The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
Device Story
System comprises flexible, steerable endoscope and camera control unit (CCU). Endoscope features plastic exterior, control lever for distal tip angulation, integrated LED light source, and CMOS imaging sensor at shaft tip. Light transmitted via fiberoptic bundles; video signal captured by CMOS sensor and transferred to CCU. CCU regulates light intensity and connects to external monitors, PCs, or recorders for visualization/documentation. Includes ventilation system for shaft protection and leakage testing capability. Used in clinical settings by physicians/specialists for airway/nasal visualization. Output displayed on video monitor to assist clinical diagnosis. Benefits include enhanced visualization of anatomy for diagnostic procedures.
Clinical Evidence
Bench testing only. Performance data demonstrated compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, and ISO 10993. Reprocessing validation performed per AAMI TIR12:2010 and AAMI TIR30:2003, including manual cleaning, high-level disinfection, and STERRAD 100S sterilization.
Technological Characteristics
Flexible steerable endoscope with plastic housing; CMOS imaging sensor; LED light source with fiberoptic transmission; ventilation system for leakage testing. Connectivity to external monitors/PCs. Reprocessing via manual cleaning, high-level disinfection, or STERRAD 100S. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, ISO 10993, ISO 14937.
Indications for Use
Indicated for endoscopic diagnosis within nasal lumens and airway anatomy in patients requiring visualization via video monitor. Restricted to use by medically qualified professionals trained in rhino/laryngoscopic techniques.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM (K103467)
{1}------------------------------------------------
Schoelly Fiberontic GmbH 510(k) Premarket Notification (Traditional)
06/21/2013 CMOS Video Nusopharyngoscope System
K132009
## Section 5-510(k) Summary
| General Information | |
|-------------------------|--------------------------------------------------------------------------|
| Preparation date: | 06/21/2013 |
| Owner's Name: | Schoelly Fiberoptic GmbH (Registration: 8043903) |
| Address: | Robert-Bosch-Str. 1-3<br>79211 Denzlingen<br>Germany |
| Telephone Number: | +49-7666-980-0 |
| Fax Number: | +49-7666-908-380 |
| Contact Person: | Dr. Sandra Baumann |
| Subject Device Name: | |
| Trade Name: | Schoelly CMOS Video Nasopharyngoscope System |
| Trade Name: | Schoelly CMOS Video Nasopharyngoscope System |
| Common/Usual Name: | Video Nasopharyngoscope System |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Predicate Device Name: | |
| Trade Name: | Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM |
| Trade Name: | Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM |
| Common/Usual Name: | Video Rhino-Laryngoscope System |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Premarket Notification: | K103467 (Karl Storz Endoscopy-America, Inc.), SE date June 28, 2012 |
| Predicate Device Name: | |
| Trade Name: | Schoelly Naso-Laryngo-Pharyngoscope |
| Trade Name: | Schoelly Naso-Laryngo-Pharyngoscope |
| Common/Usual Name: | Nasopharyngoscope |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Premarket Notification: | K083553 (Schoelly Imaging, Inc.), SE date January 16, 2009 |
#### Device Description
The Schoelly CMOS Video Nasopharyngoscope System consists of a flexible and steerable endoscope and a camera control unit (CCU) for regulation of light intensity and connection to a monitor, PC, medical video recorder or printer for image display or image documentation.
The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and an integrated LED light source. Light is transmitted through fiberoptic bundles illuminating the anatomy under investigation. The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope shaft and transferred to the CCU.
The endoscope further incorporates a ventilation system to protect the shaft. The exhaust valve at the endoscope handle can further be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories.
The Schoelly CMOS Video Nasopharyngoscope System is delivered in a non-sterile condition and is already CE marked.
{2}------------------------------------------------
06/21/2013 CMOS Video Nasopharyngoscope System
#### Predicate Devices
The Karl Storz predicate device is based on the same technology has the same essential design and dimension characteristics, the same essential optical parameters and the same principle of operations. It is made out of the same primary materials and the same standards were applied and met for design verification and validation as compared to the proposed Schoelly CMOS Video Nasopharyngoscope System.
The Schoelly predicate device differs with respect to the image transfer technology; however this device is built out of identical components/systems for light transmission, angulation of the endoscope's distal tip or ventilation. Further, the manufacturing process for major parts of the endoscope is identical as compared to the endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System.
#### Indications for Use
The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
The indications for use of the proposed devices are identical to the indications for use of the Karl Storz predicate device.
#### Non-clinical Performance Testing
Performance data demonstrated that the Schoelly CMOS Video Nasopharyngoscope System has met pre-determined acceptance criteria and is substantially equivalent to the predicate devices. The device is as safe, as effective, and performs as well as or better than the predicate devices. The risks associated with use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits of predicate devices are the same as compared to the proposed one.
Testing has been conducted as per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, and ISO 10993 to address the safety and performance aspects of the device.
#### Reprocessing
The endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System is the subject of completed reprocessing validations including manual cleaning, high-level disinfection and STERRAD® 100S sterilization.
Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilties: A Guide for Device Manufacturers) and AAMI TIR30:2003 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices). Devices that have been used for testing had been exposed to extensive clinical use including tip angulation, soiling, and reprocessing to address the FDA Draft Guidance Processing Medical Devices in Healthcare Settings (dated: May 2, 2011).
Sterilization studies have been performed in accordance with ISO 14937.
#### Conclusion
The Schoelly CMOS Video Nasopharyngoscope System meets all the pre-determined acceptance criteria of the testing performed to confirm substantial equivalence to the predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# April 9, 2014
Schoelly Fiberoptic GmbH c/o Ms. Pamela Papineau Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, MA 01432
Re: K132009
Trade/Device Name: Schoelly CMOS Video Nasopharyngoscope System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: March 3, 2014 Received: March 5, 2014
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Ms. Pamela Papineau
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman. M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132009
Device Name: Schoelly CMOS Video Nasopharyngoscope System
Indications for Use:
The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhinoflaryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
.
Concurrence of Center for Devices and Radiological Health (CDRH)
Sunny Park 2014.04.08 14:30:52 -04'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.