PERCLOT TOPICAL

K132105 · Cryolife, Inc. · QSY · Apr 3, 2014 · SU

Device Facts

Record IDK132105
Device NamePERCLOT TOPICAL
ApplicantCryolife, Inc.
Product CodeQSY · SU
Decision DateApr 3, 2014
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

PerClot Topical is intended for use under the care of a health care professional as a topical dressing for the temporary treatment of mildly bleeding wounds such as surgical wounds (postoperative, donor sites, dermatological), cuts and lacerations and for the treatment of mild bleeding from topical ENT surgical wounds and nosebleeds. It is also indicated for control of bleeding from the skin at percutaneous needle access, vascular access, and percutaneous catheter access sites.

Device Story

PerClot Topical is a hemostatic agent consisting of absorbable polysaccharide granules derived from purified potato starch; supplied with delivery applicators. Used by healthcare professionals in clinical settings to treat mild bleeding. Upon application, granules rapidly absorb water to form a gelled adhesive matrix; this matrix provides a mechanical barrier and concentrates platelets, red blood cells, and coagulation proteins at the wound site to promote the physiological clotting cascade. Granules are enzymatically degraded by alpha-amylase, glucoamylase, and macrophages. The device aids in achieving hemostasis for surgical, dermatological, and percutaneous access wounds, potentially reducing bleeding time and improving wound management.

Clinical Evidence

No clinical data. Evidence consists of bench testing (water absorption, adhesion/cohesion, particle size distribution) and pre-clinical GLP evaluation. Biocompatibility testing performed per ISO 10993.

Technological Characteristics

Absorbable polysaccharide granules derived from purified potato starch. Non-pyrogenic. Biocompatible per ISO 10993. Mechanism: mechanical barrier formation via rapid water absorption and gelled matrix creation. Delivery via applicators. No human or animal components.

Indications for Use

Indicated for health care professional use as a topical dressing for temporary treatment of mild bleeding wounds, including surgical (post-op, donor, dermatological), cuts, lacerations, ENT surgical wounds, nosebleeds, and bleeding at percutaneous needle, vascular, or catheter access sites.

Regulatory Classification

Identification

To temporarily control bleeding and cover external wounds.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # APR 0 3 2014 KI32105 ## 510(k) Summary | Date: | April 03, 2014 | |-------------------------------|-------------------------------------------------| | Submitter: | CryoLife, Inc. | | Address: | 1655 Roberts Boulevard NW<br>Kennesaw, GA 30144 | | Phone: | 770-419-3355 | | Fax: | 770-590-3783 | | Contact: | Kimberly DiCono | | Device Trade Name: | PerClot® Topical | | Common Name: | Topical hemostatic particles | | 510(k) Number: | K132105 | | Proposed Classification Name: | Dressing, Pre-Amendment Unclassified | | Product Code: | FRO - General and Plastic Surgery | #### Predicate Device Information: | Device Name | Company | 510(k) Clearance | Product Code | |-------------------------------------|----------------------------------------------------------------|------------------|--------------| | NexStat® Topical<br>Hemostat Powder | Hemostasis, LLC<br>5000 Township Parkway<br>St. Paul, MN 55110 | K102459 | FRO | ### Device Description: PerClot® Topical is a medical device composed of absorbable polysaccharide granules and delivery applicators. The granules are biocompatible, non-pyrogenic and derived from purified potato starch. The granules do not contain any human or animal components. PerClot granules have a molecular structure that rapidly absorbs water, forming a gelled adhesive matrix that provides a mechanical barrier to further bleeding and results in the concentration of platelets, red blood cells, and coagulation proteins (thrombin, fibrinogen, etc.) at the site of application. The gelled adhesive matrix thus promotes the normal physiological clotting cascade. PerClot granules are enzymatically degraded by alpha-amylase and glucoamylase and by macrophages. {1}------------------------------------------------ #### Intended Use: PerClot Topical is intended for use under the care of a health care professional as a topical dressing for the temporary treatment of mildly bleeding wounds such as surgical wounds (postoperative, donor sites, dermatological), cuts and lacerations and for the treatment of mild bleeding from topical ENT surgical wounds and nosebleeds. It is also indicated for control of bleeding from the skin at percutaneous needle access, vascular access, and percutaneous catheter access sites. #### Bench Testing: Functional testing demonstrated that PerClot is substantially equivalent when compared to the predicate device. #### Performance Testing: Studies demonstrated that PerClot meets the total water absorption, rate of water absorption, adhesion/cohesion, and particle/granule size distribution specifications. The function of the product is directly related to these performance specifications. A Pre-clinical GLP evaluation was conducted according to applicable standards. This study supports PerClot's is safety and effectiveness. #### Substantial Equivalence to Predicate Device: PerClot Topical is substantially equivalent to the predicate device. #### Clinical Testing: Evaluation in a clinical setting has not been performed. #### Conclusion: Based on the functional test results, pre-clinical, and biocompatibility studies performed in accordance with ISO 10993, these products behave similarly and are substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 3, 2014 CryoLife Incorporated Ms. Kimberly DiCono Regulatory Affairs Manager 1655 Roberts Boulevard Northwest Kennesaw, Georgia 30144 Re: K132105 Trade/Device Name: PerClot Topical Regulatory Class: Unclassified Product Code: FRO Dated: February 26, 2014 Received: February 27, 2014 Dear Ms. DiCono: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Kimberly DiCono forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ### David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K132105 Device Name: PerClot Topical Indications for Use: PerClot Topical is intended for use under the care of a health care professional as a topical dressing for the temporary treatment of mildly bleeding wounds such as surgical wounds (post-operative, donor sites, dermatological), cuts and laceration and for the treatment of mild bleeding from topical ENT surgical wounds and nosebleeds. It is also indicated for control of bleeding from the skin at percutaneous needle access, vascular access, and percutaneous catheter access sites. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Jiyoung Dang -S Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...