in clinical settings; used for diagnostic imaging of head, body, and extremities. Supports interventional procedures using MR-compatible accessories (e.g., biopsy needles, in-room displays). Output images/spectra interpreted by physicians to assist in clinical diagnosis. Prisma offered as new production; Prisma fit offered as upgrade to existing MAGNETOM Trio a Tim System (TaTS).

## Clinical Evidence

Bench testing only. Performance and safety data collected per NEMA, IEC, and ISO standards, including static field, rate of change of magnetic field, RF power deposition, acoustic noise, signal-to-noise ratio, image uniformity, geometric distortion, slice profile/thickness/gap, and high-contrast spatial resolution. No clinical data presented.

## Technological Characteristics

3T superconducting magnet; open-bore design. Software: syngo® MR VD13D. Safety/Performance standards: IEC 60601-1 series, NEMA, ISO. Connectivity: Networked/DICOM. Sterilization: N/A (non-invasive diagnostic imaging).

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Siemens MAGNETOM Skyra (3 T) MR system with syngo® MR D13D ([K123510](/device/K123510.md))
- Siemens MAGNETOM Trio a Tim System (TaTS) (3 T) MR System with syngo® MR B19 ([K123938](/device/K123938.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K132119

## NOV 22222013 SIEMENS

# Section 5 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.

### l. General Information

| Establishment                                        | Siemens Medical Solutions USA. Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA                                                                                                                                                                       |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number                                  | 2240869                                                                                                                                                                                                                                                                          |
| Date Prepared                                        | July 8, 2013                                                                                                                                                                                                                                                                     |
| Manufacturer                                         | Siemens AG<br>Henkestrasse 127<br>D-91052 Erlangen, Germany                                                                                                                                                                                                                      |
| Registration Number                                  | 3002808157                                                                                                                                                                                                                                                                       |
| Contact Person                                       | Ms. Nadia Sookdeo<br>Regulatory Affairs Technical Specialist<br>Siemens Healthcare<br><br>Siemens Medical Solutions USA, Inc.<br>Customer Solutions Group<br>51 Valley Stream Parkway<br>Mail Code G01<br>Malvern, PA 19355, USA<br>Phone: (610) 448-4918<br>Fax: (610) 448-1787 |
| Device Name                                          | Trade Names: MAGNETOM Prisma<br>MAGNETOM Prismafi                                                                                                                                                                                                                                |
| Classification Name:<br>CFR Code:<br>Classification: | Magnetic Resonance Diagnostic Device<br>21 CFR § 892.1000<br>Class II                                                                                                                                                                                                            |
| Performance Standards                                | None established under Section 514 Subpart                                                                                                                                                                                                                                       |

the Food, Drug and Cosmetic Act.

J of

{1}------------------------------------------------

## SIEMENS

Image /page/1/Picture/1 description: The image shows a black and white abstract design. The design features a scattering of small, irregular black shapes against a white background. The shapes are concentrated towards the bottom left corner of the image, with a few isolated shapes dispersed towards the top.

### II. Safety and Effectiveness Information Supporting Substantial Equivalence

### Intended Use

The MAGNETOM Prisma and the MAGNETOM Prisma™ systems are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.

The MAGNETOM Prisma and the MAGNETOM Prisma" MR systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

### Device Description

MAGNETOM Prisma (3 T) and MAGNETOM Prisma" (3 T) are similar to the previously cleared MAGNETOM Skyra (3 T) and MAGNETOM Trio a Tim System (TaTS) (3 T) systems utilizing a superconducting magnet design. The open bore, whole body scanners are designed for increased patient comfort. They focus on ergonomics and usability to reduce complexity of the MR workflow.

The MAGNETOM Prisma will be offered as ex-factory (new production) and the MAGNETOM Prismall will be offered as an upgrade to the currently installed MAGNETOM Trio a Tim System (TaTS) systems.

### Substantial Equivalence

It is Siemens opinion that the MAGNETOM Prisma systems and the upgraded MAGNETOM Prismad systems with syngo® MR VD13D software i are substantially equivalent to the following predicate devices:

| Predicate Device Name                                                              | FDA Clearance Number | FDA Clearance Date |
|------------------------------------------------------------------------------------|----------------------|--------------------|
| Siemens MAGNETOM Skyra (3 T) MR<br>system with syngo® MR D13D                      | K123510              | May 17, 2013       |
| Siemens MAGNETOM Trio a Tim System<br>(TaTS) (3 T) MR System with syngo® MR<br>B19 | K123938              | February 12, 2013  |

### Safety and Effectiveness

{2}------------------------------------------------

## SIEMENS

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product.. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards. SIEMENS adheres to recognized and established industry practice and relevant international standards, such as the IEC 60601-1 series.

Operation of the MAGNETOM Prisma (3T) and the MAGNETOM Prisma" (3T) systems with syngo® MR VD13D software is substantially equivalent to the commercially available MAGNETOM Skyra (3T) Systems with syngo® MR VD13C SW (K123510) and MAGNETOM Trio a Tim System (TaTS) (3T) with syngo® MR B19 SW (K123938)

Additionally, as specified in the FDA quidance document "Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Devices" (released Nov. 1998) the following measurements of performance and safety data have been performed following NEMA or equivalent IEC and ISO standards:

### Safety:

- · Maximum Static Field
- · Rate of Change of Magnetic Field
- · RF Power Deposition
- · Acoustic Noise Levels

### Performance:

- · Specification Volume
- · Signal to Noise
- · Image Uniformity
- · Geometric Distortion
- · Slice Profile. Thickness and Gap
- High Contrast Spatial Resolution

The MAGNETOM Prisma and MAGNETOM Prisma™ syngo® MR VD13D software will conform to the measurements of safety parameters to the international IEC, ISO and NEMA standards for safety issues with Magnetic Resonance Imaging Diagnostic Devices.

Furthermore performance testing has been completed to show that the performance of the MAGNETOM Prisma and MAGNETOM Prisma™ with syngo® MR VD 13D Software is equivalent with respect to the predicate devices.

This assures that the performance of these devices can be considered as safe and effective with respect to the currently available MAGNETOM Skyra and MAGNETOM Trio a Tim System (TaTS) MR systems.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird with outstretched wings.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2013

Siemens Medical Solutions USA, Inc. % Ms. Nadia Sookdeo Regulatory Affairs Technical Specialist 51 Valley Stream Parkway, D02 MALVERN PA 19355

Re: K132119

Trade/Device Name: MAGNETOM Prisma, MAGNETOM Prisma 11 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 5, 2013 Received: November 8, 2013

Dear Ms. Sookdeo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

### Page 2-Ms. Sookdeo

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2)CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours.

Michael D. O'Hara

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K132119

#### Device Name

. –

MAGNETOM Prisma and MAGNETOM Prisma fit

Indications for Use (Describe)

The MAGNETOM Prisms and MAGNETOM Prismalit systems are indicated for use as a magnetic resomance diagnostic device (MRDD) that produces transverse, sagital, coronal and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The MAGNETOM Prisma and MAGNETOM Prismalit systems may also be imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR Safe biopsy needles.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpant D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Michael D. O'Hara |
|------------------------------------------------------------------------------|-------------------|
|------------------------------------------------------------------------------|-------------------|

FORM FDA 3881 (9/13)

PSC Publicition Services ( VII ) 441-4740 EF

---

**Source:** [https://fda.innolitics.com/device/K132119](https://fda.innolitics.com/device/K132119)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
