PASSPORT M SERIES PATIENT MONITORING (INCLUDING MODELS PASSPORT 17M AND PASSPORT 12M)
K132075 · Mindray North America · MHX · Apr 18, 2014 · Cardiovascular
Device Facts
Record ID
K132075
Device Name
PASSPORT M SERIES PATIENT MONITORING (INCLUDING MODELS PASSPORT 17M AND PASSPORT 12M)
Applicant
Mindray North America
Product Code
MHX · Cardiovascular
Decision Date
Apr 18, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Passport M Series patient monitors including models Passport 17M and Passport 12M are intended to be used for monitoring, displaying, reviewing, alarming and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, heart rate(HR)), respiration rate(RESP), temperature (TEMP), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation(SvO2/ScvO2), carbon dioxide (CO2), Oxygen(O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS) and respiration mechanics (RM). All the parameters can be applied for single adult, pediatric and neonatal patients with the exception of the follows: the arrhythmia detection, ST Segment analysis (Mortara algorithm), BIS, RM, CCO, SvO2/ScvO2 and PAWP monitoring are intended for adult and pediatric patients only. ST Segment analysis (Mindray algorithm) is intended for adult patients only. C.O. is only for use on adult patients. The ICG is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg. The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. It must only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
Device Story
Passport M Series (17M/12M) are compact, modular patient monitors derived from BeneView T Series platform. Devices collect physiological data via various sensors/probes (ECG leads, SpO2 sensors, pressure transducers, gas analyzers, etc.). Data processed to display real-time waveforms, numeric values, and alarms for clinical staff in healthcare facilities. Features include arrhythmia detection, ST-segment analysis (Mortara or Mindray algorithms), and advanced hemodynamic monitoring (CCO, SvO2/ScvO2). Output allows clinicians to assess patient status, adjust treatment, and monitor trends. Designed for professional use in clinical settings; not for transport or home use.
Clinical Evidence
Bench testing only. No clinical data presented. Compliance with recognized safety, performance, and electromagnetic compatibility standards verified through hardware/software validation, environmental testing, and risk analysis.
Technological Characteristics
Modular patient monitor platform; supports ECG, SpO2, NIBP, IBP, CO, CCO, SvO2/ScvO2, CO2, O2, AG, ICG, BIS, RM. Connectivity via standard patient monitor interfaces. Software-based processing for arrhythmia and ST-segment analysis. Form factor: compact/modular. Sterilization: N/A (non-sterile).
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare facilities for monitoring physiological parameters (ECG, HR, RESP, TEMP, SpO2, PR, NIBP, IBP, CO, CCO, SvO2/ScvO2, CO2, O2, AG, ICG, BIS, RM). Specific exclusions: Arrhythmia, ST analysis (Mortara), BIS, RM, CCO, SvO2/ScvO2, PAWP (adult/pediatric only); ST analysis (Mindray), CO, ICG (adult only). ICG restricted to adults 122-229cm, 30-155kg.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
BENEVIEW T SERIES PATIENT MONITORS, MODELS BENEVIEW T8, T6 AND T5 (K092449)
Solar 8000 and Transport Pro with Patient Data Module (K071073)
Submission Summary (Full Text)
{0}------------------------------------------------
## K132075
### Page 1 of 4
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K132075
### Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 8188 8998 Fax: +86 755 2658 2680
#### Contact Person: .
Ms. Yanhong Bai Manager Regulatory Affairs SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China Tel: +86 755 81885635 Fax: +86 755 26582680 E-mail: baiyanhong@mindray.com
- Date Prepared:
March 21, 2014
Name of the devices:
- Trade/Proprietary Name: . Passport M Series Patient Monitors (including models Passport 17M and Passport 12M)
- Common Name: Patient Monitor
{1}------------------------------------------------
K132075 Page 2 of 4
#### . Classification
| 21 CFR 870.1025 | Arrhythmia detector and alarm<br>(including ST-segment measurement and alarm) | Class II |
|-----------------|-------------------------------------------------------------------------------|----------|
| 21 CFR 870.1025 | Detector and Alarm, Arrhythmia | Class II |
| 21 CFR 870.1025 | Monitor, ST Segment with Alarm | Class II |
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometers -Temperature Monitor with Probe | Class II |
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 870.2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868.1500 | Enflurane gas analyzer | Class II |
| 21 CFR 868.1620 | Halothane gas analyzer | Class II |
| 21 CFR 868.1700 | Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868.1720 | Oxygen gas analyzer | Class II |
| 21 CFR 882.1400 | Electroencephalograph | Class II |
| 21 CFR 870.2770 | Impedance plethysmograph | Class II |
| 21 CFR 870.1435 | Single-function, preprogrammed diagnostic computer | Class II |
## Legally Marketed Predicate Devices:
| K092449, | BENEVIEW T SERIES PATIENT MONITORS, MODELS BENEVIEW T8, T6 AND |
|----------|---------------------------------------------------------------------------|
| | T5, Shenzhen Mindray Bio-medical Electronics Co., LTD. |
| K102562, | INTELLIVUE PATIENT MONITOR, MODELS MP2, X2, MP5, MP5T, MP20, MP30, |
| | MP40, MP50, MP60, MP70, MP80, MP90, & MX800, PHILIPS MEDIZINSYSTEME |
| | BOBLINGEN GMBH. |
| K091786, | PULSION CEVOX OPTICAL MODULE (PC3015), CEVOX DISPOSABLE PROBES |
| | (PV2022-30 THRU -38, PV2022-46 THRU -48), PULSION MEDICAL SYSTEMS |
| | AG |
| K071073, | Solar 8000 and Transport Pro with Patient Data Module, GE MEDICAL SYSTEMS |
| | INFORMATION TECHNOLOGIES |
{2}------------------------------------------------
K132075 Page 3 of 4
#### Device Description:
The Passport M Series Patient Monitors (including models Passport 17M and Passport 12M) are the developed new series based on the technical platform of the BeneView T Series patient monitors. The new series, much more compact and flexible with the appearance changes, are the sub-configuration models of the BeneView T Series patient monitors by adding some parameters monitoring functions, such as continuous cardiac output (CCO) and mixed /central venous oxygen saturation (SvO2/ScvO2) monitoring.
And comparing with the cleared BeneView T Series Patient Monitors using Mortara ECG algorithm, the subject patient monitors adopt Mindray Monitoring ECG Algorithm as an optional ECG algorithm to implement ECG monitoring, arrhythmia detection, and ST Segment Analysis.
#### Statement of intended Use:
The Passport M Series Patient Monitors including models Passport 17M and Passport 12M are intended to be used for monitoring, displaying, reviewing, alarming and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis), heart rate(HR)), respiration rate(RESP), temperature (TEMP), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), continuous cardiac output (CCO), mixed /central venous oxygen saturation (SvO2/ScvO2), carbon dioxide (CO2), Oxygen(02), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS) and respiration mechanics (RM).
All the parameters can be applied for single adult, pediatric and neonatal patients with the exception of the follows: the arrhythmia detection, ST Segment analysis (Mortara algorithm), BIS, RM, CCO, SvO2/ScvO2 and PAWP monitoring are intended for adult and pediatric patients only. ST Segment analysis (Mindray algorithm) is intended for adult patients only. C.O. is only for use on adult patients. The ICG is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. It must only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
{3}------------------------------------------------
K132075 Page 4 of 4
#### Technology:
Based on the technical platform of the BeneView T Series patient monitors, which were previously cleared by FDA under K092449, on AUG 31, 2010, we developed the Passport M Series Patient Monitors including models Passport 17M and Passport 12M.
### Test Summary:
The Passport M Series Patient Monitors (including models Passport 17M and Passport 12M) comply with the recognized safety, performance and electromagnetic compatibility standards. A risk analysis has been developed to identify potential hazards and document the mitigation of the hazards. Mindray's product development process required that the following activities be completed during the development of those patient monitors:
> Requirements specification review Hardware and Software testing Code design and code reviews Environmental EMC testing Safety testing Performance testing Hardware and Software validation
#### Conclusion:
The results of all testing demonstrate that the Passport M Series Patient Monitors (including models Passport 17M and Passport 12M) are as safe, as effective, and perform as well as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of a stylized caduceus-like design with three wavy lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
April 18, 2014
Mindray North America c/o Mr. Russell Olsen VP Quality & Regulatory Affairs 800 Macarthur Blvd Mahwah, NJ 07430 US
Re: K132075
> Trade/Device Name: Passport M Series Patient Monitoring (including models Passport 17M and Passport 12M) Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (With Arrhythmia Detection Or Alarms) Regulatory Class: Class II (two) Product Code: MHX Secondary Product Codes: DSI, MLD, DRT, DXN, DSK, FLL, DQA, DPZ, CCK, CBO. CBS, CBR, CCL, DSB, DXG, OLW Dated: March 13, 2014 Received: March 14, 2014
Dear Mr. Russell Olsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{5}------------------------------------------------
Page 2 - Mr. Russell Olsen
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Image /page/5/Picture/6 description: The image shows a logo with the letters FDA. The letters are stylized and have lines drawn through them. The logo is black and white and appears to be a graphic representation of the Food and Drug Administration.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# K132075
# Indications for Use
510(k) Number (if known):
Device Name: Passport M Series patient monitors
Indications for Use:
The Passport M Series patient monitors including models Passport 17M and Passport 12M are intended to be used for monitoring, displaying, reviewing, alarming and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, heart rate(HR)), respiration rate(RESP), temperature (TEMP), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation(SvO2/ScvO2), carbon dioxide (CO2), Oxygen(O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS) and respiration mechanics (RM).
All the parameters can be applied for single adult, pediatric and neonatal patients with the exception of the follows: the arrhythmia detection, ST Segment analysis (Mortara algorithm), BIS, RM, CCO, SvO2/ScvO2 and PAWP monitoring are intended for adult and pediatric patients only. ST Segment analysis (Mindray algorithm) is intended for adult patients only. C.O. is only for use on adult patients. The ICG is only for use on adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.
The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. It must only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.