← Product Code [OHV](/productcode/OHV) · K132028

# ULTHERA SYSTEM (K132028)

_Ulthera, Inc. · OHV · Dec 11, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132028

## Device Facts

- **Applicant:** Ulthera, Inc.
- **Product Code:** [OHV](/productcode/OHV.md)
- **Decision Date:** Dec 11, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4590
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to: - lift the eyebrow . - lift lax submental (beneath the chin) and neck tissue . The Ulthera System in conjunction with the Ulthera DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: - . ensure proper coupling of the transducer to the skin - confirm appropriate depth of treatment such as to avoid bone .

## Device Story

Ulthera System is a non-invasive dermatological device using focused ultrasound for aesthetic tissue lifting. System comprises control unit, handpiece, and DeepSEE transducers. Device provides ultrasonic visualization of dermal and subdermal tissue layers up to 8 mm depth. Clinician uses real-time imaging to ensure proper transducer-to-skin coupling and to confirm treatment depth, specifically to avoid bone. Output is visual display of tissue layers; clinician uses this to guide treatment placement. Benefits include non-invasive lifting of eyebrow, submental, and neck tissue.

## Clinical Evidence

No new clinical data or performance testing performed. Substantial equivalence supported by clinical literature and ultrasound images previously provided in K121700.

## Technological Characteristics

Focused ultrasound stimulator system for aesthetic use. Includes imaging subsystem for visualization of dermal/subdermal layers up to 8 mm. Components: control unit, handpiece, and DeepSEE transducers. Complies with medical electrical safety and electromagnetic compatibility standards.

## Regulatory Identification

A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Ulthera System ([K121700](/device/K121700.md))
- Cortex Technology, DermaScan C ([K983945](/device/K983945.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 5. 510(k) Summary

This 510(k) Summary for the Ulthera "System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.

| Applicant:           | Ulthera, Inc.                                                                                                                           |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Address:             | 1840 South Stapley Drive<br>Suite 200<br>Mesa, AZ 85204                                                                                 |
| Contact Person:      | Suzon Lommel, VP of Regulatory & Quality Affairs                                                                                        |
| Telephone:           | 480-619-4069                                                                                                                            |
| Fax:                 | 480-619-4071                                                                                                                            |
| Submission Date:     | June 28, 2013                                                                                                                           |
| Device Trade Name:   | Ulthera® System                                                                                                                         |
| Common Name:         | Focused Ultrasound For Tissue Heat Or Mechanical Cellular Disruption<br>System, Imaging, Pulsed Echo, Ultrasonic                        |
| Classification:      | Regulatory Class II                                                                                                                     |
| Classification Name: | Focused Ultrasound Stimulator Use System for Aesthetic Use                                                                              |
| Regulation:          | 878.4590                                                                                                                                |
| Product Code:        | OHV<br>IYO                                                                                                                              |
| Legally Marketed     | Name: Ulthera® System                                                                                                                   |
| Predicates:          | 510(k):<br>Ulthera, Inc., Ulthera System - K121700<br>Cortex Technology, DermaScan C - K983945                                          |
| Applicable Guidance: | The Class II Special Controls Guidance Document: Focused Ultrasound<br>Stimulator System for Aesthetic Use was developed in response to |

Image /page/0/Picture/5 description: The image shows the word "Ulthera" in a stylized font. The "U" and "l" are tall and thin, while the rest of the letters are shorter and wider. There are three small circles above the "h" in "Ulthera". The word "Ulthera" is in black, and the background is white.

{1}------------------------------------------------

Ulthera's DeNovo submission and 510(k) clearance K072505 for the Ulthera System.

Guidance for Industry and FDA Staff: Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers

Device Description:

The Ulthera System consists of the following components:

- . Ulthera Control Unit
- Handpiece .
- Transducers .

#### Indications for Use:

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:

- lift the eyebrow (current cleared indication) .
- lift lax submental (beneath the chin) and neck tissue (cleared . indication)

The Ulthera System in conjunction with the Ulthera DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:

- ensure proper coupling of the transducer to the skin (requested . indication)
- confirm appropriate depth of treatment such as to avoid bone . (requested indication)

No new performance testing was performed to support the new Performance Data: indications. There have been no changes to previously cleared device (K121700, Ulthera® System) that would result in a new 510(k). Imaging capabilities of the Ulthera® System have been included in the device's profile since the original clearance. Clinical literature and ultrasound images were provided.

There have been no changes to previously cleared Ulthera® System Conclusion: (K121700) that would result in a new 510(k).

> The Ulthera System's imaging subsystem and the DermaScan C are equivalent in terms of patient contact material biocompatibility, electromagnetic compatibility, and medical electrical safety standards. The Ulthera System's imaging subsystem is equivalent in terms of general

Page | 05-2

{2}------------------------------------------------

specifications, device settings and intended use to the predicate device, DermaScan C (K983945).

Based on the design, materials, principle of operation, and intended use, the Ulthera® System is substantially equivalent to the legally marketed predicate devices.

Image /page/2/Picture/2 description: The image shows the word "ulthera" in a stylized font. The "u" and "l" are represented by three vertical lines, and the rest of the word is in a sans-serif font. Above the "thera" portion of the word are three small circles. The letters are bolded and have a textured appearance. 

Page 3 of 3

Page| 05-3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a human figure with outstretched arms.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002

Ulthera Incorporated Ms. Suzon Lommel Vice President of Regulatory & Quality Affairs 1840 South Stapley Drive, Suite 200 Mesa, Arizona 85204

December 11, 2013

Re: K132028

Trade/Device Name: Ulthera® System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused ultrasound stimulator system for aesthetic use Regulatory Class: Class II Product Code: OHV, IYO Dated: October 25, 2013 Received: October 28, 2013

Dear Ms. Lommel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Suzon Lommel

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### 4. Indications for Use Statement

510(k) Number: Unknown

Device Name: Ulthera® System

Indications for Use:

The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:

- lift the eyebrow .
- lift lax submental (beneath the chin) and neck tissue .

The Ulthera System in conjunction with the Ulthera DeepSEE transducer allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:

- . ensure proper coupling of the transducer to the skin
- confirm appropriate depth of treatment such as to avoid bone .

| Prescription Use            | ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ |  |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| (Part 21 CFR 801 Subpart D) |                                                                                                                                                                               |  |

AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C)

(Please do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |         |
|--------------------------------------------------------|---------|
| Long H. Chen - A                                       | for BSA |
| (Division Sign-off)                                    |         |
| Division of Surgical Devices                           |         |
| 510(k) Number:                                         | K132028 |

Page| 04-1

---

**Source:** [https://fda.innolitics.com/device/K132028](https://fda.innolitics.com/device/K132028)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
