← Product Code [MJN](/productcode/MJN) · K132022

# LEMAITRE 8F OCCLUSION CATHETER (K132022)

_LeMaitre Vascular, Inc. · MJN · Nov 6, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K132022

## Device Facts

- **Applicant:** LeMaitre Vascular, Inc.
- **Product Code:** [MJN](/productcode/MJN.md)
- **Decision Date:** Nov 6, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The LeMaitre 8F Occlusion Catheter is indicated for temporary vessel occlusion.

## Device Story

LeMaitre 8F Occlusion Catheter is a single-lumen catheter featuring a latex balloon (28 mm or 45 mm sizes). Used by physicians in clinical settings for temporary vessel occlusion. Device includes a stopcock for balloon inflation maintenance and radiopaque marker bands at proximal and distal balloon ends for fluoroscopic visualization. Operation involves inflating the balloon via the inflation lumen to achieve vessel occlusion. Benefits include temporary flow control during surgical procedures.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included evaluation of inflation volume, inflation pressure, radial force, contact area, inflation/deflation time, and burst volume compared to the predicate device.

## Technological Characteristics

Single-lumen catheter with latex balloon. Radiopaque marker bands for fluoroscopy. Ethylene oxide (EO) sterilization per ANSI/AAMI/ISO 11135-1:2007. Biocompatibility per ISO 10993 for limited contact (<24 hours) with circulating blood.

## Regulatory Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

## Predicate Devices

- Fogarty Occlusion Catheter ([K093911](/device/K093911.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510 (k) Summary

#### Submitter's information:

| Name:                        | LeMaitre Vascular, Inc.                      |
|------------------------------|----------------------------------------------|
| Address:                     | 63 Second Avenue<br>Burlington, MA USA 01803 |
| Phone:                       | 781-425-1727                                 |
| Fax:                         | 781-425-5049<br>NOV 0 6 2013                 |
| Contact Person:              | Bryan Cowell, MSc., RAC                      |
| Date of preparation:         | June 24, 2013                                |
| Device Name:                 | LeMaitre 8F Occlusion Catheter               |
| Trade Name.                  | LeMaitre 8F Occlusion Catheter               |
| Common/ Classification Name: | Catheter, intravascular occluding, temporary |
| Classification Panel:        | 21CFR §870.4450                              |
| Class:                       | II (2)                                       |
| Product Code:                | MJN                                          |

### Establishment Registration: 1220948

Establishment: LeMaitre Vascular, Inc., 63 Second Avenue, Burlington, MA USA 01803

Owner/Operator: 1220948

### Device Description:

The 8F Occlusion Catheters is offered in two balloon sizes; 28 mm or 45mm. It is a single lumen catheter with a latex balloon specifically designed and sized for use in the outlined general procedures. The single lumen (inflation lumen indicated by the white stopcock) is used for balloon inflation. The stop-cock is to maintain balloon inflation level throughout the procedure. The device has radiopaque marker bands at the proximal and distal ends of the balloon to enhance visibility of the balloon location when used under fluoroscopy.

### Intended Use:

The LeMaitre 8F Occlusion Catheter is indicated for temporary vessel occlusion.

{1}------------------------------------------------

## Predicate Device:

1

K093911 510(k): Device Name: Fogarty Occlusion Catheter

## Substantial Equivalence:

### Fundamental Scientific Technological Characteristics:

The LeMaitre 8F Occlusion Catheter maintains the same intended use and fundamental scientific technology as the predicate device.

## Functional/ Safety testing:

The verification activities conducted indicate that LeMuitre 8F Occlusion Catheter device meets the product performance requirements of the device specifications and does not raise any additional safety issues.

#### Sterilization:

The device is validated for ethylene oxide (EO) sterilization according to ANSI/AAMI/ISO 11135-1:2007, "Sterilization of Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization".

## Biocompatibility:

All blood contact portions of the device were subjected to biocompatibility testing according to ISO 10993 guidelines for an externally communicating device with limited contact duration (<24 hours), with circulating blood. The biocompatibility assessment established that 8F Occlusion Catheter is biocompatible.

## Summary of Product Testing:

The following tests have been completed to evaluate the safety and performance of LeMaitre 8F occlusion catheter compared with the predicate device:

- Volume Required to Inflate Balloon to IFU Indicated Diameter �
- . Inflation Pressure
- Radial Force .
- Contact Area o
- . Inflation Time
- Deflation Time .
- . Burst Volume

#### Conclusion:

LeMaitre Vascular has demonstrated that the LeMaitre 8F Occlusion Catheter is substantially equivalent to the predicate device based on its intended use and fundamental scientific technology.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

November 6, 2013

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

LeMaitre Vascular, Inc. c/o Mr. Xiang (Vic) Zhang Director of Regulatory Affairs 63 Second Ave. Burlington, MA 01803

Re: K132022

Trade/Device Name: LeMaitre 8F Occlusion Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: October 22, 2013 Received: October 29, 2013

Dear Mr. Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Xiang (Vic) Zhang

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

# Sincerely vours. Bram D. Zuckerman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

| 510(k) Number (if known) | K132022/S002                                                                |
|--------------------------|-----------------------------------------------------------------------------|
| Device Name              | LeMaitre 8F Occlusion Catheter.                                             |
| Indications              | LeMaitre 8F Occlusion Catheter is indicated for temporary vessel occlusion. |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

.............................................................................................................................................................................

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801. 109)

for Use

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K132022](https://fda.innolitics.com/device/K132022)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
