Retrospective clinical samples were used to supplement the performance evaluation of the VIDAS C. difficile GDH assay in pediatric patients (2-12 years) where prospective enrollment was limited.
Retrospective; Pediatric; Clinical samples
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis
Pediatric patients (2-12 years) submitted for routine C. difficile testing; Sample Size: 13
Not applicable for this study
Sensitivity and specificity versus CCFA bacterial culture
Indications for Use
VIDAS® C. difficile GDH (GDH) is an automated test based on the Enzyme linked Fluorescent Assay technique (ELFA), for use on the VIDAS family instruments. The VIDAS C. difficile GDH (glutamate dehydrogenase) assay is a qualitative test that detects the C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile in fecal specimens from persons suspected of having C. difficile infection (CDI). The test does not distinguish toxigenic from non-toxigenic strains of C. difficile. With the use of additional tests that detect C. difficile toxins, the test is to be used as an aid in the diagnosis of C. difficile infection. As with other C. difficile tests, results should be considered in conjunction with the patient history.
Device Story
VIDAS C. difficile GDH is an automated, qualitative enzyme-linked fluorescent assay (ELFA) for detecting C. difficile glutamate dehydrogenase (GDH) antigen in fecal specimens. The device uses a Solid Phase Receptacle (SPR) that acts as both a pipetting device and a solid phase coated with mouse monoclonal anti-GDH antibodies. The instrument automatically performs a two-step sandwich immunoassay: sample incubation, binding with alkaline phosphatase-conjugated anti-GDH antibody, and washing. The final step involves enzymatic hydrolysis of 4-methylumbelliferyl phosphate into a fluorescent product measured at 450 nm. Fluorescence intensity correlates with GDH quantity. The system automatically analyzes results and prints a report. Used in clinical laboratories to screen for C. difficile; results aid clinicians in diagnosing CDI when combined with patient history and toxin-detection assays. Benefits include automated processing, reducing manual labor and variability in screening for C. difficile presence.
Clinical Evidence
Clinical performance evaluated using 1,904 stool samples (1,891 prospective, 13 retrospective) from patients suspected of CDI. Compared against CCFA bacterial culture. Overall sensitivity 95.8% (95% CI: 92.8–97.7%) and specificity 90.0% (95% CI: 88.4–91.4%). Negative predictive value 99.1%. Performance across age groups (including pediatric) was consistent with overall findings. Method comparison with predicate showed 97.3% positive percent agreement and 94.4% negative percent agreement.
Indicated for screening fecal specimens for C. difficile antigen (glutamate dehydrogenase) in patients suspected of having C. difficile infection (CDI). Applicable to all ages. Does not distinguish toxigenic from non-toxigenic strains; must be used with additional toxin-detection tests as an aid in diagnosis.
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K132010
B. Purpose for Submission:
To obtain a substantial equivalence determination for a new device
C. Measurand:
Clostridium difficile antigen, glutamate dehydrogenase (GDH)
D. Type of Test:
Qualitative enzyme linked fluorescent assay (ELFA)
E. Applicant:
bioMérieux, SA
F. Proprietary and Established Names:
VIDAS® C. difficile GDH Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.2660 - Microorganism Differentiation and Identification Device
2. Classification:
Class I
3. Product code:
MCB – Antigen, C. difficile
4. Panel:
Microbiology (83)
{2}
H. Intended Use:
1. Intended use(s):
VIDAS® C. difficile GDH (GDH) is an automated test based on the Enzyme Linked Fluorescent Assay technique (ELFA), for use on the VIDAS family instruments. The VIDAS C. difficile GDH (glutamate dehydrogenase) assay is a qualitative test that detects the C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile in fecal specimens from persons suspected of having C. difficile infection (CDI). The test does not distinguish toxigenic from nontoxigenic strains of C. difficile. With the use of additional tests that detect C. difficile toxins, the test is to be used as an aid in the diagnosis of C. difficile infection. As with other C. difficile tests, results should be considered in conjunction with the patient history.
2. Indication(s) for use:
VIDAS® C. difficile GDH (GDH) is an automated test based on the Enzyme Linked Fluorescent Assay technique (ELFA), for use on the VIDAS family instruments. The VIDAS C. difficile GDH (glutamate dehydrogenase) assay is a qualitative test that detects the C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile in fecal specimens from persons suspected of having C. difficile infection (CDI). The test does not distinguish toxigenic from nontoxigenic strains of C. difficile. With the use of additional tests that detect C. difficile toxins, the test is to be used as an aid in the diagnosis of C. difficile infection. As with other C. difficile tests, results should be considered in conjunction with the patient history.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
VIDAS and mini-VIDAS instruments
I. Device Description:
Each VIDAS® C. difficile GDH kit contains 60 tests. The kit is comprised of the reagent strip (STR), the solid phase receptacle (SPR), the controls (C1 and C2), the standard (S1), the Pretreatment Reagent (R1), and the Master Lot Entry (MLE) card.
The interior of the SPR is coated during production with mouse monoclonal anti- C. difficile GDH antibody. Each SPR is identified by the code "GDH" code.
The reagent strip consists of 10 wells covered with a labeled, foil seal. The label contains a bar code which includes the assay code, kit lot number and expiration date. The foil of the first well is perforated to facilitate the introduction of the sample. The last well of each strip
2
{3}
is a cuvette in which the fluorometric reading is performed. The eight wells in the center section of the strip contain the various reagents required for the assay.
Description of the GDH reagent strip:
| Wells | Reagents |
| --- | --- |
| 1 | Sample well. |
| 2 - 3 - 4 | Wash solution: Buffer 0.2 mol/L (pH 7.8) + detergent + preservatives (600 μL). |
| 5 | Conjugate: mouse monoclonal anti-C. difficile GDH antibody labeled with ALP + Protein stabilizer + preservative (400 μL). |
| 6 - 7 - 8 - 9 | Wash solution: Buffer 0.2 mol/L (pH 7.8) + detergent + preservatives (600 μL). |
| 10 | Reading cuvette with substrate: 4-Methyl-umbelliferyl phosphate (0.6 mmol/L) + diethanolamine DEA* (0.62 mol/L or 6.6%) pH 9.2 + sodium azide 1g/L (300 μL). |
J. Substantial Equivalence Information:
1. Predicate device name(s):
Techlab C. Diff QUIK CHECK®
2. Predicate 510(k) number(s):
K053572
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device
VIDAS® C. difficile GDH Assay | Predicate
C. DIFF QUIK CHEK®
Assay (K053572) |
| Intended Use | VIDAS® C. difficile GDH (GDH) is an automated test based on the Enzyme Linked Fluorescent Assay technique (ELFA), for use on the VIDAS family instruments. The VIDAS C. difficile GDH (glutamate dehydrogenase) assay is a qualitative test that detects the C. difficile antigen, glutamate dehydrogenase, as a screen for the presence | The C. DIFF QUIK CHEK™ test is a rapid membrane enzyme immunoassay for use as a screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from persons suspected of having C. difficile disease. The test does not distinguish toxigenic from |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | DeviceVIDAS® C. difficile GDHAssay | PredicateC. DIFF QUIK CHEK®Assay (K053572) |
| | of C. difficile in fecal specimens from persons suspected of having C. difficile infection (CDI). The test does not distinguish toxigenic from nontoxigenic strains of C. difficile. With the use of additional tests that detect C. difficile toxins, the test is to be used as an aid in the diagnosis of C. difficile infection. As with other C. difficile tests, results should be considered in conjunction with the patient history. | nontoxigenic strains of C. difficile. With the use of additional tests that detect C. difficile toxins, the test is to be used as an aid in the diagnosis of C. difficile disease. As with other C. difficile tests, results should be considered in conjunction with the patient history. |
| Analyte | C. difficile glutamate dehydrogenase antigen | C. difficile glutamate dehydrogenase antigen |
| Interpretation | Qualitative | Qualitative |
| Differences | | |
| --- | --- | --- |
| Item | DeviceVIDAS® C. difficile GDHAssay | PredicateC. DIFF QUIK CHEK®Assay (K053572) |
| Automated | Yes | No |
| Assay Technique | Enzyme-linked fluorescent assay (ELFA) | Rapid Membrane Enzyme Immunoassay |
| Assay time | Approximately 50 minutes | Approximately 25 minutes |
| Reading method | Automated | Visual |
| Specimen type | Human stool (unpreserved) | Human stool (unpreserved and preserved in Cary Blair or C&S transport media) |
| Specimen volume | 200 μL | 25 μL unpreserved stool, 100 μL Cary Blair or C&S transport media stool |
| Detection antibodies | Alkaline phosphatase labeled monoclonal anti-C.difficile GDH antibody | Horseradish peroxidase labeled monoclonal anti-C.difficile GDH antibody |
| Capture antibodies | Monoclonal anti-C.difficile GDH antibody | Polyclonal anti-C.difficile GDH antibody |
{5}
5
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
CLSI EP07-A2: Interference Testing in Clinical Chemistry
CLSI EP12-A2: User Protocol for Evaluation of Qualitative Test Performance
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
CLSI EP24-A2: Assessment of the Diagnostic Accuracy of Laboratory Tests Using ROC Curves
CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents
L. Test Principle:
The VIDAS® C. difficile GDH assay principle combines a two-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA).
The SPR serves as the solid phase as well as the pipetting device. Reagents for the assay are ready-to-use and are pre-dispensed in the sealed reagent strips. All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times. Each step is followed by a wash cycle which eliminates unbound components.
- Specific binding of GDH present in the sample with mouse monoclonal anti-GDH antibody coated on the interior of the SPR.
- Binding between GDH and mouse monoclonal anti-GDH antibody conjugated with alkaline phosphatase (ALP).
- Detection: alkaline phosphatase catalyzes the hydrolysis of the substrate (4-Methyl-umbelliferyl phosphate) into a fluorescent product (4-Methyl-umbelliferone) the fluorescence of which is measured at 450 nm.
The intensity of the fluorescence increases according to the quantity of GDH in the sample.
When the VIDAS C. difficile GDH test is completed, the results are analyzed automatically by the instrument, a test value is generated, and a result is printed for each sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The within-laboratory precision was estimated at one site based on the recommendations of the CLSI EP5-A2. Three human samples, including two close to the clinical cut-off (one high negative and one low positive), were tested in triplicate
{6}
in two runs per day with two different operators, with two reagent lots for a total of 12 testing days (six days of test per lot) on one VIDAS instrument (N=72 test values for each sample). Two calibrations were used for each reagent lot (three days of test per calibration and lot) over the whole period of the study. Data from the study are summarized in the following table:
| Sample | N | Mean test value | Repeatability | | Total within-laboratory precision (total within-instrument, between-lot, between-calibration) | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Standard deviation | CV (%) | Standard deviation | CV (%) |
| Sample 1 High negative | 72 | 0.07 | 0.00 | 6.0 | 0.01 | 14.1 |
| Sample 2 Low positive | 72 | 0.12 | 0.01 | 5.2 | 0.01 | 11.9 |
| Sample 3 Moderate positive | 72 | 0.27 | 0.02 | 5.7 | 0.03 | 11.2 |
The within-laboratory precision of each panel member was also analyzed by determining the percentage of agreement between the test interpretation and the expected outcome (negative/positive interpretation). There was no change of interpretation for the 3 panel samples tested: all replicates of each panel member resulted in the expected interpretation. Data from the qualitative analysis are summarized in the following table:
| Sample | Expected Result | N | Observed result Lot 1 | | Observed result Lot 2 | | Total Agreement | [CI95] % |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Neg | Pos | Neg | Pos | | |
| Sample 1 High negative | Negative | 72 | 36 | 0 | 36 | 0 | 72/72 (100.0%) | [95.0 - 100.0]% |
| Sample 2 Low positive | Positive | 72 | 0 | 36 | 0 | 36 | 72/72 (100.0%) | [95.0 - 100.0]% |
| Sample 3 Moderate positive | Positive | 72 | 0 | 36 | 0 | 36 | 72/72 (100.0%) | [95.0 - 100.0]% |
The reproducibility was estimated at three sites based on the recommendations of the CLSI EP5-A2. Three human samples, including two close to the clinical cut-off (one high negative and one low positive), were tested in triplicate in two runs per day with two different operators, using two reagent lots for a total of six testing days (three days of test for each lot) on three VIDAS instruments at three different sites (N=108 test values for each sample). One calibration was used for each reagent lot over the
{7}
whole period of the study. Data from the study are summarized in the following table:
| Sample | N | Mean test value | Reproducibility (total between sample preparation/operator/run/day/lot/instrument) | |
| --- | --- | --- | --- | --- |
| | | | Standard deviation | CV (%) |
| Sample 1 High negative | 108 | 0.06 | 0.01 | 19.1 |
| Sample 2 Low positive | 108 | 0.12 | 0.02 | 12.9 |
| Sample 3 Moderate positive | 108 | 0.26 | 0.03 | 13.0 |
The reproducibility of each panel member was also analyzed by determining the percentage of agreement between the test interpretation and the expected outcome (negative/positive interpretation). Data from the qualitative analysis for all sites combined are summarized in the following table:
| Panel Sample | Expected Result | N | Observed result Site 1 | | Observed result Site 2 | | Observed result Site 3 | | Total Agreement | [CI95] % |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Neg | Pos | Neg | Pos | Neg | Pos | | |
| Sample 1 High negative | Negative | 108 | 36 | 0 | 36 | 0 | 35 | 1 | 107/108 (99.1%) | [94.95 - 99.98]% |
| Sample 2 Low positive | Positive | 108 | 0 | 36 | 1 | 35 | 3 | 33 | 104/108 (96.3%) | [90.79 - 98.98]% |
| Sample 3 Moderate positive | Positive | 108 | 0 | 36 | 0 | 36 | 0 | 36 | 108/108 (100%) | [96.64 - 100]% |
Out of the 108 results obtained for each precision sample, there were:
- 1 change of interpretation $(0.9\%)$ for the high negative sample (Sample 1),
- 4 changes of interpretation $(3.7\%)$ for the low positive sample (Sample 2),
- 0 change of interpretation $(0\%)$ for the moderate positive sample (Sample 3).
The percentages of change of interpretation observed for Sample 1 and Sample 2 were less than $5\%$ , which was considered as normal and expected for these types of samples very close to the assay decision threshold (average test value for Sample $1 = 0.06$ and average test value for Sample $2 = 0.12$ for a decision threshold at 0.10.
# b. Linearity/assay reportable range:
Not applicable
{8}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls:
One positive control (C1) and one negative control (C2) are included in each VIDAS C. difficile GDH kit. These controls must be performed each time a new lot of reagents is opened to ensure that reagent performance has not been altered. Each calibration must be checked using these controls. Results cannot be validated if the control values deviate from the expected values. The expected value range for each control is indicated on the Master Lot Entry (MLE) Card that is provided with the kit.
Calibrators:
Calibration, using the standard (S1) provided in the kit, must be performed each time a new lot of reagents is opened, after the master lot data have been entered. Calibration should then be performed every 28 days. The standard value must be within the set RFV "Relative Fluorescence Value" range as indicated on the MLE card. If this is not the case, recalibrate.
Sample stability:
Unpreserved stool specimens may be stored at 2-8°C for 3 days (from time of collection) prior to processing. If longer storage is required, freeze the specimens at -70°C +/- 10°C up to one month. Avoid repeated freezing and thawing cycles and storage at -19/-31°C.
Do not use containers which may contain detergents, preservatives or media that may interfere with the VIDAS C. difficile GDH assay results.
Samples stored in Cary Blair or C&S Transport media, or preserved 10% formalin, merthiolate formalin, sodium acetate formalin, or polyvinyl alcohol media have not been evaluated for use with the VIDAS C. Difficile GDH Assay.
Processed specimen supernatants, processed standard and reconstituted controls may be stored up to 8 hours at 18-25°C or 48 hours at 2-8°C before being tested with the VIDAS C. difficile GDH assay. Specimen supernatants and processed standard and control storage at -19/-31°C and -70°C +/-10°C was not validated and is therefore not recommended.
High dose hook effect:
No hook effect was observed up to purified native GDH antigen concentrations of 2 µg/mL.
8
{9}
d. Detection limit:
The limit of detection was evaluated using a range of dilutions of purified native C.difficile GDH C. difficile GDH in a pool of C. difficile-negative stool samples based on the recommendations of the CLSI EP17-A. The limit of detection of the VIDAS C. difficile GDH assay (at least 95% detection rate for positive samples) is 3.0 ng/mL for purified native GDH antigen.
e. Analytical specificity:
Cross reactivity:
To test for cross-reactivity, each micro-organism was diluted in a pool of C. difficile-negative stool samples, pretreated and a single replicate was tested using the VIDAS C. difficile GDH assay. To test for microbial interference, each micro-organism was diluted in a pool of C. difficile-positive stool samples, pretreated and a single replicate was tested using the VIDAS C. difficile GDH assay. The micro-organisms were tested at a concentration of 3×10^{8} CFU/mL (1 McFarland) for bacteria and 1×10^{5} PFU/mL for viruses.
None of the following micro-organisms, present in the stool samples, reacted with the VIDAS C. difficile GDH assay:
Abiotrophia defectiva, Acinetobacter baumannii, Acinetobacter lwoffii, Aeromonas hydrophila ssp hydrophila, Alcaligenes faecalis ssp faecalis, Anaerococcus tetradius, Bacillus cereus, Bacteroides caccae, Bacteroides merdae, Bacteroides stercoris, Bifidobacterium adolescentis, Bifidobacterium longum, Campylobacter jejuni ssp jejuni, Candida albicans, Candida catenulata, Cedecea davisae, Chlamydia trachomatis, Citrobacter amalonaticus, Citrobacter freundii, Citrobacter koseri, Citrobacter sedlakii, Clostridium nexile, Clostridium beijerinckii, Clostridium bifermentans, Clostridium bolteae, Clostridium butyricum, Clostridium chauvoei, Clostridium fallax, Clostridium haemolyticum, Clostridium histolyticum, Clostridium innocuum, Clostridium novyi, Clostridium paraputrificum, Clostridium perfringens, Clostridium ramosum, Clostridium scindens, Clostridium septicum, Clostridium sordellii, Clostridium sphenoides, Clostridium spiroforme, Clostridium sporogenes, Clostridium symbosum, Clostridium tertium, Clostridium tetani, Collinsella aerofaciens, Corynebacterium genitalium, Desulfovibrio piger, Edwardsiella tarda, Eggerthella lenta, Enterobacter aerogenes, Enterobacter cloacae, Enterococcus casseliflavus, Enterococcus cecorum, Enterococcus dispar, Enterococcus faecalis, Enterococcus faecium, Enterococcus gallinarum, Enterococcus hirae, Enterococcus raffinosus, Escherichia coli, Escherichia fergusonii, Escherichia hermannii, Flavonifractor plautii, Fusobacterium varium, Gardnerella vaginalis, Gemella morbillorum, Hafnia alvei, Helicobacter fenneliae, Helicobacter pylori, Klebsiella oxytoca, Klebsiella pneumoniae ssp pneumoniae, Lactobacillus acidophilus, Lactobacillus reuteri, Lactococcus lactis ssp lactis, Leminorella grimontii, Listeria grayi, Listeria innocua, Listeria monocytogenes, Peptoniphilus asaccharolyticus,
{10}
Peptostreptococcus anaerobius, Plesiomonas shigelloides, Porphyromonas asaccharolytica, Prevotella melaninogenica, Proteus mirabilis, Proteus penneri, Providencia alcalifaciens, Providencia rettgeri, Providencia stuartii, Pseudomonas aeruginosa, Pseudomonas putida, Salmonella enterica ssp arizonae, Salmonella ser.Choleraesuis, Salmonella ser.Typhimurium Serratia liquefaciens, Serratia marcescens, Shigella boydii, Shigella dysenteriae, Shigella sonnei, Staphylococcus aureus ssp aureus, Staphylococcus epidermidis, Stenotrophomonas maltophilia, Streptococcus agalactiae, Streptococcus dysgalactiae ssp dysgalactiae, Streptococcus intermedius, Streptococcus uberis, Trabulsiella guamensis, Veillonella parvula, Vibrio cholerae, Vibrio parahaemolyticus, Yersinia bercovieri, Yersinia rohdei, Adenovirus 40 et 41, Rotavirus RF, Norovirus, Enterovirus 70, Echovirus 12, Coxsackie virus, Cytomegalovirus AD169.
# Interference study:
Potential interference by commonly used drugs and other substances was determined based on the recommendations of the CLSI® EP7-A2, at 2 levels of GDH (a low positive close to the clinical cut-off and a high positive).
| Tested compound | Highest concentration tested | Result |
| --- | --- | --- |
| Hemoglobin | 3.2 mg/mL | No significant interference |
| Lipids | 20 mg/mL | No significant interference |
| Mucin | 3.33 mg/mL | No significant interference |
| Amoxicillin | 206 μmol/L | No significant interference |
| Bismuth salicylate | 8.2 mg/mL | No significant interference |
| Calcium carbonate | 13.06 mg/mL | Potential interference* |
| Ceftriaxone | 1.46 mmol/L | No significant interference |
| Benzalkonium chloride | 2 μg/mL | No significant interference |
| Ciprofloxacin | 30.2 μmol/L | No significant interference |
| Erythromycin | 81.6 μmol/L | No significant interference |
| Ethanol | 86.8 mmol/L | No significant interference |
| Fidaxomicin | 4 mg/mL | No significant interference |
| Gentamicin | 21 μmol/L | No significant interference |
| Mineral oil | 0.27 v/v | Potential interference* |
| Hydrocortisone | 0.6 mg/mL | No significant interference |
| Aluminium hydroxide | 15.3 mg/mL | No significant interference |
| Magnesium | 6.2 mg/mL | No significant interference |
| Nitric oxide | 0.2 mg/mL | Potential interference* |
| Nitric oxide | 0.01 mg/mL | Potential interference |
{11}
| hydroxide | | |
| --- | --- | --- |
| Lidocaine | 0.12 mg/mL | No significant interference |
| Loperamide | 0.08 mg/mL | No significant interference |
| Mesalazine | 19.2 mg/mL | No significant interference |
| Metronidazole | 2 mg/mL | No significant interference |
| Naproxen | 2170 μmol/L | No significant interference |
| Nystatin | 600 UI/mL | No significant interference |
| Phenylephrine | 0.225 mg/mL | No significant interference |
| Sennosides | 0.24 mg/mL | No significant interference |
| Tergitol (nonoxynol-9) | 0.5 v/v | Potential interference* |
| Tetracycline | 34 μmol/L | No significant interference |
| Vancomycin | 5 mg/mL | No significant interference |
* Calcium carbonate at 9.80 mg/mL, Mineral oil at 0.20 v/v and Tergitol at 0.125 v/v did not cause interference.
# Strain reactivity:
The VIDAS C. difficile GDH assay was evaluated using several strains of C. difficile. The VIDAS C. difficile GDH detects the following C. difficile strains at the tested concentrations of $9 \times 10^{8}$ CFU/mL (3 McFarland) and $3 \times 10^{6}$ CFU/mL:
| Toxinogenic C.difficile strains : | | Non toxinogenic C.difficile strains: |
| --- | --- | --- |
| ATCC 43255™ | ATCC 43600™ | ATCC 700057™ |
| ATCC9689™ | ATCC 43599™ | ATCC 43593™ |
| ATCC 700792™ | ATCC 43596™ | X1a IS58 |
| ATCC 17858™ | ATCC 43594™ | X1b R1 1402 |
| ATCC BAA-1805™ | ATCC 17857™ | ATCC 43601™ (3x108 |
| ATCC BAA-1382™ | ATCC 43598™ | CFU/mL only) |
| ATCC 51695™ | CCUG 20309 | |
{12}
The VIDAS C. difficile GDH detects the following C.dificile strains at the tested concentration of $9 \times 10^{8} \mathrm{CFU/mL}$ (3 McFarland):
| Cardiff ECDC collection including the following ribotypes | 001 (7 strains); 002; 003; 012; 014; 015; 017; 020; 023; 027; 029; 046; 053; 056; 070; 075; 077; 078; 081; 087; 095; 106; 126; 131; VPI 10463; 005; 010; 045; 048; 156; 174. |
| --- | --- |
| bioMerieux collection including the following ribotypes | 001 (6 strains); 002 (9 strains); 005 (2 strains); 010 (1 strain); 012 (4 strains); 014 (10 strains); 015 (1 strain); 017 (20 strains); 020 (5 strains); 023 (1 strain); 027 (24 strains); 047 (1 strain); 050 (1 strain); 053 (4 strains); 054 (2 strains); 056 (2 strains); 057 (1 strain); 058 (1 strain); 075 (1 strain); 078 (3 strains); 096 (1 strain); 097 (1 strain); 103 (2 strains); 106 (16 strains); 110 (2 strains); 118 (1 strain); 153 (1 strain); 177 (1 strain). |
# f. Assay cut-off:
The clinical cut-off was established at a Test Value of 0.10. Interpretation according to the Test Value is as follows:
| Test Value | Result | Interpretation |
| --- | --- | --- |
| < 0.10 | Negative | No detectable C. difficile GDH antigen. |
| ≥ 0.10 | Positive | Presumptive detection of C. difficile GDH antigen. The specimen must be tested with additional tests that detect C. difficile toxins. |
# 2. Comparison studies:
a. Method comparison with predicate device:
# Method comparison with a commercially available C. difficile GDH assay
One thousand nine hundred and four (1904, 1891 prospective and 13 retrospective) stool samples collected from patients suspected of having $C$ . difficile infection (CDI) were tested at three sites (USA and Europe). A single replicate of each sample was tested using VIDAS $C$ . difficile GDH on a VIDAS instrument and a commercially available $C$ . difficile GDH assay. Data from the study are summarized in the following table:
{13}
Method comparison between the VIDAS C. difficile GDH assay and the commercially available C. difficile GDH assay on prospective samples
| | Commercially available C. difficile GDH assay | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 (EU) | | Site 2 (US) | | Site 3 (US) | | Total (All Sites) | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| VIDAS Pos | 56 | 10 | 92 | 7 | 207 | 69 | 355 | 86 |
| C. difficile GDH Neg | 4 | 454 | 2 | 365 | 4 | 621 | 10 | 1440 |
| Total | 60 | 464 | 94 | 372 | 211 | 690 | 365 | 1526 |
| Performance | % | [CI95%] | % | [CI95%] | % | [CI95%] | % | [CI95%] |
| Positive Percent Agreement | 93.3% | [83.8 - 98.2]% | 97.9% | [92.5 - 99.7]% | 98.1% | [95.2 - 99.5]% | 97.3% | [95.0 - 98.7]% |
| Negative Percent Agreement | 97.8% | [96.1 - 99.0]% | 98.1% | [96.2 - 99.2]% | 90.0% | [87.5 - 92.1]% | 94.4% | [93.1 - 95.5]% |
In order to better estimate the performance of the VIDAS C. difficile GDH assay in specimens from pediatric patients (2-12 years), thirteen (13) C. difficile retrospectively collected samples were tested according to the same protocol. For all 1904 specimens and all sites combined, the VIDAS C. difficile GDH assay demonstrated a positive percent agreement of $97.3\%$ (367/377) with $95\% \mathrm{CI}$ : $95.2 - 98.7\%$ , and a negative percent agreement of $94.4\%$ (1441/1527) with $95\% \mathrm{CI}$ : $93.1 - 95.5\%$ .
b. Matrix comparison:
Not applicable
# 3. Clinical studies:
# a. Clinical Sensitivity:
One thousand nine hundred and four (1904, 1891 prospective and 13 retrospective) stool samples collected from patients suspected of having $C$ difficile infection (CDI) were tested at three sites (USA and Europe). The age groups of the patients range from one year to 100 years. A single replicate of each sample was tested using VIDAS $C$ difficile GDH on a VIDAS instrument. A bacterial culture test was performed for each sample on a CCFA medium according to the instructions for use. Data from the study are summarized in the following tables:
{14}
Performance of the VIDAS C. difficile GDH assay versus CCFA bacterial culture on prospective samples
| | CCFA bacterial culture test | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 (EU) | | Site 2 (US) | | Site 3 (US) | | Total (All Sites) | |
| | Pos | Neg | Pos | Neg | Pos | Neg | Pos | Neg |
| VIDAS Pos | 42 | 24 | 78 | 21 | 163 | 113 | 283 | 158* |
| C. difficile GDH Neg | 7 | 451 | 4 | 363 | 2 | 623 | 13** | 1437 |
| Total | 49 | 475 | 82 | 384 | 165 | 736 | 296 | 1595 |
| Performance | % | [CI95%] | % | [CI95%] | % | [CI95%] | % | [CI95%] |
| Sensitivity | 85.7% | [72.8 - 94.1]% | 95.1% | [88.0 - 98.7]% | 98.8% | [95.7 - 99.9]% | 95.6% | [92.6 - 97.6]% |
| Specificity | 94.9% | [92.6 - 96.7]% | 94.5% | [91.8 - 96.6]% | 84.6% | [81.8 - 87.2]% | 90.1% | [88.5 - 91.5]% |
| Negative Predictive Value (NPV) | 98.5% | [96.9 - 99.4]% | 98.9% | [97.2 - 99.7]% | 99.7% | [98.8 - 99.9]% | 99.1% | [98.5 - 99.5]% |
* 158 samples were found positive with the VIDAS C. difficile GDH assay and negative with the CCFA bacterial culture test, 73 of which were found positive and 85 negative with the commercially available C. difficile GDH assay.
**13 samples were found negative with the VIDAS C. difficile GDH assay and positive with the CCFA bacterial culture test, 9 of which were found negative and 4 positive with the commercially available C. difficile GDH assay.
Out of 2038 patient samples tested with the VIDAS C. difficile GDH assay, 21 (1.0%) were reported as invalid.
## Testing on retrospective samples
Thirteen (13) retrospectively collected samples from pediatric patients submitted for routine C. difficile testing (2-12 years) were assayed for C. difficile according to the same protocol. For these 13 retrospective samples alone, the VIDAS C. difficile GDH assay demonstrated a sensitivity of 100.0% (10/10) and a specificity of 33.3% (1/3).
## Performance of the VIDAS C. difficile GDH assay versus CCFA bacterial culture on all prospective and retrospective samples
For all 1904 specimens and all sites combined, the VIDAS C. difficile GDH assay demonstrated a sensitivity of 95.8% (293/306) with 95% CI: 92.8 – 97.7%, a specificity of 90.0% (1438/1598) with 95% CI: 88.4 – 91.4%, and a negative predictive value of 99.1% with 95% CI: 98.5 – 99.5%.
{15}
Sensitivity and Specificity performances versus CCFA medium by age group on prospective samples
| Age Group | VIDAS Positive /CCFA Positive | Sensitivity [CI95] % | VIDAS Negative /CCFA Negative | Specificity [CI95] % |
| --- | --- | --- | --- | --- |
| < 2 years | 1/1 | 100.0% [2.5 – 100.0]% | 2/2 | 100.0% [15.8 – 100.0]% |
| 2-12 years | 12/12 | 100.0% [73.5 – 100.0]% | 39/44 | 88.6% [75.4 - 96.2]% |
| 13-21 years | 13/13 | 100.0% [75.3 – 100.0]% | 40/45 | 88.9% [75.9 - 96.3]% |
| 22-59 years | 122/125 | 97.6% [93.1 - 99.5]% | 562/632 | 88.9% [86.2 - 91.3]% |
| ≥ 60 years | 135/145 | 93.1% [87.7 - 96.6]% | 794/872 | 91.1% [89.0 - 92.9]% |
For all 69 (56 prospective and 13 retrospective) samples from the 2-12 years pediatric population, the VIDAS C. difficile GDH assay demonstrated a sensitivity of $100.0\%$ (22/22) with $95\% \mathrm{CI}$ : $84.6 - 100.0\%$ , and a specificity $85.1\%$ (40/47) with $95\% \mathrm{CI}$ : $71.7 - 93.8\%$ .
b. Clinical specificity:
See section M3a above.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The following table summarizes the expected (observed) values for the studies performed with the VIDAS GDH Assay using a population of 1891 prospectively collected specimens submitted to the laboratory on suspicion of CDI. This population was $41.2\%$ male, $58.8\%$ female with a mean age of 58 years. The VIDAS GDH test values ranged from 0.00 to 7.98 with overall positivity rate of $23.3\%$ and overall negativity rate $76.7\%$ .
{16}
| Age Group | N | Age (Mean) | Male (%) | Female (%) | Test Values Range | Positive Results (%) | Negative Results (%) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| <2 years | 3 | 1 | 66.7 | 33.3 | (0.00 – 5.55) | 33.3% | 66.7% |
| 2-12 years | 56 | 6 | 48.2 | 51.8 | (0.00 – 7.77) | 30.4% | 69.6% |
| 13-21 years | 58 | 17 | 48.3 | 51.7 | (0.00 – 6.91) | 31.0% | 69.0% |
| 22-59 years | 757 | 44 | 41.6 | 58.4 | (0.00 – 7.98) | 25.4% | 74.6% |
| ≥ 60 years | 1017 | 73 | 40.0 | 60.0 | (0.00 – 7.63) | 20.9% | 79.1% |
| All | 1891 | 58 | 41.2 | 58.8 | (0.00 – 7.98) | 23.3% | 76.7% |
N. Instrument Name:
The VIDAS and miniVIDAS instruments.
O. System Descriptions:
1. Modes of Operation:
The VIDAS instrument (K891385) was cleared in 1989. The VIDAS instrument is attached to a computer and printer. Each instrument has five independent sections allowing five different assays to be run simultaneously. Each section can process up to six samples. Therefore, a fully loaded VIDAS can process thirty samples. A smaller compact version of the VIDAS instrument, appropriately called the miniVIDAS, was cleared under K923579 in 1993. The miniVIDAS instrument has a built-in computer, keyboard and printer. Two independent sections each accept six tests and can process up to twelve samples simultaneously. Each FDA cleared VIDAS assay can be run on either the VIDAS or the miniVIDAS instrument.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
☐ Yes ☐ X or No
3. Specimen Identification:
All assay steps are controlled automatically by the instrument. The sample is transferred into the wells. The STR strip consists of ten wells covered with labeled foil seal. The label comprises a bar code which indicates the assay code, kit lot number and expiration date.
{17}
4. Specimen Sampling and Handling:
The solid phase receptor (SPR) serves as both the solid phase and the pipetting device. The foil of the first well is perforated to allow introduction of the sample. The last well (well ten) of each strip is a cuvette in which the fluorometric reading is performed. The center wells of the strip contain the various reagents required for the assay.
5. Calibration:
The kit contains a GDH standard (S1), a positive control (C1) and a negative control (C2). Each calibration must be checked using these controls. Calibration should be performed with each new lot of reagents and then every 28 days.
6. Quality Control:
The kit contains a Master Lot Entry card (MLE) which contains specifications for the factory master data required to calibrate the test. Data from the MLE card is entered into the instrument before each new lot of reagents is used. Calibration provides instrument-specific information and compensates for possible minor variations in assay signal throughout the shelf-life of the kit.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.